DEN190023 · Asuragen, Inc. · OYV · Feb 21, 2020 · Immunology
Device Facts
Record ID
DEN190023
Device Name
AmplideX Fragile X Dx & Carrier Screen Kit
Applicant
Asuragen, Inc.
Product Code
OYV · Immunology
Decision Date
Feb 21, 2020
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 866.5970
Device Class
Class 2
Indications for Use
The AmplideX® Fragile X Dx & Carrier Screen Kit is an in vitro diagnostic device that uses polymerase chain reaction (PCR) and capillary electrophoresis to detect and identify the number of cytosine-guanine-guanine (CGG) repeats in the fragile X mental retardation-1 (FMR1) gene using genomic DNA isolated from peripheral whole blood specimens. It is solely intended as an aid in the post-natal diagnosis of fragile X syndrome, and fragile X-associated disorders [i.e., fragile X-associated tremor/ataxia syndrome (FXTAS) or fragile X-associated primary ovarian insufficiency (FXPOI)], and for carrier testing in adults of reproductive age. Assay results are solely intended to be interpreted by healthcare professionals who are board certified in molecular genetics and to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice. Reflex testing, clinical genetic evaluation, and genetic counseling should be offered as appropriate. The test is for use on the 3500Dx Series Genetic Analyzer. This test is not indicated for use for fetal diagnostic testing, newborn screening or for stand-alone diagnostic purposes.
Device Story
In vitro diagnostic kit; uses PCR and capillary electrophoresis to detect CGG repeat counts in FMR1 gene; input is genomic DNA from peripheral whole blood; processed on 3500Dx Series Genetic Analyzer; output is electropherogram showing repeat size; interpreted by board-certified molecular geneticists; used as aid in diagnosis and carrier screening; results combined with clinical findings; supports clinical decision-making regarding genetic counseling and further evaluation.
Clinical Evidence
No clinical data provided in the document. Device requires analytical validation including accuracy against a comparator, precision (repeatability/reproducibility across sites/operators/instruments), limit of detection, specificity, and stability testing.
Technological Characteristics
PCR-based nucleic acid amplification followed by capillary electrophoresis (CE) size resolution. Requires Applied Biosystems 3500 Dx Series Genetic Analyzer. Reagents: GC-Rich Amp Buffer, Polymerase Mix, Primer Mix, ROX 1000 Size Ladder. Software: AmplideX Fragile X Reporter. Connectivity: Standalone instrument. Sterilization: N/A (reagents).
Indications for Use
Indicated for post-natal diagnosis of fragile X syndrome, FXTAS, and FXPOI, and for carrier testing in adults of reproductive age. Not for fetal diagnostic testing, newborn screening, or standalone diagnostic purposes.
Regulatory Classification
Identification
An inherited nucleotide repeat disorder DNA test is a prescription in vitro diagnostic device that is intended to detect and identify the number of nucleotide repeats in a gene using genomic DNA isolated from post-natal patient specimens. It is solely intended as an aid for carrier testing and as an aid for the diagnosis of inherited nucleotide repeat-associated disorders. Assay results are solely intended to be used in conjunction with other clinical and diagnostic findings. These tests do not include those indicated for use for fetal diagnostic testing or newborn screening.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use on the device's label required under § 809.10(a)(2) of this chapter and device's labeling required under § 809.10(b)(2) of this chapter must include a statement that assay results are solely intended to be used in conjunction with other clinical and diagnostic findings, consistent with professional standards of practice, and that reflex testing, clinical genetic evaluation, and genetic counseling should be offered as appropriate.
(2) The labeling required under § 809.10(b) of this chapter must include:
(i) A warning that mosaicism detected in one tissue may not reflect mosaicism in other tissues and that the significance of mosaicism should be interpreted with caution in conjunction with other laboratory and clinical information (
*e.g.,* sex of patient, diagnostic testing or carrier screening, patient symptoms) and should include appropriate genetic counseling.(ii) A prominent statement that this test is not indicated for use for fetal diagnostic testing, newborn screening or for stand-alone diagnostic purposes.
(iii) Information that addresses how to interpret different result outputs specific to the technology, such as (peaks) in the electropherograms.
(3) Design verification and validation must include the following:
(i) Appropriate design features and control elements incorporated into the testing procedure that mitigate the risk of incorrect clinical results. These include controls as determined acceptable by FDA that:
(A) Enable the user to determine when the amplification may yield incorrect results,
(B) Enable the user to determine when cross contamination may have occurred;
(C) Software risk control measures that address device system hazards;
(D) Provide software traceability that ensures all hazards are adequately controlled and that all controls have been validated in the final device design; and
(E) Ensure the instructions for use and test reports appropriately inform the user about the limitations of the assay.
(ii) Validated and acceptable, as determined by FDA, criteria for test result interpretation and reporting, including result outputs.
(iii) Acceptable, as determined by FDA, evidence demonstrating the clinical validity of the device which supports each indicated diagnostic use, including for each genotype and associated phenotype used in providing a clinical determination for the target population.
(iv) Evidence demonstrating acceptable, as determined by FDA, analytical device performance. Patient specimens must represent the full spectrum of expected clinical results and be obtained through unbiased collection. Specimens must be representative of all categories of results and across the range of repeat sizes (
*e.g.,* categories and repeat sizes for Fragile X syndrome are: normal 1-44 repeats; intermediate 45-54 repeats; premutation 55-200 repeats, full mutation greater than 200 repeats), across a range of allelic combinations, be near decision points, and be from both male and female subjects. The number of specimens tested must be sufficient to obtain unbiased estimates of device performance. Analytical validation must include data demonstrating acceptable, as determined by FDA:(A) Agreement with a comparator method(s) determined to be acceptable by FDA. This evidence must demonstrate the accuracy for detecting the size of the nucleotide repeats and the diagnostic categorical calls in DNA in the indicated specimen type(s) from patients that are representative of the intended use population. Accuracy must be assessed for both diagnostic and carrier subsets independently.
(B) Device precision including repeatability and reproducibility, using clinical samples. The study must evaluate all possible sources of variability including, as appropriate, between-site and between operator at a minimum of three sites of which two must be external with a minimum of two operators per site, between-day on a minimum of 3 non-consecutive days, between-run, within-run, between-lot in a minimum of three lots, and between instrument on a minimum of three instruments. Precision must be demonstrated per specimen and determine for both categorical call and by the size of the repeat (
*i.e.,* the percentage of replicates for which the allele fell within the target precision size range). Precision data must be calculated and presented with and without results determined to be invalid.(C) Device performance at the limit of detection of each allele across the range of sizes and as a function of the indicated DNA input for the assay.
(D) Specificity of the reagents for their targets, absence of cross-reactivity, evaluation of sources of interference relevant to the specimen type, and a demonstration of the absence of cross contamination.
(E) Performance of the pre-analytical methods, including DNA extraction methods.
(F) Performance of the device across the range of indicated DNA input concentrations for the assay.
(G) Specimen stability throughout indicated specimen storage ranges, including under expected storage and transport conditions.
(v) Robust evidence demonstrating that the number and frequency of incorrect results due to mosaicism are clinically acceptable, as determined by FDA.
(vi) An appropriate traceability plan to minimize the risk of incorrect results over time, including a description of the molecular size standards and other reagents that may be required for result interpretation, as applicable, that demonstrate the reliable interpretation of the size of the fragments.
