DEN140032 · Cereve, Inc. · PLU · May 13, 2016 · Neurology
Device Facts
Record ID
DEN140032
Device Name
Cereve Sleep System
Applicant
Cereve, Inc.
Product Code
PLU · Neurology
Decision Date
May 13, 2016
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 882.5700
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cerêve Sleep System is indicated to reduce sleep latency to Stage 1 and Stage 2 sleep in patients with primary insomnia.
Device Story
Cerêve Sleep System is a prescription home-use device for primary insomnia. It consists of a bedside unit, Lycra headgear, and a urethane forehead pad. The bedside unit uses thermoelectric cooling to circulate chilled fluid (purified water/isopropyl alcohol) through the forehead pad at patient-selectable temperatures (14–16 °C). The device is operated by the patient at home. Firmware controls the pump and temperature regulation via pulse-width modulation (PWM) and thermistor feedback. The patient wears the pad on the forehead during sleep. By applying localized cooling, the device aims to reduce the time required to reach Stage 1 and Stage 2 sleep. Healthcare providers use the device to offer a non-pharmacological alternative for sleep latency reduction. Safety features include automatic shutdown for fault conditions and hardware-based thermal cut-outs.
Clinical Evidence
No clinical data provided in the document; classification based on non-clinical performance testing, software verification/validation, biocompatibility, electrical safety, and electromagnetic compatibility testing.
Technological Characteristics
Thermal system for skin surface application. Requires biocompatibility assessment of patient-contacting components. Includes electrical safety and electromagnetic compatibility features. Subject to mechanical testing for leakage and structural integrity under anticipated use conditions. Software-controlled thermal regulation.
Indications for Use
Indicated for patients with primary insomnia to reduce sleep latency to Stage 1 and Stage 2 sleep.
Regulatory Classification
Identification
A thermal system for insomnia is a prescription device for use in patients with insomnia that is used to apply a specified temperature to the skin surface.
Special Controls
In combination with the general controls of the FD&C Act, the Thermal System for Insomnia is subject to the following special controls: 1. The patient-contacting components of the device must be demonstrated to be biocompatible. 2. Performance testing must demonstrate electromagnetic compatibility and electrical safety. 3. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated: a. Thermal performance of the device, including maintenance of the target temperature, must be evaluated under simulated use conditions. Mechanical testing to demonstrate the device can withstand forces under b. anticipated use conditions. c. Mechanical testing to demonstrate the device is resistant to leakage under anticipated use conditions. 4. Software verification, validation, and hazard analysis must be performed. 5. Patient labeling must be provided to convey information regarding safe use of the device, including instructions for assembly.
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Performance testing must demonstrate electromagnetic compatibility and electrical safety.
(3) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
(i) Thermal performance of the device, including maintenance of the target temperature, must be evaluated under simulated use conditions.
(ii) Mechanical testing to demonstrate the device can withstand forces under anticipated use conditions.
(iii) Mechanical testing to demonstrate the device is resistant to leakage under anticipated use conditions.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Patient labeling must be provided to convey information regarding safe use of the device, including instructions for assembly.
Submission Summary (Full Text)
{0}------------------------------------------------
#### DE NOVO CLASSIFICATION REQUEST FOR CERÊVE SLEEP SYSTEM
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Thermal System for Insomnia. A thermal system for insomnia is a prescription device for use in patients with insomnia that is used to apply a specified temperature to the skin surface.
NEW REGULATION NUMBER: 882.5700
CLASSIFICATION: CLASS II
PRODUCT CODE: PLU
#### BACKGROUND
DEVICE NAME: CERÊVE SLEEP SYSTEM
SUBMISSION NUMBER: DEN140032
DATE OF DE NOVO: OCTOBER 20, 2014
CONTACT: CERÊVE, INC. 333 ALLEGHENY AVE, SUITE 201 OAKMONT, PA 15139
#### REQUESTER'S RECOMMENDED CLASSIFICATION: CLASS II
#### INDICATIONS FOR USE
The Cerêve Sleep System is indicated to reduce sleep latency to Stage 1 and Stage 2 sleep in patients with primary insomnia.
#### LIMITATIONS
The sale, distribution, and use of the device are restricted to prescription use in accordance with 21 CFR 801.109.
Other than reduction of sleep latency to Stage 1 and Stage 2 sleep, the efficacy of other sleep measures associated with insomnia has not been established by the Cerêve Sleep System in controlled clinical trials.
