Retrospective clinical samples from previously infected individuals; Pre-pandemic clinical samples (collected prior to December 2019)
Retrospective clinical samples were used to evaluate the clinical sensitivity (Positive Percent Agreement) and clinical specificity (Negative Percent Agreement) of the VITROS Anti-SARS-CoV-2 IgG test.
Clinical Agreement Study; Retrospective clinical study; Study Period: April 2020 - March 2021
Population 1: 338 individuals with prior SARS-CoV-2 positive RT-PCR result; Population 2: 304 samples collected prior to December 2019.; Sample Size: 642; Number of Sites: 3
RT-PCR test
Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA)
Indications for Use
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Reagent Pack when used in combination with the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator is a chemiluminescent immunoassay intended for the qualitative detection of IgG antibodies to SARS-CoV-2 in human serum and plasma (K2-EDTA and K3-EDTA) samples collected on or after 15 days post-symptom onset using the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems. The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test is intended for use as an aid in identifying individuals with an adaptive immune response to SARS-CoV-2, indicating recent or prior infection.
Device Story
The VITROS Anti-SARS-CoV-2 IgG test is a chemiluminescent immunoassay for laboratory professional use on VITROS automated systems. It detects IgG antibodies in human serum or plasma (K2/K3-EDTA). The test uses a two-stage immunometric reaction: patient antibodies bind to SARS-CoV-2 spike protein-coated wells; unbound sample is washed away; HRP-labeled anti-human IgG conjugate is added and binds to the complex. A luminogenic substrate is added; HRP catalyzes light emission, which is measured by the system. Results are calculated as a normalized signal (S/C) relative to a cutoff. The output aids clinicians in identifying individuals with an adaptive immune response to SARS-CoV-2, indicating recent or prior infection. It supports infection control and surveillance efforts.
Clinical Evidence
No clinical data provided in the document.
Technological Characteristics
Chemiluminescent immunoassay; utilizes VITROS ECi/ECiQ/3600 and 5600/XT 7600 systems. Reagents include Reagent Pack and Calibrator. Qualitative detection of IgG antibodies. Automated processing. Requires specific specimen types (serum, K2-EDTA/K3-EDTA plasma).
Indications for Use
Indicated for qualitative detection of IgG antibodies to SARS-CoV-2 in human serum and plasma (K2-EDTA, K3-EDTA) from individuals 15+ days post-symptom onset. Aid in identifying adaptive immune response indicating recent/prior infection. Not for acute infection diagnosis or donor screening.
Regulatory Classification
Identification
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test is a qualitative chemiluminescent immunoassay intended for the detection of IgG antibodies to SARS-CoV-2 in human serum and plasma (K2-EDTA and K3-EDTA). It is performed on the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems. The test is intended for use as an aid in identifying individuals with an adaptive immune response to SARS-CoV-2, indicating recent or prior infection.
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol features a stylized image of an eagle, while the text is written in blue, with "FDA" in a larger, bolder font than the rest of the text.
# EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Reagent Pack, VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator DECISION SUMMARY
#### Background Information: I
### A De Novo Number
DEN210038
### B Applicant
Ortho-Clinical Diagnostics, Inc.
### C Proprietary and Established Names
VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Reagent Pack, VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator
#### D Regulatory Information
| Product<br>Code(s) | Classification | Regulation<br>Section | Panel |
|--------------------|----------------|------------------------------------------------|-------------------|
| QVP | Class II | 21 CFR 866.3983 - SARS-<br>CoV-2 serology test | MI - Microbiology |
#### II Submission/Device Overview:
### A Purpose for Submission:
De Novo request for evaluation of automatic class II designation for the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test comprised of (1) the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Reagent Pack, (2) the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator, and (3) the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Controls.
#### B Measurand:
IgG antibodies to SARS-CoV-2 in human serum, K2-EDTA plasma, and K3-EDTA plasma.
### C Type of Test:
Chemiluminescent Immunoassay.
#### III Indications for Use:
{2}------------------------------------------------
### A Indication(s) for Use: Rx ONLY
For in vitro diagnostic and laboratory professional use.
# VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Reagent Pack
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Reagent Pack when used in combination with the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator is a chemiluminescent immunoassay intended for the qualitative detection of IgG antibodies to SARS-CoV-2 in human serum and plasma (K2-EDTA and K3-EDTA) samples collected on or after 15 days post-symptom onset using the VITROS ECi/ECiO/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems. The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test is intended for use as an aid in identifying individuals with an adaptive immune response to SARS-CoV-2, indicating recent or prior infection.
VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator
For use in the calibration of the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems for the in vitro qualitative detection of IgG antibodies to SARS-CoV-2 in human serum and plasma.
# B Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
### C Special Instrument Requirements:
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Reagent Pack when used in combination with the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrators uses the following VITROS Systems (instruments):
- VITROS ECi/ECiQ Immunodiagnostic Systems .
- VITROS 3600 Immunodiagnostic Systems .
- . VITROS 5600 Integrated Systems
- . VITROS XT 7600 Integrated Systems
These VITROS systems were cleared previously as part of premarket notifications: K962919, K083173, K081543, and K182063 respectively,
#### IV Device/System Characteristics:
### A Device Description:
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test is a qualitative chemiluminescent immunoassay performed on the VITROS Systems (VITROS ECi/ECiO Immunodiagnostic System, VITROS 3600 Immunodiagnostic System, VITROS 5600 Integrated
{3}------------------------------------------------
System and VITROS XT 7600 Integrated System) providing fully automated random-access testing.
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test is performed using the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Reagent Pack in combination with the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator and the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Controls on the VITROS Systems.
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Reagent Pack is supplied as ready to use and contains:
- . 100 wells coated with 100ng/well of recombinant SARS-CoV-2 spike antigen derived from human cells.
- 18.0 mL assay reagent (buffer with bovine protein stabilizers and antimicrobial agent) .
- . 20.4 mL conjugate reagent [anti-human IgG (murine monoclonal) conjugated to horseradish peroxidase, 5ng/mL] in buffer with bovine protein stabilizers and antimicrobial agent.
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator contains:
- 2 vials of VITROS Immunodiagnostic Products Anti-SARS-CoV-2 1gG Calibrator 0 (anti-SARS-CoV-2 IgG in anti-SARS-CoV-2 IgG negative human serum with antimicrobial agent, 1 mL)
- Lot calibration card .
- . Protocol card
- . 8 calibrator bar code labels
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Controls contain:
- 3 sets of VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Controls 1 and 2 . (defibrinated human plasma with anti-microbial agent, 2 mL). Control 1 is non-reactive and Control 2 is reactive
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test is designed for use on the VITROS Systems. The VITROS Systems use the following ancillary reagents (general purpose reagents):
- . VITROS Immunodiagnostic Products Signal Reagent
- VITROS Immunodiagnostic Products Universal Wash Reagent .
# B Principle of Operation
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test is performed using the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Reagent Pack and the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator on the VITROS ECi/ECiO/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems. An immunometric technique is used; this involves a two- stage reaction. In the first stage antibodies to SARS-CoV-2 present in the sample bind with SARS-CoV-2 spike protein coated on the well. Unbound sample is removed by washing. In the second stage horseradish peroxidase (HRP)-labeled murine monoclonal anti-human IgG antibodies are
{4}------------------------------------------------
added in the conjugate reagent. The conjugate binds specifically to the antibody portion of the antigen-antibody complex. If complexes are not present, the unbound conjugate is removed by the subsequent wash step.