(vii) Acceptable, as determined by FDA, device stability protocols and acceptance criteria, that are sufficient to ensure indicated analytical and clinical performance throughout the indicated device stability period. The protocols and acceptance criteria must be adequate to demonstrate that there is no degradation in signal intensity of full mutations when testing a specimen at the latest indicated time point within the indicated device stability that is comprised of the lowest indicated DNA input that can be used.
Submission Summary (Full Text)
{0}------------------------------------------------
# EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR AmplideX® Fragile X Dx & Carrier Screen Kit DECISION SUMMARY
# A. DEN Number:
DEN190023
# B. Purpose for Submission:
De Novo request for evaluation of automatic class III designation for the AmplideX® Fragile X Dx & Carrier Screen Kit
# C. Measurands:
Trinucleotide (CGG) repeats in the FMR1 gene
# D. Type of Test:
Polymerase chain reaction (PCR)-based nucleic acid amplification followed by size resolution with capillary electrophoresis
# E. Applicant:
Asuragen, Inc.
# F. Proprietary and Established Names:
AmplideX® Fragile X Dx & Carrier Screen Kit
# G. Regulatory Information:
- 1. Regulation section:
21 CFR 866.5970
- 2. Classification:
Class II
- 3. Product code(s):
OYV
- 4. Panel:
Immunology (82)
{1}------------------------------------------------
# H. Indications for use:
# 1. Indications for use:
The AmplideX® Fragile X Dx & Carrier Screen Kit is an in vitro diagnostic device that uses polymerase chain reaction (PCR) and capillary electrophoresis to detect and identify the number of cytosine-guanine (CGG) repeats in the fragile X mental retardation-1 (FMR1) gene using genomic DNA isolated from peripheral whole blood specimens. It is solely intended as an aid in the post-natal diagnosis of fragile X syndrome, and fragile Xassociated disorders [i.e., fragile X-associated tremor/ataxia syndrome (FXTAS) or fragile Xassociated primary ovarian insufficiency (FXPOI)], and for carrier testing in adults of reproductive age. Assay results are solely intended to be interpreted by healthcare professionals who are board certified in molecular genetics and to be used in conjunction. with other clinical and diagnostic findings, consistent with professional standards of practice. Reflex testing, clinical genetic evaluation, and genetic counseling should be offered as appropriate. The test is for use on the 3500Dx Series Genetic Analyzer.
This test is not indicated for use for fetal diagnostic testing, newborn screening or for standalone diagnostic purposes.
- 2. Special conditions for use statement(s):
For prescription use
For in vitro diagnostic use
- 3. Special instrument requirements:
Applied Biosystems® 3500 Dx Series Genetic Analyzer (k191030) using the AmplideX Fragile X Reporter Software v 1.0.
# I. Device Description:
The AmplideX Fragile X Dx & Carrier Screen Kit (hereafter referred to as the AmplideX Kit) includes reagents sufficient for 100 reactions and are for use on the Applied Biosystems® 3500 Dx Series Genetic Analyzer (8 and 24 capillary) with AmplideX reporter software. A description of the reagents and the storage conditions are shown in Table 1 below.
| Item | Storage Temperature |
|----------------------------|---------------------|
| GC-Rich Amp Buffer | -15 to -30°C |
| GC-Rich Polymerase Mix | -15 to -30°C |
| ROX 1000 Size Ladder | -15 to -30°C |
| Diluent | -15 to -30°C |
| Fragile X Positive Control | -15 to -30°C |
| Fragile X Primer Mix | -15 to -30°C |
Table 1. Components of the AmplideX Kit
{2}------------------------------------------------
Additional reagents required but not provided with the kit:
- Reagents for isolation and purification of genomic DNA (gDNA)
- POP-7 ™ Polymer ●
- Hi-Di™ Formamide ●
- DS-30 Matrix Standard Kit (Dye Set D) ●
# J. Standard/Guidance Document Referenced (if applicable):
Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices; May 11, 2005
General Principles of Software Validation; Final Guidance for Industry and FDA Staff; January 11,2002
Guidance for Industry - Cybersecurity for Networked Medical Devices Containing Off-the- Shelf (OTS) Software; January 14, 2005
Guidance for Industry. FDA Reviewers and Compliance on Off-the-Shelf Software Use in Medical Devices: September 9, 1999
CLSI Standard EP05-A3, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Second Edition.
CLSI Standard EP07-A2, Interference Testing in Clinical Chemistry, Approved Guideline -Second Edition.
CLSI Standard EP17-A2, Protocols for the Determination of Limits of Detection and Limits of Quantification, Approved Guideline - Second Edition.
CLSI Standard EP25-A, Evaluation of Stability on In Vitro Diagnostic Reagents, Approved Guideline.
# K. Test Principle:
Fragile X syndrome (FXS) and Fragile X associated disorders are inherited genetic conditions characterized by expanded cytosine-guanine (CGG) nucleotide repeats in the 5' untranslated region of the FMR1 gene. Expansion of the repeats is associated with hypermethylation and inactivation of gene expression and subsequent loss of protein expression. This expansion can lead to a variety of consequences depending on the length of the CGG expansion. The AmplideX Fragile X Dx & Carrier Screen Kit quantifies the number of CGG repeats in the alleles of a given sample of purified gDNA using polymerase chain reaction (PCR) followed by size resolution with capillary electrophoresis (CE).
Human gDNA is isolated from EDTA-treated whole human blood (fresh or stored at 2 to 8 °C for up to 14 days) using a DNA extraction kit capable of gDNA purity A260280 range of 1.6-2.0. A total DNA input of 20-80ng is needed to run the reaction. Purified gDNA is first added to a PCR reaction well containing a Master Mix with the GC-Rich Amp Buffer, GC-Rich Polymerase Mix, Fragile X Primer Mix, and either a specimen or control (Fragile X Diluent or Fragile X
{3}------------------------------------------------
Positive Control) in a total of 13 uL. After thermal cycling, unpurified PCR products are directly mixed with Hi-Di™ Formamide and ROX 1000 Size Ladder. Following denaturation, amplicons are resolved on an Applied Biosystems® 3500 Dx Genetic Analyzer running POP-7™ polymer and analyzed using the AmplideX Fragile X Reporter.
# Calibration
The ROX 1000 Size Ladder included in the kit is added to each sample to allow for proper fragment sizing analysis by the AmplideX® Fragile X Reporter. Spectral calibration of the 3500 Dx instrument with DS-30 Matrix Standard Kit is required in order to properly detect and analyze samples containing ROX dye and is performed prior to the assay run.
# Quality Control
The AmplideX Fragile X Reporter performs a series of OC checks on the batch controls No Template control (NTC) and Positive Control (PC) as well as on the individual samples. The software assesses three general modalities of failure: Ladder Error (LD), Low Signal (LS), and Contamination (CT). If one or more of these failures occurs in the positive control, it invalidates the run as a Batch QC failure. The NTC is expected to give a Low Signal (LS) error.
# Controls
Fragile X Positive Control and Diluent (for use as an NTC) are included in the kit. A single positive and no template control must be included in every run for the analysis software to interpret sample results correctly. The NTC should not generate any peak calls from the Reporter output (Figure 5). The Positive Control includes alleles at 18 ±1, 30 ±1, 32 ±1, 56 ±1, 85 ±3, 114 ±3, and > 200 CGG repeats, and yields a full mutation genotype (Figure 1).
