DEN200050 · Helius Medical, Inc. · QCF · Mar 26, 2021 · Neurology
Device Facts
Record ID
DEN200050
Device Name
Portable Neuromodulation Stimulator (PoNS)
Applicant
Helius Medical, Inc.
Product Code
QCF · Neurology
Decision Date
Mar 26, 2021
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 882.5889
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
DEN200050 · Mar 26, 2021
Portable Neuromodulation Stimulator (PoNS)
Helius Medical, Inc.
Retrospective clinical rehabilitation records (Canada)
The sponsor used retrospective real-world data to evaluate the effectiveness of the PoNS device in MS patients and performed a post-hoc pooled analysis combining this RWD with data from two prospective clinical trials (Tyler et al. and Leonard et al.) to support the effectiveness claim.
Retrospective analysis of real-world data (RWD) in Canadian clinical rehabilitation settings; Retrospective analysis
MS patients with gait deficits; Sample Size: 36
Not applicable for this study
Mean change from baseline to week 14 in Functional Gait Assessment (FGA) score
Indications for Use
The PoNS device is indicated for use as a short term treatment of gait deficit due to mild to moderate symptoms from multiple sclerosis and is to be used as an adjunct to a supervised therapeutic exercise program in patients 22 years of age and over by prescription only.
Device Story
Portable Neuromodulation Stimulator (PoNS) is a non-implantable electrical tongue nerve stimulator; generates electrical pulses to stimulate tongue nerves; treats motor deficits. Used as adjunct to supervised therapeutic exercise program for multiple sclerosis patients. Operated by patient under prescription. Device delivers stimulation to tongue; intended to improve gait deficit. Healthcare provider monitors progress through exercise program. Benefits include potential reduction in motor deficits associated with MS.
Clinical Evidence
No clinical data provided in the document. FDA granted De Novo classification based on identified risks and mitigation measures including biocompatibility, electrical/mechanical/thermal safety testing, EMC testing, battery safety, and software validation.
Technological Characteristics
Non-implantable tongue stimulator. Mouthpiece materials: Polyimide PCB, gold-plated nickel electrodes, thermoplastic elastomer bite pad, polycarbonate cap, polyurethane resin. Stimulation: Pulsed biphasic (unbalanced), rectangular waveform, 60 µsec primary phase, 312.5 µsec pulse duration, 3.2 kHz channel pulses in triplets at 200 Hz, repeating every 50 Hz. Connectivity: USB Type-C for charging/data transfer. Software: Moderate level of concern. Sterilization: Not sterile, single-patient reusable (14-week limit). Power: 3.7V Li-ion battery.
Indications for Use
Indicated for short-term treatment of gait deficit due to mild to moderate multiple sclerosis symptoms in patients ≥22 years old, as an adjunct to supervised therapeutic exercise.
Regulatory Classification
Identification
An electrical tongue nerve stimulator to treat motor deficits is a prescription device that consists of a non-implantable apparatus to generate electrical pulses for stimulation of the nerves in the tongue to provide treatment of motor deficits.
Special Controls
In combination with the general controls of the FD&C Act, the electrical tongue nerve stimulator to treat motor deficits is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(2) Performance data must demonstrate the electromagnetic compatibility, battery safety, and electrical, mechanical, and thermal safety of the device.
(3) Non-clinical performance testing must characterize the electrical stimulation parameters of the device.
(4) Software verification, validation, and hazard analysis must be performed. Software documentation must include an assessment of the impact of threats and vulnerabilities on device functionality and end users as part of cybersecurity review.
(5) Labeling must include:
(i) A detailed summary of the device's technical parameters;
(ii) Instructions for use;
(iii) Cleaning, storage, and charging instructions; and
(iv) Disposal instructions.
Submission Summary (Full Text)
{0}------------------------------------------------
## DE NOVO CLASSIFICATION REQUEST FOR PORTABLE NEUROMODULATION STIMULATOR (PONS)
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Electrical tongue nerve stimulator to treat motor deficits. An electrical tongue nerve stimulator to treat motor deficits is a prescription device that consists of a nonimplantable apparatus to generate electrical pulses for stimulation of the nerves in the tongue to provide treatment of motor deficits.
