ClearLLab T1, ClearLLab T2, ClearLLab B1, ClearLLab B2, ClearLLab M

DEN160047 · Beckman Coulter · PWD · Jun 29, 2017 · Hematology

Device Facts

Record IDDEN160047
Device NameClearLLab T1, ClearLLab T2, ClearLLab B1, ClearLLab B2, ClearLLab M
ApplicantBeckman Coulter
Product CodePWD · Hematology
Decision DateJun 29, 2017
DecisionDENG
Submission TypeDirect
Regulation21 CFR 864.7010
Device ClassClass 2
AttributesReal-World Evidence, Pediatric

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
DEN160047 · Jun 29, 2017ClearLLab T1, ClearLLab T2, ClearLLab B1, ClearLLab B2, ClearLLab MBeckman CoulterResidual clinical specimens (whole blood, bone marrow, lymph node) from patients undergoing routine immunophenotypingThe study evaluated the diagnostic accuracy of the ClearLLab reagents by comparing results to the final clinical diagnosis (malignant vs. non-malignant) established at the clinical sites based on routine clinical testing and WHO guidelines.Residual clinical specimens; Retrospective clinical study; Diagnostic accuracy; Hematopoietic neoplasms

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Multi-center, retrospective study of ClearLLab reagents; Retrospective, multi-center diagnostic accuracy studyPatients with hematologically abnormal specimens (suspected hematopoietic neoplasia or non-malignant hematologic abnormalities); Sample Size: 279 specimens; Number of Sites: 4Clinical outcome (malignant vs. non-malignant) based on site's final diagnosis per WHO guidelinesSensitivity, specificity, PPV, and NPV for detecting abnormal phenotypes

Indications for Use

The ClearLLab reagents are intended for in vitro diagnostic use as a panel for qualitative identification of cell populations by multiparameter immunophenotyping on an FC 500 flow cytometer. These reagents are used as an aid in the differential diagnosis of hematologically abnormal patients having, or suspected of having the following hematopoietic neoplasms; chronic leukemia, acute leukemia, non-Hodgkin lymphoma, myeloma, myelodysplastic syndrome (MDS), and/or myeloproliferative neoplasms (MPN). The reagents can be used with peripheral whole blood collected in EDTA (K2 and K3EDTA), Acid Citrate Dextrose or Heparin, bone marrow collected in K2EDTA, ACD, or Heparin, and lymph node specimens. Interpretation of the results should be confirmed by a pathologist or equivalent professional in conjunction with other clinical and laboratory findings.

Device Story

ClearLLab reagents consist of pre-cocktailed monoclonal/polyclonal antibody panels (T1, T2, B1, B2, M) conjugated to fluorochromes (FITC, PE, ECD, PC5.5, PC7). Input: peripheral whole blood, bone marrow, or lymph node specimens. Operation: specimens stained with reagents, red blood cells lysed, and analyzed on Beckman Coulter FC 500 flow cytometer. Output: qualitative multiparameter immunophenotyping data. Used in clinical laboratories by trained technicians; results interpreted by pathologists. Device aids in identifying aberrant cell surface marker expression (CD markers) to distinguish malignant from non-malignant populations. Benefits: standardized diagnostic workflow for hematolymphoid neoplasia.

Clinical Evidence

Multi-center retrospective study (n=279) comparing ClearLLab results to final clinical diagnosis. Specimens included peripheral blood (60%), bone marrow (30%), and lymph nodes (10%). Sensitivity for malignancy detection was 82.4%-85.9%; specificity was 92.7%-94.2%. PPV was 92.4%-93.6%; NPV was 83.8%-86.4%. Agreement between two independent experts was 98.4% for positive and 94.2% for negative specimens.

Technological Characteristics

Multiparameter immunophenotyping reagents for flow cytometry. Detects CD markers (CD2, CD3, CD4, CD5, CD7, CD8, CD10, CD13, CD19, CD20, CD33, CD34, CD38, CD45, Kappa, Lambda). Uses FC 500 flow cytometer. Includes polyclonal antibodies requiring Fluorescence Minus One (FMO) studies for non-specific binding evaluation.

Indications for Use

Indicated for use as an aid in the differential diagnosis of hematologically abnormal patients (adults and children) suspected of having hematopoietic neoplasms, including chronic leukemia, acute leukemia, non-Hodgkin lymphoma, myeloma, myelodysplastic syndrome (MDS), and myeloproliferative neoplasms (MPN).

Regulatory Classification

Identification

A flow cytometric test for hematopoietic neoplasms is a device that consists of reagents for immunophenotyping of human cells in relation to the level of expression, antigen density, and distribution of specific cellular markers. These reagents are used as an aid in the differential diagnosis or monitoring of hematologically abnormal patients having or suspected of having hematopoietic neoplasms. The results should be interpreted by a pathologist or equivalent professional in conjunction with other clinical and laboratory findings.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information: (i) The indications for use must indicate the clinical hematopoietic neoplasms for which the assay was designed and validated, for example, chronic leukemia or lymphoma. (ii) A detailed device description including the following: (A) A detailed description of all test components, all required reagents, and all instrumentation and equipment, including illustrations or photographs of nonstandard equipment or methods. (B) Detailed documentation of the device software including, but not limited to, standalone software applications and hardware-based devices that incorporate software. (C) A detailed description of methodology and assay procedure. (D) A description of appropriate internal and external quality control materials that are recommended or provided. The description must identify those control elements that are incorporated into the testing procedure, if applicable. (E) Detailed specifications for sample collection, processing, and storage. (F) Detailed specification of the criteria for test results interpretation and reporting including pre-established templates. (G) If applicable, based on the output of the results, a description of the specific number of events to collect, result outputs, and analytical sensitivity of the assay that will be reported. (iii) Information that demonstrates the performance characteristics of the test, including: (A) Device performance data from either a method comparison study comparing the specific lymphocyte cell markers to a predicate device or data collected through a clinical study demonstrating clinical validity using well-characterized clinical specimens. Samples must be representative of the intended use population of the device including hematologic neoplasms and the specific sample types for which the test is indicated for use. (B) If applicable, device performance data from a clinical study demonstrating clinical validity for parameters not established in a predicate device of this generic type using well-characterized prospectively obtained clinical specimens including all hematologic neoplasms and the specific sample types for which the device is indicated for use. (C) Device precision data using clinical samples to evaluate the within-lot, between-lot, within-run, between run, site-to-site and total variation using a minimum of three sites, of which at least two sites must be external sites. Results shall be reported as the standard deviation and percentage coefficient of variation for each level tested. (D) Reproducibility data generated using a minimum of three lots of reagents to evaluate mean fluorescence intensity and variability of the recovery of the different markers and/or cell populations. (E) Data from specimen and reagent carryover testing performed using well-established methods ( *e.g.,* CLSI H26-A2).(F) Specimen and prepared sample stability data established for each specimen matrix in the anticoagulant combinations and storage/use conditions that will be indicated. (G) A study testing anticoagulant equivalency in all claimed specimen type/anticoagulant combinations using clinical specimens that are representative of the intended use population of the device. (H) Analytic sensitivity data using a dilution panel created from clinical samples. (I) Analytical specificity data, including interference and cross-contamination. (J) Device stability data, including real-time stability of reagents under various storage times and temperatures. (K) For devices that include polyclonal antibodies, Fluorescence Minus One (FMO) studies to evaluate non-specific binding for all polyclonal antibodies. Each FMO tube is compared to reagent reference to demonstrate that no additional population appears when one marker is absent. Pre-specified acceptance criteria must be provided and followed. (L) For devices indicated for use as a semi-quantitative test, linearity data using a dilution panel created from clinical samples. (M) For devices indicated for use as a semi-quantitative test, clinically relevant analytical sensitivity data, including limit of blank, limit of detection, and limit of quantification. (iv) Identification of risk mitigation elements used by the device, including a detailed description of all additional procedures, methods, and practices incorporated into the instructions for use that mitigate risks associated with testing the device. (2) The 21 CFR 809.10 compliant labeling must include the following: (i) The intended use statement in the 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include a statement that the results should be interpreted by a pathologist or equivalent professional in conjunction with other clinical and laboratory findings. The intended use statement must also include information on what the device detects and measures, whether the device is qualitative, semi-quantitative, and/or quantitative, the clinical indications for which the device is to be used, and the specific population(s) for which the device is intended. (ii) A detailed description of the performance studies conducted to comply with paragraph (b)(1)(iii) of this section and a summary of the results. (3) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, product labeling and instruction manuals must include clear examples of all expected phenotypic patterns and gating strategies using well-defined clinical samples representative of both abnormal and normal cellular populations. These samples must be selected based upon the indications described in paragraph (b)(1)(i) of this section.

