Clinical study data from routine urological examinations
Clinical studies were conducted to evaluate the device's ability to reconstruct prostate images and visualize abnormalities in patients who had already undergone a standard DRE. The data confirmed the device's usability and performance in a clinical setting.
2006 Clinical Study; Retrospective/Observational clinical evaluation; Study Period: 2006
Patients undergoing standard DRE; Sample Size: 168
Not applicable for this study
Image reconstruction capability
2009 Clinical Study; Observational clinical evaluation; Study Period: 2009
56 clinical cases with prostate abnormalities detected by DRE; Sample Size: 56; Number of Sites: 5
Not applicable for this study
Imaging capability, agreement with DRE, safety, ease of use
Indications for Use
the production of an elasticity image of the prostate as an aid in documenting prostate abnormalities that were previously identified by digital rectal examination (DRE). The device utilizes a transrectal probe with pressure sensor arrays and a motion tracking system and provides real-time elasticity images of the prostate. This device is limited to use as a documentation tool and therefore is not to be used for cancer diagnosis or for any other diagnostic purpose. This device is only to be used to image and document an abnormality that was already identified by DRE. Clinical management decisions are not to be made on the basis of information from the PMI device, but rather on the basis of the DRE. If there is disagreement between the DRE and the recorded image produced by the device, patient management decisions are to be based on the DRE and other available clinical and diagnostic information (e.g., prostate-specific antigen (PSA) levels) in accordance with standard medical practice.
Device Story
Prostate Mechanical Imager (PMI) uses transrectal probe with pressure sensor arrays and motion tracking system; captures mechanical tissue properties; image reconstruction software transforms sensor data into real-time elasticity images highlighting regional stiffness differences; used by clinicians to document abnormalities already identified via DRE; device output serves strictly as documentation aid; clinical management remains based on DRE and standard diagnostic markers (e.g., PSA); device prohibits use for cancer diagnosis or influencing clinical decisions.
Clinical Evidence
Non-clinical and clinical performance testing required to demonstrate accuracy and reproducibility of constructed images; specific metrics not provided in summary document.
Technological Characteristics
Transrectal probe with pressure sensor arrays; motion tracking system; image reconstruction software; requires biocompatibility, electromagnetic compatibility, electrical, thermal, and mechanical safety testing; reprocessing validation for reusable components.
Indications for Use
Indicated for patients with prostate abnormalities previously identified by digital rectal examination (DRE). Used as a documentation tool only; not for cancer diagnosis or clinical decision-making.
Regulatory Classification
Identification
A prostate lesion documentation system is a prescription device intended for use in producing an image of the prostate as an aid in documenting prostate abnormalities previously identified during a digital rectal examination. The device uses pressure sensors and image reconstruction software to produce a prostate image that highlights regional differences in intraprostatic tissue elasticity or stiffness. The device is limited to use as a documentation tool and is not intended for diagnostic purposes or for influencing any clinical decisions.
Special Controls
In combination with the general controls of the FD&C Act, the Prostate Mechanical Imager is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical and clinical performance testing must demonstrate the accuracy and reproducibility of the constructed image.
(2) Appropriate analysis/testing must validate electromagnetic compatibility, electrical safety, thermal safety, and mechanical safety.
(3) Appropriate software verification, validation, and hazard analysis must be performed.
(4) All elements of the device that may contact the patient must be demonstrated to be biocompatible.
(5) Methods and instructions for reprocessing of any reusable components must be properly validated.
(6) The labeling must include specific information needed to ensure proper use of the device.
Submission Summary (Full Text)
{1}------------------------------------------------
#### DE NOVO CLASSIFICATION REQUEST FOR PROSTATE MECHANICAL IMAGER
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Prostate lesion documentation system. A prostate lesion documentation system is a prescription device intended for use in producing an image of the prostate as an aid in documenting prostate abnormalities previously identified during a digital rectal examination. The device uses pressure sensors and image reconstruction software to produce a prostate image that highlights regional differences in intraprostatic tissue elasticity or stiffness. The device is limited to use as a documentation tool and is not intended for diagnostic purposes or for influencing any clinical decisions.