Image /page/3/Figure/7 description: The image shows a graph with Fragment Size (bp) on the x-axis and RFU on the y-axis. There are several peaks on the graph, labeled with their corresponding fragment sizes: 18, 30, 32, 56, 85, 114, and >200. The peaks represent the relative fluorescence units (RFU) at different fragment sizes. The graph spans from approximately 150 bp to 1200 bp.
Figure 1. An electropherogram of a positive control (PC)
For a valid run, the AmplideX Fragile X Reporter output will identify each of these alleles within the expected precision range (Table 2) for their respective CGG repeat length.
{4}------------------------------------------------
| CGG Repeat Range | Precision |
|------------------|-----------|
| 1-70 | $\pm$ 1 |
| 71-120 | $\pm$ 3 |
| 121-199 | $\pm$ 5% |
| ≥ 200 | N/A |
Table 2. CGG Repeat Size Precision
# Interpretation of Results
AmplideX Fragile X Reporter software provides automated analysis of the resulting electropherogram to identify and size FMR1 alleles. It then categorizes the genotype of the sample into one of the following categories based on the size of the repeats as a Normal (NOR), Intermediate (INT), Premutation (PM), or Full mutation (FM) as shown in Table 3:
| Genotype | Size (CGG) repeats |
|---------------|--------------------|
| Normal | 1-44 |
| Intermediate | 45-54 |
| Premutation | 55-199 |
| Full Mutation | $\ge$ 200 |
Table 3. FMR1 CGG repeat length ranges and their corresponding genotypes
Individuals with normal (< 45 CGG repeats) or intermediate (45-54 CGG) FMR1 alleles are currently thought to be asymptomatic for disorders associated with the FMR1 gene. However, individuals who are carriers of a premutation allele (55-199 CGG) are at risk for passing on a full mutation to their offspring. This risk is proportional to the number of CGG repeats and to the AGG status of the expanded allele. Individuals with the FMR1 full mutation (≥ 200 CGG) typically have FXS (Monaghan et al. 2013). Since fragile X syndrome is an X-linked disorder. only mothers can pass on a full mutation to male offspring. For females, either father or mother can pass on the full mutation.
Premutation carriers can develop fragile X-associated tremor/ataxia syndrome (FXTAS) or fragile X-associated primary ovarian insufficiency (FXPOI) (Wheeler, A et al. 2017, Gallagher, A et al 2012, and Hantash, F et al. 2011). A general illustration of the biological significance of the repeat length is shown in Figure 2.
{5}------------------------------------------------
Image /page/5/Figure/0 description: The image shows a diagram of the relationship between genotype, molecular defect, biochemical effect, and potential clinical outcome. The diagram shows four different genotypes: Full Mutation, Premutation, Intermediate, and Normal. The (CGG)n values for each genotype are: ≥ 200, 55-199, 45-54, and 1-44, respectively. The potential clinical outcomes for each genotype are: FXS, FXTAS/FXPOI, and Normal.
<sup>5&#</sup>x27;-UTR FMR1
Figure 2. FMR1 CGG repeat length ranges and their corresponding genotypes and outcomes.
# Mosaicism
Mosaicism can result when a new genetic alteration arise after an embryo is formed. Both size mosaicism and methylation mosaicism occur in fragile X syndrome. The kit may also detect low levels of genetic mosaicism in CGG repeat length. The AmplideX® Fragile X Dx & Carrier Screen Kit reports all detected FMR1 alleles for each sample and determines the associated genotype category based on the longest detected allele. The assay does not distinguish between mosaic and non-mosaic alleles. The kit is not designed to detect or report methylation status.
Due to the challenges associated with size mosaicism, (e.g., due to both accurate or false results) when the assay detects more than one (1) peak for a male or more than two (2) peaks for a female, users are informed in the instructions for use to carefully review the electropherogram; and recommended to confirm the presence of size mosaicism and considered in conjunction with other laboratory and clinical information (e.g., sex of patient, diagnostic testing or carrier screening, patient symptoms) and should include consultation with a genetic professional..
# L. Performance Characteristics:
# 1. Analytical performance:
- Precision (Repeatability and Reproducibility) a.
The precision of the AmplideX Fragile X Kit was assessed based on both agreement for the categorical call (i.e., normal, intermediate, premutation, and full mutation) and for precision of the size of the CGG repeats (i.e., measurements of fragile X allele lengths). Within the precision studies, specific target criteria around CGG repeat length were applied, as shown in Table 2 above.
#### i. Precision (3 site reproducibility)
Between-laboratory precision, or reproducibility, was assessed using a panel of 11 samples consisting of 7 clinical whole blood specimens and 4 contrived samples that collectively represent allele sizes spanning all FXS genotype categories, including those at or near category boundaries. Contrived samples were genomic DNA extracted from cell lines with known gender, genotype, and expected allele sizes, spiked into leukocyte-depleted whole human blood. Table 4
{6}------------------------------------------------
describes the panel members.
| Sample ID | Sex | Source | Expected Number<br>CGG Repeats | Genotype Category |
|-----------|--------|--------|--------------------------------|-------------------|
| ASGN-109 | (b)(4) | | | Normal |
| ASGN-112 | | | | Normal |
| ASGN-005 | | | | Intermediate |
| ASGN-111 | | | | Intermediate |
| ASGN-103 | | | | Premutation |
| ASGN-113 | | | | Premutation |
| ASGN-101 | | | | Premutation |
| ASGN-016 | | | | Premutation |
| ASGN-018* | | | | Full Mutation* |
| ASGN-023 | | | | Full Mutation |
| ASGN-104 | | | | Full Mutation |
Table 4. Sample Panel used in precision and reproducibility study
* ASGN-018 has a full-mutation mosaic allele at >200 repeats and therefore this is the call made by the software, a normal allele at 20 repeats, and a premutation allele at 197 repeats. Mosaic results that do not lead to categorical change not listed above.
The study was executed at three laboratories over 5 non-consecutive days utilizing a single kit lot, two operators per site, and a different 3500Dx instrument at each site. For each run, there were 6 PCR replicates per specimen. The study yielded 1980 total sample measurements excluding controls (3 sites x 2 operators x 6 replicates x 5 days = 180 replicates per sample). Of these, 30 replicates failed QC metrics (1.1% invalid rate) and were further evaluated for genotype category agreement and allele peak precision. The agreement was determined based upon the majority call.
The reproducibility of the assay for categorical call is shown in Table 5. For 10 of 11 samples, >97% genotype agreement was observed. Sample ASGN-005, ASGN-111 and ASGN-113 were not 100% due to the appearance of an extra allele which resulted in different categorical calls for 11 replicates because the assay software is designed to yield an interpretation based on the largest repeat fragment observed. These results were not observed for these samples in the lotto-lot precision study. ASGN-018 has a full-mutation mosaic allele at >200 repeats, a normal allele at 20 repeats, and a premutation allele at 197 repeats. The mosaic allele was not found in 81/179 replicates. A limitation statement in the instructions for use alerts users to carefully review and confirm mosaic results.