Submission Summary (Full Text)

{1}------------------------------------------------ # EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR CLEARLLAB REAGENTS DECISION SUMMARY ### A. DEN Number: DEN160047 # B. Purpose for Submission: De novo request for evaluation of automatic class III designation of the ClearLLab Reagents ### C. Measurand: Cluster of Differentiation (CD) Surface Markers on white blood cells ### D. Type of Test: Immunophenotyping, qualitative, flow cytometric assay ### E. Applicant: Beckman Coulter Inc. ### F. Proprietary and Established Names: Trade Name: ClearLLab T1, T2, B1, B2, M Common Name: ClearLLab ### G. Regulatory Information: - 1. Regulation section: 21 CFR 864.7010 - 2. Classification: Class II (Special Controls) - 3. Product code: PWD - 4. Panel: 81-Hematology {2}------------------------------------------------ ### H. Indications For Use: ### 1. Indications for Use The ClearLLab reagents are intended for in vitro diagnostic use as a panel for qualitative identification of cell populations by multiparameter immunophenotyping on an FC 500 flow cytometer. These reagents are used as an aid in the differential diagnosis of hematologically abnormal patients having, or suspected of having the following hematopoietic neoplasms; chronic leukemia, acute leukemia, non-Hodgkin lymphoma, myeloma, myelodysplastic syndrome (MDS), and/or myeloproliferative neoplasms (MPN). The reagents can be used with peripheral whole blood collected in EDTA (K2 and K3EDTA), Acid Citrate Dextrose or Heparin, bone marrow collected in K2EDTA, ACD, or Heparin, and lymph node specimens. Interpretation of the results should be confirmed by a pathologist or equivalent professional in conjunction with other clinical and laboratory findings. These reagents provide multiparameter, qualitative results for the Cluster of Differentiation (CD) parameters listed below: - ClearLLab T1: CD2, CD56, CD7, CD5, CD7, CD5, CD45 - ClearLLab T2: CD8, CD4, CD3, CD45 - ClearLLab B1: Kappa, Lambda, CD19, CD5, CD45 - ClearLLab B2: CD20, CD10, CD10, CD19, CD38, CD45 - ClearLLab M: CD7, CD13, CD34, CD33, CD45 - 2. Special conditions for use statement(s): For prescription use only. For in vitro diagnostic use. - 3. Special instrument requirements: Beckman Coulter FC500 flow cytometer, CXP Software Doggintian ### I. Device Description: Itom The Clear LLab Reagents panel is used with the following reagents with their respective indicated uses. A description of the reagents provided is described below in Table 1. | Item | Description | Use | |------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | The T-Cell Lineage panels | Comprising T1 CD2-<br>FITC/CD56-PE/CD7-<br>ECD/CD5-PC5.5/CD45-PC7<br>and T2 CD8-FITC/CD4-<br>PE/CD3-PC5.5/CD45-PC7<br>Monoclonal Antibody Reagents. | For use to identify lymphocytes<br>that contain the specific cell<br>surface markers associated with<br>the T-cell lineage. | | The B-Cell Lineage panels | Comprising B1 Kappa- | For use to identify lymphocytes | | | FITC/Lambda-PE/CD19-<br>ECD/CD5-PC5.5/CD45-PC7<br>and B2 CD20-FITC/CD10-<br>PE/CD19-ECD/CD38-<br>PC5.5/CD45-PC7 Monoclonal<br>and Polyclonal Antibody<br>Reagents. | that contain the specific cell<br>surface markers associated with<br>the B-cell lineage. | | The Myeloid Lineage panel | Comprising M CD7-FITC/<br>CD13-PE/CD34-ECD//CD33-<br>PC5.5/CD45-PC7 Monoclonal<br>Antibody Reagents. | For use to identify lymphocytes<br>that contain the specific cell<br>surface markers associated with<br>the Myeloid lineage. | | Accessory Reagents Required | | | | Flow-Check Pro Fluorospheres | Flow-Check Pro Fluorospheres<br>are a suspension of fluorescent<br>microspheres used for daily<br>verification of the optical<br>alignment and fluidics for<br>Forward Scatter (FS) and FL1-<br>FL4 fluorescence parameters. | Instrument Alignment Quality<br>Control Reagent to provide<br>instrument alignment Quality<br>Control instructions. | | Flow-Set Pro Fluorospheres | Flow-Set Pro Fluorospheres are<br>a suspension of fluorescent<br>microspheres used as an aid in<br>standardizing forward scatter,<br>side scatter, and fluorescence<br>detectors on the FC 500 (FL1-<br>FL5). | Auto Setup Reagent for<br>standardization of flow<br>cytometer light scatter and<br>fluorescence intensity<br>instrument settings to provide<br>application-specific instrument<br>target ranges for<br>standardization. | | QuickComp 4 Kit | The QuickComp 4 Kit consists<br>of four single-color fluorescent<br>reagents comprised of one<br>monoclonal antibody each. Each<br>antibody is labeled with one of<br>four fluorochromes, three are<br>utilized for this test: CD45-<br>FITC, CD45-PE, and CD45-<br>ECD. | The QuickComp 4 Kit is used to<br>adjust color compensation<br>settings on a flow cytometer<br>equipped with AutoSetup<br>software, prior to multi-color<br>analysis with FITC, PE, and<br>ECD conjugated monoclonal<br>antibody reagents. | | Color Compensation Reagents | Compensation reagents CD45-<br>PC5.5 and CD45-PC7. | Used to adjust color<br>compensation settings on a flow<br>cytometer with AutoSetup<br>software.<br>Note: Color Compensation<br>Reagents are required. It is<br>recommended to use the<br>reagents with normal whole<br>blood specimens to adjust color<br>compensation settings on a flow<br>cytometer, prior to multi-color<br>analysis. | Table 1: Components of the ClearLLab Reagents {3}------------------------------------------------ {4}------------------------------------------------ | IOTest 3 Fixative Solution | The IOTest 3 Fixative Solution, which has a formaldehyde base, specially developed for the fixing of leukocytes. | Leukocyte fixative solution used on all leukocytic preparations to enable leukocytic preparations to be stored for several hours without deterioration after staining with a fluorescent antibody. It is used to fix leukocytes following immunofluorescent staining with the fluorochrome-conjugated antibodies and lysis of the red blood cells. | |-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Accessory Reagents Recommended but Not Provided | | | | VersaLyse Lysing Solution | Red cell lysing reagent intended for the lysis of red blood cells in the preparation of biological samples for flow cytometry analysis. | VersaLyse is intended for the lysis of red blood cells in the preparation of biological samples for flow cytometry analysis. | | Immutrol Control Cells | Process controls for flow cytometry. | Quality control material assayed for lymphocyte, granulocyte and monocyte specific antigens and single platform absolute counts. The light scatter, population distribution, fluorescence intensity, and antigen density mimic those of normal whole blood. | | StemTrol Control cells | An antibody-to-antigen positive control for CD34 and CD45 staining in flow cytometry studies, consisting of a preserved cell line, BK010044, | Quality control material for CD34 and CD45. | # ClearLLab Immunophenotyping Panel | Lineage | FITC | PE | ECD | PC5.5 | PC7 | |--------------------|-------|--------|------|-------|------| | T-cell tube 1 (T1) | CD2 | CD56 | CD7 | CD5 | CD45 | | T-cell tube 2 (T2) | CD8 | CD4 | | CD3 | CD45 | | B-cell tube 1 (B1) | Kappa | Lambda | CD19 | CD5 | CD45 | | B-cell tube 2 (B2) | CD20 | CD10 | CD19 | CD38 | CD45 | | Myeloid (M) | CD7 | CD13 | CD34 | CD33 | CD45 | # J. Standards/Guidance Documents Referenced: CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures, A Statistical Approach CLSI EP05-A3, Evaluation of Precision Performance of Quantitative Measurement Methods {5}------------------------------------------------ CLSI EP28-A3c, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory