Plenity is indicated to aid in weight management in overweight and obese adults with a Body Mass Index (BMI) of 25 - 40 kg/m², when used in conjunction with diet and exercise.
Device Story
Plenity is an ingested, transient, space-occupying device designed for weight management. Upon ingestion, the device material hydrates in the stomach to occupy space, potentially increasing satiety. It passes through the gastrointestinal tract naturally. The device is intended for use by adults with a BMI of 25-40 kg/m^2 as an adjunct to diet and exercise. Clinical benefits are assessed via percent total body weight loss. Safety considerations include potential gastrointestinal adverse events (obstruction, dilation, diarrhea, constipation, dehydration) and potential interactions with medications. The device requires prescription and is subject to specific labeling regarding usage, shelf life, and combination with lifestyle interventions.
Clinical Evidence
Evidence includes the pivotal GLOW trial (N=436, randomized, double-blind, sham-controlled, 24 weeks) and GLOW-EX extension study. Primary endpoint (3% TBWL superiority) not met; however, statistically significant difference in %TBWL observed (p=0.0007) and responder rate (≥5% TBWL) was 58.6% vs 42.2% (p<0.0001). Safety profile showed no significant difference in AE incidence between Plenity and Sham groups. No SAEs related to device. Bench testing confirmed hydration kinetics, elastic modulus, and disintegration properties.
Technological Characteristics
Materials: Porcine gelatin capsule, cellulose, citric acid, sodium stearyl fumarate, titanium oxide. Principle: Mechanical space-occupying hydrogel. Form factor: Oral capsule (0.75g/capsule). Non-sterile. Biocompatibility per ISO 10993-1. Shelf life supported by package integrity (ASTM D3078-02) and functional bench testing.
Indications for Use
Indicated for weight management in overweight and obese adults (BMI 25-40 kg/m²) as an adjunct to diet and exercise. Contraindicated in pregnancy and patients with history of allergic reaction to cellulose, citric acid, sodium stearyl fumarate, gelatin, or titanium oxide.
Regulatory Classification
Identification
This device is an ingested material that transiently occupies space in the stomach. The device passes from the body via the natural gastrointestinal tract.
Special Controls
In combination with the general controls of the FD&C Act, the ingested, transient, space occupying device for weight management and/or weight loss is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible for its intended use.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions for use, as follows:
(i) Performance bench testing in a simulated use model must evaluate device disintegration and device hydration state throughout the gastrointestinal tract;
(ii) Bioburden and moisture content assessments must evaluate device infection risk throughout the labeled shelf life; and
(iii) Performance data must support the shelf life of the device by demonstrating continued package integrity and device functionality over the labeled shelf life.
(3) Clinical performance testing must demonstrate the device performs as intended and evaluate the following:
(i) Weight change;
(ii) All adverse events, including obstruction, dilation, diarrhea, constipation, and dehydration; and
(iii) Interaction with representative medications.
(4) Physician and patient device labeling must state:
(i) The clinical benefit of the device as assessed by using percent total body weight loss;
(ii) Treatment must be offered in combination with diet and exercise;
(iii) Instructions on how to use the device as intended including how to avoid interaction with medication; and
(iv) The shelf life of the device.
Submission Summary (Full Text)
{1}------------------------------------------------
#### DE NOVO CLASSIFICATION REQUEST FOR PLENITY
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Ingested, transient, space occupying device for weight management and/or weight loss. This device is an ingested material that transiently occupies space in the stomach. The device passes from the body via the natural gastrointestinal tract.
#### NEW REGULATION NUMBER: 21 CFR 876.5982
CLASSIFICATION: Class II
PRODUCT CODE: QFQ
#### BACKGROUND
DEVICE NAME: Plenity
SUBMISSION NUMBER: DEN180060
DATE DE NOVO RECEIVED: November 15, 2018
#### SPONSOR INFORMATION:
Gelesis, Inc. 501 Boylston Street, Suite 6102 Boston, MA 02116
#### INDICATIONS FOR USE
Plenity is indicated to aid in weight management in overweight and obese adults with a Body Mass Index (BMI) of 25 - 40 kg/m², when used in conjunction with diet and exercise.
#### LIMITATIONS
The sale, distribution, and use of Plenity are restricted to prescription use in accordance with 21 CFR 801.109.
In the clinical study of the device, patients were required to use Plenity along with a weight management program of nutrition, diet, and exercise instruction.
Plenity should be taken as directed in the labeling to avoid adverse interaction with other oral medication.
{2}------------------------------------------------
Plenity is contraindicated for use under the following conditions:
- Pregnancy
- History of allergic reaction to cellulose, citric acid, sodium steary] fumarate, . gelatin, or titanium oxide
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS. PRECAUTIONS AND CONTRAINDICATIONS.
# DEVICE DESCRIPTION
Plenity is a porcine gelatin capsule that contains thousands of absorbent hydrogel particles (0.75 grams [g] per capsule); each particle is approximately the size of a grain of salt (see Figure 1). Plenity is non-systemic and works directly in the gastrointestinal (GI) tract. Plenity hydrogel is made from two natural ingredients, cellulose and citric acid, that form a three-dimensional matrix designed to occupy volume in the stomach and small intestine, to create a sensation of fullness. Plenity is provided non-sterile.
Image /page/1/Picture/6 description: The image shows several empty white capsules and a pile of white powder. The capsules are open, suggesting they are ready to be filled with the powder. The powder is granular and appears to be a pharmaceutical ingredient. The scene is set on a clean, metallic surface, likely in a laboratory or pharmaceutical setting.
Figure 1. Plenity capsules with hydrogel exposed
The capsules disintegrate in the stomach and release the Plenity particles, which can hydrate up to 100 times their original weight. When fully hydrated, the individual non-clustering Plenity particles from the 2.25 g/dose occupy about a quarter of the average stomach volume. The gel particles mix with ingested foods, creating a larger volume with higher elasticity and viscosity in the stomach and small intestine.
Plenity passes through the digestive system, maintaining its three-dimensional structure in the stomach and small intestine before breaking down in the colon. The water is then released and reabsorbed by the body. Plenity particles are eliminated through normal bowel movements; they are not absorbed by the body.
Patients consume three (3) capsules (2.25 g/dose) with water before both lunch and dinner. Plenity is supplied in double blister packs that, together, provide the two doses patients take
{3}------------------------------------------------
daily. Each individual blister pack holds a single dose of three (3) capsules. Seven (7) double blister packs are supplied in a weekly package.
### SUMMARY OF NONCLINICAL/BENCH STUDIES
Non-clinical/bench studies conducted on the Plenity device are summarized below.
