The Innovo is a transcutaneous electrical stimulator indicated for the treatment of stress urinary incontinence in adult females. The Innovo is indicated for prescription use only.
Device Story
Innovo is a transcutaneous electrical stimulator for treating stress urinary incontinence in adult females. Device uses cutaneous electrodes to apply external electrical stimulation to pelvic floor muscles. Operated by patient at home; prescription required. Stimulation parameters (waveforms, frequency, pulse duration, voltage/current) are controlled to induce muscle contraction. Intended to strengthen pelvic floor muscles to reduce incontinence. Healthcare providers use device output to manage patient treatment plans. Benefits include non-invasive symptom reduction.
Clinical Evidence
No clinical data provided in the document; classification based on non-clinical performance testing, software verification/validation, and risk-based special controls.
Technological Characteristics
Transcutaneous electrical stimulator; cutaneous electrodes. Characterized by electrical stimulation parameters: waveforms, output modes, max voltage/current, pulse duration, frequency, net charge per pulse, max phase charge at 500 ohms, max current density, max average current, and max average power density. Requires biocompatibility for patient-contacting materials. Subject to EMC, electrical, thermal, and mechanical safety testing. Software verification, validation, and hazard analysis required.
Indications for Use
Indicated for treatment of stress urinary incontinence in adult females.
Regulatory Classification
Identification
A transcutaneous electrical continence device consists of cutaneous electrodes that are used to apply external stimulation to reduce urinary incontinence.
Special Controls
In combination with the general controls of the FD&C Act, the transcutaneous electrical continence device is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must characterize the electrical stimulation, including the following: Waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density.
(2) The patient-contacting materials must be demonstrated to be biocompatible.
(3) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, thermal safety, and mechanical safety of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Labeling must include the following:
(i) Instructions for use, including specific instructions regarding the proper placement of electrodes;
(ii) A summary of electrical stimulation parameters; and
(iii) Cleaning instructions and reuse information.
Submission Summary (Full Text)
{1}------------------------------------------------
# DE NOVO CLASSIFICATION REQUEST FOR INNOVO
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Transcutaneous electrical continence device. A transcutaneous electrical continence device consists of cutaneous electrodes that are used to apply external stimulation to reduce urinary incontinence.
NEW REGULATION NUMBER: 21 CFR 876.5330
CLASSIFICATION: Class II
PRODUCT CODE: QAJ
#### BACKGROUND
DEVICE NAME: Innovo
SUBMISSION NUMBER: DEN170049
DATE DE NOVO RECEIVED: September 18, 2017
Bio-Medical Research Ltd. CONTACT: Parkmore Business Park West Galway H91 NHT7 Ireland
#### INDICATIONS FOR USE
The Innovo is a transcutaneous electrical stimulator indicated for the treatment of stress urinary incontinence in adult females.
The Innovo is indicated for prescription use only.
#### LIMITATIONS
The sale, distribution, and use of the Innovo are restricted to prescription use in accordance with 21 CFR 801.109.
The device is not intended for uses other than that described in the labeling.
{2}------------------------------------------------
# PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
# DEVICE DESCRIPTION
The Innovo is a single channel, rechargeable, non-implanted electrical stimulator that is intended for the treatment of stress urinary incontinence.
The Innovo is comprised of the following main components, along with accessories:
- Controller ●
- Body Garments (right and left) ●
- Gel Pads (8 surface electrodes) ●
- . Battery Charger
- Lead Wire ●
- . Neck Strap
Image /page/1/Picture/10 description: The image shows a pair of Innovo pelvic floor muscle stimulators and a controller. The stimulators are black and have the Innovo logo on them. One stimulator has the letter "R" on it, and the other has the letter "L" on it. The controller is blue and has a screen and buttons.
Figure 1: Innovo Controller and Body Garments
The Controller generates the electrical stimulation patterns for coupling the stimulation signals to the body when sued with the Gel Pad electrodes and lead wire. The Body Garments, equipped with 8 surface electrodes, are worn by the patient and cover the buttocks, lateral pelvis, and upper thighs. The electrodes have a skin conductive adhesive hydrogel laver, a current dispersing layer, and a garment conductive adhesive hydrogel layer.