Limitations on device use are also achieved through the following statements included in
{1}------------------------------------------------
the Directions for Use manual:
Warning:
- · If the patient has a skin condition on his/her forehead that may make wearing the Cereve Sleep System uncomfortable, the physician will need to evaluate the skin condition prior to use of the Sleep System.
- · Patients with cold sensitivity reactions such as Raynaud's Disease should not use cooling therapies in order to avoid cold sensitivity reactions that may include cold sensations, color changes in their skin in response to cold, and a numb, prickly feeling or stinging pain upon warming.
PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS AND PRECAUTIONS.
# DEVICE DESCRIPTION
The Cereve Sleep System is a cooling device comprised of three components: the bedside unit, the forehead pad, and headgear. The device pumps chilled fluid through the forehead pad, at patient selectable temperatures in a narrow range between 14 and 16 °C. An illustration of the device is provided in Figure 1 below.
Image /page/1/Picture/7 description: The image shows a medical device and a cartoon depiction of the device in use. On the left, the device is shown on a mannequin head next to a white box with a digital display. The device is a blue head covering with black accents. On the right, a cartoon shows a person in bed wearing the head covering, which is connected to a device on the nightstand.
Figure 1. Cerêve Sleep System (left) and schematic of device in home use environment (right).
The Cereve Sleep System bedside unit provides the means to cool the fluid and transport the fluid from the unit to the forehead pad. The bedside unit utilizes solid state thermoelectric devices to cool a thermal transfer fluid consisting of purified water and isopropyl alcohol. The unit has a user interface that allows the user to turn the unit on and off, and adjust the temperature within the range of 14 to 16 ℃. The unit contains a pump for circulating the thermal transfer fluid through the tubing and forehead pad. The bedside unit is powered by a DC electrical power supply and is controlled by an integral control unit (CU) and its firmware. The Cerêve Sleep System Headgear and Forehead pad contain the wearable portion of the sleep
{2}------------------------------------------------
system. It is comprised of a multi-use urethane forehead pad that is in contact with the patient's head, the Lycra® headgear that holds the forehead pad in place, and a 6 foot section of insulated tubing that connects to the Cerêve Bedside Unit.
### SUMMARY OF NONCLINICAL/BENCH STUDIES
# BIOCOMPATIBILITY/MATERIALS
The forehead pad is constructed of urethane and is in direct contact with the patient's skin. Each use of the device results in a limited contact duration of less than 24 hours. Therefore, per ISO 10993-1 (Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing), cytotoxicity, sensitization, and irritation testing were performed on the forehead pad and the results were found to be acceptable.
The thermal fluid is not intended to contact the patient; however, if the forehead pad is damaged, there could be potential for the fluid to come in contact with the patient during use. Therefore, the thermal fluid also was tested per ISO 10993-1 for cytotoxicity, oral toxicity, and ocular irritation and the results were found to be acceptable.
The headgear is constructed of clothing grade Lycra®, which is commonly used to fabricate clothing in contact with intact skin. In lieu of biocompatibility testing, justification was provided that the same material has a demonstrated history of safe use.
# STERILITY
The Cerêve Sleep System is a non-sterile, reusable device. It is intended only for external use. Cleaning instructions for the bedside unit and forehead pad are provided in the labeling.
### ELECTROMAGNETIC COMPATIBILITY (EMC) AND ELECTRICAL SAFETY
The Cereve Sleep System was tested for and found to be in compliance with the following standards for electromagnetic compatibility and electrical safety:
| Standard | Title |
|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AAMI ANSI ES60601-<br>1:2005/(R)2012 and<br>A1:2012, C1:2009/(R)2012<br>and a2:2010/(r)2012 | (Consolidated Text) Medical Electrical Equipment -<br>Part 1: General Requirements For Basic Safety And<br>Essential Performance |
| AAMI ANSI IEC 60601-1-<br>2:2014 | Medical Electrical Equipment -- Part 1-2: General<br>Requirements For Basic Safety And Essential<br>Performance -- Collateral Standard: Electromagnetic<br>Disturbances -- Requirements And Tests |
| IEC 60601-1-11 Edition 1.0<br>2010-04 | Medical Electrical Equipment - Part 1-11: General<br>Requirements For Basic Safety And Essential<br>Performance - Collateral Standard: Requirements For<br>Medical Electrical Equipment And Medical Electrical<br>Systems Used In The Home Healthcare Environment |
Table 1 - Summary of EMC and Electrical Safety Testing
{3}------------------------------------------------
# SOFTWARE
The Cerêve Sleep System bedside unit is powered by a DC electrical power supply and is controlled by an integral control unit (CU) and its firmware.