### BIOCOMPATIBILITY/MATERIALS
The Plenity Device is classified as mucosal membrane contacting for repeat, prolonged contact during clinical use (> 24 hours, ≤ 30 days). In accordance with ISO 10993-1, Biological evaluation of medical devices, and FDA Guidance: Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", the following biocompatibility endpoints were assessed for the Plenity Device:
- Cytotoxicity ●
- Sensitization ●
- Irritation
- Oral Irritation
- 7-day Oral Systemic Toxicity ●
- . 90-day Oral Toxicity
- Material-Mediated Pyrogenicity ●
- Mutagenicty ●
- Toxicological Risk Assessment
Results support the biocompatibility of the Plenity Device.
#### PERFORMANCE TESTING - BENCH
The integrity and performance of the Plenity Device was evaluated with the nonclinical testing summarized in Table 1
| Test | Purpose | Method | Acceptance<br>Criteria | Results |
|-------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Nonclinical Performance Testing | | | | |
| Microbiology testing | Determine device<br>total bioburden<br>and assess for<br>specific enteric<br>and pathogenic<br>microbes to<br>confirm<br>acceptable<br>bioburden | Standard<br>microbiological<br>methods (ISO<br>4833-2(2013),<br>ISO 21527-<br>2(2008), ISO<br>16449-2(2010),<br>ISO 6579-<br>1(2008), ISO<br>6888-3(2004)) | Various microbial<br>levels | Pass |
| Test | Purpose | Method | Acceptance<br>Criteria | Results |
| Water activity | Determine the<br>relationship<br>between water<br>activity and<br>moisture content | Relative humidity<br>measurements<br>under equilibrium | None | Water activity at<br>(b) (4) moisture<br>content was<br>(b) (4) |
| Media Uptake Ratio<br>(MUR) | Determine the<br>hydration capacity<br>of the hydrogel<br>particles | Weight of the<br>hydrated<br>hydrogel<br>particles in<br>simulated gastric<br>fluid, minus the<br>weight of the dry<br>hydrogel<br>particles, divided<br>by the weight of<br>the dry hydrogel<br>particles | (b) (4) | (b) (4) |
| MUR after<br>disintegration | Determine the<br>hydration capacity<br>of the hydrogel<br>particles after<br>gelatin capsule<br>disintegrates | Weight of the<br>hydrated<br>hydrogel particles<br>in simulated<br>gastric fluid,<br>minus the weight<br>of the dry<br>hydrogel<br>particles, divided<br>by the weight of<br>the dry hydrogel<br>particles | (b) (4) | (b) (4) |
| <i>In vitro</i> simulation of<br>device transit<br>through the GI tract | Ensure that<br>hydrogel particles<br>uptake and release<br>fluid at the correct<br>times during the<br>digestion process | Measure MUR<br>and elastic<br>modulus across<br>time in simulated<br>gastric, intestinal,<br>and colonic fluid | Hydration kinetics<br>and capacity must<br>adequately<br>correlate with GI<br>environment | Results are<br>acceptable |
| Moisture content<br>determination | Ensure that<br>moisture content<br>does not<br>unnecessarily<br>promote microbial<br>growth | Samples are<br>analytically<br>weighed and then<br>dried at(b) (4) for<br>minutes and<br>then weighed<br>again | (b) (4) | (b) (4) |
| Elastic modulus | Demonstrate that<br>hydrated hydrogel<br>has comparable<br>mechanical | Swollen particles<br>are put on<br>rheometer plates<br>and tested for | (b) (4) | (b) (4) |
| Test | Purpose | Method | Acceptance<br>Criteria | Results |
| | properties to food<br>in the stomach | rheological<br>properties | | |
| Disintegration time | Verify time that it<br>takes for the<br>capsule to<br>disintegrate in the<br>stomach | Capsules are<br>immersed and<br>visually checked<br>at pre-defined<br>time points | Dissolution time $\leq$<br>minutes at (b) (4) | Pass |
| Particle size<br>distribution | Confirm particle<br>dimensions | Testing is<br>conducted by<br>means of<br>analytical sieving | (b) (4) % of particles<br>are between (b) (4)<br>$\mu$ m (un-<br>hydrated particles) | (b) (4) |
| Bulk (tapped)<br>density | Support that the<br>correct amount of<br>hydrogel can be<br>encapsulated | Volume of<br>known weight of<br>powder is<br>measured (b) (4) | Hydrogel tapped<br>density (b) (4)<br>mg/ml | (b) (4) |
| Uniform hydration<br>and disintegration<br>test | Confirm absence<br>of clumps in<br>hydrated material | Hydrate material | Uniform hydration<br>without clumps | Pass |
| Packaging Integrity Testing | | | | |
| Package integrity<br>(simulated<br>distribution and<br>shipping followed by<br>associated package<br>integrity testing) | Validate<br>packaging in<br>environmental<br>conditions | Environmental<br>conditioning done<br>in accordance<br>with ASTM<br>F2825:2010<br>(2015); simulated<br>distribution in<br>accordance with<br>ASTM D4169-<br>16, followed by<br>inspection and<br>bubble emission<br>testing (ASTM<br>D3078-02) | Intact boxes and<br>non-leaking blister<br>packs | Pass |
| Performance tests on<br>packaged capsules | Ensure that<br>capsules can<br>remain intact and<br>do not<br>prematurely<br>hydrate while in<br>packaging | Pre-determined<br>test methods | Gelatin capsules<br>will visually look<br>unchanged,<br>capsules will<br>disintegrate in no<br>more than (NMT)<br>(b) (4) min, the make<br>ratio of the<br>powder in the<br>capsules will be no<br>less than (NLT) | Pass |
| Test | Purpose | Method | Acceptance<br>Criteria | Results |
| | | | (b) (4) the<br>elastic modulus of<br>the powder in the<br>capsules will be<br>NLT (b) (4) the<br>loss of drying of<br>the power in the<br>capsules will be<br>NMT (b) (4) | |
#### Table 1: Summary of Nonclinical Studies
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
# SHELF LIFE
# Table 2: Summary of Shelf-Life Testing
| 18-Month Real-Time Shelf-Life Testing | | | | | |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|------------------|--|
| Media Uptake Ratio<br>(MUR) | Determine the<br>hydration capacity<br>of the hydrogel<br>particles | Weight of the<br>hydrated<br>hydrogel particles<br>in simulated<br>gastric fluid,<br>minus the weight<br>of the dry<br>hydrogel<br>particles, divided<br>by the weight of<br>the dry hydrogel<br>particles | (b) (4) | 76.95 ± 0.48 g/g | |
| MUR after<br>disintegration | Determine the<br>hydration capacity<br>of the hydrogel<br>particles after<br>gelatin capsule<br>disintegrates | Weight of the<br>hydrated<br>hydrogel particles<br>in simulated<br>gastric fluid,<br>minus the weight<br>of the dry<br>hydrogel<br>particles, divided<br>by the weight of<br>the dry hydrogel<br>particles | (b) (4) | 73.81 ±0.79 g/g | |
| Moisture content<br>determination | Ensure that<br>moisture content<br>does not promote<br>microbial growth | Moisture loss on<br>drying method | (b) (4) % | (b) (4) % | |
