Retrospective analysis of clinical specimens tested at the Sponsor's laboratory to evaluate the accuracy of the Invitae Common Hereditary Cancers Panel by comparing results with validated orthogonal methods.
Accuracy Study with Clinical Specimens; Retrospective clinical accuracy study; Study Period: Pre-specified time period
Patients diagnosed with cancer and individuals tested for predisposition assessment; Sample Size: 6,014 samples (SNVs/Indels); 3,648 samples (CNVs); Number of Sites: 1 (Invitae Corporation)
Validated high-throughput sequencing platform (for SNVs/Indels); validated multiplexed PCR based test or validated microarray (for CNVs)
Technical Positive Predictive Value (TPPV) and Technical Negative Predictive Value (TNPV)
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Germline mutation profiling (SNVs)
Custom-built bioinformatics pipeline for variant calling
TPPV > 99%
TPPV 99.9% (95% CI 99.7->99.9%)
—
—
Clinical specimens study: 6014 samples
—
Germline mutation profiling (Indels)
Custom-built bioinformatics pipeline for variant calling
TPPV > 99%
TPPV 100% (95% CI 99.9-100%)
—
—
Clinical specimens study: 6014 samples
—
Germline mutation profiling (CNVs)
Custom-built bioinformatics pipeline for variant calling
TPPV > 99%
TPPV 99.5% (95% CI 99.2-99.7%)
—
—
Clinical specimens study: 3648 samples
—
Indications for Use
The Invitae Common Hereditary Cancers Panel is a qualitative high-throughput sequencing based in vitro diagnostic test system intended for analysis of germline human genomic DNA extracted from whole blood for detection of substitutions, small insertion and deletion alterations and copy number variants (CNV) in a panel of targeted genes. This test system is intended to provide information for use by qualified health care professionals, in accordance with professional guidelines, for hereditary cancer predisposition assessment and to aid in identifying hereditary genetic variants potentially associated with a diagnosed cancer. The test is not intended for cancer screening or prenatal testing. Results are intended to be interpreted within the context of additional laboratory results, family history, and clinical findings. The test is a single-site assay performed at Invitae Corporation.
Device Story
High-throughput DNA sequencing system; analyzes germline human genomic DNA from whole blood. Detects substitutions, small insertions/deletions, and copy number variants (CNV) in targeted cancer-related genes. Performed as single-site assay at Invitae Corporation. Workflow includes DNA extraction, library preparation, sequencing, and bioinformatic analysis. Outputs identify genetic variants; interpreted by qualified healthcare professionals alongside clinical findings and family history. Aids in hereditary cancer predisposition assessment and clinical decision-making for patients with diagnosed cancer. Benefits include identifying potential genetic associations to guide prophylactic screening, confirmatory procedures, or treatment.
Clinical Evidence
No clinical data provided in the summary. Device validation requires analytical accuracy and precision studies using clinical specimens or well-characterized cell lines. Accuracy must be evaluated against bidirectional Sanger sequencing or orthogonal methods. Performance requirements include PPA/NPA ≥99.0% for substitutions/indels/CNVs (positive) and ≥99.9% for negative agreement. Precision studies must assess multiple instruments, operators, days, and reagent lots with overall point estimates >99.0%.
Technological Characteristics
Targeted NGS panel (47 genes). Hybridization-based capture. Illumina NovaSeq 6000 sequencing-by-synthesis. Automated library prep. Unique molecular tags (TAFTs) for sample tracking. Bioinformatics pipeline for alignment, variant calling, and annotation. GRCh37 reference genome. Software-based variant classification per ACMG/AMP guidelines.
Indications for Use
Indicated for hereditary cancer predisposition assessment and identifying hereditary genetic variants associated with diagnosed cancer in patients. Not for cancer screening, prenatal testing, or as a stand-alone diagnostic. Contraindicated for patients with conditions like bone marrow transplant.
Regulatory Classification
Identification
A high throughput DNA sequencing for hereditary cancer predisposition assessment test system is a qualitative in vitro diagnostic (IVD) system intended for analysis of human DNA extracted from human specimens to detect germline mutations in a panel of targeted cancer related genes. It is intended to aid in hereditary cancer predisposition assessment by qualified health care professionals in accordance with professional guidelines. The device is not intended for screening, prenatal testing or as a stand-alone diagnostic test. The device is for prescription use only.
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
# EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR Invitae Common Hereditary Cancers Panel DECISION SUMMARY
#### I Background Information:
- A De Novo Number
DEN210011
## B Applicant
Invitae Corporation
### C Proprietary and Established Names
Invitae Common Hereditary Cancers Panel
### D Regulatory Information
| Product<br>Code(s) | Classification | Regulation<br>Section | Panel |
|--------------------|----------------|-------------------------------------------------------------------------------------------------------------------------|-----------|
| QVU | Class II | 21 CFR 866.6095— High<br>throughput DNA<br>sequencing for hereditary<br>cancer predisposition<br>assessment test system | Pathology |
#### II Submission/Device Overview:
# A Purpose for Submission:
De Novo request for evaluation of automatic class III designation for the Invitae Common Hereditary Cancers Panel
### B Measurand:
Germline substitutions, small insertion and deletion alterations and copy number variants (CNV) in a panel of 47 targeted genes
# C Type of Test:
Next generation sequencing based cancer-related germline mutation profiling
{2}------------------------------------------------
#### Indications for Use: III
## A Indication(s) for Use:
The Invitae Common Hereditary Cancers Panel is a qualitative high-throughput sequencing based in vitro diagnostic test system intended for analysis of germline human genomic DNA extracted from whole blood for detection of substitutions, small insertion and deletion alterations and copy number variants (CNV) in a panel of targeted genes.
This test system is intended to provide information for use by qualified health care professionals, in accordance with professional guidelines, for hereditary cancer predisposition assessment and to aid in identifying hereditary genetic variants potentially associated with a diagnosed cancer.
The test is not intended for cancer screening or prenatal testing. Results are intended to be interpreted within the context of additional laboratory results, family history, and clinical findings.