NEW REGULATION NUMBER: 21 CFR 882.5889
CLASSIFICATION: Class II
PRODUCT CODE: QCF
#### BACKGROUND
DEVICE NAME: Portable Neuromodulation Stimulator (PoNS)
SUBMISSION NUMBER: DEN200050
DATE DE NOVO RECEIVED: August 4, 2020
#### SPONSOR INFORMATION:
Helius Medical, Inc. 642 Newtown Yardley Road, Suite 100 Newtown, Pennsylvania 18940
#### INDICATIONS FOR USE
The PoNS device is indicated for use as a short term treatment of gait deficit due to mild to moderate symptoms from multiple sclerosis and is to be used as an adjunct to a supervised therapeutic exercise program in patients 22 years of age and over by prescription only.
#### LIMITATIONS
Prescription use only.
The device is not to be used while eating, drinking, or driving to avoid risk of physical injury.
{1}------------------------------------------------
The device should be used with caution in patients who:
- Have implanted electronic devices, including: .
- o Cardiac pacemakers
- o Cardioverter defibrillators
- o Deep Brain Stimulators
- Vagal Nerve Stimulators O
- o Sacral nerve stimulators
- o Cochlear Implants
- Have metal in the mouth (e.g. piercings, braces, retainers, or other orthodontic . appliance)
- . Have seizure disorders (e.g., epilepsy)
Use of the device does not replace the need for follow-up testing to determine the initial and ongoing effectiveness of the therapy as recommended by clinical practice guidelines.
## Warnings
The intensity should only be adjusted while the Mouthpiece is on the tongue. The patient should be the one who adjusts the intensity level. In case the patient needs assistance in pressing the buttons on the Controller, the person assisting should use care to adjust the intensity by small increments. Check with the patient each time you press the INTENSITY UP/DOWN buttons, before pressing them again.
Portable Radio Frequency (RF) communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the PoNS System, including cables supplied with the PoNS System. Otherwise, degradation of the performance of this equipment could result.
Device should not be used by persons under the age of 22.
PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
### DEVICE DESCRIPTION
The PoNS™ device is a non-implantable tongue stimulator that delivers electrical stimulation to the trigeminal and facial nerves for use as a short-term treatment of gait deficit due to mild to moderate symptoms from multiple sclerosis and is to be used as an adjunct to a supervised therapeutic exercise program. The device is for use in patients over 22 years of age by prescription only.
The PoNS™ mouthpiece is held lightly in place by the lips and teeth around the neck of the tab that goes into the mouth and rests on the anterior, superior part of the tongue. The control unit is worn around the neck during session usage. It has 5 user-controlled buttons: power, information,
{2}------------------------------------------------
intensity up, intensity down, and start/stop/pause. The control unit also records usage data including session duration and general activities performed which are captured by an accelerometer. During a patient's visit with a therapist is able to connect the control unit to a computer and, using software developed specifically for the PONSTM, view the usage data. The purpose of the usage data is to give the therapist information on how to improve a patient's execution of therapy by identifying potential areas of missed or shortened sessions. The PoNSTM System is comprised of a Controller, a Mouthpiece Retainer Case, a Charger, a Product Carry Case and the PoNSTM Software. Accessory components, also supplied, include a cable for connection to a computer running the PoNSTM software.
Controller: The Controller includes the primary user interface and drive electronics for providing the electronic stimulation waveform. Through the user interface, the intensity of the stimulation waveform can be adjusted, the stimulation can be started and stopped, and the Controller can be powered on and off. The Controller user interface consists of a visual display, audio feedback and vibration feedback.
Mouthpiece: The Mouthpiece houses the electrode array through which the stimulation waveform is applied to the patient's tongue. Electronics and software within the Mouthpiece control the timing of the stimulation waveform. The Controller sends commands to the Mouthpiece and receives status messages from the Mouthpiece via a cable that forms an integral part of the Mouthpiece assembly.
Mouthpiece Retainer Case: The Mouthpiece Retainer Case is used to store the Mouthpiece between uses. The cable of the Mouthpiece wraps around the base of the Mouthpiece Retainer Case. There is a pocket in the Carry Case sized to house the Mouthpiece Retainer Case during storage.