The four electrodes on the right side are combined into a single equivalent electrode (and similarly as are the electrodes on the left side). The electrical stimulation current is passed across the pelvic area (from the right side to the left site), thereby stimulating the pelvic floor muscles.
| No. of Output Modes | 1 |
|-------------------------------------------|-------------------|
| Number of Output Channels | 1 |
| Regulated Current or Regulated Voltage | Regulated Current |
| Software/Firmware/Microprocessor Control? | Yes |
| | Table 1: General Features |
|--|---------------------------|
| | |
{3}------------------------------------------------
| Automatic Shut Off | Yes |
|-------------------------------------------|--------------------|
| Patient Override Control? | Yes (Pause Button) |
| Indicator Display - On/Off Status? | Yes |
| Indicator Display -Low Battery? | Yes |
| Indicator Display -Voltage/Current Level? | Yes |
| Timer range (minutes) | 30 minutes |
Table 2: Output Specifications
| Waveform | Pulsed, Symmetrical, Biphasic |
|-----------------------------------------------------------------------------|---------------------------------------|
| Shape | Rectangular, with interphase interval |
| Pulse Width | 620 μS |
| Baseline to peak current @ 500 Ω | 120 mA |
| Peak Voltage | (b) (4) |
| Frequency | 50 Hz |
| Maximum Current Density | (b) (4) @ 500 Ω |
| Maximum Power Density<br>(using smallest electrode conductive surface area) | (b) (4) @ 500 Ω |
| Maximum Phase Charge | (b) (4) |
| Net Charge | (b) (4)@ 500 Ω |
| Maximum Output Voltage (RMSV) (+/-10%) | (b) (4) |
| Maximum Output Current (RMSA) (+/-10%) | (b) (4) |
| Contraction Time | 5 seconds |
| Relaxation Time | (b) (4) econds |
| Treatment Session | 30 minutes, fixed |
| Maximum Charge Current | (b) (4) |
| Power Source | Battery Pack (7.2 V) |
The Innovo delivers a symmetric biphasic, current controlled waveform. The amplitude is modulated for the contraction cycle for the pelvic floor muscles. The clinician guides the user to set the stimulation intensity. A treatment session is fixed and lasts for 30 minutes. The Controller is powered by a pre-installed rechargeable battery pack. It has a mechanical interlock to prevent connections to a charger or USB cable during treatment. The Innovo is recommended for use for one treatment session per day for a minimum 12-week treatment plan.
### SUMMARY OF NONCLINICAL/BENCH STUDIES
# BIOCOMPATIBILITY/MATERIALS
The Innovo has components that have direct or indirect patient contact. The Gel Pads (external hydrogel electrodes), were previously evaluated for biocompatibility and
{4}------------------------------------------------
cleared under K000947. Biological safety report was provided for fabrics and ink used in the Body Garments. Other patient-contacting components of the device were evaluated for biocompatibility per ISO 10993-1 and tested as a surface device with limited (<24 hours) contact with intact skin with potential for repeated exposure since the device is reusable. From the evaluations and supporting information, the components of the Innovo were found to be biocompatible for its use.
# SHELF LIFE/STERILITY
The Innovo is provided non-sterile for single-person use and does not require any of the components to be sterilized by the end user. It is intended only for external use. The Gel Pads are disposable and can be replaced as needed. Cleaning instructions are provided in the Instruction Manual for safe handling and proper care of the device.
### ELECTROMAGNETIC CAPABILITY & ELECTROMAGNETIC SAFETY
The Innovo conformed to the following electromagnetic compatibility, electrical, mechanical, and thermal safety standards:
- IEC 60601-1: Medical Electrical Equipment; Part 1: General Requirements for ● Basic Safety and Essential Performance.
- IEC 62133: Secondary Cells and Batteries Containing alkaline or Other Non-acid ● electrolytes – Safety Requirements for Portable Sealed Secondary Cells, and for Batteries Made from Them, for use in Portable Applications.
- . IEC 60601-1-2: Medical Electrical Equipment; Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic compatibility - Requirements and Tests.
- IEC 60601-2-10: Medical Electrical Equipment; Part 2-10: Particular . Requirements for the Basic Safety and Essential Performance of Nerve and Muscle Stimulators.
- . IEC 60601-1-11: Medical Electrical Equipment; General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
### MAGNETIC RESONANCE (MR) COMPATIBILITY
The Innovo has not been tested for MR Compatibility and should not be used in an MRI suite.