CLSI H26-A2, Validation, Verification, and Quality Assurance of Automated Hematology Analyzers CLSI EP17-A2, Protocols for Determination of Limits of Detection and Limits of Quantitation CLSI EP25-A, Evaluation of Stability on In Vitro Diagnostic Reagents CLSI EP09-A3 Measurement Procedure Comparison and Bias Estimation Using Patient Samples CLSI EP12-A2, User Protocol for Evaluation of Qualitative Test Performance CLSI H43-A2, Clinical Flow Cytometric Analysis of Neoplastic Hematolymphoid Cells ### K. Test Principle: Beckman Coulter's ClearLLab product is comprised of pre-cocktailed, lineage-driven combinations of the consensus Cluster of Differentiation (CD) reagents for evaluation of B and T cell neoplasia and the consensus limited CD reagents for myelomonocytic neoplasia as described in Table 3 of "2006 Bethesda International Consensus Recommendations on the Immunophenotypic Analysis of Hematolymphoid Neoplasia by Flow Cytometry: Optimal Reagents and Reporting for the Flow Cytometric Diagnosis of Hematopoietic Neoplasia" ! The ClearLLab product is intended for use on the FC 500 flow cytometer with instrument set-up performed with Flow-Check Pro Fluorospheres, Flow-Set Pro Fluorospheres, and QuickComp color compensation reagents for alignment, voltage standardization, and compensation. This test depends on the ability of a monoclonal or polyclonal antibody to bind to the surface of cells expressing discrete antigenic determinants. Specific cell staining is accomplished by incubating specimens prepared for staining with the appropriate antibody reagent. The ClearLLab Reagents are composed of five panels containing four or five monoclonal or polyclonal antibody reagents, each specific for a different cell surface antigen and conjugated to a specific fluorochrome. After sample preparation, the specimens are analyzed on the flow cytometer with manual gating. The FC 500 flow cytometer running under the control of CXP Software applies the principles of flow cytometry to analyze a whole blood, bone marrow or lymph node sample. Samples are prepared and stained with a monoclonal antibody reagents followed by lysis of red blood cells prior to introduction into the instrument. Cellular populations are identified based on the specific monoclonal and polyclonal antibodies and fluorochromes used in the different panels. Detection of fluorescent antibodies bound to cells utilizes the capability of the FC 500 flow cytometer to <sup>1 2006</sup> Bethesda International Consensus recommendations on the immunophenotypic analysis of hematolymphoid neoplasia by flow cytometry: optimal reagents and reporting for the flow cytometric diagnosis of hematopoietic neoplasia. Wood BL, Arroz M, Barnett D, DiGiuseppe J, Greig B, Kussick SJ, Oldaker T, Shenkin M, Stone E, Wallace P. Cytometry B Clin Cytom. 2007; 72 Suppl 1:S14-22 {6}------------------------------------------------ detect fluorescence with one laser, a blue laser with a 488 nm excitation and five different detection channels (FL1-FL5). The CXP SYSTEM application software (K030828) provides automated instrument setup of light scatter and fluorescence intensity standardization, color compensation, and verification when used with the quality control reagents. The CXP Software has an Auto-Set Panel which automatically standardizes the cytometer, adjusts compensation settings, passes cytometer settings to designated test protocols, and verifies cytometer setup and antibody performance. Compensation settings are determined using CD45-FITC, CD45-PE, and CD45-ECD supplied in the QuickComp 4 Kit and CD45-PC5.5 and CD45-PC7 single color reagents. The existing FC 500 CXP software will support the ClearLLab reagents for use on the FC 500 flow cytometer without any further modifications from the original clearance in K030828. ### Sample Preparation & Analysis ### Whole Blood (WB) and Bone Marrow (BM) preparation: Specimens are prepared manually with three pre-wash steps using phosphate buffered saline (PBS) containing 2% Heat Inactivated Fetal Calf Serum (HI-FCS). The remaining white blood cell pellet is then resuspended in PBS containing HI-FCS. The washed specimen is stained with the ClearLLab 5-Color Reagents. VersaLyse Lysing Solution for lysing red blood cells and the IOTest 3 Fixative Solution are used for red blood cell lysis and fixation prior to analysis. ### Lymph Node preparation: Lymph node specimens require disaggregation of the tissue into a single cell suspension. Single cell suspensions prepared from lymphoid tissues may not require washing prior to staining if the specimen was washed during the disaggregation process. If washing steps were not performed for removal of residual soluble proteins, or if the cells were re-suspended in a buffer containing human serum or serum proteins, then pre-washing is necessary. ### Sample analysis: White blood cells (WBCs) are analyzed on an FC 500 flow cytometer. Data are analyzed using a sequential gating strategy. The total WBC gate is defined in the first histogram (Side Scatter vs. CD45-PC7) by the inclusion of all CD45+ events with low, medium and high Side Scatter (SSC). A second histogram defines specific leukocyte subsets such as lymphocytes, monocytes, and granulocytes. The low SSC/bright CD45+ population identifies lymphocytes. Applicable markers are then displayed in subsequent histograms gated on lymphocytes. This process is repeated for mid SSC/medium CD45+ populations to identify monocytes and for high SSC/medium CD45+ to identify granulocytes. ### L. Interpretation of Results Flow cytometric immunophenotyping is multiparametric and relies on simultaneous identification and characterization of both normal and candidate malignant cell populations. Normal cells, which vary in type and relative numbers depending on the sample type, patient age, and clinical setting, display highly conserved and reproducible patterns of cell surface marker expression and light scatter. These normal internal control patterns may serve as both process controls for sample type. sample degradation, sample preparation and staining, data acquisition, and data analysis as well as references against which candidate malignant populations may be compared. Candidate malignant populations may display aberrant loss of a marker, aberrant gain of a marker, or aberrant over- or under-expression of a marker. Any population that does not conform to the normal internal control pattern is evaluated for {7}------------------------------------------------ malignancy. Interpretation of these patterns requires expert knowledge and training in the technique. It is not possible to predict the phenotype of a particular patient's hematolymphoid malignancy and apply fixed templates for samples for detection of specific phenotypes. Even well-characterized diseases such as chronic lymphocytic leukemia may display case-to-case immunophenotypic variability and it is therefore the role of the expert interpreter to assess the significance of findings and to interpret them together with clinical history, morphological assessment, and other ancillary techniques in order to arrive at a final diagnosis. The device labeling states that interpretation of specimens should be performed by a pathologist or equivalent professional who has the appropriate training. Finally, a small number of rare samples contain no or very few normal cells and require careful evaluation and correlation with other sources of information such as clinical history and other laboratory findings to obtain the appropriate final diagnosis. ### M. Performance Characteristics: All