Charger: The Charger consists of a mains power adaptor that connects to the Controller via a standard USB Type-C connector and provides a low voltage power supply to recharge the Controller battery. Stimulation is disabled while the Charger is connected to the Controller.
### Summary of Mouthpiece Materials of Construction and Type of Patient Contact
| Mouthpiece Component | Material | Patient Contact |
|-------------------------------------|-------------------------|-------------------------------------------|
| Electrode Printed Circuit Board | Polyimide | Mucosal Membrane<br>(mouth, tongue, lips) |
| Electrodes on Printed Circuit Board | Gold Plated Nickel | |
| Overmolded Bite Pad | Thermoplastic elastomer | |
| Mouthpiece Cap | Polycarbonate | |
| Potting Resin (internal) | Polyurethane Resin | Potentially Mucosal<br>Membrane |
{3}------------------------------------------------
Output parameters and other electrical specifications of the PoNS™ device are presented below in Table 1.
| GENERAL DEVICE CHARACTERISTICS | |
|-------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MODE OR PROGRAM NAME | PORTABLE NEUROMODULATION<br>STIMULATOR (PONSTM) |
| WAVEFORM (E.G., PULSED MONOPHASIC,<br>BIPHASIC) | PULSED BIPHASIC (UNBALANCED) |
| SHAPE (E.G., RECTANGULAR, SPIKE, RECTIFIED<br>SINUSOIDAL) | RECTANGULAR |
| MAXIMUM OUTPUT VOLTAGE (VOLTS) (+/- 5%) | 17.5V @ 500 Ω<br>20V @ 5 KΩ |
| MAXIMUM OUTPUT CURRENT (SPECIFY UNITS)<br>(+/- 5%) | 3.5 ΜΑ @ 500 Ω (CHANNEL: 26 MA)<br>440 μΑ @ 5 KΩ (CHANNEL: 3.85 MA) |
| DURATION OF PRIMARY (DEPOLARIZING) PHASE<br>(µSEC) | 60 µSEC |
| PULSE DURATION (µSEC) | 312.5 µSEC |
| FREQUENCY (HZ) [OR RATE (PPS)] | TRIPLETS AT 200 HZ INTERVALS,<br>REPEATING EVERY 50 Hz<br>CHANNEL PULSES STIMULATING<br>DIFFERENT TONGUE LOCATIONS AT<br>3.2 KHZ; GROUPS OF 16 CHANNEL<br>PULSES DELIVERED IN TRIPLETS AT<br>200 HZ; REPEATS EVERY 50 Hz |
| NET CHARGE (MICROCOULOMBS (µC) PER PULSE)<br>(IF ZERO, STATE METHOD OF ACHIEVING ZERO NET<br>CHARGE.) | -0.002 @ 500 Ω (CHANNEL: -0.016)<br>-0.0041 @ 5 KΩ (CHANNEL: -0.036) |
| Maximum Phase Charge, (µC) | 0.21 @ 500 Ω (channel: 1.56) |
| Maximum Current Density, (mA/cm², rms) | 18.79 @ 500 Ω<br>2.36 @ 5 KΩ |
| Maximum Average Current (average absolute<br>value), (mA) | 0.082 @ 500 Ω<br>0.009 @ 5 kΩ |
| Maximum Average Power Density, (W/cm²) | 0.025 @ 500 Ω<br>0.0041 @ 5 kΩ |
### SUMMARY OF NONCLINICAL/BENCH STUDIES
# BIOCOMPATIBILITY/MATERIALS
The biocompatibility evaluation for PoNS™ mouthpiece was conducted in accordance with the International Standard ISO 10993-1 :2009 "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing Within a Risk Management Process". The
{4}------------------------------------------------
PoNSTM mouthpiece and components is categorized as a surface device in prolonged (24 hours to 30 days) contact of the mucosal membrane (i.e. mouth, tongue lips). Assessment of the device included the following tests which met the prespecified acceptance criteria:
- Cytotoxicity (ISO 10993-5:2009) .
- Sensitization Test (ISO 10993-10:2010) .
- Intracutaneous Reactivity (ISO 10993-10:2010) .