The test is a single-site assay performed at Invitae Corporation.
### B Special Conditions for Use Statement(s):
For Prescription Use Only For in vitro diagnostic use
### C Special Instrument Requirements:
Illumina NovaSeq 6000 system (qualified by Invitae)
#### Device/System Characteristics: IV
#### A Device Description:
The Invitae Common Hereditary Cancers Panel uses hybridization-based capture, nextgeneration sequencing (NGS), and a custom-built bioinformatics pipeline to compare all positions in targeted regions of 47 genes to a reference sequence and identify variants, including single nucleotide variants (SNVs), insertions and deletions (Indels), and copy number variants (CNVs).
Sequence analysis covers clinically important regions of each gene, including coding exons and 10 to 20 base pairs of adjacent intronic sequence on either side of the coding exons in the transcript listed in Table 1. Genes of "high clinical significance" are defined as those for which the test result(s) may lead to prophylactic screening, confirmatory procedures, or treatment that may incur morbidity or mortality to the patient and are shown in bold text. In addition, the
{3}------------------------------------------------
analysis covers the select non-coding variants specifically defined in the table. Any variants that fall outside these regions are not analyzed.
| Gene* | SNV/Indel<br>Analysis | CNV<br>Analysis | Notes |
|---------|-----------------------|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APC | YES | YES | The 1B promoter region is covered by both SNV/Indel and CNV<br>analysis. The 1A promoter region is covered by CNV analysis.<br>SNV/Indel analysis for exon 5 is limited to cds +/-10 bp. |
| ATM | YES | YES | SNV/Indel analysis for exons 6, 24, 43 includes only cds +/- 10 bp. |
| AXIN2 | YES | YES | |
| BARDI | YES | YES | |
| BMPR1A | YES | YES | CNV analysis covers the promoter region. |
| BRCA1 | YES | YES | SNV/Indel analysis includes +/- 20 base pairs of adjacent intronic<br>sequence. |
| BRCA2 | YES | YES | SNV/Indel analysis includes +/- 20 base pairs of adjacent intronic<br>sequence. |
| BRIP1 | YES | YES | |
| CDH1 | YES | YES | |
| CDK4 | YES | YES | |
| CDKN2A | YES | YES | |
| CHEK2 | YES | YES | |
| CTNNA1 | YES | YES | |
| DICER1 | YES | YES | SNV/Indel analysis for exon 22 includes only cds +/- 10 bp. |
| EPCAM | NO | YES | SNV/Indel analysis is not offered for this gene (CNV analysis only). |
| GREMI | NO | YES | Promoter region duplication testing only. |
| HOXB13 | YES | YES | |
| KIT | YES | YES | |
| MENI | YES | YES | SNV/Indel analysis for exon 2 is limited to cds +/-10 bp. |
| MLH1 | YES | YES | CNV analysis covers the promoter region. SNV/Indel analysis for<br>exon 12 is limited to cds +/-10 bp. |
| MSH2 | YES | YES | Analysis includes the exon 1-7 inversion (Boland mutation).<br>SNV/Indel analysis for exons 2, 5 includes only cds +/- 10 bp. |
| MSH3 | YES | YES | SNV/Indel analysis of the repeat region of exon 1 (5:79950697-<br>79950765) is not offered |
| Gene* | SNV/Indel<br>Analysis | CNV<br>Analysis | Notes |
| MSH6 | YES | YES | SNV/Indel analysis for exons 7, 10 includes only cds +/- 10 bp. |
| MUTYH | YES | YES | |
| NBN | YES | YES | |
| NF1 | YES | YES | SNV/Indel analysis for exons 2, 7, 25, 41, 48 includes only cds +/-<br>10 bp. |
| NTHLI | YES | YES | |
| PALB2 | YES | YES | |
| PDGFRA | YES | YES | |
| PMS2 | YES | YES | SNV/Indel analysis for exon 7 includes only cds +/- 10 bp. |
| POLDI | YES | YES | SNV/Indel analysis for exon 22 includes only cds +/- 10 bp. |
| POLE | YES | YES | |
| PTEN | YES | NO | CNV analysis is not offered for this gene. SNV/Indel analysis for<br>exons 8 includes only cds +/- 10 bp. |
| RAD50 | YES | YES | |
| RAD51C | YES | YES | |
| RAD51D | YES | YES | |
| SDHA | YES | NO | CNV analysis is not offered for this gene. SNV/Indel analysis is not<br>offered for exon 14. SNV/Indel analysis for exons 6-8 includes only<br>cds +/- 10 bp. |
| SDHB | YES | YES | |
| SDHC | YES | NO | CNV analysis is not offered for this gene. SNV/Indel analysis for<br>exons 2, 6 includes only cds +/- 10 bp. |
| SDHD | YES | YES | |
| SMAD4 | YES | YES | |
| SMARCA4 | YES | YES | |
| STK11 | YES | YES | |
| TP53 | YES | YES | CNV analysis covers the promoter region. |
| TSC1 | YES | YES | SNV/Indel analysis for exon 21 includes only cds +/- 10 bp. |
| TSC2 | YES | YES | |
| VHL | YES | YES | |
Table 1. Variant Type Reporting per Gene
{4}------------------------------------------------
{5}------------------------------------------------
*Genes of high clinical significance are defined as those for which the test result(s) may lead to prophylactic screening, confirmatory procedures or treatment that may incur mortality to the patient and are shown in bold text.
Identified variants are assessed by clinical professionals using currently available literature and data from public genetic variant databases. Variants are assigned a score, calculated according to an algorithm that weights the available clinical evidence. Possible outcomes include the following, which are based on joint ACMG/AMP Committee guidelines:
- . Benign (not reported) - strong evidence of not being disease or risk causing
- . Likely benign (not reported) - some evidence of not being disease or risk causing
- . Likely pathogenic – some evidence in favor of causing disease or increasing risk
- . Pathogenic - strong evidence of causing disease or increasing disease risk
- Variant of Uncertain Significance insufficient information to classify in one of the other . four categories
Variants are reported using HGVS nomenclature and the human reference genome GRCh37.