The bound HRP conjugate is measured by a luminescent reaction. A reagent containing luminogenic substrates (a luminol derivative and a peracid salt) and an electron transfer agent is added to the wells. The HRP in the bound conjugate catalyzes the oxidation of the luminol derivative, producing light. The electron transfer agent (a substituted acetanilide) increases the level of light produced and prolongs its emission. The light signals are read by the system.
| Test Type | System | Incubation Time | Time to first result | Test Temperature | Reaction Sample<br>Volume |
|--------------|----------------------------------|-----------------|----------------------|------------------|---------------------------|
| Immunometric | ECi/ECiQ. 3600.<br>5600. XT 7600 | 37 minutes | 48 minutes | 37 ℃ | 20 ul |
* Not all products and systems are available in all countries.
# Figure 1: Reaction Scheme
Image /page/3/Figure/5 description: The image shows a diagram of a process involving several components. It starts with a recombinant SARS-CoV-2 antigen, followed by Anti-SARS-CoV-2, and then HRP-labeled anti-human IgG murine monoclonal. The process culminates in luminescence, which is achieved using a signal reagent with an enhancer.
# C Instrument Description Information
### 1. Instrument Name:
VITROS Systems (instruments):
- . VITROS ECi/ECiQ Immunodiagnostic System
- . VITROS 3600 Immunodiagnostic System
- VITROS 5600 Integrated System .
- VITROS XT 7600 Integrated System. .
- 2. Specimen Identification:
Not applicable
- 3. Specimen Sampling and Handling:
#### Specimen Sampling:
The specimens recommended for this assay are:
- . Serum
- K2-EDTA Plasma
- K3-EDTA Plasma
{5}------------------------------------------------
# Specimen Handling and Storage:
- . Specimens may be stored for up to 24 hours at room temperature (15-30℃) or 7 days at 2 - 8ºC.
- Specimens may be stored frozen at < 20°C for <4 weeks. .
- Specimens may be subjected to up to one freeze-thaw cvcles. .
# 4. Calibration:
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator is provided together with the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Reagent Pack.
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator contains anti-SARS-CoV-2 IgG in human serum. In addition, the calibrator contains the lot calibration card, protocol card and 8 calibrator bar code labels.
Calibration is lot-specific: reagent packs and calibrators are linked by lot number. Reagent packs from the same lot may use the same calibration. The calibrator is supplied frozen. The analyzer automatically processes the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator in duplicate. Results are calculated as a normalized signal, relative to a cutoff value. During the calibration process a lot-specific parameter is used to determine a valid stored cutoff value for the VITROS Immunodiagnostic and VITROS Integrated Systems.
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test should be calibrated:
- . When the reagent pack and calibrator lot changes
- . Every 28 days
- After specified service procedures have been performed .
- If quality control results are consistently outside of the acceptable range. .
# 5. Quality Control:
The VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Controls are provided separately from the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Reagent Pack, and contains:
- · Control I: Anti-SARS-CoV-2 IgG non-reactive human serum or defibrinated plasma.
- . Control 2: Anti-SARS-CoV-2 IgG reactive human serum or defibrinated plasma.
# Quality Control Procedure
- To verify system performance, analyze control materials: .
- O After calibration
- o If the system is turned off for more than 2 hours
- · After reloading reagent packs that have been removed from the MicroWell Supply and stored for later use
- According to local regulations or at least once each day that the test is performed
{6}------------------------------------------------
- · After specified service procedures are performed
- . If controls results fall outside the acceptance range, investigate the cause before deciding whether to report patient results.
#### V Standards/Guidance Documents Referenced:
CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition
CLSI EP07 Interference testing in Clinical Chemistry
- CLSI EP09 Measurement Procedure Comparison and Bias Estimation Using Patient Samples. Third Edition (2016).
- CLSI EP35 Assessment of Equivalence or Suitability of Specimen Type for Medical Laboratory measurement Procedure
CLSI EP37 Supplemental Tables for Interference Testing in Clinical Chemistry
#### Performance Characteristics: VI
### A Analytical Performance:
#### 1. Precision/Reproducibility:
A. Within-Laboratory Precision: The within laboratory precision of the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test was evaluated with EDTA plasma pools samples (PP4, PP5, and PP6), negative and positive quality control materials (PP1, PP2), and the calibrator (PP3) on the VITROS ECi/ECiQ and 3600 Immunodiagnostics Systems and the VITROS 5600 and XT7600 Integrated Systems following the CLSI document EP05-A3. A total of 3 lots of reagent packs, calibrators and controls were included in the study. For each reagent lot, operators ran two replicates of each precision panel sample on two occasions per day for twenty non-consecutive days.
The data presented are a representation of test performance and are provided as a guideline. Variables such as sample handling and storage, reagent handling and storage, laboratory environment, and system maintenance can affect reproducibility of test results.
Each precision panel was tested in duplicate, in two runs per day, on 20 non-consecutive days, on each instrument for a total of 80 observations per sample and lot. The study included three reagent lots and was evaluated within a single calibration cycle (2 replicates x 2 runs x 20 days x 3 lots= 240 observations total). The VITROS Anti-SARS-CoV-2 IgG's total precision %CV (from the "total" standard deviation) for the S/C values ranged from 3.4% -32.2%, depending upon the sample and instrument.
| VITROS<br>System | Panel<br>member | Number<br>of<br>Observ. | Grand<br>Mean<br>(S/C) | Repeatability<br>(Within Run) | | Between-Run | | Between-Day | | Between-Lot | | Within-<br>Laboratory | |
|------------------|-----------------|-------------------------|------------------------|-------------------------------|-----|-------------|-----|-------------|-----|-------------|-----|-----------------------|-----|
| ECi/<br>ECiQ | PP1 | 240 | 0.00 | 0.000 | N/A | 0.000 | N/A | 0.000 | N/A | 0.001 | N/A | 0.001 | N/A |
| ECi/<br>ECiQ | PP2 | 240 | 0.92 | 0.018 | 2.0 | 0.024 | 2.6 | 0.034 | 3.7 | 0.030 | 3.3 | 0.054 | 5.9 |