results provided below met the manufacturer's pre-specified acceptance criteria. - 1. Analytical Performance: - Precision and Reproducibility: Precision studies were conducted according to the a. methodology presented in CLSI EP05-A3, Evaluation of Precision Performance of Quantitative Measurement Procedures; Approved Guideline – Third Edition. - Precision (repeatability): Thirty-eight specimens across the three specimen types were i. evaluated. For Whole Blood (WB) and Bone Marrow (BM), testing was conducted at three sites, with each of the three sites using a different anticoagulant according to their current clinical testing procedures. For lymph node specimens (LN), testing was conducted at two sites. Since LN specimens are not collected using an anticoagulant, this was not a variable for lymph node specimen analysis. A minimum of one specimen for each specimen type and lineage were collected at each site. Lineage is defined by the tubes used, where T lineage is comprised of T1 and T2 tubes, B lineage is comprised of B1 and B2 tubes, and M lineage is comprised of Myeloid tube. Specimens were selected to cover the markers in each lineage tube regardless of the disease. The same specimen was analyzed twice each day with one time point in the morning and one time point in the afternoon. Six sample preparations per time point were made for a total of 12 replicates per specimen per lineage tube. A qualitative approach was applied to repeatability assessment where the presence or absence of an abnormal phenotype was reported. An analysis of the % abnormal population, if present, or pre-selected lineage-specific normal populations was assessed for repeatability and reproducibility. The specimens were tested in 12 replicates. The expected result was to obtain the same number of positive or negatives for all replicates. The total actual number positive or negative was divided by the total expected number positive or negative to calculate the % positive or % negative results. The qualitative results for both abnormal and normal populations met 100% agreement for all sites and anticoagulants for expected and actual results with respect to the presence or absence of an abnormal phenotype. {8}------------------------------------------------ | ClearLLab Lineage Tube | Cells | Markers | Population<br>of cells<br>Identified | |----------------------------------|-----------------------|-----------------------------------------------------------------|--------------------------------------| | T1 CD2/CD56/CD7/CD5/CD45 | T | CD45+ Bright, Low SS<br>CD2+CD5+CD7+ (Partial)<br>CD56- (Few +) | N#1 | | | NK | CD45+ Bright, Low SS<br>CD2+CD7+CD56+CD5- | N#2 | | T2 CD8/CD4/CD3/CD45 | T Helper | CD45+ Bright, Low SS<br>CD3+CD4+CD8- | N#3 | | | T<br>Suppressor | CD45+ Bright, Low SS<br>CD3+CD4-CD8+ | N#4 | | B1<br>KAPPA/LAMBDA/CD19/CD5/CD45 | B | CD45+ Bright, Low SS<br>CD19+ CD5-Kappa+ and<br>Lambda+ | N#5 | | B2 CD20/CD10/CD19/CD38/CD45 | B | CD45+ Bright, Low SS<br>CD19+CD20+ CD38<br>Dim/Variable CD10- | N#6 | | | Mature<br>Neutrophils | CD45+ Dim, High SS<br>CD10+ | N#7 | | | Mature<br>Neutrophils | CD45+ Dim, High SS<br>CD7-CD13+CD34-CD33+ | N#8 | | M CD7/CD13/CD34/CD33/CD45 | Monocytes | CD45+ Med, Med SS<br>CD7-CD13+ CD34-CD33+<br>Bright | N#9 | Pre-selected Lineage-Specific Normal Population | | Total Samples | | 38 | |--------|---------------|------|----| | Gender | Male | | 22 | | | Female | | 16 | | Age | Child | N | 1 | | | (2-12) | Mean | 3 | | | Adolescent | N | 0 | | | (12-21) | Mean | 0 | | | Adult | N | 19 | | | (21-65) | Mean | 53 | | | Senior | N | 18 | | | (66-89) | Mean | 77 | {9}------------------------------------------------ | | | Specimen Type | | |--------|---------------|---------------|----| | Site | Anticoagulant | WB | BM | | Site 1 | ACD | 6 | 6 | | Site 2 | K2EDTA | 4 | 5 | | Site 3 | Heparin | 6 | 7 | | | Total Samples | 16 | 18 | Repeatability for Site 1 using the ClearLLab Immunophenotyping Panel | Site 1 | Population<br>of Cells<br>Identified | Specimen<br>Type | N | Lineage<br>Tube | Mean<br>(%) | Repeatability | | Between Runs | | Within<br>Specimen | | |--------|--------------------------------------|------------------|----|-----------------|-------------|---------------|------|--------------|-------|--------------------|-------| | Sample | | | | | | SD | %CV | SD | %CV | SD | %CV | | 7 | ABN#1 | LN | 12 | B1 | 63.40 | 1.30 | 2.04 | 0.78 | 1.23 | 1.51 | 2.39 | | 7 | ABN#2 | LN | 12 | B1 | 14.99 | 1.13 | 7.55 | 0.38 | 2.56 | 1.20 | 7.98 | | 10 | N#5 | WB | 12 | B1 | 2.03 | 0.14 | 6.74 | 0 | 0 | 0.14 | 6.74 | | 10 | ABN | WB | 12 | B1 | 61.96 | 1.32 | 2.13 | 0 | 0 | 1.32 | 2.13 | | 13 | ABN#1 | LN | 12 | B1 | 69.32 | 1.25 | 1.80 | 1.96 | 2.83 | 2.33 | 3.36 | | 13 | ABN#2 | LN | 12 | B2 | 85.97 | 0.70 | 0.81 | 1.82 | 2.1 | 1.95 | 2.26 | | 14 | N#6 | BM | 12 | B2 | 7.56 | 0.45 | 5.92 | 0.22 | 2.92 | 0.50 | 6.60 | | 14 | N#7 | BM | 12 | B2 | 29.15 | 0.96 | 3.30 | 0.58 | 2.00 | 1.13 | 3.86 | | 17 | ABN | WB | 12 | B1 | 87.49 | 0.53 | 0.60 | 0.86 | 0.98 | 1.01 | 1.15 | | 18 | ABN | WB | 12 | B2 | 98.10 | 0.16 | 0.17 | 0.02 | 0.02 | 0.16 | 0.17 | | 18 | ABN | WB | 12 | B1 | 86.24 | 0.66 | 0.77 | 0 | 0 | 0.66 | 0.77 | | 19 | N#6 | BM | 12 | B2 | 83.02 | 0.46 | 0.55 | 0.33 | 0.39 | 0.56 | 0.68 | | 19 | N#7 | BM | 12 | B2 | 19.56 | 0.85 | 4.32 | 1.08 | 5.51 | 1.37 | 7.01 | | 11 | N#1 | WB | 12 | T1 | 71.28 | 0.65 | 0.91 | 0.23 | 0.32 | 0.69 | 0.97 | | 11 | N#2 | | 12 | T1 | 11.17 | 0.51 | 4.59 | 0.74 | 6.64 | 0.90 | 8.07 | | 11 | N#3 | | 12 | T2 | 35.72 | 0.44 | 1.23 | 0.09 | 0.24 | 0.45 | 1.25 | | 11 | N#4 | | 12 | T2 | 32.75 | 0.61 | 1.85 | 0 | 0 | 0.61 | 1.85 | | 12 | N#1 | BM | 12 | T1 | 76.44 | 0.83 | 1.09 | 2.74 | 3.58 | 2.86 | 3.74 | | 12 | N#2 | | 12 | T1 | 8.84 | 0.70 | 7.90 | 0 | 0 | 0.70 | 7.90 | | 12 | N#3 | | 12 | T2 | 41.61 | 1.24 | 2.99 | 1.17 | 2.82 | .71 | 4.11 | | 12 | N#4 | | 12 | T2 | 38.51 | 1.31 | 3.39 | 1.89 | 4.92 | 2.30 | 5.97 | | 15 | N#1 | BM | 12 | T1 | 81.19 | 0.83 | 1.02 | 0 | 0 | 0.83 | 1.02 | | 15 | N#2 | | 12 | T1 | 16.92 | 0.67 | 3.96 | 0 | 0 | 0.67 | 3.96 | | 5 | ABN | WB | 12 | M | 84.26 | 1.01 | 1.20 | 2.82 | 3.35 | 3.00 | 3.56 | | 8 | ABN | BM | 12 | M | 78.78 | 1.14 | 1.44 | 2.36 | 3.00 | 2.62 | 3.33 | | 9 | ABN | BM | 12 | M | 7.06 | 0.41 | 5.81 | 0.71 | 10.10 | 0.82 | 11.65 | | 16 | N#8 | WB | 12 | M | 47.31 | 1.98 | 4.18 | 1.65 | 3.49 | 2.58 | 5.45 | | 16 | N#9 | | 12 | M | 3.69 | 0.32 | 8.54 | 0.25 | 6.86 | 0.40 | 10.95 | Repeatability for Site 2 using the ClearLLab Immunophenotyping Panel | Site 2 | Population<br>of Cells | Specimen<br>Type | N | Lineage<br>Tube | Mean<br>(%) | Repeatability | | Between Runs | | Within<br>Specimen | | |--------|------------------------|------------------|----|-----------------|-------------|---------------|------|--------------|------|--------------------|------| | Sample | Identified | | | | | SD | %CV | SD | %CV | SD | %CV | | 7 | N#5 | LN | 12 | B1 | 25.04 | 0.71 | 2.84 | 0.63 | 2.53 | 0.95 | 3.80 | | | N#6 | | 12 | B2 | 25.99 | 0.50 | 1.92 | 0.91 | 3.49 | 1.04 | 3.99 | | | N#7 | | 12 | B2 | 17.14 | 1.40 | 8.18 | 0.60 | 3.47 | 1.52 | 8.89 | | 9 | N#5 | BM | 12 | B1 | 22.40 | 0.72 | 3.21 | 0.19 | 0.83 | 0.74 | 3.31 | | | N#6 | | 12 | B2 | 4.41 | 0.27 | 6.14 | 0.00 | 0.00 | 0.27 | 6.14 | | | N#7 | | 12 | B2 | 28.04 | 0.28 | 1.01 | 0.29 | 1.03 | 0.41 | 1.45 | | 11 | N#5 | WB | 12 | B1 | 17.63 | 0.27 | 1.51 | 0.10 | 0.56 | 0.28 | 1.61 | | | N#6 | | 12 | B2 | 18.37 | 0.32 | 1.72 | 0.19 | 1.02 | 0.37 | 2.00 | | | N#7 | | 12 | B2 | 73.47 | 0.69 | 0.93 | 0.30 | 0.41 | 0.75 | 1.02 | {10}------------------------------------------------ | Site 2<br>Sample | Population<br>of Cells<br>Identified | Specimen<br>Type | N | Lineage<br>Tube | Mean<br>(%) | Repeatability | | Between Runs | | Within<br>Specimen | | |------------------|--------------------------------------|------------------|----|-----------------|-------------|---------------|------|--------------|------|--------------------|------| | | | | | | | SD | %CV | SD | %CV | SD | %CV | | 12 | N#5 | LN | 12 | B1 | 25.91 | 0.49 | 1.90 | 0.76 | 2.95 | 0.91 | 3.51 | | 12 | N#6 | | 12 | B2 | 26.36 | 0.40 | 1.50 | 0.60 | 2.28 | 0.72 | 2.73 | | 12 | N#7 | | 12 | B2 | 5.05 | 0.38 | 7.61 | 0 | 0 | 0.38 | 7.61 | | 1 | N#1 | BM | 12 | T1 | 70.34 | 0.80 | 1.14 | 0.50 | 0.71 | 0.94 | 1.34 | | 1 | N#2 | | 12 | T1 | 9.83 | 0.33 | 3.38 | 0.55 | 5.60 | 0.64 | 6.54 | | 1 | N#3 | | 12 | T2 | 33.86 | 0.51 | 1.51 | 0.60 | 1.77 | 0.79 | 2.33 | | 1 | N#4 | | 12 | T2 | 37.25 | 0.86 | 2.30 | 0 | 0 | 0.86 | 2.30 | | 4 | N#1 | BM | 12 | T1 | 70.37 | 0.73 | 1.03 | 0.48 | 0.68 | 0.87 | 1.24 | | 4 | N#2 | | 12 | T1 | 8.66 | 0.57 | 6.59 | 0.62 | 7.15 | 0.84 | 9.72 | | 4 | N#3 | | 12 | T2 | 46.81 | 0.51 | 1.10 | 0.61 | 1.30 | 0.80 | 1.70 | | 4 | N#4 | | 12 | T2 | 27.61 | 0.75 | 2.73 | 0 | 0 | 0.75 | 2.73 | | 10 | ABN | WB | 12 | T1 | 7.38 | 0.16 | 2.12 | 0 | 0 | 0.16 | 2.12 | | 3 | ABN | BM | 12 | M | 58.53 | 0.40 | 0.69 | 0.17 | 0.30 | 0.44 | 0.75 | | 5 | ABN | WB | 12 | M | 3.04 | 0.15 | 4.78 | 0.13 | 4.34 | 0.20 | 6.46 | | 6 | ABN | WB | 12 | M | 62.27 | 0.48 | 0.77 | 0.09 | 0.15 | 0.49 | 0.79 | | 8 | N#8 | BM | 12 | M | 70.44 | 0.47 | 0.66 | 0.37 | 0.53 | 0.59 | 0.84 | | 8 | N#9 | | 12 | M | 6.86 | 0.16 | 2.27 | 0.09 | 1.28 | 0.18 | 2.61 | Repeatability for Site 3 using the ClearLLab Immunophenotyping Panel | Site 3 | Population<br>of Cells<br>Identified | Specimen<br>Type | N | Lineage<br>Tube | Mean<br>(%) | Repeatability | | Between Runs | | Within<br>Specimen | | |--------|--------------------------------------|------------------|----|-----------------|-------------|---------------|------|--------------|------|--------------------|------| | Sample | | | | | | SD | %CV | SD | %CV | SD | %CV | | 1 | ABN | BM | 12 | B1 | 33.95 | 0.93 | 2.73 | 1.68 | 4.94 | 1.92 | 5.65 | | 1 | N#6 | BM | 12 | B2 | 98.82 | 0.11 | 0.11 | 0.10 | 0.11 | 0.15 | 0.15 | | | N#7 | | 12 | B2 | 28.17 | 0.68 | 2.43 | 1.02 | 3.63 | 1.23 | 4.37 | | 6 | ABN | WB | 12 | B1 | 97.16 | 0.10 | 0.11 | 0.04 | 0.04 | 0.11 | 0.12 | | 6 | N#6 | WB | 12 | B2 | 99.89 | 0.02 | 0.02 | 0.01 | 0.01 | 0.02 | 0.02 | | | N#7 | | 12 | B2 | 4.18 | 0.12 | 2.85 | 0 | 0 | 0.12 | 2.85 | | 11 | N#5 | BM | 12 | B1 | 18.40 | 0.50 | 2.70 | 0.37 | 2.03 | 0.62 | 3.37 | | 11 | N#6 | BM | 12 | B2 | 78.88 | 1.82 | 2.31 | 4.26 | 5.40 | 4.63 | 5.87 | | | N#7 | | 12 | B2 | 46.24 | 0.61 | 1.31 | 0.09 | 0.20 | 0.61 | 1.33 | | 12 | ABN | BM | 12 | B1 | 53.80 | 0.66 | 1.23 | 0.41 | 0.77 | 0.78 | 1.45 | | 12 | N#6 | BM | 12 | B2 | 91.62 | 0.53 | 0.58 | 0 | 0 | 0.53 | 0.58 | | | N#7 | | 12 | B2 | 15.42 | 0.37 | 2.43 | 0.18 | 1.16 | 0.42 | 2.69 | | 15 | N#5 | WB | 12 | B1 | 10.83 | 0.39 | 3.59 | 0 | 0 | 0.39 | 3.59 | | 15 | N#6 | WB | 12 | B2 | 99.20 | 0.21 | 0.21 | 0.12 | 0.12 | 0.24 | 0.24 | | | N#7 | | 12 | B2 | 35.81 | 1.28 | 3.58 | 1.80 | 5.02 | 2.21 | 6.17 | | 2 | ABN | WB | 12 | T1 | 74.75 | 1.46 | 1.95 | 1.61 | 2.16 | 2.18 | 2.91 | | 3 | N#1 | BM | 12 | T1 | 86.56 | 0.32 | 0.37 | 0 | 0 | 0.32 | 0.37 | | 3 | N#2 | BM | 12 | T1 | 1.45 | 0.09 | 6.40 | 0 | 0 | 0.09 | 6.40 | | | N#3 | BM | 12 | T2 | 60.52 | 0.42 | 0.69 | 0 | 0 | 0.42 | 0.69 | | | N#4 | | 12 | T2 | 19.56 | 0.28 | 1.41 | 0 | 0 | 0.28 | 1.41 | | | N#1 | | 12 | T1 | 67.56 | 1.49 | 2.20 | 0 | 0 | 1.49 | 2.20 | | 8 | N#2 | WB | 12 | T1 | 10.00 | 0.17 | 1.69 | 0.13 | 1.30 | 0.21 | 2.13 | | 8 | N#3 | WB | 12 | T2 | 41.41 | 0.67 | 1.63 | 0.18 | 0.44 | 0.70 | 1.68 | | | N#4 | | 12 | T2 | 20.77 | 0.49 | 2.38 | 0 | 0 | 0.49 | 2.38 | | | N#1 | | 12 | T1 | 69.73 | 1.27 | 1.83 | 0.36 | 0.51 | 1.32 | 1.90 | | | N#2 | | 12 | T1 | 14.46 | 0.54 | 3.74 | 0.49 | 3.39 | 0.73 | 5.05 | | 9 | N#3 | BM | 12 | T2 | 43.97 | 1.01 | 2.30 | 0.12 | 0.28 | 1.02 | 2.31 | | 9 | N#4 | BM | 12 | T2 | 28.49 | 0.97 | 3.41 | 0 | 0 | 0.97 | 3.41 | | | 4 | ABN | WB | 12 | M | 60.71 | 1.36 | 2.24 | 0 | 0 | 1.36 | | 5 | N#8 | BM | 12 | M | 74.08 | 0.95 | 1.28 | 1.14 | 1.54 | 1.48 | 2.00 | | 5 | N#9 | BM | 12 | M | 7.98 | 0.17 | 2.10 | 0.11 | 1.38 | 0.20 | 2.52 | | | 7 | ABN | BM | 12 | M | 34.34 | 0.48 | 1.40 | 2.17 | 6.32 | 2.22 | {11}------------------------------------------------ | Site 3 | Population<br>of Cells | Specimen | N | Lineage<br>Tube | Mean<br>(%) | Repeatability | | Between Runs | | Within<br>Specimen | | |--------|------------------------|----------|----|-----------------|-------------|---------------|------|--------------|-----|--------------------|------| | Sample | Identified | Type | | | | SD | %CV | SD | %CV | SD | %CV | | 10 | ABN | WB | 12 | M | 42.42 | 1.30 | 3.06 | 0 | 0 | 1.30 | 3.06 | Population of Cells ABN: abnormal population identified in each sample; some samples have more than one abnormal population identified N: normal population identified based on pre-selected lineage-specific populations (see Table above) | | Positive<br>(abnormal phenotypes) | Negative<br>(normal phenotypes) | Total | % positive<br>(actual/expected) | % negative<br>(actual/expected) | |--------|-----------------------------------|---------------------------------|-------|---------------------------------|---------------------------------| | Site 1 | 11 | 17 | 28 | 100 | 100 | | Site 2 | 4 | 22 | 26 | 100 | 100 | | Site 3 | 7 | 26 | 33 | 100 | 100 | - ii. Operator-to-operator and instrument-to-instrument imprecision: Seven whole blood specimens with or without abnormal phenotypes to cover all CD markers were evaluated. The same specimen was prepared and analyzed twice in one day by three operators. Each operator made three sample preparations per time point and analyzed the samples on two FC 500 flow cytometers that had successfully passed quality control for a total of 12 determinations per operator. Considering three operators, a total of 36 acquisitions per specimen per ClearLLab reagent tube were collected. A qualitative approach was applied to the assessment of imprecision where the presence or absence of an abnormal phenotype was reported. In all cases, the operators were blinded to the patient's diagnosis. The qualitative results for both abnormal and normal populations met 100% agreement for all operators and instruments for expected and actual results with respect to the presence or absence of an abnormal phenotype. The demographics of the seven samples included six men and one woman between the ages of 21-65 years old. | Site 2 | | | Mean | Repeatability | | Between Runs | | Between Instruments | | Between Operators | | Within Specimen | | |--------|---------|----|-------|---------------|------|--------------|------|---------------------|------|-------------------|------|-----------------|-------| | Sample | Lineage | N | (%) | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | 2 | B1 | 36 | 88.44 | 0.45 | 0.51 | 0 | 0 | 0 | 0 | 0.94 | 1.06 | 1.04 | 1.18 | | 2 | B2 | 36 | 88.45 | 0.39 | 0.44 | 0.15 | 0.18 | 0 | 0 | 0.82 | 0.92 | 0.92 | 1.04 | | 4 | B1 | 36 | 9.63 | 0.65 | 6.77 | 0.47 | 4.93 | 0 | 0 | 0.57 | 5.91 | 0.99 | 10.25 | | 4 | B2 | 36 | 9.93 | 0.39 | 3.93 | 0 | 0 | 0.13 | 1.36 | 0 | 0 | 0.41 | 4.16 | | 5 | B2 | 36 | 66.59 | 1.76 | 2.64 | 1.18 | 1.78 | 0 | 0 | 0 | 0 | 2.12 | 3.18 | | 5 | B1 | 36 | 85.30 | 0.30 | 0.35 | 0.17 | 0.20 | 0 | 0 | 0.14 | 0.16 | 0.37 | 0.44 | | 5 | B2 | 36 | 85.08 | 0.26 | 0.31 | 0.12 | 0.14 | 0.07 | 0.08 | 0 | 0 | 0.29 | 0.35 | | 7 | B1 | 36 | 10.43 | 0.37 | 3.52 | 0 | 0 | 0 | 0 | 0 | 0 | 0.37 | 3.52 | | 7 | B2 | 36 | 10.43 | 0.28 | 2.71 | 0.32 | 3.05 | 0 | 0 | 0 | 0 | 0.43 | 4.08 | | 7 | B2 | 36 | 59.39 | 1.04 | 1.75 | 0.79 | 1.32 | 0 | 0 | 0.71 | 1.20 | 1.49 | 2.50 | | 1 | T1 | 36 | 96.70 | 0.11 | 0.11 | 0 | 0 | 0 | 0 | 0.12 | 0.12 | 0.16 | 0.16 | | 1 | T2 | 36 | 97.30 | 0.09 | 0.09 | 0 | 0 | 0.15 | 0.15 | 0 | 0 | 0.17 | 0.18 | | 3 | T1 | 36 | 73.39 | 0.53 | 0.72 | 0.11 | 0.15 | 0 | 0 | 0.69 | 0.95 | 0.88 | 1.20 | | 3 | T1 | 36 | 14.18 | 0.51 | 3.61 | 0 | 0 | 0.31 | 2.21 | 0 | 0 | 0.60 | 4.23 | | 3 | T2 | 36 | 52.38 | 0.62 | 1.18 | 0.11 | 0.20 | 0.43 | 0.82 | 0 | 0 | 0.76 | 1.45 | | 1 | T2 | 36 | 19.07 | 0.36 | 1.87 | 0 | 0 | 0.11 | 0.59 | 0.36 | 1.86 | 0.52 | 2.71 | | 1 | M | 36 | 96.38 | 0.12 | 0.12 | 0.09 | 0.09 | 0 | 0 | 0.05 | 0.05 | 0.16 | 0.16 | | 3 | M | 36 | 68.65 | 0.93 | 1.35 | 0.45 | 0.66 | 0 | 0 | 0.85 | 1.23 | 1.34 | 1.95 | Results for ClearLLab for all Operators and Instruments Combined {12}------------------------------------------------ | Site 2 | Lineage | N | Mean<br>(%) | Repeatability | | Between<br>Runs | | Between<br>Instruments | | Between<br>Operators | | Within<br>Specimen | | |--------|---------|----|-------------|---------------|------|-----------------|------|------------------------|------|----------------------|------|--------------------|------| | Sample | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | | M | 36 | 7.81 | 0.27 | 3.45 | 0 | 0 | 0.23 | 2.94 | 0.19 | 2.42 | 0.40 | 5.13 | | 5 | M | 36 | 64.05 | 0.41 | 0.64 | 0.31 | 0.48 | 0.43 | 0.67 | 0 | 0 | 0.67 | 1.04 | | | Positive<br>(abnormal phenotypes) | Negative<br>(normal phenotypes) | Total | % positive<br>(actual/expected) | % negative<br>(actual/expected) | |-------------|-----------------------------------|---------------------------------|-------|---------------------------------|---------------------------------| | Site 1 LHSC | 40 | 60 | 100 | 100 | 100 | - iii. Site-to-Site Reproducibility with Control Material: One level of control material (IMMUNO-TROL Cells (IT)) was prepared in duplicate with the ClearLLab reagents and analyzed twice each day for 20 days at four sites with each site using one FC 500 flow cytometer, generating a total of 80 replicates per ClearLLab reagent tube per site for an overall total of 320 replicates per ClearLLab reagent lineage tube. Eight replicates were excluded due to QC failures. An additional nine replicates were excluded from the CD34 analysis in the Myeloid Lineage tube due to CD34 pipetting and washing error, resulting in a total exclusion of 17 replicates from the overall study. To assess the precision of percent positive parameter measured for CD34-ECD marker the experiments were performed using one level of control product (Immuno-Trol) with Stem-Trol Control Cells (a preserved cell line that expresses CD34, BK010044) added. | Pop.<br>Lineage<br>** | | Marker | N | Mean<br>(%) | Repeatability | | Between<br>Runs | | Between<br>Days | | Between Sites | | Reproducibility | | |-----------------------------|---------------|-----------------------------|-----|-------------|---------------|------|-----------------|------|-----------------|------|---------------|-------|-----------------|--------| | | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | B-cell<br>Lineage<br>Tube 1 | LY+ | CD19+<br>Kappa+<br>CD19+LY | 312 | 57.38 | 1.60 | 2.78 | 1.92 | 3.35 | 1.08 | 1.88 | 1.34 | 2.33 | 3.03 | 5.29 | | | LY+ | CD19+<br>Lambda+<br>CD19+LY | 312 | 40.29 | 1.50 | 3.71 | 1.67 | 4.16 | 0 | 0 | 0.48 | 1.20 | 2.30 | 5.70 | | B-cell<br>Lineage<br>Tube 2 | | CD19+ | 312 | 14.23 | 0.37 | 2.63 | 0.39 | 2.73 | 0 | 0 | 0.67 | 4.74 | 0.86 | 6.07 | | | | CD5+ | 312 | 71.83 | 0.61 | 0.85 | 0.70 | 0.98 | 0.79 | 1.10 | 2.05 | 2.86 | 2.39 | 3.33 | | | GR+ | CD10+ | 312 | 99.96 | 0.02 | 0.02 | 0.07 | 0.07 | 0.01 | 0.01 | 0.03 | 0.03 | 0.08 | 0.08 | | | LY+ | CD19+ | 312 | 14.36 | 0.37 | 2.58 | 0.29 | 2.04 | 0.16 | 1.14 | 0.66 | 4.58 | 0.83 | 5.75 | | | | CD20+ | 312 | 14.82 | 0.43 | 2.92 | 0.51 | 3.47 | 0.60 | 4.07 | 1.22 | 8.23 | 1.52 | 10.24 | | | MO+ | CD38+ | 312 | 94.57 | 0.66 | 0.70 | 0.91 | 0.96 | 0.78 | 0.83 | 4.92 | 5.20 | 5.11 | 5.40 | | T-cell<br>Lineage<br>Tube 1 | | CD2+ | 312 | 81.75 | 0.65 | 0.79 | 0.78 | 0.95 | 0.66 | 0.81 | 1.13 | 1.38 | 1.66 | 2.03 | | | LY+ | CD5+ | 312 | 72.71 | 0.63 | 0.87 | 0.79 | 1.09 | 0.32 | 0.44 | 0.96 | 1.32 | 1.73 | 1.97 | | | | CD56+ | 312 | 13.76 | 0.46 | 3.32 | 0.97 | 7.07 | 0 | 0 | 0.18 | 1.31 | 1.09 | 7.92 | | | | CD7+ | 312 | 74.59 | 0.65 | 0.87 | 0.77 | 1.03 | 0.73 | 0.97 | 1.73 | 2.33 | 2.13 | 2.86 | | T-cell<br>Lineage<br>Tube 2 | | CD3+CD4+ | 312 | 45.47 | 0.55 | 1.22 | 0.49 | 1.07 | 0.19 | 0.42 | 0.37 | 0.82 | 0.85 | 1.87 | | | LY+ | CD3+CD8+ | 312 | 22.67 | 0.43 | 1.89 | 0.26 | 1.16 | 0.19 | 0.82 | 0.29 | 1.27 | 0.61 | 2.68 | | | | CD3+ | 312 | 71.14 | 0.54 | 0.76 | 0.64 | 0.90 | 0.23 | 0.32 | 0.62 | 0.87 | 1.07 | 1.50 | | Myeloid<br>Lineage | GR+ | CD13+ | 312 | 99.92 | 0.40 | 0.40 | 0 | 0 | 0.27 | 0.27 | 0.16 | 0.16 | 0.50 | 0.50 | | | LY+ | CD7+ | 312 | 73.91 | 0.70 | 0.95 | 1.74 | 2.35 | 2.22 | 3.01 | 4.17 | 5.64 | 5.08 | 6.88 | | | MO+ | CD33+ | 312 | 95.53 | 0.89 | 0.93 | 0.82 | 0.85 | 2.11 | 2.20 | 3.82 | 4.00 | 4.53 | 4.74 | | | Stem<br>Trol+ | CD34+ | 303 | 11.77 | 0.78 | 6.65 | 0.97 | 8.23 | 0.55 | 4.70 | 1.59 | 13.53 | 2.09 | 17.80* | Reproducibility with Control Material ** Population measured (Pop.) LY+: Lymphocyte; GR+: Granulocyte; MO+: Monocyte; StemTrol: StemTrol cells {13}------------------------------------------------ *Stemtrol Cells for CD34 measurements: The repeatability of the CD34 measurement included two additional unique variability sources. One of them was the variability contributed by the pipetting of the Stem-Trol reagent needed for staining. The other source was associated with an additional washing and decanting step. These additional sources of variability contribute to the increased imprecision. - iv. ClearLLab Reagents Lot-to-Lot Reproducibility: Lot-to-lot variability was evaluated using normal whole blood specimens. For the CD34 marker, which is not expressed in normal specimens, a specific cell line (Stem-Trol) was spiked in the specimens. Three normal whole blood specimens were tested in triplicate with three lots of each ClearLLab reagent. Mean Fluorescence Intensity (MFI) differences of the conjugates and percent positive cell recovery for each cell lineage population were evaluated between lots and all demonstrated acceptable lot to lot variability performance. | ClearLLab Reagents | Cells | Markers | |-----------------------------------------------|---------------------|---------------------------------------------------| | T1 Lineage Tube<br>CD2/CD56/CD7/CD5/CD45 | NK | CD45bright CD2+CD7+ CD5-CD56+ | | T2 Lineage Tube<br>CD8/CD4/-/CD3/CD45 | T | CD45bright CD2+CD7+ CD5+CD56- | | T2 Lineage Tube<br>CD8/CD4/-/CD3/CD45 | T Helper | CD45bright CD3+CD4+ | | T2 Lineage Tube<br>CD8/CD4/-/CD3/CD45 | T Suppressor | CD45bright CD3+CD8+ | | B1 Lineage Tube<br>Kappa/Lambda/CD19/CD5/CD45 | B#1 | CD45bright CD19+CD5-Kappa+ | | B1 Lineage Tube<br>Kappa/Lambda/CD19/CD5/CD45 | B#2 | CD45bright CD19+CD5-Lambda+ | | B2 Lineage Tube<br>CD20/CD10/CD19/CD38/CD45 | B#3 | CD45bright CD19+CD20+<br>CD38dim/variable CD10- | | B2 Lineage Tube<br>CD20/CD10/CD19/CD38/CD45 | Mature Granulocytes | CD10+ | | M Lineage Tube<br>CD7/CD13/CD34/CD33/CD45 | Mature Neutrophils | CD45+ Dim, High SS CD7-CD13+CD34-<br>CD33+ | | M Lineage Tube<br>CD7/CD13/CD34/CD33/CD45 | Monocytes | CD45+ Med, Med SS CD7-CD13+ CD34-<br>CD33+ Bright | Normal population analyzed in each Clear LLab reagent Normal population % positive recovery on Sample 1 | Sample 1 | N | Mean | Repeatability | | Between Lots | | Within Sample | | |---------------------|---|-------|---------------|------|--------------|------|---------------|------| | | | | SD | %CV | SD | %CV | SD | %CV | | NK | 9 | 2.58 | 0.13 | 5.16 | 0.00 | 0.00 | 0.13 | 5.16 | | T | 9 | 95.79 | 0.11 | 0.12 | 0.00 | 0.00 | 0.11 | 0.12 | | T Helper | 9 | 46.68 | 0.25 | 0.54 | 0.00 | 0.00 | 0.25 | 0.54 | | T Suppressor | 9 | 41.14 | 0.36 | 0.88 | 0.00 | 0.00 | 0.36 | 0.88 | | B #1 | 9 | 62.91 | 0.46 | 0.73 | 0.50 | 0.79 | 0.68 | 1.08 | | B #2 | 9 | 37.09 | 0.47 | 1.28 | 0.46 | 1.23 | 0.66 | 1.77 | | B #3 | 9 | 66.24 | 0.72 | 1.08 | 0.36 | 0.54 | 0.80 | 1.21 | | Mature Granulocytes | 9 | 56.40 | 1.61 | 2.86 | 0.60 | 1.06 | 1.72 | 3.05 | | Mature Neutrophils | 9 | 58.70 | 1.35 | 2.30 | 0.00 | 0.00 | 1.35 | 2.30 | | Monocytes | 9 | 2.49 | 0.05 | 1.84 | 0.04 | 1.72 | 0.06 | 2.52 | {14}------------------------------------------------ | Sample 2 | N | Mean | Repeatability | | Between Lots | | Within Sample | | |---------------------|---|-------|---------------|------|--------------|------|---------------|------| | | | | SD | %CV | SD | %CV | SD | %CV | | NK | 9 | 8.58 | 0.30 | 3.45 | 0.26 | 3.00 | 0.39 | 4.57 | | T | 9 | 88.26 | 0.29 | 0.33 | 0.00 | 0.00 | 0.29 | 0.33 | | T Helper | 9 | 70.56 | 0.39 | 0.55 | 0.57 | 0.80 | 0.69 | 0.98 | | T Suppressor | 9 | 23.59 | 0.48 | 2.02 | 0.51 | 2.18 | 0.70 | 2.97 | | B #1 | 9 | 64.67 | 2.20 | 3.41 | 0.62 | 0.96 | 2.29 | 3.54 | | B #2 | 9 | 35.36 | 2.27 | 6.43 | 0.65 | 1.82 | 2.36 | 6.68 | | B #3 | 9 | 59.28 | 2.24 | 3.78 | 0.76 | 1.29 | 2.37 | 4.00 | | Mature Granulocytes | 9 | 68.03 | 1.13 | 1.66 | 0.74 | 1.09 | 1.35 | 1.99 | | Mature Neutrophils | 9 | 70.37 | 2.31 | 3.28 | 0.00 | 0.00 | 2.31 | 3.28 | | Monocytes | 9 | 4.30 | 0.32 | 7.37 | 0.00 | 0.00 | 0.32 | 7.37 | Normal population % positive recovery on Sample 2 Normal population % positive recovery on Sample 3 | Sample 3 | N | Mean | Repeatability | | Between Lots | | Within Sample | | |---------------------|---|-------|---------------|------|--------------|------|---------------|------| | | | | SD | %CV | SD | %CV | SD | %CV | | NK | 9 | 13.30 | 0.26 | 1.93 | 0.23 | 1.76 | 0.35 | 2.61 | | T | 9 | 80.68 | 0.31 | 0.38 | 0.00 | 0.00 | 0.31 | 0.38 | | T Helper | 9 | 64.99 | 0.35 | 0.54 | 0.50 | 0.77 | 0.61 | 0.94 | | T Suppressor | 9 | 29.63 | 0.28 | 0.96 | 0.37 | 1.26 | 0.47 | 1.59 | | B #1 | 9 | 57.65 | 1.64 | 2.85 | 0.00 | 0.00 | 1.64 | 2.85 | | B #2 | 9 | 42.20 | 1.68 | 3.98 | 0.00 | 0.00 | 1.68 | 3.98 | | B #3 | 9 | 72.65 | 1.48 | 2.04 | 1.14 | 1.57 | 1.87 | 2.57 | | Mature Granulocytes | 9 | 63.57 | 1.55 | 2.44 | 0.43 | 0.67 | 1.61 | 2.53 | | Mature Neutrophils | 9 | 58.29 | 1.05 | 1.80 | 0.00 | 0.00 | 1.05 | 1.80 | | Monocytes | 9 | 5.02 | 0.21 | 4.19 | 0.05 | 0.97 | 0.22 | 4.30 | - v. Reagents Required as listed in Table 1 in Section I above: All tests performed for the reagents listed in Table 1 above for reproducibility and stability were acceptable. - Linearity/Reportable Range: b. Instrument Linearity : A study was performed for instrument linearity to evaluate the linearity and dynamic range of the FC500 flow cytometer's acquisition system, from the photomultiplier tubes (PMT) to the electronics, independent of the application and gating methodology. The FC 500 flow cytometer's acquisition system demonstrated linear results independent of the application and gating methodology. - Detection Limit: The study was used to establish the detection capability, i.e., the ability ﻥ to differentiate between abnormal and normal populations, of the ClearLLab reagents. - i. Limit of Blank (LoB): The LoB was determined by testing five replicates of three normal specimens run with two lots of each ClearLLab reagent (B1, B2, T1, T2 and M). The highest LoB {15}------------------------------------------------ observed from the two lots was 0.27% population of abnormal cells. - ii. Limit of Detection (LoD): The LoD was determined using one representative clinical specimen collected in K2 or K3 EDTA for each ClearLLab reagent. Each clinical specimen was spiked into a normal specimen to target low levels of abnormal population as a percentage of the total percentage of leukocytes in the sample. The percentages used were 2%, 1% and 0.5% of abnormal cells in the total population of white cells measured. The Limit of Detection defined as 1% population of abnormal cells for the ClearLLab application is therefore confirmed. - d. Analytical Specificity: - i. Interfering Substances: Not applicable. Adequate washing with phosphate buffered saline (PBS) before staining with reagents removes residual plasma and interfering substances. - ii. Assay Carryover: Carryover studies were conducted on three FC 500 flow cytometry systems according to the methodology presented in CLSI H26-A2: Validation, Verification, and Ouality Assurance of Automated Hematology analyzers; Approved Standard-Second Edition, Section 5.7 with regard to high or low value sample order. Two studies were conducted to evaluate specimen and reagent carryover. #### Specimen Carryover Three samples with high WBC counts (HTv1, HTv2, HTv3) were prepared from one clinical whole blood specimen containing a high level of WBCs. The samples were prepared to target 20,000 cells /uL WBCs. CD34 marker analysis was performed using a CD34+ clinical whole blood specimen. Three samples with low WBC counts (LTv1, LTv2, LTv3) were prepared by spiking a diluted (1:10,000) whole blood specimen with a red blood cell pool which has been depleted of white blood cells and diluted 4:1 with Immuno-Trol storage buffer. This resulted in a contrived specimen that had a red blood cell concentration similar to whole blood with very low concentrations of white blood cells. The reagent selected for specimen carryover was the ClearLLab M reagent. M reagent was chosen as a representative reagent since it contains all the different types of fluorochrome conjugates used in the device (FITC, PE, ECD, PC5.5 and PC7) as well as markers that allow identification and assessment of recovery of all leukocyte populations (lymphocytes, monocytes and granulocytes). Testing with M reagent also assesses the impact of lysing and fixative reagents on integrity of cell membranes for all relevant cell subsets and the affinity of various conjugated antibodies that is representative of all ClearLLab reagents. Three HTv samples (HTv1, HTv2, HTv3) were analyzed consecutively followed immediately by the three LTv samples (LTv1, LTv2, LTv3). This was repeated three {16}------------------------------------------------ times on three FC 500 flow cytometers for a total of nine specimen carryover runs. No significant carryover was detected as determined by percent carryover for specimen and reagent calculated from the analysis of the averages of the three runs on each flow cytometer for LTv1, LTv3, and HTv3. The specimen carryover is less than 1% for all markers. #### Reagent Carryover High concentration samples (HTv1, HTv2, HTv3): Normal whole blood specimens or clinical specimen (for CD34 measurement) stained with ClearLLab reagents were used as the high concentration samples. Low concentration samples (LTv1, LTv2, LTv3): The same normal blood specimen or clinical whole blood as in the high concentration sample was processed in triplicate using antibody dilution buffer (RD1), in place of antibody reagent. All the ClearLLab reagents (B1, B2, T1, T2 and M reagents) were selected for the reagent carryover