### NEW REGULATION NUMBER: 21 CFR 876.2050
CLASSIFICATION: CLASS II
PRODUCT CODE: OQT
#### BACKGROUND
DEVICE NAME: PROSTATE MECHANICAL IMAGER (PMI)
SUBMISSION NUMBER: DEN100016
DATE OF DE NOVO: MAY 21, 2010
- CONTACT: NOUNE SARVAZYAN, PH.D. CHIEF EXECUTIVE OFFICER ARTANN LABORATORIES, INC. 1459 LOWER FERRY ROAD TRENTON, NJ 08618
#### REQUESTER'S RECOMMENDED CLASSIFICATION: CLASS II
#### INDICATIONS FOR USE
The Prostate Mechanical Imager (PMI) is indicated for the production of an elasticity image of the prostate as an aid in documenting prostate abnormalities that were previously identified by digital rectal examination (DRE). The device utilizes a transrectal probe with pressure sensor arrays and a motion tracking system and provides real-time elasticity images of the prostate. This device is limited to use as a documentation tool and therefore is not to be used for cancer diagnosis or for any other diagnostic purpose. This device is only to be used to image and document an abnormality that was already identified by DRE. Clinical
DEN100016 De Novo Summary
{2}------------------------------------------------
management decisions are not to be made on the basis of information from the PMI device, but rather on the basis of the DRE. If there is disagreement between the DRE and the recorded image produced by the device, patient management decisions are to be based on the DRE and other available clinical and diagnostic information (e.g., prostate-specific antigen (PSA) levels) in accordance with standard medical practice.
#### LIMITATIONS
The sale, distribution, and use of the device are restricted to prescription use in accordance with 21 CFR 801.109.
Limitations on device use are also achieved through the following statements included in the user manual:
This device is limited to use as a documentation tool and therefore is not to be used for cancer diagnosis or for any other diagnostic purpose.
This device is only to be used to image and document an abnormality that was already identified by DRE.
Clinical management decisions are not to be made on the basis of information from the PMI device, but rather on the basis of the DRE. If there is disagreement between the DRE and the recorded image produced by the device, patient management decisions are to be based on the DRE and other available clinical and diagnostic information (e.g., PSA levels) in accordance with standard medical practice.
This device has not been shown to be useful as a diagnostic or management tool; therefore, the device is not to be used for such purposes, e.g., to track changes in the size of a lesion, to guide biopsy needle placement, or as a replacement for digital rectal examination.
The system automatically includes the following statements (prominently displayed with all screen images, electronically saved reports, and printed reports containing images):
Not to be used for diagnostic purposes.
Not to be used for making clinical management decisions.
The labeling includes the following instructions to not save or print a report if the images do not agree with the prior digital rectal examination findings:
CAUTION: Examination data should not be saved if the recorded prostate images do not agree with the DRE findings.
{3}------------------------------------------------
Do not save the exam or print an examination report if the recorded prostate images do not agree with the DRE findings.
The labeling includes the following additional precautions addressing limitations applicable to specific patient populations:
This device has not been shown to be effective in patients who have undergone prostate surgery, radiation therapy, cryotherapy, or chemotherapy for prior prostate cancer.
This device should not be used on patients with open wounds of the rectum or areas around the prostate, as this may increase the risk of tissue trauma, bleeding, and infection.
PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
#### DEVICE DESCRIPTION
The Prostate Mechanical Imager (PMI) is an electronic palpation device that is meant to mimic the digital rectal examination (DRE) by generating images of pressure patterns of the palpated prostate. The information provided by the device is characterized as being similar in nature to the information obtained from a standard DRE (i.e., determination of regions of relative tissue hardness within the prostate) with the utility being that the results are visually displayed and can be electronically saved, transmitted, and/or printed out for documenting in patient's medical records.
#### Principle of Operation:
- An image of the prostate is generated by the device based on the measurement of the . mechanical stress (pressure) patterns on the rectal wall over the prostate compressed by the transrectal probe.
- The device's ability to produce a color image of tissue abnormalities that were already . detected by DRE is accomplished by sensing differences in elasticity/hardness between normal and abnormal tissues. The pressure patterns obtained through the rectal probe's arrav of 128 pressure sensors are recorded and used to generate 2-D and 3-D elasticity images of the prostate.
- . The pressure sensor array, in response to the applied pressure, produces a pressure response pattern on a rectal wall covering the prostate, analogous to that sensed by the physician's finger palpating the prostate during a DRE. In general, normal tissue is less dense and more elastic, whereas cancerous tissue is firmer and less elastic. Elasticity in normal prostate tissue ranges between 15 and 30 kPa, whereas, the elasticity of most of prostate cancers ranges between 40 kPa and 300 kPa, which is an order of magnitude in hardness greater than normal tissue. With the PMI device, a series of cross-sectional images are obtained to
{4}------------------------------------------------
reconstruct the prostate and visualize the regions of varying tissue hardness within the prostate.