#### Table 1. Within-Laboratory Precision Study Data Summary (for all 4 VITROS analyzers).
{7}------------------------------------------------
| | PP3 | 240 | 0.25 | 0.007 | N/A | 0.011 | N/A | 0.011 | N/A | 0.013 | N/A | 0.022 | N/A |
|---------|-----|-----|------|-------|-----|-------|-----|-------|-----|-------|------|-------|------|
| | PP4 | 240 | 0.94 | 0.027 | 2.9 | 0.032 | 3.4 | 0.033 | 3.5 | 0.091 | 9.7 | 0.106 | 11.3 |
| | PP5 | 240 | 3.25 | 0.086 | 2.6 | 0.062 | 1.9 | 0.084 | 2.6 | 0.169 | 5.2 | 0.216 | 6.6 |
| | PP6 | 240 | 7.62 | 0.063 | 0.8 | 0.076 | 1.0 | 0.112 | 1.5 | 0.208 | 2.7 | 0.257 | 3.4 |
| | PP1 | 240 | 0.00 | 0.000 | N/A | 0.000 | N/A | 0.001 | N/A | 0.002 | N/A | 0.002 | N/A |
| | PP2 | 240 | 1.18 | 0.024 | 2.0 | 0.025 | 2.1 | 0.021 | 1.8 | 0.240 | 20.3 | 0.243 | 20.6 |
| | PP3 | 240 | 0.32 | 0.007 | N/A | 0.008 | N/A | 0.006 | N/A | 0.078 | N/A | 0.079 | N/A |
| 3600 | PP4 | 240 | 1.18 | 0.028 | 2.4 | 0.034 | 2.9 | 0.033 | 2.8 | 0.376 | 31.9 | 0.380 | 32.2 |
| | PP5 | 240 | 3.94 | 0.083 | 2.1 | 0.092 | 2.3 | 0.072 | 1.8 | 1.064 | 27.0 | 1.073 | 27.2 |
| | PP6 | 240 | 8.56 | 0.099 | 1.2 | 0.099 | 1.2 | 0.105 | 1.2 | 1.610 | 18.8 | 1.619 | 18.9 |
| | PP1 | 240 | 0.00 | 0.000 | N/A | 0.000 | N/A | 0.000 | N/A | 0.001 | N/A | 0.001 | N/A |
| | PP2 | 240 | 1.03 | 0.020 | 1.9 | 0.027 | 2.6 | 0.027 | 2.6 | 0.058 | 5.6 | 0.072 | 7.0 |
| | PP3 | 240 | 0.28 | 0.006 | N/A | 0.010 | N/A | 0.007 | N/A | 0.025 | N/A | 0.028 | N/A |
| 5600 | PP4 | 240 | 1.02 | 0.026 | 2.5 | 0.036 | 3.5 | 0.026 | 2.5 | 0.140 | 13.7 | 0.149 | 14.6 |
| | PP5 | 240 | 3.43 | 0.073 | 2.1 | 0.074 | 2.2 | 0.095 | 2.8 | 0.288 | 8.4 | 0.321 | 9.4 |
| | PP6 | 240 | 7.51 | 0.097 | 1.3 | 0.105 | 1.4 | 0.119 | 1.6 | 0.276 | 3.7 | 0.333 | 4.4 |
| | PP1 | 240 | 0.00 | 0.000 | N/A | 0.000 | N/A | 0.000 | N/A | 0.001 | N/A | 0.001 | N/A |
| | PP2 | 240 | 1.21 | 0.043 | 3.6 | 0.042 | 3.5 | 0.016 | 1.3 | 0.153 | 12.6 | 0.165 | 13.6 |
| | PP3 | 240 | 0.32 | 0.009 | N/A | 0.009 | N/A | 0.010 | N/A | 0.052 | N/A | 0.055 | N/A |
| XT 7600 | PP4 | 240 | 1.19 | 0.034 | 2.9 | 0.042 | 3.5 | 0.037 | 3.1 | 0.239 | 20.1 | 0.248 | 20.8 |
| | PP5 | 240 | 3.94 | 0.078 | 2.0 | 0.093 | 2.4 | 0.088 | 2.2 | 0.604 | 15.3 | 0.622 | 15.8 |
| | PP6 | 240 | 8.46 | 0.113 | 1.3 | 0.117 | 1.4 | 0.098 | 1.2 | 0.896 | 10.6 | 0.916 | 10.8 |
N/A: Not Applicable, %CV is not meaningful for S/C results <0.5.
B. Reproducibility (Between-Laboratory Precision): The reproducibility of the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test was determined with EDTA plasma pools samples (GRP4, GRP5, and GRP6), quality controls (GRP1 and GRP2) and customer calibrator (GRP3) using the VITROS ECi/ECiQ and 3600 Immunodiagnostics Systems, and the VITROS 5600 and XT7600 Integrated Systems. Samples were measured in triplicate using 3 reagent lots, in 2 runs per day over 5 days at 3 sites, according to CLSI EP05-A3. Variance components were calculated. The VITROS Anti-SARS-CoV-2 IgG test total reproducibility %CV (from the "total" standard deviation) for the S/C values ranged from 6.1% - 22.8%, depending upon the sample and instrument. The following results were observed per analyzer:
Table 2. Reproducibility of VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test evaluated on VITROS ECi/ECiO Immunodiagnostic Systems
| Panel<br>Member<br>r | Number<br>of<br>Observ. | Grand<br>Mean<br>(S/C) | Repeatability<br>(Within Run) | | Between Run | | Between Day | | Between Lota | | Within-<br>Laboratoryb | | Between-Sitec | | Reproducibility<br>(Overall)d | |
|----------------------|-------------------------|------------------------|-------------------------------|-----------|-------------|-----------|-------------|-----------|--------------|-----------|------------------------|-----------|---------------|-----------|-------------------------------|-----------|
| | | | SD | CV<br>(%) | SD | CV<br>(%) | SD | CV<br>(%) | SD | CV<br>(%) | SD | CV<br>(%) | SD | CV<br>(%) | SD | CV<br>(%) |
| GRP1 | 270 | 0.01 | 0.002 | N/A+ | 0.000 | N/A+ | 0.000 | N/A+ | 0.003 | N/A+ | 0.003 | N/A+ | 0.003 | N/A+ | 0.006 | N/A+ |
| GRP2 | 270 | 0.98 | 0.058 | 5.9 | 0.000 | 0.0 | 0.021 | 2.2 | 0.000 | 0.0 | 0.062 | 6.3 | 0.018 | 1.8 | 0.095 | 9.6 |
| GRP3 | 270 | 0.28 | 0.016 | N/A+ | 0.007 | N/A+ | 0.010 | N/A+ | 0.021 | N/A+ | 0.029 | N/A+ | 0.000 | N/A+ | 0.034 | N/A+ |
| GRP4 | 270 | 1.08 | 0.059 | 5.5 | 0.011 | 1.0 | 0.035 | 3.2 | 0.206 | 19.2 | 0.217 | 20.2 | 0.000 | 0.0 | 0.225 | 21.0 |
| GRP5 | 270 | 3.59 | 0.164 | 4.6 | 0.040 | 1.1 | 0.071 | 2.0 | 0.479 | 13.3 | 0.513 | 14.3 | 0.024 | 0.7 | 0.545 | 15.2 |
| GRP6 | 270 | 7.84 | 0.220 | 2.8 | 0.000 | 0.0 | 0.091 | 1.2 | 0.425 | 5.4 | 0.487 | 6.2 | 0.311 | 4.0 | 0.665 | 8.5 |
11 Between lot: Variability of the test performance from lot to lot.
{8}------------------------------------------------
b Within-Laboratory variability contains the Within Run, Between Run, Between Day and Between Lot variance components.
Between site: Variability of the test performance from site to site.
d Reproducibility: Variability of the test incorporating factors of site, lot, run and day.