studies. The potential for carryover of these products and the subsequent impact on performance were assessed for FS, SS, and FL1-FL5, in all seven of the detection channels. The testing consisted of a high concentration sample analyzed consecutively three times (HTv1. HTv2. HTv3) followed immediately by testing the low concentration sample three times (LTv1, LTv2, LTv3). This run was repeated three times on three FC 500 flow cytometers for a total of nine reagent carryover runs. The reagent carryover is less than 1% for all markers demonstrating acceptable scatter and fluorescent carryover performance. - e. Traceability and Stability (Reagent and Specimen): - i. Traceability: For monoclonal antibodies. Leukocvte Typing Workshop information for each antibody clone was provided. Each clone identity is controlled and is traced back to the original source. Polyclonal antibodies for identification of Kappa or Lambda cell markers are supported by reference literature. | # | Marker | Clone/Ig Chain<br>(species) | Leukocyte<br>Typing<br>International<br>Workshop* | Reactivity | |----|--------|-----------------------------------|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------| | 1 | CD2 | 39C1.5 IgG2a (rat) | Second | T cells and most of the NK cells | | 2 | CD3 | UCHT1 IgG1<br>(mouse) | First | Mature T cell (cytoplasmic<br>expression in immature T cells) | | 3 | CD4 | SFCI12T4D11<br>IgG1 (mouse) | First | Helper / inducer T cells,<br>monocytes, immature myeloid | | 4 | CD5 | BL1a IgG2a<br>(mouse) | Third | Thymocytes, mature T cells,<br>subpopulation of B cells | | 5 | CD7 | 8H8.1 IgG2a<br>(mouse) | Second | T cells, NK cells, subpopulation<br>of immature myeloid cells | | 6 | CD8 | SFCI21Thy2D3<br>(T8) IgG1 (mouse) | Second | Cytotoxic / suppressor T cells,<br>subpopulation of NK cells | | 7 | CD10 | ALB1 IgG1<br>(mouse) | Third | Common acute leukemia antigen<br>(CALLA), lymphatic precursor<br>cells, neutrophils, subpopulation<br>of mature B cells | | 8 | CD13 | SJ1D1 366 IgG1<br>(mouse) | Third | Mature and immature myeloid<br>cells | | 9 | CD19 | J3-119 IgG1<br>(mouse) | Fourth | Precursor and mature B cells | | 10 | CD20 | B9E9 (HRC20)<br>IgG2a (mouse) | Fifth | B cells and a subpopulation of B<br>precursor cells | | 11 | CD33 | D3HL60.251 IgG1<br>(mouse) | Fifth | Monocytes, myeloid precursor<br>cells, weak on neutrophils | | 12 | CD34 | 581 IgG1 (mouse) | Fifth | Myeloid and lymphoid precursor<br>cells | | 13 | CD38 | LS198-4-3 IgG1<br>(mouse) | Fifth | Activated lymphocytes,<br>subpopulation of B cells,<br>plasma cells | | 14 | CD45 | J.33 IgG1 (mouse) | Third | All leukocytes | | 15 | CD56 | N901/NKH-1 IgG1<br>(mouse) | Second | NK cell subset and activated T-<br>cells | | 16 | Kappa | N/A: Polyclonal | N/A | Subpopulation of immature B<br>Lymphocytes, subpopulation<br>of mature B cells | | 17 | Lambda | N/A: Polyclonal | N/A | Subpopulation of immature B<br>Lymphocytes, subpopulation<br>of mature B cells | {17}------------------------------------------------ {18}------------------------------------------------ | Workshop | | CD2 assigned | Number of CDs assigned | |-----------|--------------|--------------|------------------------| | I First | Paris 19822 | CD1-CDw15 | 15 | | II Second | Boston 19843 | CD16-CDw26 | 11 | | III Third | Oxford 19874 | CD27-CD45 | 19 | | IV Fourth | Vienna 19895 | CD46-CDw78 | 33 | | V Fifth | Boston 19936 | CD79-CDw109 | 31 | *Human Leukocyte Differentiation antigens Workshops ### ii. Reagent Stability: Testing of reagent stability was based on CLSI EP-25A (Evaluation of Stability of In vitro Diagnostic Reagents, Approved Guideline. Three lots of each ClearLLab reagent were evaluated for real-time open- and closed-vial stability. This study evaluated recovery and percent positive cells for each marker of the different lineage reagents (T1/T cell Lineage Markers (TL) 1, T2/TL2, B1/B cell Lineage markers (BL) 1, B2/BL2. M/Myeloid Lineage markers). Closed-vial stability testing was performed with stabilized control cells: Immuno-Trol and Immuno-Trol spiked with Stem-Trol to assess CD34 marker expression. Immuno-Trol Control Cells were used as a standardized assay system to enable traceability throughout the length of the study. Open-vial stability testing was performed with normal whole blood samples spiked with Stem-Trol cell line to assess the CD34+ cell population. Open-vial testing was performed at six time points with ClearLLab reagents. Percent difference from Time zero percent positive are measured for each marker of the different lineages tubes. The open- and closed-vial stability claims at refrigerated temperature (2-8℃) conditions are 90 and 365 days respectively. ### iii. Whole Blood and Bone Marrow Specimen Stability: A total of 45 clinical specimens consisting of 10 bone marrow specimens at each of three sites per anticoagulant (n = 30) and five whole blood specimens at each of three sites per anticoagulant (n = 15) were collected with different anticoagulants. Specimens provided from Site 1 were collected in ACD, from Site 2 were collected in Heparin, and from Site 3 were collected in K2EDTA. Each specimen was stained with the ClearLLab reagents for the B-cell Lineage Tube 1 and 2, T-cell Lineage Tube 1 and 2 and Myeloid Lineage reagent panels. Specimens were held at room temperature for the indicated time and prepared and analyzed at the following target time points post specimen collection: T0, T28, and T52 hours. The study examined the combined effect <sup>2</sup> Bernard AR, Boumsell L, Dausset J, et al. eds. Leucocyte Typing: Human Leucocyte Differentiation Antigens Detected by Monoclonal antibodies. Berlin: Springer-Verlag, 1984. <sup>3</sup> Reinherz EL, Haynes BF, Nadler L, Bernstein ID, eds. Leukocyte Typing II. New York: Springer-Verlag, 1985. <sup>4</sup> McMichael AJ, Beverley PCL, Cobbold S, et al. eds. Leucocyte Typing III. White Cell Differentiation Antigens. Oxford: Oxford University Press, 1987. <sup>3</sup> Knapp W, Dorken B, Gilks W et al., eds. Leucocyte Typing IV. Oxford University Press, 1989. Schlossman SF, Boumsell L, Gilks W, et al. eds. Leucocyte Typing V: White cell differentiation antigens. Oxford: Oxford University Press, 1995. {19}------------------------------------------------ of specimen and prepared sample stability drift of ClearLLab reagents' CD marker panel percent positive recoveries. The results at room temperature (RT, 20-25°C) are presented in the table below: | | Whole Blood | | Bone Marrow | | | |---------------|--------------------------------|---------------------------------|--------------------------------|---------------------------------|-----------| | Anticoagulant | Specimen<br>stability<br>C(RT) | Prepared<br>Sample<br>stability | Specimen<br>stability<br>C(RT) | Prepared<br>Sample<br>stability | | | | EDTA | 24 hours | < 3 hours | 24 hours | < 3 hours | | | ACD | 48 hours | < 3 hours | 48 hours | < 3 hours | | Heparin | 48 hours | < 3 hours | 48 hours | < 3 hours | | C(RT): Condition (Room Temperature 20-25°C) #### iv. Specimen Stability (Lymph Node): A total of twelve clinical specimens consisting of four lymph node specimens per site were collected at three external sites. The prepared sample stability was assessed for the applicable markers CD19+, CD19+Kappa+, CD19+Lambda+, CD5+, CD20+, CD19+CD38+, CD7+, CD2+, CD56+, CD3+CD4+,CD3+CD8+, and CD3+. These lymphocyte-specific CD markers are detected on lymph node specimen types. Specimens were prepared and analyzed in single replicate on a total of three FC 500 flow cytometers. Specimens were…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...