### SOFTWARE
The Innovo Controller operates on software and allows the user to select the stimulation level and program via the use of the following: on/off/pause, increase or decrease intensity, program, information, up intensity, down intensity, mute and keylock. The Controller has a display screen to provide information for Intensity Level, Load Sense,
{5}------------------------------------------------
Program Number, Low Battery, Battery, Level, Pause Function, Recharge Status, and Countdown Timer/Treatment Time. The software for the Controller has a "moderate" level of concern, as discussed in the 2005 FDA "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices," and is addressed by supporting software documentation.
# PERFORMANCE TESTING - BENCH
All features and output specifications of the device, including those identified in Tables 1 and 2, were verified by individual pulse output waveform tracings for loads of 500, 2k, and 10k ohms, to simulate conditions that the device could encounter during use.
# SUMMARY OF CLINICAL INFORMATION
Clinical data for the Innovo primarily consisted of a randomized non-inferiority US pivotal study and a prematurely terminated German sham-controlled study. There was also a series of smaller feasibility studies conducted during the development of the device. The two clinical studies are briefly summarized below with a discussion of their findings.
# Study 1 - Sham Controlled Trial Germany:
Study Design: Randomized (1:1), double-blinded, sham-controlled study to evaluate the safety and performance of the Innovo device ("high dose") compared to a modified version ("low dose") of the device to be representative of sham treatment.
- Patient Population: The study reported on measures for 50 women clinically ● diagnosed with stress urinary incontinence (SUI).
- Co-Primary Endpoints: Reduction from baseline to 12 weeks in the 1-hour Pad Weight Test and improvement in Incontinence Quality of Life Questionnaire (iQoL).
- Secondary Endpoints: There were many secondary endpoints. Data was provided ● for the proportion of subjects with greater than 50% reduction in 24-hour Pad Weight Test.
Image /page/4/Figure/10 description: The image contains two bar charts comparing baseline and week 12 data for control and active groups. The left chart, titled "Post Intervention iQoL Score," shows iQoL scores for both groups at baseline and week 12. For the control group, the baseline iQoL score is around 53, while the week 12 score is approximately 59. The right chart, titled "1-HR Stress Test Score," displays urine leakage in grams for both groups at baseline and week 12. For the control group, the baseline urine leakage is about 32g, and the week 12 leakage is around 14g.
Figure 2: Primary Endpoint Results at Week 12
{6}------------------------------------------------
| Outcome Measure | Group | Baseline | Week 4 | Week 8 | Week 12 |
|--------------------------------------------------------------------|-----------|----------|--------|--------|---------|
| Significant<br>Improvement - 1hr<br>pad test (50% ↓ v<br>baseline) | Low Dose | 0% | 37% | 42% | 47% |
| | High Dose | 0% | 53% | 61% | 75% |
| Significant<br>Improvement- 24hr<br>pad test (50% ↓ v<br>baseline) | Low Dose | 0% | 25% | 42% | 41% |
| | High Dose | 0% | 26% | 56% | 44% |
Table 3: Secondary Endpoint Results
Study 1 compared the safety and effectiveness of the Innovo to a sham version of the device. The sham version of the device was designed to not treat stress incontinence from the contraction/strengthening of the pelvic muscles because the current path did not cross the pelvic floor region. The sham version of the device used a single electrode pair on one leg at a low level to illicit sensory response at a distal location on the patient's leg. which may have caused a significant difference in sensation between the study arms, thereby questioning whether blinding was effectively maintained in the study.
This study was prematurely stopped, which limited the treatment arm to N=24 subjects and control (sham) arm to N=26 subjects that completed 12 weeks of treatment. From the data available, both the active and sham groups had improved iQOL scores versus baseline after 12 weeks. The 1-hour pad weight test scores also showed improvement in both groups after 12 weeks compared to baseline. The one-hour pad weight test showed improvement in the percentage of subjects who experienced greater than 50% reduction in pad weight for the treatment over sham at 12 weeks (75% treatment versus 47% control). The 24-hour pad weight measures for treatment and control at 12 weeks showed little difference for the same outcome measure (44% treatment versus 41% control). Additional secondary endpoint data were not reported. This study had a smaller sample size than originally intended, but it provided some evidence that there is an effect of the device over the sham in the one-hour pad weight test.
# Study 2 - Randomized Non-Inferiority Study USA:
Study Design: Multicenter, randomized (1:1), non-inferiority clinical study involving 180 subjects at 12 US sites to assess the safety and performance of the Neurotech Vital Compact Device (Innovo device) compared to the iTouch Sure Pelvic Floor Exerciser (comparator device) for the treatment of stress urinary incontinence in females.