The PMI is a PC-based system consisting of the following components, which are pictured and described in more detail below:
- a hand-held transrectal probe with two pressure sensor arrays and a motion tracking system; (1)
- (2) a probe calibration chamber;
- a data acquisition unit; (3)
- (4) a central processing unit consisting of an off-the-shelf computer with proprietary software;
- an off-the-shelf touch screen monitor; (ર)
- (6) an off-the-shelf color printer (optional);
- an off-the-shelf medical grade keyboard; (7)
- a disposable probe sheath; (8)
- (9) a disposable system cover;
- (10) a 510(k) cleared lubricating jelly to be applied to the sheathed probe; and
- (11) a mobile cart.
The first pressure sensor array located at the tip of the probe is used to palpate the prostate. The second pressure sensor array located on the shaft of the transrectal probe shows the position of the probe relative to the sphincter, thereby providing an anatomical landmark helping to guide the probe manipulation. The motion tracking system built into the probe handle provides real time probe orientation data.
Image /page/3/Picture/14 description: The image shows a white ultrasound probe on a blue background. The probe has a long, curved handle and a smaller, rectangular head. There is a gray section between the head and the handle. A cord extends from the back of the handle.
Hand-Held Transrectal Probe
{5}------------------------------------------------
All components reside on the ergonomic mobile cart (pictured below) which also contains a medical grade isolation transformer to which all components are connected. There is a single power cord from this transformer that plugs into a standard electrical outlet.
Image /page/4/Picture/1 description: This image shows a medical device on a mobile cart. The device includes a touch screen monitor, calibration chamber, transrectal probe, master power button, data acquisition and processing unit, and keyboard. The mobile cart also has a color printer, power supply indicator, and isolation transformer. The device appears to be used for medical imaging or diagnostics.
PMI System Components on Mobile Cart
In order to perform the procedure, the user must first place a disposable sheath on the end of the probe and apply a lubricant to the sheath. The tip of the probe is then inserted into the patient's rectum and the operator applies pressure with the probe tip while scanning over the prostate. As the prostate pressure patterns are being detected, a color image of the prostate is displayed on the computer monitor, along with an indication of the amount of pressure being applied to the probe. Procedure time is typically 1 minute, but up to 5 minutes.
{6}------------------------------------------------
The generated images show gradations in detected pressenting regional differences in prostatic tissue elasticity/hardness) by mapping to a range of colors, i.e., a color-coded elasticity image. For example, areas of the image depicted in deep blue represent the areas of tissue with the lowest firmness levels, progressing to higher firmness levels represented by the colors green, yellow, orange, and then red (highest firmness), in that order.
Image /page/5/Figure/1 description: This image shows a medical imaging software interface. The interface displays patient information such as Patient ID: P1.3, Patient Name: Prostate Phantom, and Exam Time: 1/16/2009 4:04 PM. It also presents 2-D and 3-D images, with controls for adjusting the color map and layer visibility, along with options to open, save, and print the exam report. The system message indicates that the user can review stored examination data and print the examination report.
Sample of information displayed on LCD monitor, including 2-D and 3-D images
Description of Individual System Components:
## Hand-Held Transrectal Probe
The PMI probe is ergonomically designed based on the rectal anatomy to minimize patient discomfort and to fit comfortably in the examiner's hand. The probe is equipped with two pressure arrays and a motion tracking system. The first pressure sensor array located at the head of the probe is used to palpate the prostate. The second pressure sensor array located on the shaft of the trans-rectal probe shows the position of the probe relative to the sphincter, thereby providing an anatomical landmark helping to guide the probe manipulation. The probe also has an off-the-shelf motion tracking system including a micro assembly of 3-D gyroscope, a 3-D accelerometer and a 3-D magnetometer built into the handle to provide real time probe orientation angles (elevation, azimuth and rotation). The orientation information in conjunction with the sphincter sensors information is used to map the probe location. The probe is intended and labeled for single use only.