| Elastic modulus | Demonstrate that<br>hydrated hydrogel<br>has comparable<br>mechanical | Swollen particles<br>are put on<br>rheometer plates<br>and tested for | (b) (4) Pa | (b) (4) Pa | |
{7}------------------------------------------------
| | properties to food<br>in the stomach | rheological<br>properties | | |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|------|
| Disintegration time | Verify time that it<br>takes for the<br>capsule to<br>disintegrate in the<br>stomach | Capsules are<br>immersed and<br>visually checked<br>at pre-defined<br>time points | Dissolution time $≤$<br>(b) (4) minutes at (b) (4) | Pass |
| Microbiology<br>testing | Confirm that<br>device is not<br>contaminated with<br>microbial species | Standard<br>microbiological<br>methods | Various microbial<br>levels | Pass |
| Package integrity<br>(blister pack bubble,<br>leak testing) | Evaluate the<br>packaging<br>integrity to ensure<br>that capsules<br>remain intact and<br>do not prematurely<br>hydrate throughout<br>the shelf-life | Testing done in<br>accordance with<br>ASTM D3078-02<br>Standard Test<br>Method for<br>Determination of<br>Leaks in Flexible<br>Packaging by<br>Bubble Emission | Seal intact | Pass |
# SUMMARY OF CLINICAL INFORMATION
Clinical data from a pivotal study (GLOW), small extension study (GLOW-EX), gastric emptying study, and drug-device interaction study were leveraged to evaluate the safety and effectiveness of the Plenity Device.
# Pivotal Study (GLOW)
The Gelesis Loss Of Weight (GLOW) trial (ClinicalTrials.gov, NCT02307279) was a multicenter, randomized, double-blind, sham-controlled, parallel-group study assessing the safety and efficacy of 2.25 g of Plenity on body weight over 24 weeks at 33 study sites (US and Europe) in 436 overweight and obese subjects (with and without type 2 diabetes). All subjects were prescribed reduced caloric intake and exercise. Enrollment included subjects ages 22-65 and with BMI 27-40 kg/m2. Those with BMI <30 kg/m² needed to have at least one of the following comorbidities: type 2 diabetes, dyslipidemia, or hypertension. Fasting glucose was required to be between ≥90 mg/dL and ≤145 mg/dL (≥5.0 mmol/L and ≤8.1 mmol/L). The sham used in this study was a placebo consisting of capsules filled with raw cane sugar.
# Endpoints
The intent-to-treat (ITT) population is the set of all randomized subjects. The ITT-multiple imputation (ITT-MI) population was the primary group analyzed for primary and secondary endpoints and includes all randomized subjects with multiple imputation performed for missing primary and secondary endpoint data. The ITT-observed (ITT-Obs) population is the set of all randomized subjects who completed the study. The ITT-Obs population was used to analyze the tertiary endpoints. The per protocol (PP) population is the ITT-Obs population but excludes subjects with major protocol deviations or < 80% compliance with taking the dispensed Plenity capsules.
{8}------------------------------------------------
### Primary endpoints
The co-primary effectiveness endpoints looked at the change in body weight from baseline to Day 171 (week 25).
- . Superiority margin of 3% of the percent total body weight loss (%TBWL) for the Plenity group compared to the Sham group
- More than 35% of subjects on Plenity achieving at least 5% TBWL (performance goal) .
### Secondary endpoints
- Body weight in subjects with impaired plasma glucose status at baseline (percent change ● from baseline to Day 171);
- Plasma glucose status (normal, impaired, diabetic) in subjects with impaired plasma ● glucose status at baseline (change from baseline to Day 171);
- Plasma glucose in subjects with impaired plasma glucose status at baseline and T2D ● (percent change from baseline to Day 171);
- BMI (change from Baseline to Day 171); and ●
- HbA1c in subjects with T2D (change from baseline to Day 171). ●
### Assessed Tertiary endpoints
- Body weight responders (> 10% weight loss from baseline to Day 171); ●
- Estimated excess body weight (percent change from baseline to Day 171); ●
- Waist circumference (change from baseline to Day 171);
- Serum insulin (percent change from baseline to Day 171); ●
- Homeostasis model assessment-insulin resistance (HOMA-IR) (percent change from ● baseline to Day 171);
- Serum C-reactive protein (CRP) (change from baseline to Day 171);
- Serum TC (percent change from baseline to Day 171);
- Serum LDL-C (percent change from baseline to Day 171);
- Serum high-density lipoprotein-cholesterol (HDL-C) (percent change from baseline to Day 171);
- Serum TC/HDL-C ratio (change from baseline to Day 171);
- Serum TG (percent change from baseline to Day 171);
- Supine and standing SBP and DBP (change from baseline to Day 171);
- Satisfaction (overall impression on Day 169); and
- . Impact of weight on quality of life (IWOOL) (change from baseline to Day 171) at selected sites.
Safety endpoint
- All Adverse Events (AEs) and Serious AEs (SAEs), including Adverse Device Effects . (ADEs) and Serious ADEs (SADEs)
To assess safety, the following were conducted or measured:
- Full physical examination (excluding pelvic and rectal examination) ●
- Supine and standing SBP and DBP .
{9}------------------------------------------------
- Supine and standing heart rate
- . Hematology including hemoglobin, hematocrit, red blood cell count, reticulocyte count, white blood cell count with differential, and platelet count
- Blood chemistry including plasma glucose and serum sodium, potassium, chloride, ● calcium, phosphorous, magnesium, blood urea nitrogen, creatinine, uric acid, total cholesterol, LDL cholesterol, HDL cholesterol, triglycerides, total protein, albumin, total bilirubin, alkaline phosphatase, lactate dehydrogenase, alanine aminotransferase, aspartate aminotransferase, and gamma glutamyltransferase
# Study population demographics
The majority of subjects were Caucasian (369 [84.6%] subjects), obese Class I (BMI 30.0 - 34.9 kg/m²; 237 [54.3%] subjects), middle aged (48.0 ± 10.4 years, mean ± SD) and distributed fairly equally across gender (245 [56.2%] females, 191 [43.8%] males). Dyslipidemia was the most common co-morbidity (69.1% of subjects in the Plenity group: 72.3% of subjects in the Sham group). Most subjects had normal fasting plasma glucose (FPG) at baseline, some were classified as prediabetic (66 [29.6%] subjects in the Plenity group: 66 [31.0%] subjects in the Sham group), and few had type 2 diabetes (T2D) (21 [9.4%] subjects in the Plenity group; 25 [11.7%] subjects in the Sham group).