- Oral Mucosa Irritation (ISO 10993-10:2010) .
- Acute Systemic Toxicity (ISO 10993-11:2006) .
- Material-mediated pyrogenicity (ISO 10993-11:2006) .
- Extractables & Leachables Study (ISO 10993-17:2002 .
- Chemical Characterization of medical device materials (ISO 10993-18:2005). .
- Genotoxicity (ISO 10993-3:2014) .
### SHELF LIFE/REPROCESSING/STERILITY
The PoNSTM device is not provided sterile and is not intended to be sterilized prior to use. The device is a single- patient, reusable device and instructions have been provided to dispose of the mouthpiece portion of the device after fourteen weeks from first use.
The PoNS Controller is programmed such that it ceases to provide stimulation after [014] weeks and a patient must return to their Healthcare Professional to reset the Controller for additional time. This mechanism is intended to control long-term and unchecked patient usage, and to ensure Healthcare Professional oversight.
The PoNS™ device consists of the following components: a mouthpiece, a controller, and a retainer case. Cleaning instructions have been provided for each of the three components.
# ELECTROMAGNETIC CAPABILITY & ELECTROMAGNETIC SAFETY
The PoNSTM system was tested according to the following FDA-recognized consensus standards:
- . IEC 60601-1:2005 (Modified to be equivalent to (AAMI/ANSI ES60601-1:2005/(R)2012 and C1:2009/(R)2012 and, A2:2010/(R)2012) "Medical Electrical Equipment; Part 1: General requirements for basic safety and essential performance." Results demonstrated that the device is compliant to this standard.
- · IEC 60601-1-2:2014 "Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance-Collateral Standard: Electromagnetic disturbances - Requirements and Tests."
- IEC 60601-1-11:2015 "Medical electrical equipment: General requirements for . basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment."
{5}------------------------------------------------
- · IEC 60601-2-10: 2012 and A1:2016 "Medical Electrical Equipment Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators."
#### SOFTWARE
A failure or latent flaw in the software of the PoNS™ device could indirectly result in minor injury to the patient or operator: therefore, the software of this device is considered to have a "Moderate" level of concern.
The submission contained all the elements of software documentation corresponding to a "Moderate" level of concern, as outlined in the FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." issued May 11, 2005 (https://www.fda.gov/media/73065/download). Adequate documentation describing the software, firmware, software specifications, architecture design, software development environment, traceability, revision level history, unresolved anomalies and cybersecurity provides the foundation that the software will operate in a manner as described in the specifications. A hazard analysis was performed to characterize software risks including device malfunction and measurement related errors. The submission included verification and validation (V&V) testing to address the potential hazards with satisfactory result.
#### PERFORMANCE TESTING - BENCH
### LITHIUM-ION BATTERY TESTING:
The PoNS™ controller is powered by a rechargeable 3.7V Lithium-Ion battery. The battery complies with IEC 62133-2 (Standard for lithium batteries), and IEC 60601-1 (Basic Safety).
#### ELECTRICAL STIMULATION OUTPUT
Testing was performed to characterize the stimulation output waveform, the functionality of the PoNSTM as a system, and the requirements of the output stimulation parameters. Results demonstrated that the system meets specifications.
#### ELECTRODE BENCH TESTING:
Testing was performed to assess the mechanical measurements, the design of the electrodes (and tolerances) and the electrical characteristics (impedance and current distribution) of the electrodes under the expected worst-case conditions of normal operation. Results demonstrated that the electrodes passed all testing.
### SUMMARY OF CLINICAL INFORMATION
The following clinical information was provided to evaluate the safety and effectiveness of the PoNS™ device:
{6}------------------------------------------------
- · Two Clinical Studies:
- A randomized, double blind controlled trial in I subjects with gait deficits due to Multiple Sclerosis (Tyler et. al, 2014)
- A pilot study intended to investigate the effects of PoNS™M with cognitive rehab i and physical rehab, on the primary outcome of improvement in sensory organization tasks (Leonard et. al, 2017)
- A retrospective analysis of real-world data ("RWD") collected with the PoNS™ in MS . patients in clinical rehabilitation settings in Canada
- . Post-hoc re-analysis of studies by Tyler et. al, 2014 and Leonard et. al, 2017, pooling results with RWD.