# B Principle of Operation
## Materials and Equipment
The reagents, consumables, and instruments needed to perform the Invitae Common Hereditary Cancers Panel test are used exclusively at, and qualified by the Invitae Corporation clinical laboratory (1400 16th Street, San Francisco, CA 94103). Instrumentation includes integrated, automated systems that incorporate liquid handling instruments, incubators, thermal cvclers, sonicators, and centrifuges, as well as Illumina NovaSeq 6000 sequencers. Reagents include extraction reagents, custom-made molecular barcode (TAFT) plates, buffers, PCR Cleanup Mag Beads, ER/AT Mix, Ligase Mix, PCR Master Mix, custom baits and blockers, PCR primer mix, streptavidin beads, and sequencing reagent kits.
### Specimen Collection and Preparation
The test includes a specimen collection kit, which is sent to ordering physicians. The shipping kit contains the following components:
- Plastic whole blood collection tube with K2EDTA .
- Patient information card with IB code sticker
- Test requisition form
- Biohazard bag and absorbent pouch, for submission of specimen
- . Handling instructions for shipping the specimen
The healthcare provider is directed to collect a 3 mL whole blood specimen, label the tube in the space provided with the patient's full name and at least one additional unique identifier, then ship the specimen to the Sponsor's laboratory at room temperature. When the specimen is received by the Sponsor, the accessioning operators inspect it for the following acceptance criteria:
- Minimum testing volume of 1.5 mL ●
- No evidence of clotting ●
{6}------------------------------------------------
- No evidence of leaks ●
- Specimen collection date <90 days from receipt of the kit
- Paperwork matching tube label ●
Samples are processed as soon as they are received.
## DNA Extraction, Normalization and Shearing
Genomic DNA is extracted from whole blood specimens per protocol using automated extraction and extraction kit qualified by Invitae. Unique molecular tags (TAFTs, Tagged Amplicons For Tracking) are appended to a region of non-clinical significance and processed with the samples from the time of extraction throughout the wet lab process. This allows sample identification to be maintained when samples are processed multiplexed. Extracted DNA is quantified and normalized to 23ng/uL to create batches for assay processing. The DNA is then sonically sheared into consistently sized pieces. Following DNA shearing, the DNA goes through a series of small enzyme-mediated steps that repair the sheared ends of the DNA sample and attach specific oligonucleotides to the ends of the sample. These oligonucleotides are specific for primers that are used in subsequent steps to anneal to them. This includes end-repair, A-tailing, adapter ligation, and dual-indexing of each sample.
## DNA Processing and Library Preparation
The specific fragments that contain targeted regions for the assay (i.e., the exonic regions for all 47 genes that are sequenced in this test) can be isolated from the remaining gDNA in the sample. Library production is fully automated end-to-end with integrated robotic systems; the system can create approximately 1,600 libraries every 8 hours with no human intervention. Libraries are pooled by row to create 8 pools of 12 libraries each.
# Hybridization
Following library preparation, the barcoded fragments are mixed, and the relevant DNA segments annealed to biotinylated probes for magnetic bead capture. The probes hybridize to the patient DNA samples and these fragments are separated from the rest of the genomic DNA using a magnet that interacts with ferromagnetic beads attached to avidin bound to biotin, which enables all non-bound DNA to be washed away. The beads are manufactured by a third-party vendor per the Sponsor's specification. Both ends of the fragments are sequenced, the sequence of the links is deconvoluted using the molecular barcode to identify which sample each fragment was derived from, and the sequence is aligned to the reference.
### Sequencing
The system uses Illumina's NovaSeq, a high throughput sequencing system that employs Sequencing-by-Synthesis chemistry, to perform paired-end reads of 150 bases in length and dual-index barcode reads of 8-nucleotides on the adapters. Equimolar mixing of hybridization reactions forms sequencing pools that allow each library to be sequenced with a minimum of
{7}------------------------------------------------
4.000.000 clusters on the Illumina flow cell. The system's average read depth (i.e., coverage) is 450x; any region with coverage below 50x is flagged as "low" and manually reviewed in the bioinformatics pipeline. Any region below 20x coverage is flagged as "very low," and the sample is failed and either re-run or a new sample is requested.
# Sequence Alignment/Mapping and Variant Calling
Following sequencing, raw BCL files are demultiplexed, producing individual fastg files for each sample, then mapped to the reference sequence, producing BAM files. The Invitae bioinformatics software then identifies "Active Regions", i.e., regions that may contain a putative variant, and different variant types are then called by the software.
# Variant Interpretation and Review
Variant calls are annotated based on evidence from published literature, public databases, prediction programs, and an internal curated variants database. Regions with low read depth, split reads, or low CNV quality scores are flagged for manual review. All other variants are processed through a combination of automated processes and manual review. For all variants that have been seen previously, the most recent interpretation and associated evidence is automatically placed in the report. For novel variants, variants that have not been seen recently, and variants that have new clinical evidence available, the bioinformatics software has tools to aid the genetics professional in variant interpretation/classification. Variants are assigned one of the following classifications, which are based on joint ACMG/AMP Committee guidelines:
- Pathogenic (Odds >99:1 in favor of causing disease or increasing disease risk) ●
- Likely Pathogenic (Odds >9:1 but <99:1 in favor of causing disease or increasing ● risk)
- Variant of Uncertain (Unknown) Significance (VUS) - Insufficient information to classify in one of the other four categories
- Likely Benign (Odds >9:1 but <99:1 in favor of not being disease or risk causing) ●
- Benign (Odds >99:1 in favor of not being disease or risk causing) ●
Refer to Section VI. C for more information on the interpretation and curation process.
# Controls
- a) No Template Control (NTC): A NTC is processed as a negative control through the DNA extraction process and post-extraction DNA quantification process. NTC is checked for position and post-extraction DNA quantification, to ensure the plate is being processed in the correct orientation and that the plate is not contaminated. The NTC is not included in the test steps post DNA quantification.