# Subject disposition
A total of 904 subjects were screened in the GLOW study. Of these subjects, 436 were randomized with 223 subjects assigned to the Plenity group and 213 subjects assigned to Sham (intent-to-treat (ITT) cohort). 112 subjects failed to complete the treatment phase citing personal reasons as the most common cause for withdrawal (51 [23%] subjects in the Plenity group and 61 [29%] subjects on Sham). A total of 15 out of 436 (1.1%) subjects withdrew from the study due to adverse events (8 out of 223 [3.6%] subjects in the Plenity group and 7 out of 213 [3.3%] subjects on Sham). A total of 56 out of 436 (12.8%) subjects withdrew from the study based on lifestyle reasons or other personal choices (22 out of 223 [9.9%] subjects in the Plenity group and 34 out of 213 [16%] subjects on Sham). A total of 324 subjects (172 Plenity and 152 Sham) completed the entire treatment phase.
# Safety results
Subjects who received treatment after randomization were included in the analysis of safety (n = 223 for Plenity and n = 211 for Sham). Overall, there were 540 (64.3%) mild AEs (282 in 124 [55.6%] subjects in the Plenity group and 258 in 117 [55.5%] subjects in the Sham group) and 276 (32.8%) moderate AEs (143 in 88 [39.5%] subjects in the Plenity group and 133 in 83 [39.3%] subjects in the Sham group) (Table 3). The overall incidence of treatment-emergent adverse events in the Plenity treatment group was no different than Sham (71.3% versus 70.6%, respectively) (Table 4). In both groups, most (>95%) adverse events were assessed by the investigator as mild or moderate in intensity (Table 5). There were no serious adverse events (SAE) in the Plenity group, whereas there was one (1) non-device related SAE in the Sham group (colon adenoma). There were two cases of severe abdominal distension and severe nausea for which discontinuation of the device resulted in resolution. No deaths occurred during the trial.
Table 3: Summary of adverse events by treatment group – Safety Population
{10}------------------------------------------------
| | Plenity (N=223) | | Sham (N=211) | |
|------------------------------------------------------|-----------------|-----------------------------------------------|---------------|-----------------------------------------------|
| | Number Events | Number<br>Subjects with<br>Event<br>[% (n/N)] | Number Events | Number<br>Subjects with<br>Event<br>[% (n/N)] |
| Number of Subjects with<br>any AE | 436 | 71.3%<br>(159/223) | 404 | 70.6%<br>(149/211) |
| Grade 3 (Severe) | 11 | 3.6% (8/223) | 13 | 4.7% (10/211) |
| Grade 2 (Moderate) | 143 | 39.5% (88/223) | 133 | 39.3% (83/211) |
| Grade 1 (Mild) | 282 | 55.6%<br>(124/223) | 258 | 55.5%<br>(117/211) |
| Number of Subjects with<br>any SAE | 0 | 0.0% (0/223) | 1 | 0.5% (1/211) |
| Number of Subjects with<br>AEs leading to withdrawal | 29 | 3.6% (8/223) | 21 | 3.3% (7/211) |
| Death | 0 | 0.0% (0/223) | 0 | 0.0% (0/211) |
Table 4: Summary of Treatment-Emergent Adverse Events by Severity (≥5% by SOC in Either Treatment Group) by Preferred Term, and Severity—Safety Population
| | Plenity (N=223) | | Sham (N=211) | |
|----------------------------------------------------------------------------------------------|------------------|-----------------------------------------------|------------------|-----------------------------------------------|
| | Number | Number<br>Subjects with<br>Event<br>[% (n/N)] | Number | Number<br>Subjects with<br>Event<br>[% (n/N)] |
| All Adverse Events | 436 | 71.3%<br>(159/223) | 404 | 70.6%<br>(149/211) |
| Gastrointestinal disorders | | | | |
| Abdominal distension | 27 | 11.7%<br>(26/223) | 14 | 6.6% (14/211) |
| Mild | 20 | 8.5% (19/223) | 12 | 5.7% (12/211) |
| Moderate | 6 | 2.7% (6/223) | 2 | 0.9% (2/211) |
| Severe | 1 | 0.4% (1/223) | 0 | 0.0% (0/211) |
| Abdominal pain | 12 | 5.4% (12/223) | 6 | 2.8% (6/211) |
| Mild | 8 | 3.6% (8/223) | 5 | 2.4% (5/211) |
| Moderate | 4 | 1.8% (4/223) | 1 | 0.5% (1/211) |
| Constipation1 | 13 | 5.4% (12/223) | 11 | 5.2% (11/211) |
| Mild | 10 | 4.0% (9/223) | 6 | 2.8% (6/211) |
| Moderate | 3 | 1.3% (3/223) | 5 | 2.4% (5/211) |
| Diarrhea | 31 | 12.6%<br>(28/223) | 20 | 8.5% (18/211) |
| Mild | 19 | 7.6% (17/223) | 14 | 6.2% (13/211) |
| | Plenity (N=223) | | Sham (N=211) | |
| | | Number | | Number |
| | | Subjects with | | Subjects with |
| | Number | Event | Number | Event |
| | Events | [% (n/N)] | Events | [% (n/N)] |
| Moderate | 12 | 4.9% (11/223) | 5 | 1.9% (4/211) |
| Severe | 0 | 0.0% (0/223) | 1 | 0.5% (1/211) |
| Flatulence | 21 | 8.5% (19/223) | 14 | 5.2% (11/211) |
| Mild | 19 | 8.1% (18/223) | 14 | 5.2% (11/211) |
| Moderate | 2 | 0.4% (1/223) | 0 | 0.0% (0/211) |
| Infrequent bowel movements1 | 24 | 9.4% (21/223) | 12 | 4.7% (10/211) |
| Mild | 21 | 8.1% (18/223) | 9 | 3.8% (8/211) |
| Moderate | 3 | 1.3% (3/223) | 3 | 0.9% (2/211) |
| Nausea | 12 | 4.9% (11/223) | 12 | 5.2% (11/211) |
| Mild | 8 | 3.6% (8/223) | 9 | 3.8% (8/211) |
| Moderate | 3 | 0.9% (2/223) | 2 | 0.9% (2/211) |
| Severe | 1 | 0.4% (1/223) | 1 | 0.5% (1/211) |
| Infections and infestations | | | | |
| Nasopharyngitis | 31 | 11.7%<br>(26/223) | 37 | 14.2%<br>(30/211) |
| Mild | 25 | 9.0% (20/223) | 30 | 10.9%<br>(23/211) |