The Level of Concern of the software for the Cerêye Sleep System was assessed using the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and resulted in a determination of a Moderate level of concern.
- The software regulates the temperature of thermal fluid at the output of the bedside . unit to the temperature set point selected by the patient, and maintains the temperature set point without requiring additional patient adjustment by pulse-width modulation (PWM).
- The software will allow for variation of the fluid temperature set point at the output of . the bedside unit in preset increments based on patient adjustment.
- The software also includes additional safety features that provide for automatic shutdown for fault conditions. All of the fault conditions are backed up by hardware controls in the case of software loss of control.
Software validation was performed on all key performance characteristics for temperature control. Additionally, software validation was performed on all safety critical performance features for the following requirements: unit temperature (thermistor control), unit temperature (thermal cut-out), fluid temperature (thermistor control), and fluid temperature (thermal cut-out).
### PERFORMANCE TESTING - BENCH
The Cerêve Sleep System was tested to characterize the device outputs at each temperature setting on the controller of the bedside unit. The non-clinical bench tests performed to support the performance and safety of the Cerêve Sleep System are summarized in the table below:
| Test | Test Summary | | |
|-----------------------|-------------------------------------------------------------------|--|--|
| Fluid flow rate | The fluid flow rate through the forehead pad was evaluated and | | |
| | verified to be within specification. | | |
| Time to reach target | The time for the device to reach the target temperature was | | |
| temperature | verified to be within specification. | | |
| Maintenance of target | The ability of the device to maintain the target temperature | | |
| temperature | setting for 8 hours was verified while operating at an ambient | | |
| | temperature of (4) and applying heat to the forehead pad. | | |
| Sound level | The sound level generated from the device during use was | | |
| | verified to be in an acceptable range. | | |
| Temperature drop | The temperature difference between the outlet of the forehead | | |
| across forehead pad | pad and inlet of the forehead pad was verified to be less than or | | |
| | equal to b(4 while applying heat to the forehead pad. | | |
Table 2 - Summary of Non-clinical Testing
{4}------------------------------------------------
| Test | Test Summary |
|------------------------|-------------------------------------------------------------------|
| Temperature drop | The temperature difference between the forehead pad and |
| from bedside unit and | bedside unit was verified to be less than or equal to b(4) |
| forehead pad | |
| Tubing and connector | The tubing connection strength was verified to withstand a |
| strength | minimum tensile force. |
| Puncture load on | The forehead pad integrity was verified to withstand a minimum |
| forehead pad | puncture load force without rupture. |
| Kinked/occluded | The device was verified not to leak when the tubing was |
| tubing and leakage | simulated in a kinked or occluded condition. The forehead pad |
| testing | assembly was also verified not to leak when exposed to tensile |
| | forces beyond those anticipated during use conditions. |
| Simulated distribution | The packaged device was confirmed to be undamaged and in |
| and conditioning | operable condition following simulated distribution conditions in |
| | accordance with ASTM D4169-14 (Standard Practice for |
| | Performance Testing of Shipping Containers and Systems) and |
| | conditioning in accordance with ASTM D4332-14 (Standard |
| | Practice for Conditioning Containers, Packages, or Packaging |
| | Components for Testing). |
# SUMMARY OF CLINICAL INFORMATION
Three clinical studies were conducted with the Cereve Sleep System in patients with primary insomnia. The studies are further summarized below.
### 1. CIP-006:
Design: This was the pivotal study in support of the Cerêve Sleep System. This was a randomized, multi-center (7 U.S. sites), sham-controlled study of 116 primary insomnia adult subjects.