- . 6-patient case series poster presentation.
The clinical information provided is summarized below.
## 1. Summary of the study by Tyler et. al (2014):
A Randomized, double blind controlled trial in subjects with gait deficits due to Multiple sclerosis. I subjects used Active PoNSTM device, subjects used a control device that did not deliver stimulation. The intervention consisted of gait training in the laboratory while simultaneously using the device to provide stimulation during gait training. This was followed by [ ] weeks of home use of the device while performing gait training for a total of weeks of treatment. The primary outcome measure was the Dynamic Gait Index (DGI, clinician scored index of @gait tasks). DGI was performed at baseline week Results showed the (b)(4) active group on average achieved improvement in DGI score of [ ] a] at [b] weeks while the control group improved [0] points. Compared to baseline, the active group achieved a statistically significant and clinically significant ( the study, while the control group did not.
Table 19.4: DGI Over Time in Tyler Active vs Control Groups
Image /page/6/Figure/10 description: The image shows a table with the headers "Active" and "Control". Under each of these headers are the sub-headers "Week", "N", "Mean", "SD", "Lower Cl", and "Upper Cl". The table appears to be empty, with the exception of the text "(b)(4)" in the middle of the table.
{7}------------------------------------------------
Image /page/7/Figure/0 description: The image shows a figure and a table. The figure is titled "Figure 19-1: Change in DGI Score from Baseline". The table is titled "Table 19-5: Responder Analysis (Tyler, et al. DGI Data)" and contains the columns "Change in DGI at Week 14 versus Baseline", "Active Group", and "Control Group".
# 2. Summary of the study by Leonard et. al (2017):
Patient blinded, randomized, controlled trial in subjects with MS assessing effects of PoNS+cognitive and physical rehabilitation (n=7), vs sham device+cognitive and physical rehabilitation (n=7). Baseline evaluations including structural and fMRI, balance tests (sensory organization tests ("SOT")), DGI, and a neuropsychological assessment. Participants received physical therapy and working memory training for [] weeks. For the first weeks, an in- lab training program was completed with twice-daily training sessions, followed by weeks of at- home training with the same exercises. During both stages, subjects completed three sessions per day: morning, afternoon and evening. Participants returned every weeks for re-evaluation and to progress their training. Results: SOT scores measured baseline stability under six progressively difficult conditions. The composite scores showed a trend of improvement over time for both the Control and Active groups, although more consistent for the Active group. The Active group showed a statistically significant improvement at be weeks compared to baseline values. The DGI tasks included: steady state walking, walking with changing speeds, walking with head turns horizontally and vertically, stepping over and around obstacles, pivoting and stair climbing. Analysis of DGI scores after weeks revealed a nonsignificant trend towards the Active group .. This study also examined functional magnetic resonance imaging (fMRI) during a gait imagery task.
{8}------------------------------------------------
Figure 6. Dynamic Gait Index showing average scores for baseline and following DIC weeks of training
Image /page/8/Picture/1 description: The image is a gray rectangle with a thin red border. There is some text in the upper center of the rectangle, but it is illegible. The rectangle takes up most of the image, and the red border is visible on all four sides.
Series 2 = Sham Group in &
# 3. Real World Data
The sponsor also provided a retrospective analysis of real-world data ("RWD") collected with the PoNS™ in MS patients in clinical rehabilitation settings in Canada. Patients who agreed to treatment were given 1-hr consultation, consented, and given baseline assessments of gait function using FGA (Functional Gait Assessment (FGA). The primary outcome for effectiveness was defined as the mean change from baseline to week or in FGA. At Week the mean improvement in baseline functional gait assessment (FGA) score was points ( (0) 4) (n=36). No serious safety AE were reported. Of note, Do of the of real world patients had baseline Expanded disability status scale (EDSS) scores These results suggest that the PoNS™ may not be as effective for subjects with more severe baseline disability.