| Moderate | 6 | 2.7% (6/223) | 7 | 3.3% (7/211) |
| Upper respiratory tract infection | 9 | 3.6% (8/223) | 14 | 5.7% (12/211) |
| Mild | 8 | 3.1% (7/223) | 14 | 5.7% (12/211) |
| Moderate | 1 | 0.4% (1/223) | 0 | 0.0% (0/211) |
| Musculoskeletal and connective<br>tissue disorders | | | | |
| Arthralgia | 9 | 3.1% (7/223) | 13 | 6.2% (13/211) |
| Mild | 6 | 2.2% (5/223) | 4 | 1.9% (4/211) |
| Moderate | 3 | 0.9% (2/223) | 7 | 3.3% (7/211) |
| Severe | 0 | 0.0% (0/223) | 2 | 0.9% (2/211) |
| Nervous system disorders | | | | |
| Headache | 23 | 7.2% (16/223) | 26 | 8.5% (18/211) |
| Mild | 19 | 5.4% (12/223) | 12 | 3.8% (8/211) |
| Moderate | 3 | 1.3% (3/223) | 12 | 3.8% (8/211) |
| Severe | 1 | 0.4% (1/223) | 2 | 0.9% (2/211) |
| [1] Coded by investigator verbatim terms using Medical Dictionary for Regulatory Authorities | | | | |
| [MedDRA] version 17.1 | | | | |
| | Plenity (N=223) | | Sham (N=211) | |
| | Number<br>Events | Number<br>Subjects with<br>Event<br>[% (n/N)] | Number<br>Events | Number<br>Subjects with<br>Event<br>[% (n/N)] |
| Gastrointestinal Disorders [1] | 158 | 37.7%<br>(84/223) | 105 | 27.5%<br>(58/211) |
| Mild | 119 | 28.3%<br>(63/223) | 83 | 20.4%<br>(43/211) |
| Moderate | 35 | 8.1% (18/223) | 20 | 6.6% (14/211) |
| Severe | 4 | 1.3% (3/223) | 2 | 0.5% (1/211) |
| [1] Subjects with more than one AE are counted only once, at the worst severity. | | | | |
{11}------------------------------------------------
Table 5: Summary of Gastrointestinal AEs by Severity Deemed Possibly or Most Probably Related to Investigational Product - Safety Population
{12}------------------------------------------------
A sub-analysis at four study sites measured vitamin A. B1, B2, B12, B6, B9, D, and E levels at baseline, Day 85, and Day 171. No significant differences were noted from baseline measurements or from the Sham group. No significant difference was observed for serum electrolytes or hematocrit in either group. No signals of altered absorption of medications were observed based on thyroid stimulating hormone levels of subjects on thyroid replacement therapy, blood pressure management while on antihypertensives, low density lipoprotein cholesterol (LDL-C) while on lipid lowering agents.
# Efficacy results
The mean ± standard deviation (SD) %TBWL measured in kilogram [kg] from baseline to Day 171 for the ITT-MI population were -6.41 ± 5.79 and -4.39 ± 5.52 in the Plenity and Sham study groups, respectively. The adjusted mean ± standard error (SE) of difference in %TBWL from baseline to Day 171 (week 25) was -2.07 ± 0.59 (95% confidence interval (CI): -3.24, -0.90)
The difference did not meet the pre-defined threshold of 3% and the lower limit of the CI did not surpass 3%; therefore, super-superiority had not been established. However, while the difference in weight reduction did not meet the super-superiority threshold, the difference between the Plenity and Sham groups (-6.41 ± 5.79 vs. -4.39 ± 5.52) was statistically different (p = 0.0007). Table 6 provides the study results for the %TBWL in the ITT-MI and PP study populations. The ITT analyses could be influenced by the choice of statistical models, so the PP population analysis without statistical analysis is presented for transparency purposes.
| ITT-MI Analysis Population | Plenity (N=223) | Sham (N=213) |
|--------------------------------|----------------------|-----------------------|
| Percent TBWL [1] | | |
| Mean ± SD | -6.41 ± 5.79 | -4.39 ± 5.52 |
| Median (min, max) | -5.80 (-26.40, 7.74) | -3.97 (-22.31, 15.90) |
| LS Mean Difference [2] | | |
| Mean ± SE | -2.07 ± 0.59 | |
| 95% CI [3] | (-3.24, -0.90) | |
| p-value: Super Superiority [4] | 0.1193 | |
#### Table 6: Percent TBWL from baseline to Day 171
{13}------------------------------------------------
| ITT-MI Analysis Population | Plenity (N=223) | Sham (N=213) |
|-------------------------------------------------------------------------------------------------|----------------------|-----------------------|
| p-value: Superiority [5] | 0.0007 | |
| PP Analysis Population | Plenity (N=154) | Sham (N=141) |
| Percent TBWL | | |
| Mean ± SD | -6.31 ± 6.01 | -4.89 ± 5.40 |
| Median (min, max) | -5.73 (-26.40, 7.74) | -4.15 (-19.25, 10.42) |
| Difference (95% CI) | -1.42 (-2.73, -0.10) | |
| [1] Endpoint data imputed for 22.9% (51/223) in Plenity group and 28.6% (61/213) in Sham group. | | |
[2] Difference in adjusted means taken for comparability between the two groups.
[3] 95% Confidence Interval for the difference in LS means.
[4] p-value from ANCOVA model adjusted for stratification factors and baseline weight, testing for super superiority (> 3% difference).
[5] p-value from ANCOVA model adjusted for stratification factors and baseline weight, testing for superiority (difference > 0).
Of the 436 randomized subjects, 245 of them were from European study sites and 191 from US study sites. A pooling analysis of the pivotal study was done for treatment difference on the mean of the co-primary endpoint of percent TBWL in the ITT-MI population. The unadjusted difference in treatment mean (treatment - control) and the 95% CI of this difference is -2.39% (-3.94%, -0.84%) and -1.50% (-3.16%, 0.16%) for the European pooled region and the US pooled region respectively.