+ % CVs are not meaningful for S/C results < 0.50.
| PanelMember | Number of Observ. | Grand Mean (S/C) | Repeatability (Within Run) | | Between Run | | Between Day | | Between Lota | | Within-Laboratoryb | | Between-Sitec | | Reproducibility (Overall)d | |
|-------------|-------------------|------------------|----------------------------|--------|-------------|--------|-------------|--------|--------------|--------|--------------------|--------|---------------|--------|----------------------------|--------|
| | | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| GRP1 | 270 | 0.00 | 0.001 | N/A+ | 0.000 | N/A+ | 0.001 | N/A+ | 0.001 | N/A+ | 0.002 | N/A+ | 0.001 | N/A+ | 0.002 | N/A+ |
| GRP2 | 270 | 1.05 | 0.046 | 4.4 | 0.000 | 0.0 | 0.029 | 2.8 | 0.051 | 4.9 | 0.074 | 7.1 | 0.010 | 1.0 | 0.089 | 8.5 |
| GRP3 | 270 | 0.29 | 0.012 | N/A+ | 0.003 | N/A+ | 0.012 | N/A+ | 0.026 | N/A+ | 0.032 | N/A+ | 0.008 | N/A+ | 0.038 | N/A+ |
| GRP4 | 270 | 1.08 | 0.043 | 4.0 | 0.012 | 1.1 | 0.017 | 1.5 | 0.230 | 21.4 | 0.235 | 21.9 | 0.033 | 3.1 | 0.245 | 22.8 |
| GRP5 | 270 | 3.57 | 0.108 | 3.0 | 0.040 | 1.1 | 0.049 | 1.4 | 0.575 | 16.1 | 0.588 | 16.5 | 0.082 | 2.3 | 0.615 | 17.3 |
| GRP6 | 270 | 7.53 | 0.132 | 1.8 | 0.031 | 0.4 | 0.055 | 0.7 | 0.445 | 5.9 | 0.468 | 6.2 | 0.000 | 0.0 | 0.509 | 6.8 |
Table 3. Reproducibility of VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test evaluated on VITROS 3600 Immunodiagnostic Systems
a Between lot: Variability of the test performance from lot to lot.
b Within-Laboratory variability contains the Within Run, Between Run, Between Day and Between Lot variance components.
Between site: Variability of the test performance from site to site.
d Reproducibility: Variability of the test incorporating factors of site. lot, run and day.
+ %CVs are not meaningful for S/C results < 0.50
#### Table 4. Reproducibility of VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test evaluated on VITROS 5600 Integrated Systems
| PanelMember | Number<br>of<br>Observ. | Grand<br>Mean<br>(S/C) | Repeatability<br>(Within Run) | | Between Run | | Between Day | | Between Lota | | Within-<br>Laboratoryb | | Between-Sitec | | Reproducibility<br>(Overall)d | |
|-------------|-------------------------|------------------------|-------------------------------|-----------|-------------|-----------|-------------|-----------|--------------|-----------|------------------------|-----------|---------------|-----------|-------------------------------|-----------|
| | | | SD | CV<br>(%) | SD | CV<br>(%) | SD | CV<br>(%) | SD | CV<br>(%) | SD | CV<br>(%) | SD | CV<br>(%) | SD | CV<br>(%) |
| GRP1 | 270 | 0.00 | 0.000 | N/A+ | 0.000 | N/A+ | 0.001 | N/A+ | 0.001 | N/A+ | 0.001 | N/A+ | 0.002 | N/A+ | 0.002 | N/A+ |
| GRP2 | 270 | 1.05 | 0.085 | 8.1 | 0.000 | 0.0 | 0.000 | 0.0 | 0.045 | 4.3 | 0.097 | 9.2 | 0.000 | 0.0 | 0.119 | 11.3 |
| GRP3 | 270 | 0.29 | 0.015 | N/A+ | 0.002 | N/A+ | 0.004 | N/A+ | 0.020 | N/A+ | 0.025 | N/A+ | 0.002 | N/A+ | 0.031 | N/A+ |
| GRP4 | 270 | 1.08 | 0.050 | 4.6 | 0.009 | 0.8 | 0.027 | 2.5 | 0.204 | 18.9 | 0.212 | 19.7 | 0.000 | 0.0 | 0.221 | 20.5 |
| GRP5 | 270 | 3.58 | 0.103 | 2.9 | 0.017 | 0.5 | 0.099 | 2.8 | 0.483 | 13.5 | 0.504 | 14.1 | 0.000 | 0.0 | 0.538 | 15.1 |
| GRP6 | 270 | 7.55 | 0.137 | 1.8 | 0.006 | 0.1 | 0.161 | 2.1 | 0.251 | 3.3 | 0.328 | 4.3 | 0.000 | 0.0 | 0.484 | 6.4 |
al Between lot: Variability of the test performance from lot to lot.
b Within-Laboratory variability contains the Within Run, Between Run, Between Day and Between Lot variance components.
· Between site: Variability of the test performance from site to site.
d Reproducibility: Variability of the test incorporating factors of site, lot, run and day.
+ % CVs are not meaningful for S/C results < 0.50.
| Table 5. Reproducibility of VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG |
|----------------------------------------------------------------------------------|
| test evaluated on VITROS XT 7600 Integrated Systems |
| PanelMember | Number of Observ. | Grand Mean (S/C) | Repeatability (Within Run) | | Between Run | | Between Day | | Between Lota | | Within-Laboratoryb | | Between-Sitec | | Reproducibility (Overall)d | |
|-------------|-------------------|------------------|----------------------------|--------|-------------|--------|-------------|--------|--------------|--------|--------------------|--------|---------------|--------|----------------------------|--------|
| mber | Observ. | Mean (S/C) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| GRP1 | 270 | 0.00 | 0.001 | N/A+ | 0.000 | N/A+ | 0.001 | N/A+ | 0.001 | N/A+ | 0.001 | N/A+ | 0.001 | N/A+ | 0.002 | N/A+ |
| GRP2 | 270 | 1.06 | 0.035 | 3.4 | 0.000 | 0.0 | 0.010 | 0.9 | 0.039 | 3.7 | 0.053 | 5.1 | 0.009 | 0.9 | 0.061 | 5.7 |
| GRP3 | 270 | 0.29 | 0.012 | N/A+ | 0.000 | N/A+ | 0.007 | N/A+ | 0.023 | N/A+ | 0.027 | N/A+ | 0.003 | N/A+ | 0.029 | N/A+ |
| GRP4 | 270 | 1.09 | 0.037 | 3.4 | 0.000 | 0.0 | 0.013 | 1.2 | 0.220 | 20.2 | 0.224 | 20.6 | 0.013 | 1.1 | 0.229 | 21.1 |
| GRP5 | 270 | 3.59 | 0.097 | 2.7 | 0.000 | 0.0 | 0.000 | 0.0 | 0.510 | 14.2 | 0.519 | 14.5 | 0.018 | 0.5 | 0.537 | 15.0 |
| GRP6 | 270 | 7.54 | 0.118 | 1.6 | 0.000 | 0.0 | 0.032 | 0.4 | 0.380 | 5.0 | 0.399 | 5.3 | 0.000 | 0.0 | 0.457 | 6.1 |
a Between lot: Variability of the test performance from lot to lot
b Within-Laboratory variability contains the Within Run, Between Run, Between Lot variance components.
{9}------------------------------------------------
Between site: Variability of the test performance from site to site.
d Reproducibility: Variability of the test incorporating factors of site, lot, run and day.
+ % CVs are not meaningful for S/C results < 0.50.