A post-hoc re-analysis of Tyler and Leonard studies, pooling results with RWD was also provided by pooling the DGI results from Tyler and Leonard studies with the RWD to provide a larger data set on which to evaluate the evidence supporting the effectiveness of PoNSTM for the proposed indications for use. Comparison was facilitated by creating from the RWE an "adjusted DGI" with the same range as the 8-item DGI; this entailed summing the 7 items in common across the DGI and FGA scales and multiplying by 8/7. Although the sponsor provides a rationale for why one can multiply the FGA results by 8/7ths to convert the score to a surrogate DGI score, this practice is not validated according to any published literature.
The mean improvement from baseline to Week pooling over all three groups is 10000 when time points in common (Weeks br(4) and P) are (b)4) used. The mean improvement is [b](4) (0)(4) I when all possible
{9}------------------------------------------------
data at all weeks are used. However, as noted in the Tyler and Leonard studies, there are differences in the sub-type of MS included in Tyler, et al, Leonard et al, and RWD, differences in baseline DGI score, and differences in medication's used by subjects in each group. Furthermore, he demonstrated improvement in the RWE pooled with Tyler and Leonard et al (0) improvement from baseline), is not much better than the demonstrated improvement in the control group of the Tyler et. al, study [[0](4] improvement).
#### Figure 19-5: Adjusted DGI Scores in RWE Analysis Dataset
Image /page/9/Picture/2 description: The image is a gray rectangle with a red border. There is text in the middle of the rectangle that says "(b)(4)". The text is also in red.
Figure 19-6: Pooling Analysis - RWE + Tyler
Image /page/9/Picture/4 description: The image is a gray rectangle with a red border. In the top center of the rectangle, there is the text "(b)(4)". The text is small and in black font. The rectangle takes up most of the image.
Pediatric Extrapolation
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
### LABELING
The labeling (User Instructions) meets the requirements of 21 CFR Part 801.109 for prescription devices.
The following contraindications, warnings, and precautions are included within device labeling:
{10}------------------------------------------------
# Contraindications:
The PoNS device delivers electrical stimulation directly to the surface of the tongue. Precautions for use are similar to those for transcutaneous electrical nerve stimulation (TENS).
Electrical stimulation SHOULD NOT be used:
- if there is an active or suspected malignant tumor .
- in areas of recent bleeding or open wounds .
- in areas that lack normal sensation .
The PoNS has not been tested on, and thus should not be used by individuals who are pregnant. Do not use the PoNS if you are sensitive to nickel, gold or copper.
### Precautions:
The device should be used with caution in patients who:
- . Have implanted electronic devices, including:
- o Cardiac pacemakers
- o Cardioverter defibrillators
- O Deep Brain Stimulators
- Vagal Nerve Stimulators o
- o Sacral nerve stimulators
- o Cochlear Implants
- . Have metal in the mouth (e.g. piercings, braces, retainers, or other orthodontic appliance)
- Have seizure disorders (e.g., epilepsy) .
### Warnings:
- DO NOT connect the Mouthpiece to the Controller before seeing your PoNSTM therapist, . to avoid premature expiration of the Mouthpiece.
- . DO NOT place the Mouthpiece in your mouth while the Controller is charging or connected to your PoNSTM therapist's computer.
- . DO NOT use a power adapter other than the one provided with your PoNSTM to charge the Controller. Always charge the Controller indoors.
- . DO NOT wet the Controller or use your PoNSTM in the rain or snow.
- . DO NOT wet the connector on the end of the Mouthpiece cord when cleaning the Mouthpiece. Always check and ensure the connector is dry before connecting and using the Controller.
- . DO NOT use the PoNS™ if you are sensitive to nickel, gold or copper. The PoNS™ Mouthpiece contains nickel, gold and copper.
- . DO NOT use your PoNS™ if you have open mouth sores, ulcers or tongue piercings.
- . DO NOT eat, drink or smoke while using your PoNSTM.
- . If the materials on the Mouthpiece cause any allergic reactions or undesirable effects, discontinue using the PoNS™ immediately and contact your healthcare provider.
- . DO NOT bite down on the Mouthpiece with excessive force.
- DO NOT use your PoNSTM if it has been dropped and broken. .
{11}------------------------------------------------
- DO NOT use your PoNSTM while driving or operating machinery. .