- Patient Population: Women clinically diagnosed with stress urinary incontinence. . All subjects scored 9 or less out of 18 for urge incontinence and were confirmed as having predominant stress urinary incontinence on the Medical Epidemiologic and Social Aspects of Aging Urinary Incontinence (MESA) Questionnaire.
{7}------------------------------------------------
- Primary Endpoint: Proportion of subjects with significant improvement (defined . as at least 50% reduction) in the Provocative Pad Weight Test at 12 weeks.
- . Follow-Up: After screening and baseline, follow-ups occurred in Week 1 (via telephone), Week 4, and Week 12 for evaluations.
| | Vital Compact<br>(N=89) | | itouch sure<br>(N=91) | |
|------------------------------------------------------|-------------------------|----------|-----------------------|----------|
| System Organ Class<br>Preferred Term | Subjectsa<br>n (%) | # Events | Subjectsa<br>n (%) | # Events |
| Any Adverse Event | | | | |
| Overall | 44 (49.4%) | 76 | 43 (47.3%) | 87 |
| Gastrointestinal disorders | | | | |
| Abdominal pain | 0 | 0 | 3 (3.3%) | 3 |
| Constipation | 0 | 0 | 2 (2.2%) | 2 |
| General disorders and administration site conditions | | | | |
| Chest pain | 1 (1.1%) | 1 | 2 (2.2%) | 2 |
| Medical device discomfort | 8 (9.0%) | 10 | 1 (1.1%) | 1 |
| Medical device pain | 4 (4.5%) | 4 | 1 (1.1%) | 1 |
| Infections and infestations | | | | |
| Bronchitis | 3 (3.4%) | 3 | 1 (1.1%) | 1 |
| Gastroenteritis viral | 1 (1.1%) | 1 | 2 (2.2%) | 2 |
| Sinusitis | 3 (3.4%) | 4 | 4 (4.4%) | 4 |
| Upper respiratory tract infection | 3 (3.4%) | 3 | 6 (6.6%) | 6 |
| Urinary tract infection | 2 (2.2%) | 3 | 4 (4.4%) | 4 |
| Vaginal infection | 1 (1.1%) | 1 | 3 (3.3%) | 3 |
| Vulvovaginal mycotic infection | 1 (1.1%) | 1 | 2 (2.2%) | 2 |
| Musculoskeletal and connective tissue disorders | | | | |
| Pain in extremity | 2 (2.2%) | 2 | 1 (1.1%) | 1 |
| Nervous system disorders | | | | |
| Headache | 0 | 0 | 2 (2.2%) | 2 |
| Psychiatric disorders | | | | |
| Depression | 2 (2.2%) | 2 | 0 | 0 |
| Respiratory, thoracic and mediastinal disorders | | | | |
| Cough | 1 (1.1%) | 1 | 2 (2.2%) | 2 |
| Oropharyngeal pain | 2 (2.2%) | 2 | 1 (1.1%) | 1 |
| Skin and subcutaneous tissue disorders | | | | |
| Skin irritation | 4 (4.5%) | 4 | 1 (1.1%) | 1 |
Table 4: Summary of Adverse Events by Treatment in >1 Subject per Treatment Group
{8}------------------------------------------------
| | Vital Compact<br>(N=89) | | itouch sure<br>(N=91) | |
|------------------------------------------------------|-------------------------|-------------|-----------------------|-------------|
| System Organ Class<br>Preferred Term | Subjectsa<br>n (%) | #<br>Events | Subjectsa<br>n (%) | #<br>Events |
| Any Adverse Event | | | | |
| Overall | 17 (19.1%) | 24 | 11 (12.1%) | 13 |
| Gastrointestinal disorders | | | | |
| Overall | 0 | 0 | 1 (1.1%) | 1 |
| Abdominal pain | 0 | 0 | 1 (1.1%) | 1 |
| General disorders and administration site conditions | | | | |
| Overall | 12 (13.5%) | 15 | 2 (2.2%) | 2 |
| Medical device discomfort | 8 (9.0%) | 10 | 1 (1.1%) | 1 |
| Medical device pain | 4 (4.5%) | 4 | 1 (1.1%) | 1 |
| Pain | 1 (1.1%) | 1 | 0 | 0 |
| Infections and infestations | | | | |
| Overall | 0 | 0 | 7 (7.7%) | 7 |
| Urinary tract infection | 0 | 0 | 3 (3.3%) | 3 |
| Vaginal infection | 0 | 0 | 2 (2.2%) | 2 |
| Vulvovaginal mycotic infection | 0 | 0 | 2 (2.2%) | 2 |
| Musculoskeletal and connective tissue disorders | | | | |
| Overall | 2 (2.2%) | 2 | 0 | 0 |