- b) Positive control: A reference cell line sample is processed as a positive control from DNA extraction through sequencing. The positive control is checked for quality metrics such as library concentration, sequence coverage, and gap rates. Failure of the positive control to meet the pre-defined quality metrics will result in a plate failure.
{8}------------------------------------------------
## Result Reporting
The Invitae Common Hereditary Cancers Panel test results are for professional use only; the final clinical reports generated from the assay summarize the clinical findings for the ordering physician. Test reports are generated and reviewed by PhD level scientists, genetic counselors, as well as licensed, board-certified clinical molecular geneticists or licensed, board-certified molecular pathologists before signing out.
# C Instrument Description Information
- 1. Instrument Name:
Illumina NovaSeq 6000 system (qualified by Invitae)
- 2. Specimen Identification:
Whole blood
## 3. Specimen Sampling and Handling:
Refer to Section IV. B, Specimen Collection and Preparation
- 4. Calibration:
Not applicable
- 5. Quality Control:
Quality metrics are evaluated throughout the bench workflow, including library preparation, hybridization, and sequencing, to verify that samples have appropriate DNA concentrations at key points and that the sequencing run generates reads with adequate quality and depth. Failure to meet the prespecified quality thresholds results in manual review and possible resequencing. Post-sequencing quality metrics are assessed at the sample and batch level to determine sample and batch quality (Table 2). Samples are held for review whenever the value for a metric exceeds the thresholds in the QC Service. Batch quality is determined by reviewing the number of samples that failed for an individual given metric and a batch is held when the number of individual failures exceeds a predetermined threshold for number of failures. Samples and/or batches with holds are reviewed by trained professionals for quality according to Standard Operating Procedure (SOP) documents.
{9}------------------------------------------------
| Metric | Description | Sample<br>Threshold | Batch<br>Threshold<br>(# failed<br>samples) |
|----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|---------------------------------------------|
| Mean target coverage | The mean depth of coverage (i.e., number of reads)<br>over a target region | ≥300x | ≤21 |
| Very low coverage | Number of targeted bases (assay-wide) with <20x<br>depth of coverage | ≤400 | ≤42 |
| Percent Selected Bases | The fraction of bases in aligned reads that map<br>directly to or within a fixed interval containing the<br>baited region. | ≥68% | ≤21 |
| Pass Filter and High Quality<br>aligned read base calling error rate | The fraction of bases that mismatch the genomic<br>reference sequence in reads that pass quality filters<br>and align with mapping quality Phred score of 20<br>or higher. | ≤0.01 | ≤21 |
| AT dropout (bias) | A measure of how undercovered regions with<br><50% GC content are relative to the mean. A<br>higher score implies that a higher percentage of<br>total reads that should have mapped to regions<br>with ≤50% GC content mapped elsewhere. | ≤6% | ≤10 |
| GC dropout (bias) | A measure of how undercovered regions with<br>≥50% GC content are relative to the mean. A<br>higher score implies that a higher percentage of<br>total reads that should have mapped to regions<br>with ≥50% content mapped elsewhere | ≤15% | ≤21 |
| Median insert size | The median size of sections of DNA being<br>sequenced | 200 to 320 bp | ≤21 |
| Percent duplication | Percent of paired and unpaired reads marked as<br>duplicates | ≤0.2% | ≤21 |
| Contamination estimate | A maximum likelihood estimate of the percentage<br>of the DNA in a sample that is from a foreign<br>source. | *** | *** |
| Consensus variant rate | The fraction of processed loci that are classified as<br>variant loci. | 0.0006 to<br>0.0029 | ≤21 |
| Consensus<br>heterozygous/homozygous ratio | The ratio of heterozygous to homozygous variant<br>loci. | 1 to 5.8 | ≤21 |
| Consensus heterozygous rate | The fraction of processed loci that are classified as<br>heterozygous variant loci. | ≤0.0015 | ≤21 |
| Sample Goodness-of-fit to a<br>Baseline sample set | A per-sample measure of how well the observed<br>data at each CNV target are described by the mean<br>model of predicted sample read counts as a | *** | *** |
| Metric | Description | Sample<br>Threshold | Batch<br>Threshold<br>(# failed<br>samples) |
| | function of the read counts for a set of baseline<br>samples. | | |
| CNVitae Q-score <35 | Called copy-number error probability p-value<br>represented as a Phred-scaled quality score. | *** | *** |
| TAFT (molecular barcode) -<br>minimum expected counts | The number of reads that align to the molecular<br>barcode sequence | *** | *** |
| TAFT - proportion of classified | The proportion of reads that align to the correct<br>molecular barcode sequence | *** | *** |
| CNVitae_XY_<br>ploidy_match | Whether the X and Y chromosome count observed<br>through sequencing match the biological sex<br>reported by the provider or patient. | *** | *** |
# Table 2. Post-Sequencing Sample and Batch QC Metrics
{10}------------------------------------------------
*** Pre-specified threshold values provided. Data not shown
In addition, based on per variant QC metrics, any variant whose data does not reflect a minimum level of confidence that the variant was correctly called is assigned a "FILTER" flag. The criteria for SNV and Indel variants and CNVs are detailed in Table 3 and Table 4 below, respectively.
