← Product Code [QTG](/productcode/QTG) · DEN210039

# SNOO Smart Sleeper (DEN210039)

_Happiest Baby, Inc. · QTG · Mar 30, 2023 · General Hospital · DENG_

**Canonical URL:** https://fda.innolitics.com/device/DEN210039

## Device Facts

- **Applicant:** Happiest Baby, Inc.
- **Product Code:** [QTG](/productcode/QTG.md)
- **Decision Date:** Mar 30, 2023
- **Decision:** DENG
- **Submission Type:** Direct
- **Regulation:** 21 CFR 880.5690
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| DEN210039 · Mar 30, 2023 | SNOO Smart Sleeper | Happiest Baby, Inc. | User-reported data; Video recordings of infants in home environments; CDC WONDER database (historical control data); Manufacturer-collected real-world user data | The sponsor used real-world data to demonstrate the device's ability to maintain infants in a supine position and to compare the incidence of SIDS/SUID in SNOO users against historical population data to ensure the device does not increase risk. | SIDS/SUID incidence; Supine sleep position; Real-world user cohort; Historical control comparison |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| SNOO User Real-World Evidence Analysis; Retrospective observational cohort study | SNOO Smart Sleeper users (infants 0-6 months); Sample Size: 1,012 users (supine position study); 106,723 users (SIDS/SUID safety analysis); Number of Sites: Not specified (Home use) | Historical data from the CDC WONDER database | Rate of supine sleep position; incidence of SIDS/SUID |

## Indications for Use

The SNOO Smart Sleeper bassinet plus the SNOO Sleep Sack are jointly intended to facilitate a supine position during sleep. Infants who are placed in a supine sleep position are at lower risk of SIDS/SUID. The device is intended for home use by caregivers of infants from birth to 6 months of age, who are not yet able to roll over consistently.

## Device Story

SNOO Smart Sleeper consists of a bassinet and integrated sleep sack; designed to facilitate supine sleep position for infants. Device used in home environment by caregivers. System provides securement to maintain infant on back; intended to reduce risk of SIDS/SUID by preventing rolling. Operates via mechanical structure; includes components for infant containment. Caregivers place infant in sleep sack, which secures to bassinet. Device requires adherence to safe sleep practices and supervision. Benefits include consistent supine positioning for infants unable to roll over independently.

## Clinical Evidence

No clinical data provided in the document. FDA requires premarket clinical information to demonstrate the device holds the infant on the back, provide data on adverse events/malfunctions, and demonstrate that use does not increase the rate of SIDS/SUID.

## Technological Characteristics

Infant supine sleep system; includes bassinet and sleep sack. Requires biocompatibility evaluation for patient-contacting components. Mechanical stability testing required to prevent tipping. Material compatibility testing required for cleaning durability.

## Regulatory Identification

The SNOO Smart Sleeper is an infant supine sleep system comprised of a bassinet and a sleep sack that jointly constitute an infant securement device intended to facilitate a supine position during sleep. It is intended for home use by caregivers of infants from birth to 6 months of age who are not yet able to roll over consistently.

## Special Controls

In combination with the general controls of the FD&C Act, the infant supine sleep system is subject to the following special controls:

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a De Novo](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# DE NOVO CLASSIFICATION REQUEST FOR SNOO SMART SLEEPER

### REGULATORY INFORMATION

FDA identifies this generic type of device as:

Infant supine sleep system. An infant supine sleep system is a device intended to facilitate a supine position during sleep for use in infants that are not yet able to roll over consistently. Infants placed in a supine sleep position are at lower risk of sudden infant death syndrome (SIDS) or sudden unexpected infant death (SUID).

NEW REGULATION NUMBER: 21 CFR 880.5690

CLASSIFICATION: Class II

PRODUCT CODE: OTG

### BACKGROUND

DEVICE NAME: SNOO Smart Sleeper

SUBMISSION NUMBER: DEN210039

DATE DE NOVO RECEIVED: September 20, 2021

### SPONSOR INFORMATION:

Happiest Baby, Inc. 3115 S. La Cienega Blvd. Los Angeles, California 90016

### INDICATIONS FOR USE

The SNOO Smart Sleeper is indicated as follows:

The SNOO Smart Sleeper bassinet plus the SNOO Sleep Sack are jointly intended to facilitate a supine position during sleep. Infants who are placed in a supine sleep position are at lower risk of SIDS/SUID. The device is intended for home use by caregivers of infants from birth to 6 months of age, who are not yet able to roll over consistently.

### LIMITATIONS

The SNOO is demonstrated to facilitate a supine position during sleep; however, the SNOO has not directly demonstrated a reduction in the incidence of SIDS/SUID.

{1}------------------------------------------------

# This device is only indicated for use with infants who cannot consistently roll over.

The device is not a substitute for adult supervision.