The co-primary endpoint of body weight responders was successfully achieved. The percent of responders with >5% weight loss was 58.6% (95% CI: 51.8. 65.4), significantly exceeding the performance goal of 35% (p<0.0001). Table 7 provides the study results for responder rates in the ITT-MI and PP study populations. The ITT analyses could be influenced by the choice of statistical models, so the PP population analysis without statistical analysis is presented for transparency purposes.
| ITT-MI Analysis Population | Plenity (N=223) | Sham (N=213) |
|------------------------------------------|-----------------|--------------|
| Percent of body weight responders [1, 2] | 58.6 | 42.2 |
| 95% CI | (51.8, 65.3) | (35.2, 49.1) |
| p-value [3] | <0.0001 | |
| PP Analysis Population | Plenity (N=154) | Sham (N=141) |
| Percent of body weight responders [2] | 57.1 | 44.0 |
| 95% CI | (48.9, 65.1) | (35.6, 52.6) |
| p-value [3] | <0.0001 | |
Table 7: Body Weight Responders > 5% TBWL from baseline to Day 171
[1] Endpoint data imputed for 22.9% (51/223) in Plenity group and 28.6% (61/213) in Sham group.
[2] Body weight responders defined as patients with ≥5% reduction in body weight.
[3] p-value from binomial proportion test for % of responders in treatment group compared to 35% performance goal.
A tipping point analysis was conducted for the performance goal of 5% weight loss and the 35% responder analysis. If all 51 missing values in the Plenity group were failures, there would be 100 (44.8%) 5% responders (95% CI: 38.2, 51.6). With all missing values set to failures, the study would have still shown success against the performance goal of 35%.
{14}------------------------------------------------
The first secondary effectiveness endpoint did not achieve statistical significance. Since the first secondary endpoint did not achieve p<0.05 significance level all the other secondary endpoints were evaluated only as descriptive statistics. Notably, at the Day 171 assessment, BMI means (SD) in the ITT-MI population were 31.43 (3.66) kg/m² and 32.57 (3.72) kg/m² for the Plenity and Sham treatment groups, respectively. The mean changes (SD) in BMI from baseline to Day 171 were -2.12 kg/m2 (1.92) and -1.51 kg/m² (1.90) for the Plenity and Sham treatment groups, respectively. The adjusted mean (SE) change in BMI from baseline to Day 171 for the treatment difference for Plenity versus Sham was -0.60 kg/m2 (0.20) (95% CI; -1.00, -0.20) from an ANCOVA model adjusting for stratification factors and baseline BMI.
The initial statistical testing of the tertiary endpoints did not achieve significance, and therefore all subsequent evaluations were based on descriptive statistics. Multiple prespecified tertiary endpoints were measured in the ITT-Obs population, including 10% total body weight responders, estimated excess body weight loss, and change in waist circumference (Table 8). All differences between treatment groups and 95% CI's shown are from analyses adjusting for stratification factors and the corresponding baseline value for the respective endpoint. For 10% body weight responders, the Plenity arm had 26% (45/172) while the Sham arm had 16% (25/152), with the odds of being a 10% responder in the Plenity arm 1.88 (95% CI; 1.07, 3.30) times the odds in the Sham arm. For excess body weight loss, the Plenity arm had -28.96 (30.14) percent change while the Sham arm had -20.98 (25.69) percent change. The patients in the Plenity arm achieved more excess body weight loss than those in the Sham arm, the adjusted difference between groups was -6.44 (2.94) (95% CL; -12.2, -0.64). Similarly, the patients in the Plenity arm achieved greater reduction in waist circumference than those in the Sham arm: - 2.64 inches (2.19) and -1.98 inches (2.32). respectively. The adjusted difference in waist circumference between the two groups was -0.73 in (0.25) with a 95% CI of (-1.22, -0.24).
| Tertiary Endpoints | Plenity<br>(N = 172) | Sham<br>(N = 152) | Difference [1] | 95% CI [2] |
|--------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|-------------------|----------------|-----------------|
| Estimated excess body weight<br>(% change) Mean ± SD | -28.96 ± 30.14 | -20.98 ± 25.69 | -6.44 ± 2.94 | (-12.23, -0.64) |
| Waist circumference (change in<br>inches) Mean ± SD | -2.64 ± 2.19 | -1.98 ± 2.32 | -0.73 ± 0.25 | (-1.22, -0.24) |
| [1] Difference in adjusted means taken for comparability between the two groups (T - C)<br>[2] 95% Confidence Interval for the difference in LS means. | | | | |
Table 8: Tertiary Endpoints for Change or Percent Change from baseline to Dav 171 - ITT-Obs Population
Patient perspectives considered for the Plenity during the review were GLOW study tertiary endpoints of subject satisfaction and Impact of Weight on Quality of Life (IWQOL). Measured satisfaction levels were increased from baseline and comparable between the treatment and control study arms.
# Extended Study (GLOW-EX)
{15}------------------------------------------------
This study assessed the effect of Plenity on body weight after an additional exposure of 24 weeks in subjects in the Plenity and Sham groups who completed the 24-week treatment period of the GLOW study and had at least 3% weight loss. Measurements were made on Day 339 from baseline (~48 weeks). This is an open-label, one arm extension study and all subjects were treated with Plenity 2.25 g, twice daily. The GLOW-EX study was conducted at 15 clinical sites in 5 countries in the US and Europe. Of the 15 clinical sites, 11 enrolled subjects into the study.
The following study groups were assessed:
- active-active (subjects who received Plenity in the GLOW study) .
- control-active (subjects who received the Sham in the GLOW study) .
Of the 324 subjects who completed the treatment phase of the GLOW study, 18 subjects who were assigned to Sham and 21 subjects who were assigned to Plenity were enrolled in GLOW-EX. Of these, 17 and 18 subjects, respectively, completed the extended treatment of the study.
