BK261351 · NanoEntek, Inc. Korea · GKZ · Jul 20, 2026 · Hematology
Device Facts
Record ID
BK261351
Device Name
ADAMII CD34 System
Applicant
NanoEntek, Inc. Korea
Product Code
GKZ · Hematology
Decision Date
Jul 20, 2026
Decision
SESE
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
ADAMII CD34 System includes ADAMII-CD34 Kit which is designed for use with ADAMII (Instrument), a benchtop image-based fluorescence cell counter. ADAMII CD34 System provides enumeration of viable CD34+ cells, viable CD45+ cells, and calculates percentage of viable CD34+ cells out of viable CD45+ cells. ADAMII CD34 System can be used for mobilized peripheral blood (MPB) collected in Na-Heparin or EDTA, fresh and thawed hematopoietic progenitor cell – apheresis (HPC-A) collected in ACD or ACD+Heparin, fresh cord blood (FCB) collected in CPD, and thawed frozen cord blood (TFCB) collected in CPD and stored with 10% DMSO, 1% Dextran 40. ADAMII CD34 System is intended for use in clinical laboratories and for in vitro diagnostic use only. It is not intended for use in point-of-care settings.
Device Story
ADAMII CD34 System is a benchtop image-based fluorescence cell counter for clinical laboratories; utilizes fluorescence-labeled antibodies and nucleic acid dyes to stain samples; processes samples via 10X RBC lysis buffer; loads into ADAMII Assay Slides. Instrument uses 4-channel (Bright field, PE, FITC, PerCP) LED-induced fluorescence imaging; CCD camera captures images; software analyzes fluorescence to quantify viable CD34+ and CD45+ cells. Outputs include absolute counts (cells/μL) and viability percentages. Results assist clinicians in assessing hematopoietic progenitor cell samples for transplantation. System is not for point-of-care use.
Clinical Evidence
Bench-only testing. Method comparison study (N=1089) against predicate showed mean relative bias of 1.14% for CD34+ counts and -0.17% for CD45+ counts. Regression analysis for pooled data yielded R²=0.99 for CD34 and CD45 counts. Precision testing (CV%) across three sites for thawed HPC-A samples ranged from 5.26% to 10.46%. Analytical sensitivity for thawed HPC-A established LoB 1.38 cells/μL, LoD 8.15 cells/μL, and LoQ 40.06 cells/μL.
Technological Characteristics
Benchtop image-based fluorescence cell counter; 4-channel (Bright field, PE, FITC, PerCP) LED excitation; high-sensitivity monochrome CCD camera; 10X objective lens. Materials include ADAMII Assay Slides and reagent kit (antibodies/dyes). Connectivity: laptop-based software interface. Power: 100-240 VAC. Standards: EN 61010-1, EN 61010-2-081, EN 61010-2-101, EN 61326-1, EN 61326-2-6.
Indications for Use
Indicated for clinical laboratory use to enumerate viable CD34+ and CD45+ cells and calculate the percentage of viable CD34+ cells out of viable CD45+ cells in mobilized peripheral blood, fresh/thawed hematopoietic progenitor cell-apheresis (HPC-A), fresh cord blood, and thawed frozen cord blood.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
{0}
Individuals using assistive technology may not be able to fully access the information contained in this file. For assistance, please call 800-835-4709 or 240-402-8010, extension 1. CBER Consumer Affairs Branch or send an e-mail to: ocod@fda.hhs.gov and include 508 Accommodation and the title of the document in the subject line of your e-mail.
{1}
# Traditional 510(k) Summary
# A. Applicant
Company Name: NanoEntek, Inc.
Address: 12F, 5, Digital-Ro 26-Gil
Guro-Gu
Seoul, KR 08389
Contact Person: Jason Jeong
Phone Number: +82-2-6220-7887
Facsimile Number: +82-2-6220-7999
# B. Contact Person
Company Name: NanoEntek, America, Inc.
Address: 220 Bear Hill Road, Suite
102, Waltham, MA 02451,
USA
Contact Person: Chan Young Park
Phone Number: (b) (6)
Facsimile Number: +1-781-790-5649
# C. Trade name: ADAMII CD34 System
Classification name: Automated Differential Cell Counter
Classification: 21 CFR §864.5220, Product code GKZ, Class II
- Secondary Classification
Classification name: Flow Cytometric Reagents and Accessories
Classification: 21 CFR §864.5220, Product code OYE, Class II
# D. Predicate device: BD Stem Cell Enumeration Kit (BK210652) for use on BD flow cytometers (FACS Lyrics)
# E. Indications for Use:
ADAMII CD34 System includes ADAMII-CD34 Kit which is designed for use with ADAMII (Instrument), a benchtop image-based fluorescence cell counter. ADAMII CD34 System provides enumeration of viable CD34+ cells, viable CD45+ cells, and calculates percentage of viable CD34+ cells out of viable CD45+ cells. ADAMII CD34 System can be used for mobilized peripheral blood (MPB) collected in Na-Heparin or EDTA, fresh and thawed hematopoietic progenitor cell – apheresis (HPC-A) collected in ACD or ACD+Heparin, fresh cord blood (FCB) collected in CPD, and thawed frozen cord blood (TFCB) collected in CPD and stored with 10% DMSO, 1% Dextran 40.