Do not use this product if the infant can push up on hands and knees, can roll over consistently, or has reached 6 months of age, whichever comes first.

## DEVICE DESCRIPTION

The SNOO Smart Sleeper (SNOO) is comprised of the SNOO bassinet and the SNOO Sleep Sack that iointly constitute an infant securement device intended to facilitate a supine position during sleep. The device is intended for home use by caregivers of infants from birth to 6 months of age, who are not yet able to roll over consistently.

The SNOO Sleep Sack is a cotton swaddle with fixed wings made of woven cotton fabric that extend to the right and left of the infant's body. Three sizes of sleep sacks (small 5-12lbs, medium 12-18lbs. and large 18-25lbs) are provided with each bassinet to accommodate the growing baby. Small loops at the end of each SNOO Sleep Sack wing are slid over the safety clips of the SNOO bassinet to securely position the swaddled infant on the back. The SNOO Sleep Sack has corresponding wings that are designed to be physically attached to those clips after the baby is wrapped in the sleep sack - thereby keeping the infant securely positioned on the back during sleep and to prevent rolling.

Image /page/1/Picture/6 description: The image shows a baby swaddled in a white blanket, lying on a black and white patterned surface. The baby is lying on its back, with its head turned to the left. The swaddle is wrapped tightly around the baby's body, with only its head and face visible. The baby appears to be content and relaxed. The black and white patterned surface appears to be a baby lounger.

The SNOO bassinet also provides sound and motion. White noise and rocking can be automatically increased in response to a baby's cries. There is a SNOO App which can be used to adjust the levels of motion.

## SUMMARY OF NONCLINICAL/BENCH STUDIES

## BIOCOMPATIBILITY/MATERIALS

{2}------------------------------------------------

The patient-contacting components of the SNOO Smart Sleeper (SNOO) include the SNOO bassinet fitted sheet and the SNOO Sleep Sack. The SNOO bassinet fitted sheet is made of 100% cotton. and the SNOO Sleep Sack is made of 95% cotton and 5% elastane. and features mesh panels made of 100% polyester. Biocompatibility testing was performed on the patient contacting components of the SNOO. Testing included Cytotoxicity testing per ISO 10993-5:2009 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity, sensitization testing per ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization, and irritation testing per ISO 10993-23:2021 Biological Evaluation of Medical Devices, Part 23: Tests for Irritation on the final finished device.

# SHELF LIFE

Durability of the SNOO Sleep Sack's zipper and safety loops were evaluated. The effect of washing over the expected lifecycle of the SNOO Sleep Sack and SNOO bassinet fitted sheet was assessed. Use life testing was completed to demonstrate a use life of 6 months.

# ELECTROMAGNETIC CAPABILITY & ELECTRICAL SAFETY

Electrical safety testing was performed in accordance with the following standards:

- . BS EN 61558-1 Safety of transformers, reactors, power supply units and combinations thereof - General requirements and tests
- . BS EN 61558-2-16 Safety of transformers, reactors, power supply units and combinations thereof - Part 2-16: Particular requirements and tests for switch mode power supply units and transformers for switch mode power supply units for general applications.
- . UL 1310 - Class 2 Power Units CAN/CSA C22.2 No. 223 - Power Supplies with Extra-low Voltage Class 2 Outputs.

Electromagnetic compatibility (EMC) testing was conducted per the FDA Recommended Standard IEC 60601-1-2. The SNOO Smart Sleeper demonstrated that it complies with the emission limits conducted for CISPR 11. Additionally, conducted and radiated emissions testing was performed per CFR 47 FCC Part 15. Subpart B. which included EMC testing to the same frequencies (0.15 MHz- 30MHz) and acceptance limits as IEC 60601-1-2.

# PERFORMANCE TESTING - BENCH

The SNOO Smart Sleeper bassinet was tested in accordance with ASTM F2194-16e1 (Standard Consumer Safety Specification for Bassinets and Cradles) and 16 CFR 1218 (Requirements for bassinets and cradles).

The SNOO Sleep Sack was tested as described in the table below.