{7}------------------------------------------------
#### Pressure Sensor Arrays
The first pressure sensor array located on the head of the probe consists of 8 by 16 capacitive sensors with the grid 2.0 mm by 2.0 mm. The second pressure sensor array consists of 3 by 16 capacitive sensors placed along the probe shaft. The dynamic range of the sensors (0-100 kPa) far exceeds that needed for the PMI functionality (0-60 kPa). The pressure measuring sensor is a capacitor with two electrodes separated by a matrix of air and molded RTV silicone rubber. This creates an air gap whose width and hence capacitance changes with externally applied surface pressure. The electrodes are excited by a low voltage AC signal (5 V). The orientation sensor provides information on the probe angular motion.
Each sensor is created by intersection of the orthogonal conductive strips. The capacitance of individual sensors is in the range of tens of picoFarads (pF) and the relative change of the capacitance under the 100 kPa pressure is over 5% of the range.
The sensors are calibrated by placing the probe into the calibration chamber and applying the pressure sequence: 0 kPa, 5 kPa, 10 kPa, 15 kPa, 20 kPa, 25 kPa and 30 kPa. The range of 0 - 30 kPa corresponds to the operational range for the pressure sensor arrays. The sensor signals corresponding to known pressure values (calibration data) are stored in a calibration file. To translate a pressure sensor signal into an actual pressure, PMI software uses a poly-linear function. It converts an incoming signal into a pressure value expressed in kPa using the calibration data. If the pressure signal value exceeds 30 kPa, linear extrapolation is used based on the last calibration data for 20 kPa. 25 kPa and 30 kPa. Estimated maximum error in sensor pressure readings due to the linear extrapolation of the calibration data into the extended range up to 60 kPa is about 5%, which was determined to be too low to have any significant impact on PMI image features because its level is lower than the average accuracy calculated during PMI performance bench testing.
#### Orientation Sensor (Motion Tracking System)
The 3-D orientation sensor used in the PMI system provides a full 360 degrees measurement range on three axes (elevation, rotation and azimuth) with the accuracy of ±1.0 degree and with the sample rate of 180 Hz. The 180 Hz sample rate for the output data for 3-D orientation sensor is sufficient for synchronization of the orientation data and pressure sensor data because the latter is acquired at much lower rate at 30 frames per second.
#### Probe Calibration Chamber
The calibration chamber is designed to test and calibrate the probe each time the PMI device is turned on. The calibration system includes a pressure chamber, an air pump, two air valves, an off-the-shelf air pressure sensor, and a printed circuit board with a microchip to provide its functionality. The pressure chamber becomes air-tight after its cover is securely closed with the probe inside. When the pump is activated it applies a known pressure measured by the air pressure sensor simultaneously to every sensor of the probe pressure sensor arrays.
{8}------------------------------------------------
#### Data Acquisition Unit
The data acquisition unit is designed for signal amplification, sampling, digital conversion, and transmission to the central processing unit. It is composed of a printed circuit board (PCB) that connects through a USB port to the central processing unit. A cable is used to connect the probe sensors and orientation sensors to the data acquisition unit. The data from the probe and orientation sensor are serialized into data packets by the central processing unit. A microprocessor on the data acquisition unit runs a scanning algorithm that selects analog channels via analog multiplexers and samples them with internal analogto digital converter. The probe and orientation data packets are combined and sent to the host for processing. The USB supplies the power needs for the data acquisition unit.
### Central Processing Unit with Proprietary Software
The central processing unit processes the digital data from the probe sensors using proprietary PMI software. The computer is used to acquire the prostate pressure response signals and to calculate and compose 3-D prostate elasticity images. Once the prostate scan is completed and a 3-D prostate image is composed, selected coronal and transverse 2-D cross-sectional prostate views can be displayed on the monitor. The system uses a mini PC with a Pentium T7200 2 GHz processor with 1 GB memory.
### Touch Screen Monitor
The computer communicates with a 1024x768 pixel 15-inch LCD touch screen monitor that is attached by a movable arm to the top of the system cart and is used as the graphical user interface to control the PMI functionality and visualize the resulting prostate images.
#### Color Printer
The computer is connected through a USB port to an optional compact color printer. The printer produces color hard copies of the prostate examination report that includes prostate images.
## Kevboard
A commercially available, water-resistant keyboard is provided to enable data input by the user.
#### Disposable Probe Sheath
A clear, non-sterile, disposable elastic polyurethane sheath is used to cover the probe before each calibration and prostate imaging procedure, protecting the sensor array on the probe head, the probe shaft, and the probe handle. 510(k)-cleared sheaths are provided with each PMI system, are labeled for single use only, and are available for reorder.