{7}------------------------------------------------
| Sample | Genotype<br>Category | Total<br>number of<br>Replicates<br>(of 180) | Number of<br>Replicates<br>in<br>Agreement | Agreement<br>(%) | CI (95%) |
|----------|----------------------|----------------------------------------------|--------------------------------------------|------------------|------------|
| ASGN-109 | Normal | 178 | 178 | 100 | 97.9, 100 |
| ASGN-112 | Normal | 178 | 178 | 100 | 97.9, 100 |
| ASGN-005 | Intermediate | 179 | 177 | 98.9 | 96, 99.7 |
| ASGN-111 | Intermediate | 175 | 170 | 97.1 | 93.5, 98.8 |
| ASGN-103 | Premutation | 178 | 178 | 100 | 97.9, 100 |
| ASGN-113 | Premutation | 179 | 177 | 98.9 | 96, 99.7 |
| ASGN-101 | Premutation | 174 | 174 | 100 | 97.8, 100 |
| ASGN-016 | Premutation | 178 | 178 | 100 | 97.9, 100 |
| ASGN-018 | Full Mutation | 179 | 98 | 54.7 | 47.4, 61.9 |
| ASGN-023 | Full Mutation | 177 | 177 | 100 | 97.9, 100 |
| ASGN-104 | Full Mutation | 175 | 175 | 100 | 97.9, 100 |
Table 5. Three sites reproducibility-Categorical Genotype Agreement by Sample Excluding Invalids.
The reproducibility of the assay for allele size (number of repeats) is shown in Table 6 for each allele in the sample. Reproducibility was based on identifying the mode for each repeat. Replicates results were considered to agree if they fell within the precision specifications described in Table 2. As noted in the reproducibility for categorical calls, 11 replicates had artifacts and are shown below. The reproducibility of the expected allele lengths with the exception of the alleles identified as "mosaic" showed >98% reproducibility based on the prespecified precision criteria. The samples with contaminants/artifacts are listed for information purposes.
Table 6. Three sites Reproducibility- Percent of Allele Replicates within Target Precision Range Excluding Invalids.
| Sample ID | Mode Expected<br>CGG Length<br>(± Target<br>Precision) | Measured<br>CGG<br>Length | No of<br>replicates<br>at that<br>size | Total<br>Number of<br>replicates | % within<br>Target<br>Precision<br>Range |
|-----------|-------------------------------------------------------------|---------------------------|----------------------------------------|----------------------------------|------------------------------------------|
| ASGN -109 | $30 \pm 1$ | 29 | 1 | 178 | 100.0 |
| | | 30 | 177 | | |
| ASGN-112 | $30 \pm 1$ | 30 | 177 | 178 | 100.0 |
| | | 31 | 1 | | |
| | $29 \pm 1$ | 29 | 177 | 178 | 100.0 |
| Sample ID | Mode Expected<br>CGG Length<br>( $\pm$ Target<br>Precision) | Measured<br>CGG<br>Length | No of<br>replicates<br>at that<br>size | Total<br>Number of<br>replicates | % within<br>Target<br>Precision<br>Range |
| | | 30 | 1 | | |
| | Contamination/<br>artifact | 34 | 1 | 1 | 0.6% |
| ASGN-005 | 29 $\pm$ 1 | 29 | 179 | 179 | 100.0 |
| | 45 $\pm$ 1 | 45 | 141 | | 100.0 |
| | | 46 | 38 | | |
| | Contamination/<br>artifact | > 200 | 2 | 2 | 1.1 |
| ASGN-111 | 29 $\pm$ 1 | 29 | 173 | 175 | 100.0 |
| | | 30 | 2 | | |
| | 50 $\pm$ 1 | 50 | 87 | 175 | 98.9 |
| | | 51 | 86 | | |
| | | 52 | 1 | | |
| | | 53 | 1 | | |
| | Contamination/<br>artifact | 74<br>> 200 | 1<br>4 | 5 | 2.9 |
| ASGN-103 | 30 $\pm$ 1 | 30 | 177 | 178 | 100.0 |
| | | 31 | 1 | | |
| | 56 $\pm$ 1 | 55 | 81 | 178 | 100.0 |
| | | 56 | 96 | | |
| | | 57 | 1 | | |
| ASGN-113 | 30 $\pm$ 1 | 30 | 179 | 179 | 100.0 |
| | 103 $\pm$ 3 | 102<br>103<br>104<br>105 | 23<br>82<br>73<br>1 | 179 | 100.0 |
| | Mosaic Result* | 84<br>88<br>89 | 1<br>7<br>7 | 15 | 8.4 |
| | Contamination/<br>artifact | >200 | 2 | 2 | 1.1 |
| | | | | | |
| ASGN-101 | 35 $\pm$ 1 | 35 | 174 | 174 | 100 |
| | 93 $\pm$ 3 | 92 | 30 | 174 | 100.0 |
| | | 93 | 132 | | |
| Sample ID | Mode Expected<br>CGG Length<br>(± Target<br>Precision) | Measured<br>CGG<br>Length | No of<br>replicates<br>at that<br>size | Total<br>Number of<br>replicates | % within<br>Target<br>Precision<br>Range |
| | | 94 | 12 | | |
| | Mosaic result* | 97 | 2 | 9 | 5.2 |
| | | 98 | 7 | | |
| | 18 ± 1 | 18 | 178 | 178 | 100.0 |
| | | 113 | 37 | | |
| | | 114 | 115 | | |
| ASGN-016 | 114 ± 3 | 115 | 23 | 178 | 99.4 |
| | | 116 | 2 | | |
| | | 119 | 1 | | |
| | | | | | |
| | 20 ± 1 | 20 | 179 | 179 | 100.0 |
| | 198 ± 5%<br>(10 CGG) | 195 | 12 | 179 | 100.0 |
| | | 196 | 20 | | |
| | | 197 | 57 | | |
| | | 198 | 65 | | |
| | | 199 | 22 | | |
| ASGN-018 | Mosaic result | > 200 | 98 | 109 | 60.1 |
| | | 184 | 1 | | |
| | | 183 | 1 | | |
| | | 182 | 2 | | |
| | | 181 | 6 | | |
| | | 180 | 1 | | |
| | Contamination/ | 30 | 1 | 2 | 1.1 |
| | artifact | 103 | 1 | | |
| | | | | | |
| ASGN-023 | > 200 | > 200 | 177 | 177 | 100.0 |
| | | | | | |
| ASGN-104 | 24 ± 1 | 24 | 175 | 175 | 100.0 |
| | > 200 | > 200 | 175 | 175 | 100.0 |
{8}------------------------------------------------
{9}------------------------------------------------
* Represents potential mosaic allele. Mosaic results may be below the limit of detection for mosaicism.
#### ii. Precision (lot-to-lot reproducibility)_
Within-laboratory precision was assessed running the same panel in the 3-site reproducibility study and described in Table 4. Each sample was tested in duplicate with 3 kit lots, 3 operators, and for over 12 non-consecutive days. The study was executed on 3 instruments and yielded 2376 total sample measurements excluding controls (2 replicates X 3 lots X 3 operators X 12 days = 216 per sample). Of the 2376 total sample measurements, 60 were invalid (Invalid rate 2.5%). Samples were evaluated for genotype category agreement and allele peak precision. The agreement was determined based upon the majority call. Excluding invalids, greater than >99%
{10}------------------------------------------------
categorical genotype agreement was observed for all samples as shown in Table 7.). Samples ASGN-112 and ASGN-016 were not 100% due to the appearance of third alleles (3 replicates) which resulted in different categorical calls. These results were not observed in the 3-site reproducibility result ASGN-018 has a full-mutation mosaic allele at >200 repeats, a normal allele at 20 repeats, and a premutation allele at 197 repeats. The mosaic allele was not found in 100/212 replicates. Allowing this sample to have a categorical call of a pre or full mutation results in 100% (212/212) agreement between replicates (95% CI: 98.2%-100%) and overall agreement of 99.9% (2313/2316, 95% CI: 99.6%-100%).In addition, as shown in Table 8, the reproducibility of the all alleles with the exception of the alleles identified as "mosaic" showed 100% reproducibility with the specifications). Precision across 3 lots was acceptable. Specimens with artifacts are listed for information purposes and calculated in the categorical section.