| Arthralgia | 1 (1.1%) | 1 | 0 | 0 |
| Myalgia | 1 (1.1%) | 1 | 0 | 0 |
| Renal and urinary disorders | | | | |
| Overall | 2 (2.2%) | 2 | 0 | 0 |
| Dysuria | 1 (1.1%) | 1 | 0 | 0 |
| Micturition urgency | 1 (1.1%) | 1 | 0 | 0 |
| Reproductive system and breast disorders | | | | |
| Overall | 0 | 0 | 1 (1.1%) | 1 |
| Vaginal discharge | 0 | 0 | 1 (1.1%) | 1 |
| Skin and subcutaneous tissue disorders | | | | |
| Overall | 5 (5.6%) | 5 | 2 (2.2%) | 2 |
| Erythema | 1 (1.1%) | 1 | 0 | 0 |
| Pruritus | 0 | 0 | 1 (1.1%) | 1 |
| Rash | 1 (1.1%) | 1 | 0 | 0 |
| Skin irritation | 3 (3.4%) | 3 | 1 (1.1%) | 1 |
Table 5: Summary of Device Related Adverse Events by Treatment >1 Subject by Treatment Group
The most common adverse events observed were pain, medical device discomfort, skin irritation, and urinary tract and vaginal infections. There were 24 adverse events in 17 subjects (19.1%) for the Innovo group and 13 events in 11 subjects (12.1%) for itouch sure group that were considered to be related to the device. Most of the adverse events were considered to be mild or moderate in both groups and were resolved by stopping the treatment and/or reducing the stimulation intensity of the device. Overall, the Innovo (identified as Vital Compact in the above table) had a low adverse event profile.
{9}------------------------------------------------
| Outcome Measure | Treatment Device | Control Device |
|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|------------------------------|
| Primary Endpoint: Proportion of Patients who attained significant improvement (50% reduction) in Provocative Pad Test<br>(95% CI) at Week 12 | | |
| ITT Population/Multiple Imputation | N=89<br>56.3% (45.4%, 66.8%) | N=91<br>63.0% (52.2%, 72.9%) |
| ITT Population/LOCF | N=89<br>58.4% (47.5%, 68.8%) | N=91<br>61.5% (50.8%, 71.6%) |
| Per Protocol Population/Observed Case | N=72<br>59.7% (47.5%, 71.1%) | N=70<br>70% (57.9%, 80.4%) |
| Key Secondary Endpoints: Change from Baseline at Week 12 (LOCF) | | |
| Provocative pad weight test (Mean $\pm$ SD) | -8.48 $\pm$ 25.053 | -9.66 $\pm$ 22.876 |
| 24-hour pad weight test (Mean $\pm$ SD) | -13.07 $\pm$ 21.531 | -9.89 $\pm$ 19.989 |
| Number of incontinence episodes/day (Mean $\pm$ SD) | -1.24 $\pm$ 1.564 | -1.43 $\pm$ 4.120 |
| Incontinence Quality of Life Questionnaire (I-QOL) | 13.41 $\pm$ 16.463 | 15.42 $\pm$ 18.376 |
| Number of pads used/day | -0.30 $\pm$ 0.998 | -0.44 $\pm$ 0.984 |
| Other Secondary Endpoints: Change from Baseline at Week 12 (Observed Case) | | |
| Dryness (<1g leakage on the provocative pad weight test) | 17 (19.1%)<br>(11.5%, 28.8%) | 29 (31.9%)<br>(22.5%, 42.5%) |
| Improvement on Global Impression of Improvement (result of<br>1, 2 or 3 indicating improvement after treatment) | 70.7% | 63.0% |
Table 6: Summary of Efficacy Outcomes
For the primary effectiveness endpoint of the proportion of patients that had significant (at least 50%) improvement in the provocative pad weight test, the Innovo treatment arm failed to demonstrate non-inferiority against the active comparator (iTouch Sure). The Innovo performed better than the active comparator for two of the secondary endpoints, the Global Impression of Improvement and the 24-hour pad weight test. Overall, the primary outcome of 56.3 percent of the patients experiencing a clinically significant (at least 50% reduction in the provocative pad weight test) is clinically meaningful. Moreover, all of the secondary outcomes showed evidence of clinical improvement.