- 2. Linearity:
Not applicable
- 3. Analytical Specificity:
- A. Cross-Reactivity: Cross-reactivity of the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test was evaluated by testing serum samples with antibodies to other microorganisms or underlying conditions which could cause false positive results. Cross-reactivity was evaluated in the VITROS 5600 Integrated system, with samples tested in singlicate. No cross-reactivity was observed with any of the cross-reactants evaluated. The results for cross-reactivity are presented in Table 6 below.
| Sample Category | Number of<br>Samples Tested | VITROS Immunodiagnostic<br>Products<br>Anti-SARS-CoV-2 IgG test<br>Results | |
|------------------------------------------------|-----------------------------|----------------------------------------------------------------------------|----------|
| | | Non-reactive | Reactive |
| Influenza A Antibody | 10 | 10 | 0 |
| Influenza B Antibody | 12 | 12 | 0 |
| Hepatitis C Virus (HCV) Antibody | 15 | 15 | 0 |
| Hepatitis B Virus (HBV) Antibody | 12 | 12 | 0 |
| Haemophilus influenzae Antibody | 12 | 12 | 0 |
| Anti-Nuclear Antibody (ANA) | 10 | 10 | 0 |
| Rheumatoid Factor | 11 | 11 | 0 |
| Human Anti-Mouse Antibody (HAMA) | 10 | 10 | 0 |
| Adenovirus Antibody | 12 | 12 | 0 |
| Coxsackie B Virus Antibody | 10 | 10 | 0 |
| Echovirus Antibody | 10 | 10 | 0 |
| Poliovirus Antibody | 9 | 9 | 0 |
| Respiratory Syncytial Virus (RSV)<br>Antibody | 10 | 10 | 0 |
| Anti-SARS-coronavirus | 13 | 13 | 0 |
| Anti-MERS-coronavirus | 14 | 14 | 0 |
| Human coronavirus HKU1 Antibody | 11 | 11 | 0 |
| Human coronavirus NL63 Antibody | 10 | 10 | 0 |
| Human coronavirus OC43 Antibody | 21 | 21 | 0 |
| Human coronavirus 229E Antibody | 24 | 24 | 0 |
| Human Immunodeficiency Virus (HIV)<br>Antibody | 17 | 17 | 0 |
| Human Parainfluenza Virus (HPIV)<br>Antibody | 5 | 5 | 0 |
| Human Metapneumovirus (HMPV)<br>Antibody | 10 | 10 | 0 |
#### Table 6. Cross-reactivity study samples and results
{10}------------------------------------------------
| Sample Category | Number of<br>Samples Tested | VITROS Immunodiagnostic<br>Products<br>Anti-SARS-CoV-2 IgG test<br>Results | |
|--------------------------------------------------------|-----------------------------|----------------------------------------------------------------------------|----------|
| | | Non-reactive | Reactive |
| Enterovirus Antibody | 10 | 10 | 0 |
| Rhinovirus Antibody | 11 | 11 | 0 |
| Epstein-Barr Virus (EBV) Antibody | 11 | 11 | 0 |
| Epstein-Barr Virus Nuclear Antigen<br>(EBVNA) Antibody | 11 | 11 | 0 |
| Rubella Virus Antibody | 10 | 10 | 0 |
| Legionella pneumophila Antibody | 10 | 10 | 0 |
| Bordetella pertussis Antibody | 15 | 15 | 0 |
| Mycoplasma pneumoniae Antibody | 14 | 14 | 0 |
| Chlamydophila pneumoniae Antibody | 10 | 10 | 0 |
| Streptococcus pneumoniae Antibody | 10 | 10 | 0 |
| Streptococcus pyogenes Antibody | 10 | 10 | 0 |
| Candida albincans Antibody | 14 | 14 | 0 |
| Pneumocystis jirovecii Antibody | 10 | 10 | 0 |
| Pseudomonas aeruginosa Antibody | 14 | 14 | 0 |
| Staphylococcus epidermidis | 10 | 10 | 0 |
| Cytomegalovirus (CMV) Antibody | 11 | 11 | 0 |
- B. Interference: The VITROS Immunodiagnostic Products Anti SARS-CoV-2 IgG test was evaluated for interference consistent with CLSI document EP07. Commonly encountered substances were tested on one lot of reagent using the VITROS 5600 Integrated System in 3 reactive (low positive) and 3 non-reactive (negative) samples.
(0) +
To evaluate interference, the percent bias was calculated as:
All substances evaluated have been shown to not interfere (<10% bias for low reactive samples and bias <0.22 S/C for non-reactive samples) with the test performance at the concentration listed in the table below.
Table 7. Endogenous interferants evaluated with VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test
| | Test Concentration | | | | | |
|-------------------------|--------------------|-----------------|--|--|--|--|
| Substance | Conventional | SI Units<br>N/A | | | | |
| Anti-Nuclear Antibody | > 8 AI | | | | | |
| Bilirubin, conjugated | 40 mg/dL | 475 umol/L | | | | |
| Bilirubin, unconjugated | 40 mg/dL | 684 umol/L | | | | |
| Cholesterol | 400 mg/dL | 10.3 mmol/L | | | | |
| Hemoglobin | 1000 mg/dL | 10 g/L | | | | |
| Human anti-Mouse | 3600 ng/mL | 0.024 umol/L | | | | |
| Ig (total) | 6 g/dL | 60 g/L | | | | |
{11}------------------------------------------------
| | Test Concentration | |
|-------------------|--------------------|--------------|
| Substance | Conventional | SI Units |
| Rheumatoid Factor | 35.7-61.7 IU/mL | N/A |
| Total Protein | 15 g/dL | 150 g/L |
| Triglycerides | 1500 mg/dL | 16.94 mmol/L |
N/A: Not Applicable (alternative units are not provided)
Table 8. Exogenous interferants evaluated with VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test
| Substance | Test Concentration | |
|------------------|--------------------|---------------|
| | Conventional | SI Units |
| Abacavir | 1.27 mg/dL | 44.4 µmol/L |
| Acetaminophen | 15.6 mg/dL | 1.03 mmol/L |
| Amoxicillin | 1.35 mg/dL | 37 µmol/L |
| Aspirin | 3 mg/dL | 0.167 mmol/L |
| Atorvastatin | 75 µg/dL | 1.34 µmol/L |
| Azithromycin | 1.1 mg/dL | 14.8 µmol/L |
| Biotin | 3510 ng/mL | 14.3 µmol/L |
| Cefoxitin | 660 mg/dL | 15.5 mmol/L |
| Ceftriaxone | 84 mg/dL | 1.51 mmol/L |
| Dextromethorphan | 1.56 µg/dL | 0.0575 µmol/L |
| EDTA | 0.099 mg/dL | 3.39 µmol/L |
| Gentamicin | 3.0 mg/dL | 62.8 µmol/L |
| Guaifenesin | 450 µg/dL | 22.7 µmol/L |
| Heparin | 330 units/dL | 330 units/dL |
| Ibuprofen | 21.9 mg/dL | 1.06 mmol/L |
| Intralipid | 2000 mg/dL | N/A |
| Levofloxacin | 3.6 mg/dL | 99.7 µmol/L |
| Levothyroxine | 429 µg/dL | 0.552 µmol/L |
| Lisinopril | 24.6 µg/dL | 0.607 µmol/L |
| Lopinavir | 57.17 µg/mL | 90.89 µmol/L |
| Loratadine | 8.7 µg/dL | 0.271 µmol/L |
| Losartan | 1155 ng/mL | 2.505 µmol/L |
| Meropenem | 33.9 mg/dL | 884 µmol/L |
| Metformin | 1.2 mg/dL | 92.9 µmol/L |
| Metoprolol | 150 µg/dL | 5.61 µmol/L |
| Naproxen | 36.0 mg/dL | 1.56 mmol/L |
| Omeprazole | 840 µg/dL | 24.3 µmol/L |
| Oseltamivir | 39.9 µg/dL | 1.28 µmol/L |
| Peramivir | 183600 ng/mL | 559 µmol/L |
| Prednisone | 10 µg/dL | 0.276 µmol/L |
| Ritonavir | 10.98 mg/dL | 126.42 mmol/L |
| Theophylline | 6.0 mg/dL | 333 µmol/L |
| Vancomycin | 12.0 mg/dL | 82.8 µmol/L |
| Zanamivir | 1089 ng/mL | 3.28 µmol/L |
N/A = Not Applicable (alternative units are not provided)
{12}------------------------------------------------
When amlodipine was tested at a concentration of 1.88 ug/dL, a negative bias (-11.1% change of the S/C value) was observed in reactive samples.