{3}------------------------------------------------

| Test<br>Performed             | Device<br>Description /<br>Sample Size                | Test<br>Method                                                                                                                                                                                 | Acceptance<br>Criteria                                                                          | Results<br>(PASS/ FAIL) |
|-------------------------------|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|-------------------------|
| Zipper<br>Durability<br>Test  | 10 zippers /<br>batch                                 | Open and close the zipper 500 times at<br>an approximate speed of 10cm/sec.                                                                                                                    | 500 cycles without<br>any jamming or<br>breaking issues.                                        | PASS                    |
| Loop Hook<br>&<br>Unhooking   | 3 SNOO<br>Sacks<br>(1 Small.<br>1 Medium,<br>1 Large) | Perform 500 hook and unhook cycles<br>on SNOO Sleep Sack elastic loops to<br>the safety clips, per the IFU<br>instructions.                                                                    | Visual Inspection<br>for deterioration of<br>components after<br>500 hook and<br>unhook cycles. | PASS                    |
| Washing<br>Durability<br>Test | 10 pieces /<br>batch                                  | Utilizing any commercial large<br>washing machine and household<br>clothes drying machine, wash and dry<br>the Sleep Sack for 100 cycles. After<br>every 10 cycles, measure the<br>dimensions. | Dimension<br>deviates less than<br>10%, except for<br>safety clip width<br>(<5%).               | PASS                    |
| Velcro<br>Durability<br>Test  | 10 pieces /<br>batch                                  | After a wash and dry cycle, perform<br>open and close cycles on the hook<br>(male) and loop (female) Velcro strips.                                                                            | Velcro adheres at<br>360 uses.                                                                  | PASS                    |

# HUMAN FACTORS USABILITY TESTING

A Human Factors validation study was completed in accordance with the FDA 2016 guidance "Applying Human Factors and Usability Engineering to Medical Devices". The study encompassed a series of scenarios in a simulated use environment that contained tasks related to use of the system by its intended users. The scenarios were tied to the risk analysis process and included user tasks that are typically done when using the system. Fifteen test participants were evaluated on tasks that included unpacking the SNOO, setting up the system, swaddling a baby (doll) in the SNOO Sleep Sack, securing the baby (doll) in the SNOO bassinet, and properly removing the baby (doll) from the device. The validation study met the acceptance criteria and was deemed acceptable.

# SUMMARY OF CLINICAL INFORMATION

Happiest Baby provided real world evidence (RWE) showing a rate of supine sleep position of approximately 98.7% in 1012 users (confidence interval [97.8%,99.2%]). In addition, over 93 hours of video recordings showed all observed infants remaining in a supine position during their time in the SNOO. In addition, Happiest Baby provided real-world evidence on 106,723 SNOO users and compared the incidence of reported SIDS/SUID in SNOO users to historical data from the CDC Wide-ranging Online Data for Epidemiological Research (WONDER) database. Although the data was not of sufficient quality to determine a precise SIDS/SUID rate, it did demonstrate that the device did not increase the risk of SIDS/SUID in the study population.

{4}------------------------------------------------

The following adverse events, including serious adverse events, have been reported at or below expected rates when using the SNOO Smart Sleeper: death, gastroesophageal reflux, plagiocephaly, aspiration, contusion, and dermatitis.

## POSTMARKET SURVEILLANCE

The clinical data described above were collected in the United States but did not evaluate the device in a population that is reflective of the entire US population, including the US subpopulations at highest risk of SIDS/SUID. A section 522 postmarket surveillance study will therefore be required to obtain a more comprehensive safety profile of the device by collecting safety data associated with use of the device in US subpopulations at higher risk of SIDS/SUID to determine the rates of serious injury and death for high-risk infants when they use the SNOO Smart Sleeper. Safety data will be collected and provided in updated labeling.

# LABELING

The device labeling includes the following:

- On the packaging and in directions for use, a prominent warning that the device has not a. been demonstrated to reduce the risk of SIDS/SUID:
- b. A summary of available clinical data with the device, including a discussion of adverse events:
- c. A warning that the device is only indicated for use with infants who cannot consistently roll over:
- d. The level of supervision necessary to monitor a sleeping infant in the intended use environment:
- e. Recommendations for safe sleep environments;
- f. Instructions and guidelines to ensure proper fit of the infant into the sleep system; and
- g. Instructions for maintaining and reprocessing the device.

## RISKS TO HEALTH

The table below identifies the risks to health that may be associated with use of an Infant Supine Sleep System.

## Risk and Mitigation Table

| Risk to Health                                                                                              | Mitigation Measures                                                                      |
|-------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Increased risk of death, including from<br>inadequate securement or inadequate<br>positioning of the infant | Clinical data<br>Postmarket surveillance<br>Non-clinical performance testing<br>Labeling |
| Inappropriate securement leading to<br>Injuries, contusions, or bruising Entrapment                         | Clinical data<br>Postmarket surveillance<br>Labeling                                     |

{5}------------------------------------------------

| Risk to Health                                                                                                            | Mitigation Measures                                   |
|---------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| • Respiratory compromise or suffocation<br>• Gastroesophageal reflux<br>• Plagiocephaly (“flat head syndrome”)<br>• Death |                                                       |
| Inappropriate or inadequate securement due to<br>device degradation over time (wear and tear,<br>laundering)              | Non-clinical performance testing<br>Labeling          |
| Inappropriate use or inadequate securement due<br>to use error and/or improper fit                                        | Clinical data<br>Human factors assessment<br>Labeling |
| Injury due to unstable device (tipping, rocking,<br>improper placement)                                                   | Non-clinical performance testing<br>Labeling          |
| Infection                                                                                                                 | Labeling                                              |
| Adverse tissue reaction (e.g., dermatitis)                                                                                | Biocompatibility evaluation<br>Labeling               |