1
{2}
ADAMII CD34 System is intended for use in clinical laboratories and for in vitro diagnostic use only. It is not intended for use in point-of-care settings.
## Device Description
ADAMII-CD34 Kit utilizes fluorescence-labeled antibodies and nucleic acid staining dye to quantify absolute counts (cells/μL) of viable CD34+ hematopoietic cells and the percentage of viable CD34+ out of viable CD45+ cells. The purpose of this 510(k) is to add new sample types to BK231031; thawed HPC-A.
ADAMII-CD34 Kit includes the following components:
- ADAMII CD34 Reagent
- ADAMII Calibration bead
- ADAMII 10X RBC Lysis Buffer
- ADAMII Assay slides
ADAMII (Instrument) is an image-based fluorescence cell counter supplied with a laptop pre-installed with ADAMII-CD34 software. ADAMII (Instrument) is a 4-channel (Bright field, PE, FITC and PerCP) bench-top imaging platform equipped with state-of-the-art optics and a slide holder that accepts ADAMII Assay Slide. ADAMII (Instrument) is based on quantitative fluorometric assay technology capable of quantifying single or multiple fluorophore(s) by measuring LED-induced fluorescence from stained cells. ADAMII Software controls graphical user interface, communication with hardware, database management and data analysis. The software also controls the mechanical components including motors, light sources, and acquisition of images from CCD camera. After completion of image acquisition, ADAMII Software displays and saves results. Final results include (1) Viable CD34+ cells [/μL], (2) Viable CD45+ cells [/μL], (3) Total CD34+ cells [/μL], (4) Total CD45+ cells [/μL], (5) CD34 Viability [%], (6) CD45 Viability [%], (7) the ratio of Viable CD34+ out of Viable CD45 [%].
### F. Substantial Equivalence Discussion
An overview of the similarities and differences between the ADAMII-CD34 Kit and the predicate is provided in the Tables 1 – 3 below.
2
{3}
Table 1 General
| | Subject Device ADAMII CD34 System | Predicate Device BD Stem Enumeration Kit BK210652 |
| --- | --- | --- |
| Indications for Use | ADAMII CD34 System includes ADAMII-CD34 Kit which is designed for use with ADAMII (Instrument), a benchtop image-based fluorescence cell counter. ADAMII CD34 System provides enumeration of viable CD34+ cells, viable CD45+ cells, and calculates percentage of viable CD34+ cells out of viable CD45+ cells. ADAMII CD34 System can be used for mobilized peripheral blood (MPB) collected in Na-Heparin or EDTA, fresh and thawed hematopoietic progenitor cell – apheresis (HPC-A) collected in ACD or ACD+Heparin, fresh cord blood (FCB) collected in CPD, and thawed frozen cord blood (TFCB) collected in CPD and stored with 10% DMSO, 1% Dextran 40. ADAMII CD34 System is intended for use in clinical laboratories and for in vitro diagnostic use only. It is not intended for use in point-of-care settings. | The BD® Stem Cell Enumeration Kit is intended for enumeration of viable dual positive CD45+/CD34+ hematopoietic stem cell populations to determine absolute counts (cells/μL) of viable CD34+ and the percentages of viable CD45+/CD34+ hematopoietic stem cells (%CD34). The following cellular-based products (specimens) can be analyzed with this kit: •Normal and mobilized peripheral blood •Fresh and thawed leukapheresis products •Fresh and thawed bone marrow •Fresh and thawed cord blood The kit is intended for in vitro diagnostic (IVD) use on any of the following flow cytometer systems: •BD FACSLyric™ flow cytometer using BD FACSuite™ Clinical application •BD FACSCanto™ II flow cytometer using BD FACSCanto™ clinical software •BD FACSCalibur™ flow cytometer using BD CellQuest™ or BD CellQuest™ Pro software |