#### Disposable System Cover
A clear disposable system cover is used to cover the PMI system before each examination. The system cover protects the cart handles, monitor, and the working areas on the cart and keyboard. The cover is used during an exam to reduce the risk of cross contamination of the system. The cover is not patient contacting. The labeling provides reordering information and requires the user to dispose of the system cover after each use to help
{9}------------------------------------------------
reduce the possibility of cross contamination. System functionality and usability has been validated with the system cover in place.
### Probe Lubricant
Bacteriostatic, non-allergenic, water soluble, sterile lubricating jelly is required to lubricate the probe sheath and body orifice to facilitate entry of the probe into the rectal cavity and to optimize sensor head translation to help ensure proper image generation. A 510(k)-cleared lubricating jelly is provided with the PMI system and is available for reorder.
### Mobile Cart
A mobile cart is provided to enable the computer console, printer and associated hardware to be housed and moved if necessary from room to room. The mobile cart is a commercially available medical cart with adjustable height. The cart has an electronics enclosure that is used to house the electronics and PC components.
#### SUMMARY OF NONCLINICAL / BENCH STUDIES
Nonclinical performance data were provided or referenced to address the device's biocompatibility, electromagnetic compatibility, electrical, and thermal safety, and software. In addition, bench testing was performed to assess the device's ability to visualize abnormalities and produce images showing elasticity differences in prostate phantoms in a pelvic simulator.
#### BIOCOMPATIBILITY
The only patient-contacting materials of the Prostate Mechanical Imager are the probe sheath and the probe lubricant, both of which are 510(k)-cleared devices. From a biocompatibility perspective, the cleared uses of these devices are consistent with their uses as part of the PMI system (i.e., direct mucosal contact for 1-5 minutes).
The type of tissue contact (rectal mucosa) and duration of contact (< 5 minutes) do not raise the level of biocompatibility concern relative to the cleared uses. Thus, the biocompatibility of the probe sheath and probe lubricant (i.e., sensitization, and cytotoxicity) has been demonstrated by the premarket clearance of these two patientcontacting PMI system components.
#### STERILITY / SHELF LIFE / REUSE
Other than the 510(k)-cleared probe lubricant, none of the PMI system components is sold sterile, and there are no components that are to be sterilized by the user. The 510(k)cleared, disposable probe sheath is not provided sterilization for this use is unnecessary because it is not being used in a sterile environment.
The system does not have a stated shelf life. Based on the nature of the system components (i.e., electrical and computer hardware, medical cart, printer, keyboard, monitor), the absence of a shelf life is acceptable.
{10}------------------------------------------------
The PMI probe, probe sheath, and system cover are all single-use disposable components. The PMI probe and probe sheath are considered semi-critical devices. Because these components are not labeled for reuse, validation of the reprocessing methods and instructions (including cleaning and high-level disinfection) was not necessary. The electrical and computer hardware, calibration chamber, medical cart, printer, keyboard, and monitor are all reusable, non-critical components. To help mitigate the potential for cross-contamination between patients (e.g., due to the operator contacting the keyboard or touch-screen monitor with a soiled glove), a disposable system cover is used during the procedure and replaced afer each use.
# ELECTROMAGNETIC COMPATIBILITY
The de novo includes adequate test reports and labeling to support the claim of conformity to IEC 60601-1-2 (edition 2.1, Amendment 1: 2004) "Medical Electrical Equipment – Part 1: General Requirements for Safety – Collateral Standard 2; Electromagnetic Compatibility - Requirements and Tests." Immunity Tests included for IEC 60601-1-2 testing included the following collateral standards:
- Immunity Standards EN61000-4-2 (ESD) ●
- EN 61000-4-3 (Radiated Susceptibility) .
- EN 61000-4-4 (EFT/Burst) .
- EN 61000-4-5 (Surge) .
- EN 61000-4-6 (Conducted Susceptibility) .
- EN 61000-4-8 (Magnetic Fields) .
- EN 61000-4-11 (Short Interruptions and Voltage Variations)
Test reports and labeling were provided demonstrating the PMI device's compliance with the above standards.
## ELECTRICAL, MECHANICAL, AND THERMAL SAFETY
The device was tested to address compliance with the requirements of IEC 60601-1 (1998: Amendment 1 and Amendment 2) "Medical Electrical Equipment – Part 1: General Requirements for Safety." The de novo includes adequate test reports and labeling to support the electrical, mechanical, and thermal safety of the device. The results, with accompanying justifications, were determined to be sufficient to provide reasonable assurance of the PMI system's electrical, mechanical, and thermal safety.