- 4. Assay Reportable Range:
Not applicable
- 5. Traceability. Stability. Expected Values (Controls, Calibrators, or Methods):
Specimen Stability: The stability of SARS-CoV-2 antibodies in serum. K2-EDTA plasma. and K -- EDTA plasma was evaluated with the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test after various storage conditions with and without freeze-thawing cycles in the VITROS 5600 Integrated Systems. The storage conditions evaluated were room temperature, 2 - 8°C, and <- 20°C with 5 freeze-thaw cycles.
Freshly drawn whole blood from individual donors was collected. Of the whole blood donors, (1) were unaltered and were spiked using a 15/41
The whole blood was distributed among the collection tubes used in the study. Each spiked sample was unique.
After centrifugation, each sample from each collection tube was tested in duplicate using one reagent lot on one VITROS 5600 Integrated System, denoted as the fresh timepoint. Aliquots of each sample were prepared and stored at room temperature, 2-8°C. and <- 20°C. For the < 20℃ samples were prepared. 1 subjected to 5 freeze-thaw cycles (F/T) and the other subjected to (0) F/F/T cycles.
For each sample type and storage temperature the percent difference to baseline was calculated as follows where baseline is considered as the fresh sample before storing at any temperature condition:
(D (4)
The results support the following specimen storage conditions for serum, K-- EDTA and K --EDTA plasma:
| Sample Type | Temperature | Stability | F/T |
|-----------------------------------------------|-------------------------------|------------|-----|
| Serum and plasma<br>(K2-EDTA and K3-<br>EDTA) | Room Temperature<br>(15-30°C) | ≤ 24 hours | N/A |
| | Refrigerated<br>(2-8°C) | ≤ 7 days | N/A |
| | Frozen<br>(≤-20°C) | ≤ 4 weeks | N/A |
Table 9. Summary of specimen stability
{13}------------------------------------------------
| | N/A | N/A | 1 cycle |
|------------------------------------------------------|-----|-----|---------|
| A THE CHART COLLECT CHE CASE CONSTITUTION CONSULTION | | | |
N/A= Not Applicable
Based upon this study design and the results thereof, the specimen stability data support, storage of all matrices at: 15-30 ℃ for 24 hours, 2-8℃ for up to 7 days, and ≤-20℃ or below for up to five freeze-thaw cycle
- 6. Detection Limit:
Not applicable
- 7. Assay Cut-Off:
The study was performed to determine the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test S/C cutoff. The study included the testing of a collection of [0] megative samples collected prior to the COVID-19 pandemic, and samples collected from individuals with a prior SARS-CoV-2 RT-PCR positive result. Additionally, promogative samples collected prior to the pandemic were tested with reagent lots. All samples were tested in singlicate using one VITROS Anti-SARS-CoV-2 IgG reagent pack and calibrator lot. A Receiver Operating Characteristic (ROC) curve analyses was performed to optimize for those cutoff values that maximize both sensitivity specificity. At the cutoff S/C = 1.00 the resultant ROC was [b] confirming high sensitivity and specificity of the established S/C cutoff value (Figure 2).
Image /page/12/Figure/7 description: The image shows a figure labeled as "Figure 2:". There is also the text "(D)(4)" in the upper right corner of the figure. The image appears to be a gray rectangle with a red border. The text "Analytical sensitivity at the cutoff using a Certified Reference Material" is at the bottom of the image.
Analytical sensitivity at the cutoff using a Certified Reference Material
The analytical sensitivity of the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test was determined using a series of serial dilutions of the WHO First International Standard for Anti-SARS-CoV-2 Immunoglobulin (human) code 20/136 (Certified Reference
{14}------------------------------------------------
Material or CRM) in negative patient sodium citrate plasma. Samples included in the study were prepared covering a will BAU/mL to | BAU/mL range (Table 10). Each sample was tested in duplicate using of the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test on two VITROS analyzers (VITROS ECi/ECiQ and VITROS 3600 Immunodiagnostic Systems). In addition, an internal reference calibrator was also included in the assay.
Ortho-Clinical Diagnostics collected both ALU and S/C for each of the CRM samples tested. Ortho-Clinical Diagnostics analyzed the data using the least-squares regression analysis, where the [14] represented the BAU/mL and the [10] the S/C values for each sample. Using the regression equation, Ortho-Clinical Diagnostics calculated the BAU/mL at the cutoff (S/C = 1.00). The data analysis shows that a S/C of 1.00 corresponds to 100 |BAU/mL using the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test.
Table 10: CRM serial dilutions S/C results and calculation of BAU/mL at the cutoff
| CRM<br>concentrations<br>(BAU/mL) | VITROS ECi/ECiQ<br>(S/C) | VITROS 3600<br>(S/C) | VITROS System<br>Average S/C |
|-----------------------------------|--------------------------|----------------------|------------------------------|
| | (0)(4) | | |
Figure 3: Analytical sensitivity regression analysis.
Image /page/13/Picture/5 description: The image is a gray rectangle with a thin red border. In the top center of the rectangle is the text "(b)(4)". The rectangle takes up most of the image. The image is simple and contains no other objects.
{15}------------------------------------------------
# B Comparison Studies:
- 8. Method Comparison:
Not Applicable
- 9. Matrix Comparison:
A matrix equivalency study was conducted to support equivalent performance of the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 1gG test between serum. K3-EDTA plasma, and K3-EDTA plasma. The study was conducted using 1 lot for the reagent pack, calibrator and controls in the VITROS 5600 Integrated Systems.
For this study unique paired clinical samples were evaluated (spiked samples prepared using a high titer convalescent plasma pool and non-spiked samples) with SARS-CoV-2 IgG antibodies levels representing the whole assay range (from low negative to high positive, including a 1046 of samples near the cutoff). After centrifugation, the matched serum, Ko-EDTA, and K3-EDTA samples were tested in duplicate using one reagent lot on 100 VITROS 5600 Integrated System. Sample distribution is summarized in Table 11 below.