| Sample | Genotype<br>Category | Total<br>Number of<br>Replicates<br>from 216 | Replicates<br>in<br>Agreement | Agreement<br>(%) | CI<br>(95%) |
|----------|----------------------|----------------------------------------------|-------------------------------|------------------|-------------|
| ASGN-109 | Normal | 200 | 200 | 100 | 98.1, 100 |
| ASGN-112 | Normal | 214 | 213 | 99.5 | 97.4, 100 |
| ASGN-005 | Intermediate | 209 | 209 | 100 | 98.2, 100 |
| ASGN-111 | Intermediate | 213 | 213 | 100 | 98.2, 100 |
| ASGN-103 | Premutation | 213 | 213 | 100 | 98.2, 100 |
| ASGN-101 | Premutation | 212 | 212 | 100 | 98.2, 100 |
| ASGN-113 | Premutation | 208 | 208 | 100 | 98.2, 100 |
| ASGN-016 | Premutation | 213 | 211 | 99.1 | 96.6, 99.7 |
| ASGN-018 | Full Mutation | 212 | 112 | 52.8 | 46.1, 59.4 |
| ASGN-023 | Full Mutation | 212 | 212 | 100 | 98.2, 100 |
| ASGN-104 | Full Mutation | 210 | 210 | 100 | 98.2, 100 |
Table 7. Lot to lot reproducibility - Genotype Category Agreement by Sample Excluding Invalids.
Table 8: Lot to lot reproducibility - Percent of Allele Replicates within Target Precision Range Excluding Invalids.
| Sample ID | Mode Expected<br>CGG Length<br>( $\pm$ Target Precision) | Reported<br>CGG<br>Length | No of<br>replicates<br>at that size | Total<br>Number of<br>Replicates | % within<br>Target<br>Precision<br>Range |
|-----------|----------------------------------------------------------|---------------------------|-------------------------------------|----------------------------------|------------------------------------------|
| ASGN-109 | 30 $\pm$ 1 | 29<br>30 | 4<br>196 | 200 | 100.0 |
| ASGN-112 | 29 $\pm$ 1 | 28 | 2 | 214 | 100.0 |
| Sample ID | Mode Expected<br>CGG Length<br>(± Target Precision) | Reported<br>CGG<br>Length | No of replicates<br>at that size | Total<br>Number of<br>Replicates | % within<br>Target<br>Precision<br>Range |
| ASGN-101 | 30 ± 1 | 29 | 212 | 214 | 100.0 |
| | | 29 | 2 | | |
| | | 30 | 212 | | |
| | Contamination/<br>artifact | 114 | 1 | 1 | 0.05 |
| | 35 ± 1 | 35 | 212 | 212 | 100.0 |
| | 92 ± 3 | 92 | 187 | 212 | 100 |
| | | 93 | 24 | | |
| | | 94 | 1 | | |
| | Mosaic result* | 96 | 10 | 22 | 10.4 |
| | | 97 | 12 | | |
| | Contamination/<br>artifact | 30 | 1 | 3 | 1.4 |
| | | 55 | 1 | | |
| | | 79 | 1 | | |
| ASGN-005 | 29 ± 1 | 29 | 209 | 209 | 100.0 |
| | 45 ± 1 | 45 | 209 | | |
| ASGN-111 | 29 ± 1 | 29 | 213 | 213 | 100.0 |
| | 50 ± 1 | 50 | 208 | | |
| | | 51 | 5 | | |
| ASGN-103 | 30 ± 1 | 30 | 213 | 213 | 100.0 |
| | 55 ± 1 | 55 | 207 | | |
| | | 56 | 6 | | |
| ASGN-101 | 92 ± 3 | 92 | 187 | 212 | 100.0 |
| | | 93 | 24 | | |
| | | 94 | 1 | | |
| | 30 ± 1 | 30 | 208 | 208 | 100.0 |
| | 102 ± 3 | 101 | 5 | 208 | 100.0 |
| | | 102 | 123 | | |
| | | 103 | 70 | | |
| ASGN-113 | | 104 | 9 | | |
| | | 105 | 1 | | |
| | Mosaic result* | 87 | 25 | 45 | 21.6 |
| | | 88 | 20 | | |
| | 18 ± 1 | 17 | 21 | 213 | 100.0 |
| | | 18 | 192 | | |
| ASGN-016 | 113 ± 3 | 113 | 176 | 213 | 100.0 |
| | | 114 | 32 | | |
| | | 115 | 4 | | |
| | | 116 | 1 | | |
| | Contamination/<br>artifact | > 200 | 2 | 2 | 9.4 |
| ASGN-018 | 20 ± 1 | 19 | 3 | 212 | 100.0 |
| Sample ID | Mode Expected<br>CGG Length<br>(± Target Precision) | Reported<br>CGG<br>Length | No of<br>replicates<br>at that size | Total<br>Number of<br>Replicates | % within<br>Target<br>Precision<br>Range |
| | 196 ± 5% (10 CGG) | 20 | 209 | 212 | 100.0 |
| | | 195 | 19 | | |
| | | 196 | 112 | | |
| | | 197 | 60 | | |
| | | 198 | 17 | | |
| | | 199 | 3 | | |
| | | > 200 | 1 | | |
| | Mosaic allele | 178 | 2 | 131 | 52.8 |
| | | 179 | 6 | | |
| | | 180 | 5 | | |
| | | 181 | 5 | | |
| | | 182 | 1 | | |
| | | >200 | 112 | | |
| ASGN-023 | > 200 | > 200 | 212 | 212 | 100.0 |
| ASGN-104 | 24 ± 1 | 23 | 1 | 210 | |
| | | 24 | 209 | | 100.0 |
| | > 200 | > 200 | 210 | 210 | 100.0 |
{11}------------------------------------------------
{12}------------------------------------------------
* Represents potential mosaic allele. Mosaic results may be below the limit of detection for mosaicism.
#### iii. Precision (Mosaicism)
Mosaicism is well known for Fragile X and was observed in 6 panel members out of 11 in the precision study. These results were not present in 100% of the replicates. To assess the reproducibility at the claimed limit of detection for mosaicism, a mosaicism reproducibility/LOD study was conducted with minimal DNA input (see below in the Detection of Limit Section).
A sample panel representing different mosaic/major allele combinations was generated by mixing DNA from 7 clinical specimens of different repeat sizes. The targeted MAFs are at or higher than the estimated LoD. The samples were tested at the lowest total DNA input level (20 ng). Note, one clinical specimen used to create the panel was later determined to be an outlier due to reproducibly unusual peak pattern. Subsequently, panel members involving this specimen were removed from data analysis. Each panel member was tested by 3 lots of AmplideX Fragile X Dx & Carrier Screen Kit in duplicates by 2 operators across 3 days. The study was executed on 2 instruments (one 8-capillary and one 24capillary) and vielded a total of 36 replicate measurements per specimen (2 replicates x 2 operators x 3 lots x 3 days = 36 replicates per sample). Greater than 95% hit rate were demonstrated for the claimed LoD listed in Table 9.