- . DO NOT connect the Mouthpiece to any other device or modify any part of the equipment.
- . Keep your PoNSTM out of Magnetic Resonance Imaging (MRI) field.
- Keep your PoNSTM out of the sight and reach of children. .
- . BE AWARE: the long-term effects of electrical stimulation are unknown.
- . Only use specified charger and cable.
- . The PoNSTM Controller contains a rechargeable lithium battery.
The labeling also includes:
- Information on how the device operates and the typical sensations experienced during . use
- . Information in the Instructions for Use regarding how to place the device on the user
- . Storage and cleaning instructions for the device
- . Physical limitations of the device including suitable temperature range, duration of use, and expected product life.
- . Charging instructions; and
- Disposal instructions. .
#### RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of the electrical tongue nerve stimulator to treat motor deficits and the measures necessary to mitigate these risks.
| Identified Risk | Mitigation Measures |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Adverse tissue reaction | Biocompatibility evaluation |
| Thermal, electrical, or mechanical<br>fault, or system malfunction resulting<br>in tissue damage due to<br>overstimulation or thermal injury (e.g.<br>burn/shock) to user | Electrical, mechanical, and thermal safety testing<br>Electromagnetic compatibility (EMC) testing<br>Battery safety testing<br>Non-clinical performance testing<br>Software validation, verification & hazard analysis<br>Labeling |
| Use error that may result in user<br>discomfort or injury | Labeling |
| Device contamination resulting in<br>patient illness | Labeling |
| Adverse events involving the mouth,<br>tongue, or gums such as irritation and<br>discomfort | Labeling |
### SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the electrical tongue nerve stimulator to treat motor deficits is subject to the following special controls:
{12}------------------------------------------------
- (1) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
- (2) Performance data must demonstrate the electromagnetic compatibility, battery safety, and electrical, mechanical, and thermal safety of the device.
- (3) Non-clinical performance testing must characterize the electrical stimulation parameters of the device.
- (4) Software verification, validation, and hazard analysis must be performed. Software documentation must include an assessment of the impact of threats and vulnerabilities on device functionality and end users as part of cybersecurity review.
- (5) Labeling must include:
- a) A detailed summary of the device's technical parameters;
- b) Instructions for use;
- c) Cleaning, storage, and charging instructions; and
- d) Disposal instructions.
## BENEFIT-RISK DETERMINATION
The known probable risks of the device are based on the data collected in the clinical study described above. The device exhibited an acceptable safety profile in the clinical studies which were conducted, and any adverse events that occurred were temporary and had complete resolution. No device-related serious adverse events were observed, such as electrical injury, burns, or permanent injury to tongue nerve or muscle function. The results of the nonclinical testing demonstrated that the device performed as per specifications and the results did not raise concerns regarding risks to the patients.
The probable benefits of the device are also based on data collected in clinical studies as described above. From the Tyler et al. study, there is evidence of a clinically meaningful treatment effect on Dynamic Gait Index (DGI) scores, a clinician scored index of I gait tasks. DGI was performed at baseline week | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | active group on average achieved improvement in DGI score of (6) at 190 weeks while the control group improved [ b] points. Compared to baseline, the active group achieved a clinically significant (4-point change in DGI) by the end of the study, while the control group did not.
Thus. PoNS™ device provides clinically meaningful relief improvement treatment of gait deficit due to mild to moderate symptoms from multiple sclerosis when used as an adjunct to a supervised therapeutic exercise program.
### Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
{13}------------------------------------------------
### Benefit/Risk Conclusion
In conclusion, given the available information above, for the following indication statement:
The PoNSTM device is indicated for use as a short-term treatment of gait deficit due to mild to moderate symptoms from multiple sclerosis and is to be used as an adjunct to a supervised therapeutic exercise program in patients over 22 years of age by prescription only.
The probable benefits outweigh the probable risks for the Portable Neuromodulation Stimulator (PoNSTM). The device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.
### CONCLUSION
The De Novo request for the Portable Neuromodulation Stimulator (PoNS™) is granted and the device is classified as follows:
Product Code: QCF Device Type: Electrical tongue nerve stimulator to treat motor deficits Regulation Number: 21 CFR 882.5889 Class: II
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.