| Table 3. FILTER Flag Criteria for SNVs/Indels | | | |
|-----------------------------------------------|--|--|--|
| Metric | Description | Acceptance<br>Criteria |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| QUAL (quality) score | Phred-scaled quality score measures the error<br>probability of the called variant. | < 30 |
| QD (quality by depth) score if variant<br>length is ≤3 bp | The QD is the QUAL score normalized by allele<br>depth (AD) for a variant. | < 2.0 |
| Allele balance if variant length is ≤7 bp | In a location where a variant has been detected, the<br>proportion of reads that indicate a variant | < 0.15 |
| Repeat Unit Wobble Filter | The detected variant is in a region with a high<br>amount of variability between reads with respect to<br>the number of short tandem repeats (STRs - short<br>repeated sequences of DNA) | True |
| Variant is on curated "block list" of known<br>artifacts | The detected variant has been repeatedly identified<br>in the past as being a result of the sequencing<br>process or something other than an actual variant. | True |
| Variant was removed following manual<br>review of data | Used in specific cases which are selected using<br>additional criteria | True |
{11}------------------------------------------------
| Metric | Description | Acceptance Criteria |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Excluded Target | Targets are excluded due to very low counts, as<br>well as inability to confidently call copy number<br>(at that target) in too many of the baseline<br>samples. Baseline samples are used to establish<br>what the expected profile of a sample with a<br>normal number of copies of a target region<br>looks like. | *** |
| BaselineCNV | High proportion of copy number variants were<br>called in the baseline OR there was a failure to<br>confidently call copy number in the baseline at<br>this target. | *** |
| CNVLowQual: Quality score | The quality score is derived from the likelihood<br>that the call is normal for copy number normal<br>events and at least part of the called segment is<br>a deletion/duplication for deletion/duplication<br>events. Failing values indicate that there<br>remains significant uncertainty in the call. | *** |
| Variant is on curated "block list" of<br>known artifacts | The detected variant has been repeatedly<br>identified in the past as being a result of the<br>sequencing process or something other than an<br>actual variant. | True |
| Variant was removed following<br>manual review of data | Used in specific cases which are selected using<br>additional criteria | True |
Table 4. FILTER Flag Criteria for CNVs
*** Pre-specified threshold values provided. Data not shown
Variants with a FILTER flag are later discarded. All other variants are assigned one of three flags (WARN, LIGHTWARN or CONFIDENT), which are later used to guide whether manual inspection may be needed.
#### V Standards/Guidance Documents Referenced:
- . Considerations for Design, Development, and Analytical Validation of Next Generation Sequencing (NGS)-Based In Vitro Diagnostics (IVDs) Intended to Aid in the Diagnosis of Suspected Germline Diseases
- Guidance for the Content of Premarket Submissions for Software Contained in Medical ● Devices
- General Principles of Software Validation .
- . Recommended Content Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices ●
- Statistical Guidance on Reporting Results from Studies ●
{12}------------------------------------------------
- CLSI standard EP05-A3 Evaluation of Precision of Quantitative Measurement ● Procedures (3rd ed)
- CLSI standard EP07 Interference Testing in Clinical Chemistry (3rd ed)
- CLSI standard EP37 Supplemental Tables for Interference Testing in Clinical Chemistry ●
#### VI Performance Characteristics:
# A Analytical Performance:
# 1. Precision/Reproducibility:
The precision of the Invitae Common Hereditary Cancers Panel was evaluated with two sets of clinical samples. The first set included 25 samples, with a total of 2,047 SNVs across 40 genes (including 11 genes of high clinical significance), 33 Insertions across 7 genes (including 2 genes of high clinical significance), 38 Deletions across 5 genes (including 1 gene of high clinical significance), 5 CNV duplications across 4 genes (including 1 gene of high clinical significance). and 16 CNV deletions across 12 genes (including 5 genes of high clinical significance). The second set included 18 samples enriched for Indels and CNVs, with a total of 19 Insertions across 4 genes (including 1 genes of high clinical significance), 12 Deletions across 7 genes (including 1 gene of high clinical significance), 7 CNV duplications across 6 genes (including 4 genes of high clinical significance) and 16 CNV deletions across 6 genes (including 5 genes of high clinical significance). Each sample was tested with 14 replicates across 3 different operators, 3 different sequencers, and 3 different sequencing reagent lots. The assay runs were performed on 3 different, non-consecutive days. The repeatability of the assay was assessed by analyzing the concordance between sample replicates within the same run. The reproducibility of the assay was assessed by analyzing the concordance between sample replicates across runs, overall and across sequencing reagent lots, sequencing operators, and sequencers.
Results of the precision study are summarized in Table 11. The overall PPA is 99.95% (95% CI 99.92-99.97%) for SNVs, 99.57% (95% CI 99.07-99.80%) for Indels and 99.67% (95% CI 98.80-99.91%) for CNVs. The overall NPA is >99.99% for all variant types (95% CI >99.99-100%) (Table 5).
| Type | # Total Positive<br>Variants* | # Detected<br>Positive<br>Variants** | # Total Negative<br>Variants* | # Detected<br>Negative<br>Variants** | PPA<br>(95% CI) | NPA<br>(95% CI) |
|------------|-------------------------------|--------------------------------------|-------------------------------|--------------------------------------|---------------------------|---------------------------|
| SNVs | 28633 | 28619 | 35645541 | 35645506 | 99.95%<br>(99.92-99.97%) | >99.99% (>99.99-<br>100%) |
| Indels | 1405 | 1399 | 14854187 | 14854180 | 99.57% (99.07-<br>99.80%) | >99.99% (>99.99-<br>100%) |
| Insertions | 728 | 728 | 3973866 | 3973866 | 100%<br>(99.48-100%) | 100%<br>(>99.99-100%) |
| Deletions | 677 | 671 | 10880321 | 10880314 | 99.11%<br>(98.08-99.59%) | >99.99% (>99.99-<br>100%) |
Table 5. Overall Precision by Variant Type
{13}------------------------------------------------
| CNVs | 602 | 600 | 467326 | 467323 | 99.67% (98.80-<br>99.91%) | >99.99% (>99.99-<br>100%) |
|---------------------|-----|-----|--------|--------|---------------------------|---------------------------|
| CNV<br>duplications | 157 | 157 | 467326 | 467326 | 100%<br>(97.61-100%) | 100%<br>(>99.99-100%) |
| CNV<br>deletions | 445 | 443 | 467326 | 467323 | 99.55%<br>(98.38-99.88%) | >99.99% (>99.99-<br>100%) |
** # Detected positive variants: the total number of positive/negative variants that were detected across replicates of the study samples.