{13}------------------------------------------------
| Allele Combination | | LOD<br>(%MAF) | Valid<br>Replicates<br>(n) | %<br>Detected | 95% CI |
|-----------------------|-----------------------|---------------|----------------------------|---------------|------------|
| Mosaic | Major | | | | |
| Intermediate | Normal | 2.0 | 34 | 100.0 | 90.0-100.0 |
| Premutation | Normal | 2.0 | 36 | 94.4 | 82.0-98.0 |
| Full Mutation | Normal | 6.1 | 36 | 100.0 | 90.0-100.0 |
| Full Mutation | Premutation<br>(Low) | 7.0 | 36 | 100.0 | 90.0-100.0 |
| | Premutation<br>(High) | 7.0 | 35 | 100.0 | 90.0-100.0 |
| Premutation<br>(Low) | Premutation<br>(High) | 7.0 | 36 | 100 | 90.0-100.0 |
| Premutation<br>(High) | Full Mutation | 10.0 | 36 | 100 | 90.0-100.0 |
| Premutation<br>(Low) | Premutation<br>(High) | 5.0 | 34 | 97.1 | 85.0-99.0 |
| | | | 36 | 100.0 | 90.0-100.0 |
| Premutation<br>(High) | Premutation<br>(Low) | 5.0 | 36 | 100.0 | 90.0-100.0 |
Table 9. Reproducibility of Mosaic Samples
#### iv. DNA extraction reproducibility/equivalence
Three common commercial DNA extraction method for extracting genomic DNA from whole human blood were assessed using a 5-member sample panel representing all FMR1 genotype categories. The methods evaluated included: (D) (4) solution precipitation, manual silica spin column, and manual magnetic bead. Each extraction method was performed times per panel member. The extracted DNA was amplified in duplicate and analyzed on a single instrument.
All three methods yielded sufficient concentrations of DNA (on average) to meet the recommended input (b)(4) for the test kit. Categorical genotype agreement was 100% across all samples and 100% of all measured alleles were within target precision range. These results support the claim that the test can be used on gDNA generated using generally available DNA extraction methods capable of yielding at least 20ng/uL and A260-280 1.6-2.0.
- Thermal Cycler equivalence V.
A 5-member sample panel representing all FMR1 genotype categories was amplified on three separate thermal cyclers and then sized on a single CE instrument. Four PCR runs were tested on each thermal cycler. In each PCR run
{14}------------------------------------------------
four replicates of each sample were tested.
For all sample replicates, across and within all thermal cvcler types, genotype category calls were in 100% agreement. 100% of all identified independent alleles were within their respective target precision ranges. The study results support the use of commonly available commercial thermal cyclers that can run the specified protocol for the test kit.
- b. Linearity/assay Reportable Range:
Not applicable.
- c. Detection Limit
- i. Limit of Detection
The FMR1 gene is present on the X chromosome with one copy in males and two copies in females. The detection of the repeats is influenced by the DNA concentration and the size of the repeats. The "Limit of Detection" for detecting minor or mosaic alleles was assessed in combination with the reproducibility study.
The limit of detection (LoD) for mosaic alleles in different mosaic/major allele combinations were first estimated by testing mixed cell line DNA with known genotypes at 6 mosaic allele frequencies (MAF 10, 5, 2, 1, 0.1, and 0.01% genotypes at 6 mosaic allele frequencies (MAF 10, 5, 2, 1, 0.1, and 0.01%). The
allele frequency demonstrating >000 correct categorical call rate within befor
replicates were replicates were identified. The estimated LoD were then confirmed by testing a panel of 30 samples representing different mosaic/major allele combinations. The panel was generated by mixing DNA from clinical specimens of different repeat sizes with targeted MAFs at or higher than the estimated LoD. The samples were tested at the lowest total DNA input level (20 ng). Note, one clinical specimen used to create the panel was later determined to be an outlier due to an unusual peak pattern. Subsequently 6 panel members involving this specimen were removed from data analysis. Each panel member was tested by 3 lots of AmplideX Fragile X Dx & Carrier Screen Kit in duplicates by 2 operators across 3 days. The study was executed on 2 instruments (one 8-capillary and one 24capillary) and yielded a total of 36 replicate measurements per specimen (2 replicates X 2 operators x 3 lots x 3 days = 36 replicates per sample). Greater than 95% hit rate were demonstrated for the claimed LoD listed in Table 10.
Table 10. Limit of Detection for mosaic alleles in different major allele backgrounds
| Mosaic Allele | Major Allele | LoD for Mosaic Allele (%MAF) |
|---------------|--------------|------------------------------|
| Intermediate | Normal | 2.0 |
| Premutation | Normal | 2.0 |
| Full Mutation | Normal | 6.1 |
{15}------------------------------------------------
| Normal | Premutation | 2.0 |
|---------------|---------------|------|
| Premutation | Premutation | 5.0 |
| Full Mutation | Premutation | 7.0 |
| Normal | Full Mutation | 2.0 |
| Premutation | Full Mutation | 10.0 |
#### ii. DNA Input
The DNA input range for the assay is 20-80ng total (or 10-40 ng/uL). A study was conducted to assess a range of DNA input to span (6) fold below and " fold higher than this range to determine the robustness of the assay to imprecision of DNA measurements that might occur in a laboratory. DNA concentration and purity were quantified using standard spectrophotometric methods. Input levels were established using a panel of 8 samples (6 clinical samples and 2 contrived), representing all FMR1 genotype categories, with DNA mass inputs corresponding to 1, 10, 20, 40, 80, and 160 ng per reaction. Each reaction was performed in duplicate. The study was executed by a single operator using two kit lots, and generated 20 replicates per dilution level per sample, resulting in 160 replicates per dilution level, and a total of 960 sample measurements excluding controls.
Categorical genotype agreement was 100% for all valid samples and all input levels. All expected allele peaks were within the target precision range. The OC failure rate at 1 ng, 10 ng, 20 ng, 40 ng, 80 ng, and 160 ng were 6.25%, 1.88%, 0.6%, 1.3%, 0%, and 0.63% respectively. These results support the recommended input range of 20-80 ng of gDNA for the assay as determined by standard absorbance or spectrophotometric methods.
### d. Traceability
The assay uses molecular standards for use with fragment analysis on the instrument, and positive control that includes a range of CGG repeats. The assay is not traceable to any international standard, however an assessment of the accuracy of the molecular ladder used in the assay was conducted by testing the international standards. Refer to Clinical Section below.
### e. Interfering Substances
### Endogenous Interference
Endogenous interference was established by testing the interferents in accordance with CLSI EP07-A2 guidelines on a panel of "" samples ("" clinical and(b)(4) contrived). Interferents including conjugated bilirubin, unconjugated bilirubin, triglycerides, hemoglobin, and EDTA were tested. Each interferent, at its appropriate concentration, was added directly to an aliquot of a blood sample. Sodium hydroxide (diluent for hemoglobin and EDTA) as well as water were also tested to observe any solvent effects. An aliquot of untreated blood (no interferents added) was also extracted. Each condition was tested in triplicate.
{16}------------------------------------------------
Effects on DNA concentration yield were observed with hemoglobin (reduced yield) and EDTA (enhanced yield). Technical issues with the interference included low volume, low yield, sample loss, high screening QC failure rate, and instrument deviations, leading to some samples not completing processing. A limitation regarding the impact of excess hemoglobin and EDTA on the assay failure rate is included in the instructions for use. For samples that could be processed, the genotype category agreement was > 95% and the number of alleles within the expected precision range was 100%.
rv-Over and Cross- Contamination
### g. Stability
- i. Specimen Handling Stability
The ability of the assay to utilize DNA extracted from whole human blood stored at 2 to 8 °C for up to 2 weeks was determined by testing a panel of 14 clinical specimens representing all FMR1 genotype categories. For all samples, triplicate extractions were performed using independent aliquots on the day each sample was received to generate a T0 measurement. For subsequent timepoints, duplicate extractions were done. Extracted samples were stored at 2 to 8 ℃ for the duration of the study. A single operator using a single kit lot executed the study.