Table 11. Sample distribution: negative, near the cutoff and positive
| Group | S/C Result | Number of<br>specimens (% total) |
|-------------------|------------|----------------------------------|
| Negative | | (b)(4) |
| Near cutoff | | |
| Moderate Positive | | |
| High Positive | | |
| Total | | (b)(4) |
A weighted Deming regression analysis comparing K2-EDTA plasma (the comparator matrix) to K2-EDTA plasma and to serum was conducted. The Deming regression analysis did not demonstrate significant deviation from the comparator matrix (Figure 4 - 5). Therefore, the study demonstrated equivalency between serum. K->EDTA plasma, and K-EDTA plasma.
Figure 4: Weighted Deming linear regression analysis between K2-EDTA plasma and serum
{16}------------------------------------------------
# C Clinical Studies:
- 10. Clinical Sensitivity:
Not applicable
# 11. Clinical Specificity:
Not applicable
# 12. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
{17}------------------------------------------------
### Clinical Agreement Study:
The clinical performance of VITROS Immunodiagnostic Products anti SARS- CoV-2 IgG test was evaluated at three testing sites using 642 unique retrospective clinical samples acquired from two populations. Population 1 consisted of 338 samples collected from individuals previously infected with SARS-CoV-2 with a prior SARS-CoV-2 positive test result using a comparator that FDA determined is appropriate (RT PCR test). Population 2 consisted of 304 samples collected prior to December 2019 (before the widespread outbreak of COVID-19). Of the 304 samples, 30% of the tested samples were collected from blood donor centers.
Of the 338 samples in Population 1, 24 samples were collected 0 - 7 days from COVID-19 symptom onset, 21 samples collected between 8 - 14 days from symptom onset, and 293 samples were collected >15 days from COVID-19 symptom onset. Table 12 below shows the sample distribution and the respective percentages from the total of samples tested per "Days post-symptoms onset" time bin.
| Time bin<br>(n= total number<br>samples tested) | Days post-<br>symptom<br>onset | Percent Samples<br>(per time bin) |
|-------------------------------------------------|--------------------------------|-----------------------------------|
| 0-7 days (n = 24) | 0 - 4 | 70.83% |
| | 5 - 7 | 29.17% |
| 8-14 days (n = 21) | 8 - 11 | 47.62% |
| | 12 - 14 | 52.38% |
| ≥ 15 days (n = 293) | 15 - 21 | 10.58% |
| | 22 - 30 | 10.92% |
| | 31 - 60 | 40.61% |
| | 61 - 90 | 26.62% |
| | 91 - 195 | 11.26% |
Table 12. Sample distribution within each "Days post symptoms onset" time bin.
Tables 13 and 14 below represents sample distribution per matrix tested for each study population (Population 1 and Population 2).
| Table 13. Distribution of Population I samples by matrix and days post-symptom onset. | | | | |
|---------------------------------------------------------------------------------------|-------------------------|-------------|----------|-------|
| Matrix | Days post-symptom onset | | | |
| | ≤7 days | 8 - 10 days | ≥15 days | Total |
| EDTA plasma | | (b)(4) | | 169 |
| Serum | | | | 169 |
| Total | | | | 338 |
Table 14. Distribution of Population 2 samples by matrix.
{18}------------------------------------------------
| EDTA plasma | 138 |
|-------------|-----|
| Serum | 166 |
| Total | 304 |
Ortho conducted the clinical agreement study at 3 testing sites: one internal site and 2 external sites.
# Positive Percent Agreement (PPA)
A total of 338 samples collected from individual patients confirmed to have a prior SARS-CoV-2 positive result by RT-PCR were tested. Blood samples were collected within the United States between April 2020 and March 2021. Samples were tested with the VITROS Immunodiagnostic Products Anti SARS-CoV-2 IgG test in each analyzer (except with the VITROS XT7600 Integrated System where one sample was not tested due to limited volume). Of the 338 samples included in the study, 169 were EDTA plasma samples and 169 were serum samples.
The performance of the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test and 95% Confidence Interval for each VITROS analyzer is summarized in the tables below.
PPA performance of the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test and 95% Confidence Interval by system:
| Days from<br>Symptom<br>Onset | Number of<br>Subjects Tested<br>(with prior RT-<br>PCR Positive) | Reactive | PPA | 95% CI<br>(Wilson score) |
|-------------------------------|------------------------------------------------------------------|----------|--------|--------------------------|
| 0-7 days | 24 | 11 | 45.83% | 27.89%-64.93% |
| 8-14 days | 21 | 11 | 52.38% | 32.37% - 71.66% |
| ≥15 days | 293 | 275 | 93.86% | 90.50%-96.10% |
| Total | 338 | -- | -- | -- |
Table 15. In the VITROS ECi/ECiO Immunodiagnostic Systems.
Table 16. In the VITROS 3600 Immunodiagnostic Systems and the VITROS 5600 Integrated Systems
| | | VITROS Immunodiagnostic Products<br>Anti-SARS-CoV-2 IgG test Results | | |
|----------------------------|------------------------------------------------------------------|----------------------------------------------------------------------|--------|--------------------------|
| Days from<br>Symptom Onset | Number of<br>Subjects Tested<br>(with prior RT-<br>PCR Positive) | Reactive | PPA | 95% CI<br>(Wilson score) |
| 0-7 days | 24 | 10 | 41.67% | 24.47% - 61.17% |
| 8-14 days | 21 | 11 | 52.38% | 32.37% - 71.66% |
| ≥15 days | 293 | 274 | 93.52% | 90.10% - 95.81% |
| Total | 338 | -- | -- | -- |
{19}------------------------------------------------
| Days from<br>Symptom<br>Onset | Number of<br>Subjects Tested<br>(with prior RT-<br>PCR Positive) | Reactive | PPA | 95% CI<br>(Wilson score) |
|-------------------------------|------------------------------------------------------------------|----------|--------|--------------------------|
| 0-7 days | 24 | 10 | 41.67% | 24.47% - 61.17% |
| 8-14 days | 21 | 11 | 52.38% | 32.37% - 71.66% |
| ≥15 days | 292 | 273 | 93.49% | 90.10% - 95.80% |
| Total | 337 | -- | -- | -- |
Table 17. In the VITROS XT 7600 Integrated Systems
# Negative Percent Agreement (NPA)
Three hundred and four (304) presumed SARS-CoV-2 negative samples collected prior to the COVID-19 pandemic within the United States were tested. Of the 304 samples, 138 were EDTA plasma samples and 166 serum samples. All samples were tested using VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems "(except the samples that were not tested on one or more analyzers due to limited volume). The performance of the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test and 95% Confidence Interval for each VITROS analyzer is summarized in the table below:
Table 18: NPA performance of the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test and 95% confidence interval in all VITROS analyzers
| | | VITROS Immunodiagnostic Products<br>Anti-SARS-CoV-2 IgG test Results | | |
|--------------------------------------------|--------------------------------------------|----------------------------------------------------------------------|--------|--------------------------|
| Analyzer | Presumed Negative<br>(Collected Pre-COVID) | Non-<br>Reactive | NPA | 95% CI<br>(Wilson score) |
| VITROS ECi/ECiQ/VITROS<br>3600/VITROS 5600 | 304 | 301 | 99.01% | 97.14%-<br>99.66% |
| VITROS XT 7600 | 303 | 300 | 99.01% | 97.13% -<br>99.66% |
# D Clinical Cut-Off:
Not applicable
# E Expected Values/Reference Range:
Not applicable
### F Other Supportive Performance Characteristics Data:
{20}------------------------------------------------
# Calibration Cvcle Stability
The purpose of this study was to establish the calibration interval, or how frequently the assay should be calibrated. The calibration interval was established by testing samples on the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test using a single calibration over time.