The PPAs are >99% across different zygosity (for SNVs), variant sizes (for Indels and CNVs) or genomic contexts, except for deletions of 1-5bp, for which the PPA is 97.53% (95% CI 94.72-98.86%) (Table 6 and Table 7). This was due to 6 false negative results for the same variant on the SDHA gene, which were in a low mappability/complexity region.
| | Variant Type<br>Stratified by Zygosity and<br>Size | # Total<br>Positive<br>Variants* | PPA | 95% CI | # Total<br>Negative<br>Variants* | NPA | 95% CI |
|---------------------|----------------------------------------------------|----------------------------------|--------|--------------|----------------------------------|---------|-------------|
| SNVs | Homo | 10248 | 100% | 99.96-100% | N/A | N/A | N/A |
| | Hetero | 17181 | 99.92% | 99.86-99.95% | N/A | N/A | N/A |
| | Unknown<br>zygosity | 1204 | 100% | 99.68-100% | N/A | N/A | N/A |
| Insertions | 1-5 bp | 588 | 100% | 99.35-100% | 3525432 | 100% | >99.99-100% |
| | 6-10 bp | 112 | 100% | 96.68-100% | 315417 | 100% | >99.99-100% |
| | 11-20 bp | 14 | 100% | 78.47-100% | 131247 | 100% | >99.99-100% |
| | 21+ bp | 14 | 100% | 78.47-100% | 1770 | 100% | >99.99-100% |
| Deletions | 1-5 bp | 243 | 97.53% | 94.72-98.86% | 9706966 | >99.99% | >99.99-100% |
| | 6-10 bp | 196 | 100% | 98.08-100% | 813027 | 100% | >99.99-100% |
| | 11-20 bp | 196 | 100% | 98.08-100% | 359433 | 100% | >99.99-100% |
| | 21+ bp | 42 | 100% | 91.62-100% | 888 | 100% | >99.99-100% |
| CNV<br>Duplications | ≤ Single Exon | 40 | 100% | 91.24-100% | N/A | N/A | N/A |
| | 2-5 Exons | 70 | 100% | 94.80-100% | N/A | N/A | N/A |
| | 6-9 Exons | 12 | 100% | 75.75-100% | N/A | N/A | N/A |
| | 10+ Exons | 14 | 100% | 78.47-100% | N/A | N/A | N/A |
Table 6. Precision by Zygosity and Variant Size
{14}------------------------------------------------
| Variant Type<br>Stratified by Zygosity and<br>Size | | # Total<br>Positive<br>Variants* | PPA | 95% CI | # Total<br>Negative<br>Variants* | NPA | 95% CI |
|----------------------------------------------------|-----------------------------------------------------|----------------------------------|--------|--------------|----------------------------------|-----|--------|
| | Entire Coding<br>Sequence | 21 | 100% | 84.54-100% | N/A | N/A | N/A |
| | Other<br>(intronic, non-<br>coding,<br>combination) | 0 | N/A | N/A | N/A | N/A | N/A |
| CNV<br>Deletions | ≤ Single Exon | 235 | 99.15% | 96.95-99.77% | N/A | N/A | N/A |
| | 2-5 Exons | 116 | 100% | 96.79-100% | N/A | N/A | N/A |
| | 6-9 Exons | 61 | 100% | 94.08-100% | N/A | N/A | N/A |
| | 10+ Exons | 33 | 100% | 89.57-100% | N/A | N/A | N/A |
| | Entire Coding<br>Sequence | 0 | N/A | N/A | N/A | N/A | N/A |
| | Other<br>(intronic, non-<br>coding,<br>combination) | 0 | N/A | N/A | N/A | N/A | N/A |
** N/A: data not available or not calculated.
## Table 7. Precision by Genomic Context
| Variant Type<br>Stratified by Genomic<br>Context | | # Total Positive<br>Variants* | PPA | 95% CI of<br>PPA | # Total<br>Negative<br>Variants* | NPA | 95% CI of<br>NPA |
|--------------------------------------------------|---------------------|-------------------------------|-------|------------------|----------------------------------|------|------------------|
| High GC<br>Content<br>(>70%) | SNVs | 504 | 100% | 99.24-100% | 118395 | 100% | >99.99-100% |
| | Insertions | 0 | N/A** | N/A | 12297 | 100% | 99.97-100% |
| | Deletions | 182 | 100% | 97.93-100% | 12951 | 100% | 99.97-100% |
| | CNV<br>duplications | 56 | 100% | 93.58-100% | 25223 | 100% | 99.98-100% |
| | CNV<br>deletions | 126 | 100% | 97.04-100% | 25223 | 100% | 99.98-100% |
| Low GC Content<br>(<30%) | SNVs | 5320 | 100% | 99.93-100% | 7191969 | 100% | >99.99-100% |
| | Insertions | 14 | 100% | 78.47-100% | 946545 | 100% | >99.99-100% |
| | Deletions | 168 | 100% | 97.76-100% | 3087513 | 100% | >99.99-100% |
{15}------------------------------------------------
| Variant Type<br>Stratified by Genomic<br>Context | | # Total Positive<br>Variants* | PPA | 95% CI of<br>PPA | # Total<br>Negative<br>Variants* | NPA | 95% CI of<br>NPA |
|--------------------------------------------------|---------------------|-------------------------------|--------|------------------|----------------------------------|---------|------------------|
| | CNV<br>duplications | 76 | 100% | 95.19-100% | 129043 | 100% | >99.99-100% |
| | CNV<br>deletions | 209 | 100% | 98.17-100% | 129043 | 100% | >99.99-100% |
| Low<br>Mappability/<br>Low Complexity | SNVs | 9075 | 99.84% | 99.74-99.91% | 9738339 | >99.99% | >99.99-100% |