Genotype category agreement was 100% across all timepoints for each sample and all expected allele peaks were within target precision for all timepoints across all samples. These results support the use of sample DNA extracted from whole human blood that has been stored for up to 2 weeks at 2 to 8 ℃.
- ii. DNA freeze/thaw Stability
DNA freeze/thaw stability was tested using a " contrived sample panel across (6)(4) freeze/thaw cycles. (b) (4) (D)(4) (4) This study was
executed by a single operator using a single lot of kit reagents and a single instrument.
Genotype category agreement was 100% across all timepoints for each sample
{17}------------------------------------------------
and all expected allele peaks were within target precision for all timepoints across all samples. These results support that extracted DNA is useable for at least (b)(4) freeze-thaw cycles to support (b) (4)
- PCR & Capillary Electrophoresis (CE) Reagent Product Stability iii. PCR &CE product stability was assessed in two phases using a 5-member sample panel (4 clinical + 1 contrived) representing all FMR1 genotype categories. The time first phase evaluated post-amplification DNA stability at 4 ℃ across points for up to "" days ((b)(4) hours). With " replicates per sample at each time point, a total of (b)(4) measurements were collected.
The second phase evaluated the on-instrument, ambient temperature stability of samples prepared for CE analysis. CE deck time stability was established by preparing a (b) (4)
(b) (4
For PCR product stability, categorical genotype agreement was 100% for all timepoints across all samples, and > 95% of all measured alleles were within target precision range. The study results demonstrated that PCR products generated by the assay were stable for up to 3 days when stored at 2 to 8 ℃.
For CE product study, categorical genotype agreement was 100% for all timepoints across all samples and 100% of all measured alleles were within target precision range. The study results further demonstrated that samples prepared for CE analysis were stable at ambient temperature for up to 24 hours.
#### iv. Real Time Kit Stability
The real time kit stability was assessed using a 5-member sample panel representing all FMR1 genotype categories. Three lots of reagents were stored at -30°C to -15°C(b) (4)
(b) (4)
(b) (4) The study is still ongoing for (b) (4) timepoints.
Categorical genotype agreement was 100% for all timepoints across all samples and 100% of all measured alleles were within target precision range. The study results support at least a 1-year storage stability at this time. The study will carry on until 25-month to support 24-month stability.
#### Freeze/Thaw Kit Stability V.
The freeze/thaw kit stability was assessed using the same 5-member sample panel oresenting all FMR1 genotype categories as in real time kit stability study.
{18}------------------------------------------------
Categorical genotype agreement was 100% for all timepoints across all samples and 100% of all measured alleles were within target precision range. The study results support the stability of the kit for up to eight freeze/thaw cycles.
- vi. Kit Shipping Stability
The kit shipping stability was assessed using the same 5-member sample panel representing all FMR1 genotype categories as in real time kit stability study. The temperature profile tested was based on the International Safe Transit Association (ISTA) 7D, 24-hour summer temperature profile. This profile (Table 11) was repeated for(b) (4) days to create the domestic and international shipping simulations, respectively.
| Table 11. Shipping simulation based on the ISTA 7D 24-hour "Summer Profile". | | | | | |
|------------------------------------------------------------------------------|---------------------|--|--|--|--|
| | CHILL THE COLL COLL | | | | |
Image /page/18/Figure/5 description: The image shows the title "Summer Profile [24-hr cycle]" and the text "All the kits tested across all the different shipping configurations showed 100% of their". The text (b)(4) is present twice in the image. The image appears to be a snippet of a larger document.
tested across all the different shipping configurations showed 100% of their measured alleles within the target precision range. In addition, all sample replicates showed 100% genotype category agreement.
# 2. Clinical Performance
Two multi-center clinical validation studies were conducted to validate a diagnostic claim and a carrier testing claim for the AmplideX Fragile X Dx & Carrier Screen Kit on the Applied Biosystems 3500 Dx series of capillary electrophoresis instruments.
# a. Diagnostic Performance
- i. Full Mutation
Diagnostic performance was assessed for Full Mutation classification using Southern
{19}------------------------------------------------
Blot as the reference method. Leftover specimens were obtained from patient samples representing the intended use population submitted for routine FMR1 5 UTR mutation testing across ? linical sites. Specimens were selected using a specific
protocol so as to minim ''e bias. Sixty-nine (%(4) specimens were enrolled into each of the three categories (full mutation, premutation, and normal/intermediate) based on available Southern Blot result, and then tested on the AmplideX Fragile X Dx & Carrier Screen Kit. A total of 207 specimens were evaluated, including 111 specimens from female patients and 96 from male patients. The results demonstrated agreement for full mutation positive percent agreement (PPA), 95.71%, negative percent agreement (NPA) 99.3%, and overall percent agreement (OPA) 97.6%. One of the 3 discordant samples was borderline sample, 194-repeat by AmplideX and >200 peak by Southern Blot.
The agreement between the AmplideX and Southern blot for full mutation are shown in Table 12.
| | Reference Method (Southern Blot) | | | |
|-----------------------------------------------|----------------------------------|---------------------|---------------------|-----------------|
| | | Positive (≥<br>200) | Negative (<<br>200) | Total |
| AmplideX Fragile X Dx<br>& Carrier Screen Kit | Positive<br>(≥ 200) | 67 | 1 | 68 |
| | Negative<br>(< 200) | 3* | 137 | 139 |
| | Total | 70 | 138 | 208 |
| | | Percent | Lower 95% CI | Upper 95%<br>CI |
| | PPA | 95.7 | 88.1 | 98.5 |
| | NPA | 99.3 | 96.0 | 99.9 |
| | OPA | 97.6 | 94.5 | 99.0 |
Table 12. Agreement comparing the AmplideX Fragile X Dx & Carrier Screen Kit with Southern Blot Analysis to identify full mutation alleles (≥ 200 CGGs).
*One of the 3 discordant samples was borderline sample, 194 repeat by AmplideX and >200 peak by Southern Blot.
- ii. Premutation vs. Normal or Intermediate assessment Agreement between Southern blot analysis for premutation and normal/intermediate were also compared to the Amplide X. The results demonstrated PPA 100%, NPA 97.1% and OPA 98.6%, all exceeded 95% and all two-sided Wilson score 95% confidence intervals were at or above 90%. The results are shown in Table 13.
{20}------------------------------------------------
| | Reference Method (Southern Blot) | | | |
|--------------------------------------------|----------------------------------|-----------------|--------------|--------------|
| | Positive (55-199) | Negative (< 55) | Total | |
| AmplideX Fragile X Dx & Carrier Screen Kit | Positive (55-199) | 69 | 2 | 71 |
| | Negative (< 55) | 0 | 67 | 67 |
| | Total | 69 | 69 | 138 |
| | | Percent | Lower 95% CI | Upper 95% CI |
| | PPA | 100.0 | 94.7 | 100.0 |
| | NPA | 97.1 | 90.0 | 99.2 |
| | OPA | 98.6 | 94.9 | 99.6 |
# Table 13. Premutation vs. Normal or Intermediate assessment comparing the AmplideX Fragile X Dx & Carrier Screen with Southern Blot Analysis
# b. Carrier Screening Performance
Carrier screening performance was assessed by determining the percent agreement between the AmplideX Fragile X Dx & Carrier Screen Kit and a reference method for categorization into Full Mutation/Premutation, Intermediate, or Normal genotypes. A total of (0)(4) specimens from females ((b)(4) years of age) were enrolled across "" clinical sites. (80) evaluable subjects were obtained.