## SOFTWARE
The software for the Prostate Mechanical Imager presents a minor level of concern based on FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (issued May 11, 2005). Considering the nature of the device and the limitations on its use (i.e., solely to be used as a documentation tool, not for diagnostic purposes or for influencing any clinical decisions, and not to be used as a replacement for digital rectal examination), there is no potential for a failure or design flaw in the software to result in injury, erroneous diagnosis, or a delay in delivery of appropriate medical care.
The PMI software provides data acquisition from the pressure and orientation sensors located in the probe and transforms the acquired data into 2-D and simulated 3-D prostate
{11}------------------------------------------------
elasticity images. Once the prostate examination is completed and a 3-D prostate elasticity image is composed, the user can select 2-D cross sectional views of the 3-D prostate elasticity image to document the details in a hard-copy color printout and/or electronically stored examination report. The PMI system software runs on a standard computer running Windows XP software platform with Microsoft .NET Framework.
Described at a high level, the primary functions of the PMI software are to:
- 1. Collect prostate examination data (pressure and orientation data),
- 2. Display the collected data,
- 3. Provide real time prostate images and 2-D prostate image construction,
- 4. Store the collected data and prostate image,
- 5. Document the prostate images,
- 6. Present the results of stored examinations, and
- 7. Generate a printed customizable report.
The de novo includes a device hazard analysis, which contains a tabular description of identified hardware and software hazards, including severity assessment and mitigation measures.
The sponsor's verification and validation testing documentation (PMI Software Test Plan and PMI Software Test Report) includes the following expected elements for a minor level of concern: functional test plan, pass/fail criteria, and summary of results. The documentation provides sufficient evidence that the specified requirements have been fulfilled.
All expected elements of software documentation (per FDA's Guidance for Off-the-Shelf Software Use in Medical Devices; Final, issued September 9, 1999, and Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff) are included in the de novo and are in sufficient detail to provide reasonable assurance that the software performs as intended and all softwarerelated risks have been adequately mitigated.
## PERFORMANCE TESTING - BENCH
Performance of the device was initially designed and evaluated using soft tissue phantoms with a set of hard inclusions of different sizes prepositioned at different depths. The experiments demonstrated that the technology enabled detection of hard nodules.
The device's ability to image the prostate and detect areas of hardness was also addressed by a comparison of the device against manual palpation and pathology results on removed prostate glands in a 1998 published study. The study was performed in collaboration with the Robert Wood Johnson University Medical Center. Histological findings from all posterior segments of nine excised prostates were compared to PMI generated images of the gland. Study findings confirmed that all PMI depicted abnormalities correlated closely with the palpated nodules and pathology findings.
{12}------------------------------------------------
An anatomical model of the male pelvis with an anal canal, rectal wall, and interchangeable silicone models of the prostate gland was designed to accurately represent elasticity of real tissue. Several sets of prostate models of varied dimensions, shapes and elasticity, with structural inclusions mimicking nodules of various diameters and elasticity, were manufactured for the bench tests. Hard nodules were positioned in specific locations within the prostate models to simulate various prostate elasticity distributions. Performance testing was conducted (by 5 operators using 5 PMI systems on 24 prostate models, for a total of 720 examinations) to determine imaging accuracy, reproducibility over time, inter-system reproducibility, and inter-operator reproducibility for generating a real-time digital image. It was demonstrated through extensive bench testing that the PMI can reliably visualize abnormalities and produce images of nodules in prostate models in the pelvic simulator.
# SUMMARY OF CLINICAL INFORMATION
In 2006, clinical testing of an earlier generation of the PMI device was conducted to study the factors affecting successful PMI image reconstruction capabilities. Following a standard DRE performed by an urologist, the PMI examination was conducted. The examiner was able to observe and inspect, in real-time, two orthogonal cross-sections of the prostate. The PMI scan data were saved in a digital format. As an outcome of this study, the PMI data were sufficient for image reconstruction of the prostate in 84% (141/168) of study cases. Based on a patient survey, the level of discomfort of the PMI was judged to be similar to that of a DRE examination.