Eligible subjects were provided with information about the study design and risks and signed the informed consent document. Criteria for inclusion included:
- Age > 22 years
- Diagnosis of insomnia that meets criteria for Diagnostic and Statistical Manual of . Mental Disorders, 4th Edition (DSM IV) diagnosis of primary insomnia and International Classification of Sleep Disorders (ICSD) general insomnia and Research Diagnostic Criteria (RDC) insomnia disorder criteria
- Agreement to remain alcohol free and avoid drugs that could affect sleep during the ● studv
- >14 on the Insomnia Severity Index ●
- Sleep-Wake diary that demonstrated sleep efficiency <85% on at least 50% of nights ● over a 7 day consecutive period
Criteria for exclusion included:
- . Neuropsychiatric disorders that may affect sleep, brain function or cognition, such as current major syndromal psychiatric disorders, including DSM-IV mood, anxiety, psychotic, and substance use disorders
{5}------------------------------------------------
- . Specific exclusionary diagnoses included major depressive disorder, dysthymic disorder, bipolar disorder, panic disorder, obsessive compulsive disorder, generalized anxiety disorder, any psychotic disorder, and any current substance use disorder. Unstable medical conditions including severe cardiac, liver, kidney, endocrine (e.g., diabetes), hematologic (e.g., porphyria or any bleeding abnormalities), other impairing or unstable medical conditions or impending surgery, central nervous system disorders (e.g., head injury, seizure disorder, multiple sclerosis, tumor), active peptic ulcer disease, inflammatory bowel disease, and arthritis (if the arthritis impacts sleep)
- Raynaud's disease ●
- Irregular sleep schedules including shift workers
- Latency to persistent sleep < 15 minutes on either the sleep disorder screening night or baseline polysomnograph (PSG)
- A sleep efficiency >85% on either the sleep disorder screening night or baseline PSG ●
- An AHI (apnea hypopnea index) >10 and/or a periodic limb movement arousal index . (PLMAI)>15 from screening night
- Body mass index >34
- Use of medications known to affect sleep or wake function
- Consumption of more than one alcoholic drink per day
- Consumption of caffeinated beverages >4 per day ●
Following completion of 2 nights baseline PSG recordings in a sleep lab setting, subjects were randomized to receive 2 additional nights PSG sleep studies using either the Cerêve Sleep System or the sham device. There were 2 co-primary endpoints for the study, including (1) latency to persistent sleep (time to the first epoch of sleep sustained for 10 minutes) based on PSG and (2) sleep efficiency (total sleep time / time in bed) based on PSG. The two secondary endpoints for the study were Stage 3 non-rapid eve movement (NREM) sleep based on PSG and subjective sleep quality based on response to a self-reported questionnaire (Pittsburgh Sleep Diary) following each night of sleep.
Results: There were no statistically significant differences between the Cerêve and sham groups for the primary endpoints (latency to persistent sleep and sleep efficiency), nor for the secondary endpoints (Stage 3 NREM sleep and sleep quality). However, additional data analysis of the study showed there was a 12 minute reduction in the sleep latency to Stage 1 sleep (p=0.004) and Stage 2 sleep (p=0.008) in the Cereve group compared to the sham. The measures of sleep latency to Stage 1 and Stage 2 sleep did not require sustained sleep. There was no statistically significant difference between the Cerêve and sham groups for reduction in sleep latency to Stage 3 sleep.
Adverse Events: There were a total of 3 adverse events (AEs) in the Cerêve group and 1 in the sham group that were deemed possibly or probably related to the device. The 3 AEs in the Cerêve group included headache. There were an additional 2 AEs in the Cerêve group and 1 AE in the sham group that were not device related. No serious AEs were observed in either group.
{6}------------------------------------------------
# 2. CIP-003:
Prior to conducting CIP-006, the sponsor conducted a multi-center, controlled, within subjects cross-over study comparing the effects of 2 different temperature modes of the Cerêve Sleep System on subjects with primary insomnia. This was a multi-center (9 U.S. sites) study of 145 subjects with inclusion and exclusion criteria that were similar to those in CIP-006. Following screening and baseline PSG measurements, subjects had two sequential nights in each condition with the Cereve Sleep System separated by 3-5 nights at home. There were 2 co-primary endpoints for the study, including (1) latency to persistent sleep based on PSG and (2) sleep efficiency based on PSG.
Similar to CIP-006, there were numerical improvements in latency to persistent sleep and sleep efficiency in both groups, but there were no statistically significant differences between the two modes. The rate of adverse events was low and further supported the overall safety of the Cerêve Sleep System. The most common AE was headache, with a total of 11 headache AEs in 10 subjects that were deemed possibly, probably or definitely related to the device. Other AEs were deemed not to be device-related. No serious AEs were observed.