# Endpoints
Co-primary endpoints:
- Body weight (percent change from baseline of the GLOW study to Day 339) ●
- Body weight (percent change from baseline of the GLOW-EX study to Day 339) ●
- Body weight responders (≥ 5% weight loss from baseline of the GLOW study to Day . 339)
- . Body weight responders (≥ 5% weight loss from baseline of the GLOW-EX study to Day 339)
# Assessed Secondary endpoints
- Plasma glucose (percent change from baseline of the GLOW study to Day 339) ●
- Plasma glucose (percent change from baseline of the GLOW-EX study to Day 339) ●
- HOMA-IR (percent change from baseline of the GLOW study to Day 339)
- HOMA-IR (percent change from baseline of the GLOW-EX study to Day 339)
- Body mass index (BMI) (change from baseline of the GLOW study to Day 339) ●
- BMI (change from baseline of the GLOW-EX study to Day 339) ●
# Assessed Tertiary endpoints
- Body weight responders (≥ 10% weight loss from baseline of the GLOW study to Day . 339)
- Body weight responders (≥ 10% weight loss from baseline of the GLOW-EX study to . Day 339)
- Estimated excess body weight (percent change from baseline of the GLOW study to Day ● 339)
- Estimated excess body weight (percent change from baseline of the GLOW-EX study to Day 339)
- Waist circumference (change from baseline of the GLOW study to Day 339) ●
- . Waist circumference (change from baseline of the GLOW-EX study to Day 339)
{16}------------------------------------------------
### Safety endpoint
- All AEs and SAEs (including ADEs and SADEs)
To assess safety, the following were conducted or measured:
- . Full physical examination (excluding pelvic and rectal examination)
- Supine and standing SBP and DBP
- Supine and standing heart rate
- Hematology including hemoglobin, hematocrit, red blood cell count, reticulocyte count, white blood cell count with differential, and platelet count
- Blood chemistry including plasma glucose and serum sodium, potassium, chloride, calcium, phosphorous, magnesium, blood urea nitrogen, creatinine, uric acid, total cholesterol, LDL cholesterol, HDL cholesterol, triglycerides, total protein, albumin, total bilirubin, alkaline phosphatase, lactate dehydrogenase, alanine aminotransferase, aspartate aminotransferase, and gamma glutamyltransferase
#### Safety results
Adverse events in the last 24 weeks (GLOW-EX) were compared to the first 24 weeks (GLOW study) for the overall safety population (n = 223) and for the GLOW-EX subjects only.
The overall incidence of AEs was no different between the Plenity subjects who were exposed for an entire vear (29 AEs in 10 [47.6%] subjects) versus the Sham subjects who were switched to Plenity and exposed to Plenity for the first time for a period of 24 weeks (37 AEs in 12 [66.7%] subjects). There were 11/29 AEs in 4 (19.0%) subjects deemed related (Possibly or Most Probably) to Plenity. Most of these were mild gastrointestinal (GI) events which were similar to what was observed in the first 24 weeks (GLOW study). There were no deaths and no SAEs.
The most common AEs (based on MedDRA's version 17.1) in descending order of frequency for the Plenity-Plenity group during GLOW-EX were:
- Infrequent bowel movements: 5 AEs in 3 (14.3%) subjects; .
- Headache: 4 AEs in 3 (14.3%) subjects; ●
- Abdominal distension: 2 AEs in 2 (9.5%) subjects; and ●
- Constipation: 2 AEs in 2 (9.5%) subjects.
The most common AEs (based on MedDRA's version 17.1) in descending order of frequency in the Sham-Plenity group during GLOW-EX were:
- Nausea: 3 AEs in 2 (11.1%) subjects: .
- Headache: 2 AEs in 2 (11.1%) subjects;
- Nasopharyngitis: 2 AEs in 2 (11.1%) subjects; and ●
- . Oropharyngeal pain: 2 AEs in 2 (11.1%) subjects.
#### Efficacv results
At the time of enrollment in GLOW-EX, the subjects treated with Plenity during GLOW (21 subjects) had already reached a mean ± SD %TBWL of 7.1% ± 2.8%. The additional six (6) months of exposure to Plenity provided an additional 0.8 ± 3.0% weight loss for a total of 7.9%
{17}------------------------------------------------
weight loss after 48 weeks. At entry in the GLOW-EX extension study, 15 of these 21 subjects had reached at least 5% weight loss, and, among that subset, 12 out of 15 maintained that weight threshold over the following 6 months. All 5 of the 21 subjects who had achieved the 10% threshold by entry into the GLOW-EX extension study maintained that threshold through 1 year with continuation of Plenity. The primary endpoint of weight maintenance with continuation of Plenity was demonstrated as the 95% confidence intervals at day 171 (-8.37 to -5.78) and day 339 (-10.54 to -5.22) overlap.
The 18 subjects assigned to Sham in GLOW had already reached a mean ± SD % TBWL of 7.1% ± 4.1% during the GLOW study before starting on Plenity for 24 weeks. These subjects, who successfully lost with lifestyle modification, were able to lose an additional 2.5% ± 4.1% over the subsequent 24 weeks with the addition of Plenity. At entry in the GLOW-EX extension study, 11 of the 18 subjects had reached at least 5% weight loss, and, among that subset, 10 out of these 11 subjects maintained their 5% TBWL status in the subsequent 24 weeks.
#### Gastric Emptying Study (IMAGES)
The IMAGES Study was a single center partially-blinded, randomized, four-way, crossover, two-part study involving eight overweight/obese male volunteers in each part designed to assess and compare the gastric emptying kinetics and gastrointestinal (GI) transit times of a radiolabeled meal following a single administration of either 2.25g or 3.75g of Plenity or Sham.
No differences were observed in the rate of gastric emptying between Plenity and Sham arms of the study.
#### Drug-Device Interaction Study
A drug-device interaction study was conducted to evaluate the effect of Plenity capsules on the pharmacokinetics (PK) of metformin, administered as single dose in healthy, overweight or obese subjects (N=24). This was a single center, randomized, single-dose, open-label. 4-period, 4-way, crossover, device-drug interaction, performed under fasting and fed conditions.
Results support that Plenity affects metformin absorption similar to food. Plenity should be administered as directed under the directions for use.
#### Pediatric Extrapolation
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
#### LABELING
The labeling comprises patient and physician labeling that includes the device indications for use, a description of the device, warnings and precautions, clinical data on the device, and instructions for the safe and effective use of the device. The labeling satisfies the requirements of 21 CFR 801.109 Prescription devices.
Per the Special Controls for this generic type of device, labeling includes the following:
{18}------------------------------------------------
- Safety instructions intended to minimize the risks of improper device use, including when to take the device with medication
- Contraindications and warnings to ensure usage of the device for the intended patient . population
- A detailed summary of the clinical testing including device effectiveness and device ● related adverse events and a summary of baseline demographic information
- A shelf life for the device. .
# RISKS TO HEALTH
The table below (Table 9) identifies the risks to health that may be associated with use of an ingested, transient, space occupying device for weight management and/or weight loss, and the measures necessary to mitigate these risks.
| Table 9: Identified Risks to Health and Mitigation Measures | |
|---------------------------------------------------------------|----------------------------------|
| Identified Risks to Health | Mitigation Measures |
| Device related gastrointestinal<br>adverse events, including: | Clinical performance testing |
| • Obstruction | Non-clinical performance testing |
| • Dilation | Labeling |
| • Diarrhea | Shelf life testing |
| • Constipation | |
| • Dehydration | |
| Weight gain | Clinical performance testing |
| | Labeling |
| Interaction with medication | Clinical performance testing |
| | Non-clinical performance testing |
| | Labeling |
| Adverse tissue reaction | Biocompatibility evaluation |
| Infection | Non-clinical performance testing |
| | Shelf life testing |
# Toble 0; Idantified Dicks to Haalth and Mitigation M
# SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the ingested, transient, space occupying device for weight management and/or weight loss is subject to the following special controls:
- (1) The patient-contacting components of the device must be demonstrated to be biocompatible for its intended use.