In 2009, a clinical study was performed to test the imaging capability of the latest generation of the device by multiple users and to confirm that the incremental device modifications (introduced after the completion of the 2006 clinical study) did not negatively impact the performance and usability of the device. The main differences in the device are that the newer generation has a slightly smaller probe head, automated calibration of probe pressure sensors, and updated electronics. Five clinical investigators from five clinical sites collected PMI data for 56 clinical cases with prostate abnormalities detected by DRE. The following conclusions were derived from the study:
- PMI was shown to be capable of visualizing the prostate in 98% (55/56) of patients. .
- There was agreement between DRE and PMI determinations regarding the presence of an . abnormality in 89% of the studied cases.
- No safety concerns or side effects directly associated with the PMI examination procedure . were found by any of the five clinical investigators participating in the study.
- All clinical investigators who participated in the study rated the PMI ease of use on a scale . from 1 "not easy" to 5 "very easy". The average score for all investigators participating in the study was 4.4.
## LABELING
Labeling has been provided which includes instructions for use and a prescription use statement as required by 21 CFR 801.109. The labeling has been written to help ensure that sufficient
{13}------------------------------------------------
instructions are provided to limit the likelihood of (1) user error, (2) misinterpretation of displayed images, (3) failure to produce an accurate image, and (4) microbial contamination from reusable components.
The risk of user error is mitigated through labeling by including instructions stating that:
The device is not to be used for any diagnostic purpose, as a replacement for digital rectal examination, or for making any clinical management decisions;
The device is only to be used to image and document an abnormality that was already identified by DRE;
Clinical management decisions are not to be made on the basis of information from the PMI device, but rather on the basis of the DRE. If there is disagreement between the DRE and the recorded image produced by the device, patient management decisions are to be based on the DRE and other available clinical and diagnostic information (e.g., PSA levels) in accordance with standard medical practice;
This device has not been shown to be effective in patients who have undergone prostate surgery, radiation therapy, cryotherapy, or chemotherapy for prior prostate cancer; and
This device should not be used on patients with open wounds of the rectum or areas around the prostate, as this may increase the risk of tissue trauma, bleeding, and infection.
The risk of misinterpretation of displayed images is mitigated through labeling by including instructions explaining the meaning of the displayed colors and statements explaining that the images are not to be used for any diagnostic purpose or for making any clinical decisions.
The risk of failure to produce an accurate image is mitigated through labeling by including instructions explaining how to initialize the system and properly calibrate the probe prior to use, how to apply the proper amount of pressure to the probe head, how to redo the examination if the acquired image is not satisfactory, and when not to save and print the results (i.e., if the recorded images do not agree with the DRE findings).
The risk of microbial contamination from reusable components is mitigated through labeling by including instructions stating that the probe, probe sheath, and system cover are limited to single patient use; the reusable components are to be covered with the disposable system cover; a contaminated probe sheath must never be inserted into the calibration chamber; and specific reprocessing instructions must be followed.
{14}------------------------------------------------
### RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of a Prostate Lesion Documentation System and the measures recommended to mitigate these risks.
| Identified Risk | Mitigation Measure |
|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Failure to consistently produce an<br>accurate image | Performance Testing (non-clinical and clinical);<br>Software Verification, Validation, and Hazard<br>Analysis<br>Labeling |
| Misinterpretation of displayed images | Labeling |
| User error | Labeling |
| Microbial contamination from<br>reusable components | Labeling<br>Validation of Reprocessing Methods and Instructions |
| Adverse tissue reaction | Biocompatibility Testing |
| Electromagnetic incompatibility | Electromagnetic Compatibility Testing |
| Electrical injury | Electrical Safety Testing |
| Thermal injury | Thermal Safety Testing |
| Mechanical injury | Mechanical Safety Testing |
#### SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the Prostate Mechanical Imager is subject to the following special controls:
- (1) Non-clinical and clinical performance testing must demonstrate the accuracy and reproducibility of the constructed image.
- (2) Appropriate analysis/testing must validate electromagnetic compatibility (EMC), electrical safety, thermal safety, and mechanical safety.
- (3) Appropriate software verification, validation, and hazard analysis must be performed.
- (4) All elements of the device that may contact the patient must be demonstrated to be biocompatible.
{15}------------------------------------------------
- (5) Methods and instructions for reprocessing of any reusable components must be properly validated.
- (6) The labeling must include specific information needed to ensure proper use of the device.
## CONCLUSION
The de novo for the Prostate Mechanical Imager is granted and the device is classified under the following:
Product Code: OQT Device Type: Prostate lesion documentation system Class: II Regulation: 21 CFR 876.2050
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.