| | Insertions | 238 | 100% | 98.41-100% | 1163808 | 100% | >99.99-100% |
| | Deletions | 607 | 99.01% | 97.86-99.55% | 3817896 | >99.99% | >99.99-100% |
| | CNV<br>duplications | 43 | 100% | 91.80-100% | 28710 | 100% | 99.99-100% |
| | CNV<br>deletions | 104 | 100% | 96.44-100% | 28710 | 100% | 99.99-100% |
**N/A: data not available or not calculated.
#### Table 8. Precision by Sources of Variance and Variant Type
| Variable | Average Positive Agreement (95% CI) | | | | |
|-----------------------------------------------|-------------------------------------|-----------------------|--------------------------|--------------------------|--------------------------|
| | SNVs | Insertions | Deletions | CNV duplications | CNV deletions |
| Repeatability | 99.91%<br>(99.88-99.94%) | 100%<br>(>99.99-100%) | 98.08%<br>(97.48-99.52%) | 99.83%<br>(99.83-100%) | 99.43%<br>(98.29-99.82) |
| Reproducibility –<br>Instrument-to-Instrument | 99.93%<br>(99.91-99.96%) | 100%<br>(>99.99-100%) | 97.72%<br>(96.59-99.05%) | 96.69%<br>(94.17-98.80%) | 98.99%<br>(98.06-99.68%) |
| Reproducibility - Lot-to-Lot | 99.87%<br>(99.83%-99.90%) | 100%<br>(>99.99-100%) | 99.24%<br>(99.16-99.85%) | 98.37%<br>(95.14-99.16%) | >99.99%<br>99.13-100% |
| Reproducibility -<br>Operator-to-Operator | 99.89%<br>(99.85-99.92%) | 100%<br>(>99.99-100%) | 99.85%<br>(98.90-99.93%) | 98.72%<br>(96.82-100%) | 99.85%<br>(99.16-100%) |
| Reproducibility - Run-to-Run/Day-to-Day | 99.83%<br>(99.74-99.89%) | 100%<br>(>99.99-100%) | 95.17%<br>(90.55-96.49%) | 100%<br>(95.63-100%) | 100%<br>(>99.99-100%) |
### Table 9. No Call Rate
| No call rate* | Repeatability | Reproducibility<br>- Instrument-<br>to-Instrument | Reproducibility<br>- Lot-to-Lot | Reproducibility<br>- Operator-to-<br>Operator | Reproducibility -<br>Run-to-Run/Day-<br>to-Day |
|---------------|---------------|---------------------------------------------------|---------------------------------|-----------------------------------------------|------------------------------------------------|
| SNVs/Indels | 0.05% | 0.05% | 0.05% | 0.05% | 0.18% |
| CNVs | 1.78% | 1.99% | 1.99% | 1.95% | 1.82% |
*For SNVs and Indels, a "no call" is an area that could not be called due to a sample or region-specific limitation like an unfilled coverage gap. The no call (NC) rate for SNVs and Indels is calculated as the proportion of variants with a no call result in the total number of SNV/Indel called variants, assessed across all sample comparisons.
{16}------------------------------------------------
For CNVs, a "no call" is a region that could not be called due to a sample or region-specific limitation like an unfilled coverage gap. The NC rate for CNVs is calculated as the proportion of CNV target regions with a no call result in the total number of target regions, assessed across all sample comparisons.
| Specimen Stratified by Variant Type | # Total Positive Variants* | # Detected Positive Variants* | PPA | 95% CI of PPA | # Total Negative Variants* | # Detected Negative Variants* | NPA | 95% CI of NPA | |
|-------------------------------------------|----------------------------------|------------------------------------------|------------------------------------------|------------------|----------------------------------|------------------------------------------|------------------------------------------|------------------|------------------|
| Study 1 | | | | | | | | | |
| Specimen 1 SNVs | 1204 | 1204 | 100% | 99.68-100% | 776217 | 776217 | 100% | >99.99-100% | |
| Specimen 1 Indels | 56 | 56 | 100% | 93.58-100% | 197977 | 197977 | 100% | >99.99-100% | |
| Specimen 1 CNVs | 14 | 14 | 100% | 78.47-100% | 10874 | 10874 | 100% | 99.96-100% | |
| Specimen 2 SNVs | 1106 | 1106 | 100% | 99.65-100% | 776217 | 776217 | 100% | >99.99-100% | |
| Specimen 2 Indels | 42 | 42 | 100% | 91.62-100% | 198016 | 198016 | 100% | >99.99-100% | |
| Specimen 2 CNVs | 14 | 14 | 100% | 78.47-100% | 10878 | 10878 | 100% | 99.96-100% | |
| Specimen 3 SNVs | 1512 | 1512 | 100% | 99.75-100% | 776035 | 776035 | 100% | >99.99-100% | |
| Specimen 3 Indels | 56 | 56 | 100% | 93.58-100% | 198003 | 198003 | 100% | >99.99-100% | |
| Specimen 3 CNVs | 28 | 28 | 100% | 87.94-100% | 10892 | 10892 | 100% | 99.96-100% | |
| Specimen 4 SNVs | 1008 | 1008 | 100% | 99.62-100% | 776308 | 776308 | 100% | >99.99-100% | |
| Specimen 4 Indels | 28 | 28 | 100% | 87.94-100% | 198029 | 198029 | 100% | >99.99-100% | |
| Specimen 4 CNVs | 0 | 0 | N/A*** | N/A | 10888 | 10888 | 100% | 99.96-100% | |
| Specimen 5 SNVs | 882 | 882 | 100% | 99.57-100% | 776295 | 776295 | 100% | >99.99-100% | |