{19}------------------------------------------------
- (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions for use, as follows:
- (i) Performance bench testing in a simulated use model must evaluate device disintegration and device hydration state throughout the gastrointestinal tract.
- Bioburden and moisture content assessments must evaluate device infection risk (ii) throughout the labeled shelf life.
- (iii) Performance data must support the shelf life of the device by demonstrating continued package integrity and device functionality over the labeled shelf life.
- (3) Clinical performance testing must demonstrate the device performs as intended and evaluate the following:
- Weight change; (i)
- All adverse events, including obstruction, dilation, diarrhea, constipation, and (ii) dehydration; and
- (iii) Interaction with representative medications.
- (4) Physician and patient device labeling must state:
- The clinical benefit of the device as assessed by using percent total body weight (i) loss:
- Treatment must be offered in combination with diet and exercise; (ii)
- (iii) Instructions on how to use the device as intended including how to avoid interaction with medication; and
- (iv) The shelf life of the device.
#### BENEFIT-RISK DETERMINATION
The risks of the device are based on nonclinical laboratory studies as well as data collected in the clinical studies described above.
The overall incidence of adverse events in the pivotal clinical study (GLOW) in the Plenity treatment group was not different than Sham (71.3% versus 70.6%, respectively). The majority of the events were mild in nature. Only 3.6% of the events were considered severe. 39.5 % were moderate, and 55.6% were mild. There were slightly higher rates of GI-related AEs in the Plenity treatment arm relative to the Sham arm. The majority of GI AEs (37.7%) were device-related. The AEs were diarrhea, abdominal distension, flatulence, constipation, abdominal pain and nausea. This is expected for this device type. There were no serious adverse events (SAEs) or
{20}------------------------------------------------
deaths. There were, however, two cases of severe abdominal distension and severe nausea for which discontinuation of the device resulted in resolution. In the extended clinical study (GLOW-EX), the overall incidence of AEs was no different between the Plenity subjects who were exposed for an entire year (29 AEs in 10 [47.6%] subjects) versus the Sham subjects who had sham capsules for 24 weeks, and then Plenity for 24 weeks (37 AEs in 12 [66.7%] subjects).
The GLOW pivotal clinical study showed minimal weight loss with 2.25 g of Plenity daily relative to Sham therapy (sugar filled capsules) in patients who used the device for 171 days. Subjects taking Plenity lost on average 6.4% TBWL in subject taking the Sham sugar pill. Although the co-primary endpoint of a 3% TBWL superiority margin was not met, there was statistically significant greater %TBWL in the treated group versus the control group. The co-primary endpoint of percent responders with ≥ 5% TBWL was 58.6% with Plenity vs 42.2% with Sham and was of statistical significance. There were secondary and tertiary endboint analyses in the pivotal clinical study that showed some benefit of Plenity use.
An extended study (GLOW-EX) assessed the effect of Plenity on body weight after an additional exposure of 24 weeks in subjects who completed the 24-week treatment period of the GLOW study and had at least 3% TBWL (48 weeks total device exposure). Twenty-one (21) subjects taking Plenity kept taking Plenity, and 18 subjects taking Sham took Plenity during this extended study. Weight loss was assessed at 48 weeks (339 days) post baseline of the GLOW study.
The GLOW-EX enrolled subjects treated with Plenity had reached a mean ± SD %TBWL of 7.1% ± 2.8% at the end of the GLOW study. These subjects maintained that loss ending with 7.9% TBWL. The 18 subjects assigned to Sham in GLOW had lost 7.1% ± 4.1% TBWL during the GLOW study before starting on Plenity for 24 weeks. The subjects who successfully lost with lifestyle modification were able to lose an additional 2.5% ± 4.1% TBWL over the subsequent 24 weeks with the addition of Plenity.
There are several items that lead to uncertainty about the applicability of the study results across the broad US patient population: treatment withdrawal rate, diversity of study subjects, non-US subjects.
- The treatment withdrawal rate was 23% (51/223) for the Plenity arm and 29% (61/213) ● for the Sham arm of the study. The large amount of missing data made the primary ITT analysis more dependent on the statistical models used.
- The baseline parameters were appropriately balanced between the Plenity arm and the Sham arm; however, there was a bias toward Caucasian subjects with rates: 84.8% white, 11.7% African-American, and 4.9% Hispanic. The study population may not represent the patient population who are obese or overweight in the US.
- Of the 436 randomized subjects, 245 of them were from European study sites and 191 from US study sites. A pooling analysis of the pivotal study was done for treatment difference on the mean of the co-primary endpoint of percent TBWL in the ITT-MI population. The unadiusted difference in treatment mean (treatment - control) and the 95% CI of this difference is -2.39% (-3.94%, -0.84%) and -1.50% (-3.16%, 0.16%) for
{21}------------------------------------------------
the European pooled region and the US pooled region, respectively. Thus, patients in the US may see less benefit from the use of Plenity than may European patients.
The level of effectiveness of the Plenity device demonstrated through clinical studies is not appropriate for an indication of weight loss. However, the Plenity device has demonstrated effectiveness when used as an aid to weight management.
#### Patient Perspectives
Patient perspectives considered for the Plenity during the review were GLOW study tertiary endpoints of subject satisfaction and Impact of Weight on Quality of Life (IWQOL). Measured satisfaction levels were increased from baseline and comparable between the treatment and control study arms.
#### Benefit/Risk Conclusion
In the clinical study, one of the co-primary endpoints was met, demonstrating a modest clinical benefit in weight loss. There is moderate uncertainty in the benefit. However, considering the very low adverse event profile, the low level of uncertainty associated with the risks of the device, and the fact that the overall incidence of adverse events in the Plenity group were similar to the Sham group, the benefit-risk profile is favorable.
In conclusion, given the available information above, for the following indication statement:
Plenity is indicated to aid in weight management in overweight and obese adults with a Body Mass Index (BMI) of 25 - 40 kg/m², when used in conjunction with diet and exercise
the probable benefits outweigh the probable risks for the Plenity. The device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.
#### CONCLUSION
The De Novo request for Plenity is granted and the device is classified as follows:
Product Code: OFO Device Type: Ingested, transient, space occupying device for weight management and/or weight loss Regulation Number: 21 CFR 876.5982 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.