# SPECIAL CONTROLS

In combination with the general controls of the FD&C Act, the infant supine sleep system is subject to the following special controls:

- Premarket clinical information and, as determined by FDA, postmarket surveillance data (1) acquired under anticipated conditions of use must be collected to fulfill the following:
	- Demonstrate that the device holds the infant on the back; (i)
	- Provide data on adverse events (including deaths and injuries) and malfunctions to (ii) demonstrate the device can be safely used in the intended use population; and
	- (iii) Provide data to demonstrate that use of the device does not increase the rate of SIDS/SUID in the intended use population.
- Human factors testing must demonstrate that the user can safely and correctly use the (2) device.
- The patient-contacting components of the device must be demonstrated to be (3) biocompatible.
- (4) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following must be conducted:
	- (i) Testing to ensure the mechanical and structural stability of the device and demonstrate that the device does not present a tipping hazard due to mechanical failures; and
	- Material compatibility testing to demonstrate that the cleaning instructions provided (ii) by the manufacturer do not cause crazing, cracking, or deterioration of the device.
- Labeling must include: (5)

{6}------------------------------------------------

- (i) Unless clinical performance data demonstrates that it can be removed or modified, a prominent warning that the device has not been demonstrated to reduce the risk of SIDS/SUID. Such warning must appear prominently on all labeling:
- (ii) A summary of available clinical information with the device, including a discussion of adverse events:
- A warning that the device is only indicated for use with infants who cannot (111) consistently roll over;
- (iv) Instructions to ensure proper fit;
- (v) Instructions for cleaning the device; and
- (vi) Information regarding safe sleep practices to ensure the safe use of the device, including:
	- (A) Recommendations for safe sleep environments; and
	- The level of supervision necessary to monitor a sleeping infant. (B)

## BENEFIT-RISK DETERMINATION

The probable benefits of the device are based on clinical data, as described above. The data demonstrated that the SNOO Smart Sleeper secures the infant in a supine position during sleep. Infants who are placed in a supine sleep position are at lower risk of SIDS/SUID (Moon, R. Y .. et al. (2022). "Sleep-Related Infant Deaths: Updated 2022 Recommendations for Reducing Infant Deaths in the Sleep Environment." Pediatrics 150(1).).

The probable risks of the device are also based on clinical data. A securement device that is misused or poorly fitted presents risks to the infant such as entrapment, suffocation, and SIDS/SUID. Plagiocephaly, also known as flat head syndrome, may occur as a result of placement in the supine position for an extended period of time. Gastroesophageal reflux was also observed. These two were the most common risks. Contusion, aspiration and dermatitis were also observed, but to a lesser extent.

The data collected for the SNOO Smart Sleeper did not demonstrate that use of the device decreased the risk of SIDS/SUID, and the device was not extensively studied in infants demographically at highest risk of SIDS/SUID. However, the data did demonstrate that the device could secure an infant on its back. Safe to Sleep guidelines state that placing infants on the back to sleep is associated with lower rates of SIDS/SUID (https://safetosleep.nichd.nih.gov/). The risks reported from the study were well understood and in line with historical data and clinical experience. FDA has determined that while additional data are necessary to further assess the benefits and risks of the device in the most vulnerable infants, the benefits of the device outweigh its risks as studied.

In order to mitigate the risks of a securement device, caregivers should read the directions for use carefully and ensure that the device is used accordingly and with appropriate supervision. Safe to Sleep guidelines continue to recommend that an infant be placed on its back to sleep with no other objects in the crib.

### Patient Perspectives

{7}------------------------------------------------

This submission did not include specific information on patient perspectives for this device.

### Benefit/Risk Conclusion

In conclusion, given the available information above, for the following indication statement:

The SNOO Smart Sleeper bassinet plus the SNOO Sleep Sack are jointly intended to facilitate a supine position during sleep. Infants who are placed in a supine sleep position are at lower risk of SIDS/SUID. The device is intended for home use by caregivers of infants from birth to 6 months of age, who are not yet able to roll over consistently.

The probable benefits outweigh the probable risks for the SNOO Smart Sleeper. The device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.

### CONCLUSION

The De Novo request for the SNOO Smart Sleeper is granted and the device is classified as follows:

Product Code: QTG Device Type: Infant supine sleep system Regulation Number: 21 CFR 880.5690 Class: II

---

**Source:** [https://fda.innolitics.com/device/DEN210039](https://fda.innolitics.com/device/DEN210039)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a De Novo](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