| Specimen 5 Indels | 70 | 70 | 100% | 94.80-100% | 198003 | 198003 | 100% | >99.99-100% | |
| Specimen 5 CNVs | 21 | 21 | 100% | 84.54-100% | 10810 | 10808 | 99.98% | 99.93-99.99% | |
| Specimen 6 SNVs | 997 | 994 | 99.70% | 99.12-99.90% | 776228 | 776222 | >99.99% | >99.99-100% | |
| Specimen<br>Stratified by<br>Variant Type | | # Total<br>Positive<br>Variants* | # Detected<br>Positive<br>Variants*<br>* | PPA | 95% CI<br>of PPA | # Total<br>Negative<br>Variants* | # Detected<br>Negative<br>Variants*<br>* | NPA | 95% CI<br>of NPA |
| | Indels | 30 | 28 | 93.33% | 78.68-98.15% | 197972 | 197969 | >99.99% | >99.99-100% |
| | CNVs | 22 | 22 | 100% | 85.13-100% | 10805 | 10805 | 100% | 99.96-100% |
| | SNVs | 1190 | 1190 | 100% | 99.68-100% | 776152 | 776152 | 100% | >99.99-100% |
| Specimen 7 | Indels | 14 | 14 | 100% | 78.47-100% | 198042 | 198042 | 100% | >99.99-100% |
| | CNVs | 14 | 14 | 100% | 78.47-100% | 10850 | 10850 | 100% | 99.96-100% |
| | SNVs | 1036 | 1036 | 100% | 99.63-100% | 776295 | 776295 | 100% | >99.99-100% |
| Specimen 8 | Indels | 56 | 56 | 100% | 93.58-100% | 198042 | 198042 | 100% | >99.99-100% |
| | CNVs | 14 | 14 | 100% | 78.47-100% | 10877 | 10877 | 100% | 99.96-100% |
| | SNVs | 1120 | 1120 | 100% | 99.66-100% | 776139 | 776139 | 100% | >99.99-100% |
| Specimen 9 | Indels | 14 | 14 | 100% | 78.47-100% | 198068 | 198068 | 100% | >99.99-100% |
| | CNVs | 0 | 0 | N/A | N/A | 10888 | 10888 | 100% | 99.96-100% |
| | SNVs | 1206 | 1205 | 99.92% | 99.53-99.99% | 776128 | 776125 | >99.99% | >99.99-100% |
| Specimen<br>10 | Indels | 14 | 14 | 100% | 78.47-100% | 198055 | 198055 | 100% | >99.99-100% |
| | CNVs | 14 | 14 | 100% | 78.47-100% | 10874 | 10874 | 100% | 99.96-100% |
| | SNVs | 1050 | 1050 | 100% | 99.64-100% | 776334 | 776330 | >99.99% | >99.99-100% |
| Specimen<br>11 | Indels | 63 | 63 | 100% | 94.25-100% | 197994 | 197990 | >99.99% | >99.99-100% |
| | CNVs | 14 | 14 | 100% | 78.47-100% | 10864 | 10864 | 100% | 99.96-100% |
| Specimen<br>12 | SNVs | 1456 | 1456 | 100% | 99.74-100% | 776063 | 776057 | >99.99% | >99.99-100% |
| | Indels | 70 | 70 | 100% | 94.80-100% | 198003 | 198003 | 100% | >99.99-100% |
| Specimen<br>Stratified by<br>Variant Type | # Total<br>Positive<br>Variants* | # Detected<br>Positive<br>Variants* | PPA | 95% CI<br>of PPA | # Total<br>Negative<br>Variants* | # Detected<br>Negative<br>Variants* | NPA | 95% CI<br>of NPA | |
| CNVs | 14 | 14 | 100% | 78.47-100% | 10878 | 10878 | 100% | 99.96-100% | |
| SNVs | 882 | 882 | 100% | 99.57-100% | 776438 | 776438 | 100% | >99.99-100% | |
| Specimen<br>13 | Indels | 14 | 14 | 100% | 78.47-100% | 198042 | 198042 | 100% | >99.99-100% |
| CNVs | 14 | 14 | 100% | 78.47-100% | 10869 | 10869 | 100% | 99.96-100% | |
| SNVs | 1106 | 1106 | 100% | 99.65-100% | 776256 | 776256 | 100% | >99.99-100% | |
| Specimen<br>14 | Indels | 28 | 28 | 100% | 87.94-100% | 198016 | 198016 | 100% | >99.99-100% |
| | CNVs | 14 | 14 | 100% | 78.47-100% | 10864 | 10864 | 100% | 99.96-100% |
| SNVs | 1080 | 1074 | 99.44% | 98.79-99.75% | 776244 | 776238 | >99.99% | >99.99-100% | |
| Specimen<br>15 | Indels | 14 | 14 | 100% | 78.47-100% | 198029 | 198029 | 100% | >99.99-100% |
| CNVs | 0 | 0 | N/A | N/A | 10892 | 10892 | 100% | 99.96-100% | |
| SNVs | 952 | 952 | 100% | 99.60-100% | 776373 | 776373 | 100% | >99.99-100% | |
| Specimen<br>16 | Indels | 56 | 56 | 100% | 93.58-100% | 198003 | 198003 | 100% | >99.99-100% |
| CNVs | 0 | 0 | N/A | N/A | 10892 | 10892 | 100% | 99.96-100% | |
| SNVs | 1302 | 1302 | 100% | 99.71-100% | 776113 | 776113 | 100% | >99.99-100% | |
| Specimen<br>17 | Indels | 14 | 14 | 100% | 78.47-100% | 198016 | 198016 | 100% | >99.99-100% |
| CNVs | 0 | 0 | N/A | N/A | 10892 | 10892 | 100% | 99.96-100% | |
| SNVs | 1694 | 1694 | 100% | 99.77-100% | 775788 | 775788 | 100% | >99.99-100% | |
| Specimen<br>18 | Indels | 84 | 84 | 100% | 95.63-100% | 197977 | 197977 | 100% | >99.99-100% |
| CNVs | 14 | 14 | 100% | 78.47-100% | 10892 | 10892 | 100% | 99.96-100% | |
| Specimen<br>Stratified by<br>Variant Type | # Total<br>Positive<br>Variants* | # Detected<br>Positive<br>Variants* | PPA | 95% CI<br>of PPA | # Total<br>Negative<br>Variants* | # Detected<br>Negative<br>Variants* | NPA | 95% CI<br>of NPA | |
| Specimen<br>19 | SNVs | 924 | 924 | 100% | 99.59-100% | 776334 | 776334 | 100% | >99.99-100% |
| | Indels | 14 | 14 | 100% | 78.47-100% | 198016 | 198016 | 100% | >99.99-100% |
| | CNVs | 0 | 0 | N/A | N/A | 10891 | 10891…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.