# 3. CIP-004:
Following CIP-003, the sponsor conducted a non-randomized home-use study to evaluate long-term use effects of the Cerêve Sleep System in 32 subjects from CIP-003. Subjects used the Cereve Sleep System in their home for 30 days and up to 5 months for safety assessments. The subjects had similar numerical improvements in latency to persistent sleep and sleep efficiency measures at 30 days as those observed in prior measurements obtained in CIP-003. The subjects reported improved self-reported sleep quality scores after 30 days use. These results demonstrate that the Cereve Sleep System effect is durable over 30 days of use in the home in these subjects. No serious device-related AEs were reported and further supported the safety of the Cerêve Sleep System for extended use.
# LABELING
The Cerêve Sleep System User Manual is consistent with the clinical data and covers all of the hazards and other clinically relevant information that may impact use of the device. Patient labeling is also provided, which includes a warning for patients to consult with their physician if they have a skin condition that may make use of the device uncomfortable. The labeling includes the following caution statement noting the limitation of the established device effectiveness:
Other than reduction of sleep latency to Stage 1 and Stage 2 sleep, the efficacy of other sleep measures associated with insomnia has not been established by the Cerêve Sleep System in controlled clinical trials.
The labeling satisfies the requirements of 21 CFR & 801.109 Prescription devices.
{7}------------------------------------------------
# RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of the Thermal System for Insomnia and the measures necessary to mitigate these risks.
| Identified Risk | Mitigation Method |
|-----------------------------------------|----------------------------------------|
| Adverse Skin Reaction | Biocompatibility Assessment |
| | Labeling |
| Electromagnetic Interference with Other | Electromagnetic Compatibility Testing |
| Devices | Labeling |
| Electrical Safety (e.g., shock) | Electrical Safety Testing |
| | Labeling |
| Thermal Injury | Non-clinical Performance Testing |
| | Software Verification, Validation, and |
| | Hazard Analysis |
| | Labeling |
| Table 3 - Identified Risks to Health and Mitigation Measures | |
|--------------------------------------------------------------|--|
|--------------------------------------------------------------|--|
# SPECIAL CONTROLS:
In combination with the general controls of the FD&C Act, the Thermal System for Insomnia is subject to the following special controls:
- 1. The patient-contacting components of the device must be demonstrated to be biocompatible.
- 2. Performance testing must demonstrate electromagnetic compatibility and electrical safety.
- 3. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
- a. Thermal performance of the device, including maintenance of the target temperature, must be evaluated under simulated use conditions.
- Mechanical testing to demonstrate the device can withstand forces under b. anticipated use conditions.
- c. Mechanical testing to demonstrate the device is resistant to leakage under anticipated use conditions.
- 4. Software verification, validation, and hazard analysis must be performed.
- 5. Patient labeling must be provided to convey information regarding safe use of the device, including instructions for assembly.
### BENEFIT/RISK DETERMINATION
The risks of the device are based on data collected in the clinical studies described above. There were no serious adverse events associated with the use of the device.
The probable benefits of the device are also based on data collected in the clinical studies described above. Although the clinical studies do not support that the device is effective for improvement of primary insomnia symptoms in general (including latency to persistent sleep,
{8}------------------------------------------------
sleep efficiency, and sleep quality), the probable benefit is a reduction in the time to reach stage 1 and stage 2 sleep.
Additional factors to be considered in determining probable risks and benefits for the Cerêve Sleep System include: risks, such as headache, will be apparent to the individual user of the device with use. Similarly, if there is benefit to the individual user it should be apparent. One alternative low-risk treatment includes cognitive behavioral therapy. A second alternative therapy includes pharmacologic therapy which may carry greater risk. There are no similar medical devices on the market for this proposed indication for a reduction in the time to reach Stage 1 and Stage 2 sleep.
In conclusion, given the available information above, the data support that for reducing latency to Stage 1 and Stage 2 sleep in patients with primary insomnia, the probable benefit outweighs the probable risks for the Cerêve Sleep System. The device provides probable benefit for reducing time to Stage 1 and Stage 2 sleep and the risks can be mitigated by the use of general and the identified special controls.
# CONCLUSION
The de novo request for the Cerêve Sleep System is granted and the device is classified under the following:
Product Code: PLU Device Type: Thermal System for Insomnia Class: II Regulation: 21 CFR 882.5700
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.