In the study, Ortho-Clinical Diagnostics included a panel of 6 precision pool samples. Each panel member was prepared as follows:
- Precision Pool 1 (PP1): Control level 1 (Non-Reactive) .
- . Precision Pool 2 (PP2): Control level 2 (Reactive)
- Precision Pool 3 (PP3): Calibrator .
- Precision Pool 4 (PP4): High negative EDTA plasma sample .
- . Precision Pool 5 (PP5): Low Positive EDTA plasma sample
- Precision Pool 6 (PP6): High Positive EDTA plasma sample .
After a single initial calibration on Day samples were tested on Day and up to Day on the VITROS ECi/ECiO, VITROS 3600, VITROS 5600, and VITROS XT7600.
For each positive sample (PP2, PP4, PP5, and PP6) linear regression analysis was conducted. In addition, percent difference from baseline was calculated as follows:
| (b)(4) |
|--------|
|--------|
Percent difference to baseline should be (b)(4).The percent difference to baseline should be
The study results support a calibration interval stability of 28 days for the VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG test.
#### VII Proposed Labeling:
The labeling supports the decision to grant the De Novo request for this device.
#### VIII Identified Risks and Mitigations:
| Risks to Health | Mitigation Measures |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Risk of false test results | Certain labeling information including<br>limitations, device descriptions, explanations<br>of procedures and performance information<br>identified in special controls (1), (3), and (5).<br>Use of certain specimen collection devices<br>identified in special control (2).<br>Certain design verification and validation<br>including documentation of device<br>descriptions, certain analytical studies and<br>clinical studies, and risk analysis strategies<br>identified in special control (4).<br>Testing of characterized samples and labeling<br>information identified in special control (6). |
10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov DEN210038 -
{21}------------------------------------------------
| Risks to Health | Mitigation Measures |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Failure to correctly interpret the test results | Certain labeling information including<br>limitations, device descriptions, explanations<br>of procedures and performance information<br>identified in special controls (1), (3), and (5).<br>Use of certain specimen collection devices<br>identified in special control (2).<br>Certain design verification and validation<br>including documentation of device<br>descriptions, certain analytical studies and<br>clinical studies, and risk analysis strategies<br>identified in special control (4).<br>Testing of characterized samples and labeling<br>information identified in special control (6). |
| Failure to correctly operate the device | Certain labeling information including<br>limitations, device descriptions, explanations<br>of procedures and performance information<br>identified in special controls (1), (3), and (5).<br>Use of certain specimen collection devices<br>identified in special control (2). |
#### IX Benefit/Risk Assessment:
### A Summary of the Assessment of Benefit:
The benefit of the assay is the ability to detect Anti-SARS-CoV-2 IgG antibodies as an aid in identifying individuals with an adaptive immune response to SARS-CoV-2, indicating recent or prior infection. The device could also provide a tool for tracking possible patient exposure.
True positive test results provide additional support for the diagnosis of recent or past SARS-CoV-2 infection. Results could be used in conjunction with clinical and epidemiological information, as well as other laboratory results to guide patient management. The test results may improve infection control measures and may aid in tracking and reducing transmission of infection. There is currently no SARS-CoV-2 antibody test that has undergone full FDA premarket review to most definitively determine clinical truth of the presence or absence of detectable antibodies for the method comparison study, however this uncertainty could be acceptable, particularly because the sponsor used a comparator that FDA determined is appropriate (SARS-CoV-2 RT-PCR devices), which represents the most reasonable alternative to establish clinical truth in the clinical study. This is an acceptable source of uncertainty regarding the benefits of the test.
#### Summary of the Assessment of Risk: B
The risks associated with the device, when used as intended, are those related to the risk of false test results, which have essentially the same impacts as the risks related to failure to correctly
{22}------------------------------------------------
interpret the test results and failure to correctly operate the device as all would cause the user to rely on incorrect information.
A false negative result could be interpreted as indicating that a person did not recently have COVID-19, which may lead a person to take fewer necessary precautions against spreading the virus to others if they are still shedding the virus from a recent infection. This may increase the risk of transmission. In the context of the current public health emergency, incorrect serological test results used to guide infection control activities could lead to misallocation of resources used for surveillance and prevention. The positive percent agreement performance point estimate of the device observed in the clinical study indicates that false negative results are not likely to occur when the device is used in the intended use population.
A false positive SARS-CoV-2 antibody result could be interpreted as a diagnosis of recent COVID-19, and a clinician may assume a patient may still be shedding the virus, which may result in unnecessary additional testing, quarantine, or self-isolation to prevent the spread of the virus to others. False positive serology test results can lead to an incorrect assessment that the tested person had an immune response to SARS-CoV-2, which may lead the person to take fewer necessary precautions against virus exposure. This may increase the individual's risk of infection and may lead the person to not seek testing if later infected, likely increasing the spread of the disease. In the context of the current public health emergency, incorrect serological test results used to guide infection control activities could lead to misallocation of resources used for surveillance and prevention.
A positive result could be wrongly interpreted as a diagnosis of acute COVID-19 to explain an individual's symptoms and delay correct diagnosis and initiation of appropriate treatment for the actual cause of patient illness. A positive test result could be wrongly interpreted as indicating that the tested person has immunity to SARS-CoV-2, which may lead the person to take fewer precautions against virus exposure. This may increase the individual's risk of infection and may lead the person to not seek testing if later infected, likely increasing the spread of the disease. A negative result may be misinterpreted as ruling out SARS-CoV-2 infection, with a concomitant delay in the correct diagnosis and treatment.
# C Patient Perspectives:
This submission did not include specific information on patient perspectives for this device.
# D Summary of the Assessment of Benefit-Risk:
The risks associated with the device (risk of false test results, failure to correctly interpret the results, and failure to correctly operate the device) are mitigated by labeling information, which will assist the operator in correctly performing the test and will assist healthcare providers in understanding the intended use of the test and evaluating the predictive value of a result based on the analytical and clinical performance of the test. In addition, those risks are mitigated by the use of certain validated specimen collection devices. Further, the risk of false test results and failure to correctly interpret the results are mitigated by certain design verification and validation, including analytical and clinical studies and risk analysis strategies to reduce the likelihood of such errors. Finally, the risk of false test results due to a disease or disorder that Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov DEN210038 -
{23}------------------------------------------------
presents a public health emergency, or a public health emergency that otherwise exists are addressed by special controls requiring certain testing of characterized samples and labeling information in those situations. The special controls help to ensure that errors will be uncommon and will facilitate accurate assay implementation of results. In addition, the device's performance observed in the clinical study suggests that errors will be uncommon and that the assay will provide benefits to patients as an aid in identifying individuals with an adaptive immune response to SARS-CoV-2, indicating recent or prior infection. While general controls alone are insufficient to mitigate the risks associated with the device, the benefits outweigh the risks given the special controls.
#### Conclusion: X
The De Novo request is granted, and the device is classified under the following regulation and subject to the special controls identified in the letter granting the De Novo request:
| Product Code(s): | QVP |
|------------------|--------------------------|
| Device Type: | SARS-CoV-2 serology test |
| Class: | Class II |
| Regulation: | 21 CFR 866.3983 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.