# Clinical Results Discussion:
In Study 1, the Innovo device was superior to the sham control in a responder analysis for both the 1-hour pad weight test at 12 weeks (75% treatment versus 47% sham control) and the 24 -hour pad weight even though neither was statistically significant. While the Study 2 (US study) failed to demonstrate non-inferiority to an active control, this study did confirm a clinically meaningful (56.3%) rate of improvement from baseline for the provocative pad weight test at 12 weeks.
{10}------------------------------------------------
# Pediatric Extrapolation:
In this De Novo request, existing clinical information was not leveraged to support the use of the device in a pediatric patient population.
# LABELING
Labeling has been provided that includes instructions for use and an appropriate prescription statement as required by 21 CFR 801.109. The labeling includes:
- Instruction Manual: The manual is the primary labeling material for the device. It . provides information about the device and its components, indications, contraindications, precautions, warnings, possible adverse reactions, device functions, and guidelines for use. The manual includes instructions and diagrams that explain the user interface and describe where and how the device and its electrodes should be placed on the patient. The manual also includes cleaning instructions, care instructions for the included gel pads, and information about how the device can be washed and reused. Finally, the manual includes the electrical stimulation parameters of the device, in alignment with Table 2 above.
- Package Label: This provides sizing information, manufacturer information, and product ● summary.
# RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of the transcutaneous electrical continence device and the measures necessary to mitigate these risks.
| Identified Risks to Health | Mitigation Measures |
|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Pain or tissue damage due to<br>overstimulation | Non-clinical performance testing<br>Software verification, validation, and hazard analysis<br>Electrical safety testing<br>Labeling |
| Adverse tissue reaction | Biocompatibility evaluation |
| Electrical shock or burn | Electrical safety testing<br>Software verification, validation, and hazard analysis<br>Labeling |
| Device failure due to<br>electromagnetic interference | Electromagnetic compatibility (EMC) testing<br>Software verification, validation, and hazard analysis<br>Labeling |
| Use error that may result in<br>user discomfort, injury, or<br>delay in treatment | Software verification, validation, and hazard analysis<br>Labeling |
|--|
{11}------------------------------------------------
### SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the transcutaneous electrical continence device is subject to the following special controls:
- 1. Non-clinical performance testing must characterize the electrical stimulation, including the following: waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density.
- 2. The patient-contacting materials must be demonstrated to be biocompatible.
- 3. Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, thermal safety, and mechanical safety of the device.
- 4. Software verification, validation, and hazard analysis must be performed.
- 5. Labeling must include the following:
- a. Instructions for use, including specific instructions regarding the proper placement of electrodes;
- b. A summary of electrical stimulation parameters; and
- c. Cleaning instructions and reuse information.
#### BENEFIT-RISK DETERMINATION
The risks of the device are based on nonclinical testing and data collected in clinical studies described above. The probable risks with the use of the device include: pain, discomfort, skin irritation, urinary symptoms, and unintended injury (from electrical effects), all of which are minor. short-lived and reversible. Based on the available performance data, the probability of such harmful events is low, and occurrence can be managed or treated.
The probable benefits of the device are also based on nonclinical testing and data collected in clinical studies as described above. Probable benefits include improvement in symptoms related to stress urinary incontinence and improved quality of life.
#### Patient Perspectives
In Study 1 the treatment group had at least a 10-point improvement in the incontinence quality of life (iQOL) score at 12 weeks from baseline. In Study 2, the global impression of improvement was 70.7% for the Innovo at 12 weeks as compared to baseline.
#### BENEFIT/RISK CONCLUSION
In conclusion, given the available information above, for the following indication statement:
{12}------------------------------------------------
The Innovo is a transcutaneous electrical stimulator indicated for the treatment of stress urinary incontinence in adult females.
The Innovo is indicated for prescription use only.
The probable benefits outweigh the probable risks for the Innovo. The device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.
### CONCLUSION
The De Novo request for the Innovo is granted and the device is classified as follows:
Product Code: QAJ Device Type: Transcutaneous electrical continence device Regulation Number: 21 CFR 876.5330 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.