Because Southern blot is not expected to have accurate sizing in the premutation range, an alternate orthogonal independently validated PCR assay (FMR1 Dual-PCR Test) was used as the comparator. The assay is comprised of two different sets of primers that target the FMR1 gene for amplification. The FMR1 Dual-PCR Reference Method result is the average of the longest CGG repeat counts from each of the two sets of primers. The results are shown in Table 14 and Table 15.
# Table 1. Classification results comparing the AmplideX Fragile X Dx & Carrier Screen with FMR1 Dual-PCR
| Reference Method ( <i>FMR1</i> Dual-PCR) | | | |
|---------------------------------------------|-------------------------|-----------------|-------|
| Premutation<br>/ Full<br>mutation (><br>54) | Intermediate<br>(45-54) | Normal<br>(<45) | Total |
{21}------------------------------------------------
| AmplideX Fragile X Dx & Carrier Screen Kit | Premutation / Full mutation | 68 | 10* | 1 | 79 |
|--------------------------------------------|-----------------------------|----|-----|----|-----|
| | Intermediate | 0 | 60 | 0 | 60 |
| | Normal | 0 | 0 | 68 | 68 |
| | Total | 68 | 70 | 69 | 207 |
* Note: Eight of these samples had allele peaks of 54 and 55 by the FMR1 Dual-PCR Reference Method and the AmplideX Fragile X Dx & Carrier Screen Kit, respectively. Note the established precision of the assay in this range is ±1 CGG repeat.
# Table 2. Agreement of the AmplideX® Fragile X Dx & Carrier with FMR1Dual-PCR
| Category | Percent<br>Agreement | Lower 95% CI | Upper 95% CI |
|--------------------------------|----------------------|--------------|--------------|
| Premutation /<br>Full mutation | 100.0 | 94.7 | 100.0 |
| Intermediate* | 85.7* | 75.7* | 92.1* |
| Normal | 98.6 | 92.2 | 99.7 |
* Note: When taking into account the assay precision and count the 8 borderline samples as concordance, the percent agreement of the Intermediate category is 97.1%, 95% CI 90.2-99.2%.
The agreement of Premutation/Full Mutation, Intermediate and Normal categories are 100%, 85.7%, and 98.6% with 95% confidence intervals of 94.7-100%, 75.7-92.1%, 100%, 2017/04/2016 9010-01-2019 11:20:00 am samples were borderline samples, 55
and 92.2-99.7%. Eight (8) of the (8)
A secondary objective was to compare methods of reporting CGG repeat length between the AmplideX test and the reference method. Method comparison was done using a Passing-Bablok regression analysis. The analysis yields a slope of 1.0 with a 95% confidence interval of 1.0-1.0 (i.e., no measurable deviation from the line of identity) and an intercept of 0 with a 95% confidence interval of 0.0-0.0. The AmplideX Fragile X Dx and Carrier Kit detected an additional peak with low signal strength for 3 samples which may not be consistently detected and annotated by an alternate analysis method as noted in Figure 3.
{22}------------------------------------------------
Image /page/22/Figure/0 description: The image is a scatter plot that compares the longest CGG length reported by an investigational device (AmplideX) with a reference method. The x and y axes both range from 50 to 150. The plot includes a Passing Bablok Regression Fit line, represented by the equation 0 + 1 * Method1, and an identity line. The Passing Bablok Regression Fit is based on a sample size of n=177.
#### Longest Allele in the 5-199 CGG range
Longest CGG length reported by reference method
# Figure 1. Scatter plot of the longest CGG repeat lengths reported from the investigational device (AmplideX) test (y-axis) compared to those reported by the FMR1 Dual-PCR Reference Method.
The regression analysis above truncated the data set at (b)(4) length due to the range. A contingency table for allele greater than (b) poor resolution in (b)(4) (b)(4) length is provided below to further demonstrate the agreement between AmplideX Fragile X Dx and Carrier Kit with FMR1 Dual-PCR.
#### Table 16: Dual FMR1 Ref PCR vs AmplideX Contingency Table (b)(4)
{23}------------------------------------------------
- c. Mosaic allele calling in clinical studies and further analysis.
# Accuracy of the mosaic alleles
Mosaicism was detected in 49 specimens by Amplide X in the two clinical studies. 27 out of 207 (13%) in the diagnostic study and 22 out of 207 (~11%) in the carrier screening study. The mosaic results were not reported for 35 of the specimens by the orthogonal PCR method. To determine the accuracy for calling mosaic results, the results of the orthogonal PCR method were manually evaluated for mosaicism. Upon manual examination of the 35 cases, 31 mosaic cases detected by Asuragen, were also detected by the orthogonal method. Three (3) are suspected to have been contaminated during PCR setup, and 1 was due to a capillary electrophoresis artifact.
To ensure an unbiased resolution of the mosaic results, 40 randomly selected samples evaluated. After eliminating the cases already concordant for mosaicism or evaluated in the discordant resolution above, there were 34 samples: of these 31 showed no evidence of mosaicism by either method and three vielded mosaic results by both methods. Of the 83 independent specimens, 4 were discordant after manual evaluation.
- d. Testing of the International World Health Organization (WHO) Standards
Accuracy was determined by running a panel of 5 samples (Table 17) from the WHO International Standard Fragile X Syndrome Reference Panel and the National Institute for Biological Standards and Control (NIBSC code: 08/158) (Hawkins et al. 2011). A single operator performed each reaction in triplicate across three different kit lots. The study yielded 135 sample measurements excluding control (b)(4) replicates per sample). Of these measurements, 134 passed QC metrics and were further analyzed to determine consensus in size ranges for the standards. The accuracy results for the WHO panel are summarized in Table 18.
| Sample ID | Gender | Genotype | Mean Repeat<br>Length (CGG) | Range of Repeat<br>Lengths (CGG) |
|-----------|--------|---------------|-----------------------------|----------------------------------|
| 07/120 | Female | Normal | 22, 31 | 19-24, 28-33 |
| 07/122 | Female | Premutation | 33, 113 | 30-36, 100-132 |
| 07/168 | Female | Full Mutation | 38, 346 | 33-41, 300-401 |
| 07/170 | Male | Full Mutation | 754 | 353-960 |
| 07/174 | Male | Premutation | 114 | 97-127 |
Table 17. The sample panel as characterized by the WHO study consortium
{24}------------------------------------------------
| Sample ID | Expected<br>CGG Range | Measured<br>CGG Length | Allele Count | Total<br>Alleles | Total<br>Measurements | % in<br>Expected<br>Range |
|-----------|-----------------------|------------------------|--------------|------------------|-----------------------|---------------------------|
| 07/120 | 19-24,<br>28-33 | 22 | 27 | 27 | 54 | 100.0 |
| | | 31 | 27 | 27 | | |
| | | 34 | 27 | 27 | | |
| 07/122 | 30-36,<br>100-132 | 102 | 1 | 27 | 109 | 96.3 |
| | | 103 | 23 | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.