| Principle Method/Technology | This assay is performed by staining the sample with the reagent and loaded into an individual ADAMII assay slide for absolute counts. When a sample is mixed with reagent, the fluorochrome labeled antibodies and nucleic acid dyes bind specifically to the cell surface and dead cell nucleic acids. Lysis buffer is added to lyse erythrocytes before the sample is loaded into an assay slide. The slide is placed on the precision stage of ADAMII (Instrument), an image-based fluorescence cell counter. After acquiring images and analyzing them, the concentrations of viable CD34+ cells and viable CD45+ cells, and the percentage of viable CD34+ cells in viable CD45+ cell population are calculated. | The single-tube assay is performed by staining the sample with the reagent in individual BD Trucount™ tubes for absolute counts. When a sample is added to the reagent, the fluorochrome-labeled antibodies in the reagent bind specifically to the cell surface. Additionally, the lyophilized pellet in the BD Trucount™ tubes dissolves, releasing a known number of fluorescent beads. The dye 7-AAD is added to assess viability of the cells. Cells that are 7-AAD+ are not viable. Ammonium chloride is added to lyse erythrocytes before the sample is acquired on a flow cytometer. During analysis of the sample, the concentration of viable CD34+ cells and viable CD45+ cells, and the percentage of viable CD34+ cells in the viable CD45+ cell population, are calculated. |
3
{4}
Table 2 Reagent Kits and Sample
| | Subject Device ADAMII-CD34 Kit | Predicate Device BD Stem Cell Enumeration Kit |
| --- | --- | --- |
| Sample type/anticoagulants | • Mobilized peripheral blood (MPB) collected in Na-Heparin or EDTA • Fresh and thawed Hematopoietic progenitor cell – apheresis (HPC-A) collected in ACD or ACD+Heparin • Fresh cord blood collected in CPD • Thawed frozen cord blood collected in CPD and stored with 10% DMSO and 1% Dextran 40. | • Normal and mobilized peripheral blood collected with either EDTA, ACD-A, heparin, or CPD anticoagulants • Fresh and thawed leukapheresis products collected with a mixture or single source of EDTA, ACD-A, or heparin anticoagulants • Fresh and thawed bone marrow collected with either EDTA, ACD-A, or heparin anticoagulants • Fresh and thawed cord blood collected with either EDTA, ACD-A, heparin, or CPD anticoagulants |
| Parameters in: ADAMII Indications for Use/BD Principle/Methods | Viable CD34 cells/μL Percentage of viable CD34 of viable CD45 Viable CD45 cells/μL | Viable CD34 cells/μL Percentage of viable CD34 of viable CD45 Viable CD45 cells/μL |
| Other parameters measured and outputted^{1} | Total CD34 cells/μL CD34 viability % Total CD45 cells/μL CD45 viability % | Total CD34 cells/μL CD34 viability % Total CD45 cells/μL CD45 viability % |
| Kit components | CD34-PE/CD45-PerCP/Sytox Blue reagent Calibration bead solution Ammonium chloride lysing solution ADAMII Assay Slides | • CD45 FITC/CD34 PE reagent • 7-aminoactinomycin-D (7-AAD) reagent • Ammonium chloride lysing solution • BD Trucount tubes |
| Measuring range | CD34+ cells: 1 – 1,000 cells/μL | For FACSLyric flow cytometer: CD34+ cells: 1 – 1,000 cells/μL For FACSCalibur flow cytometer: CD34+ cells: 0 – 1,000 cells/μL |
| Measuring time | ~ 7 min/test | |
| Sample volume | MPB or HPC-A : 20 μL Cord blood : 50 μL | 100 μL |
| Measuring volume | ~ 8 μL | Not reported |
4
{5}
| Sample stability | • MPB: Stain specimens within 24 hours of collection • Fresh HPC-A Products: Stain specimens within 24 hours of collection • Thawed HPC-A Products: Stain immediately after thawing • Fresh Cord Blood: Stain specimens within 48 hours of collection • Thawed Frozen Cord Blood: Stain immediately after thawing | • Normal Peripheral Blood: Stain specimens within 24 hours of collection • Mobilized Peripheral Blood: Stain specimens within 24 hours of collection • Fresh Leukapheresis Products: Stain specimens within 24 hours of collection • Fresh Cord Blood: Stain specimens within 48 hours of collection • Fresh Bone Marrow: Stain specimens within 24 hours of collection • Thawed Leukapheresis Products: Stain immediately after thawing • Thawed Cord Blood: Stain immediately after thawing • Thawed Bone Marrow: Stain immediately after thawing |
| --- | --- | --- |
| Stain sample stability | • MPB: Keep prepared samples on wet ice and measure within 1 hour of lysing • Fresh HPC-A: Keep prepared samples on wet ice and measure within 1 hour of lysing • Thawed HPC-A: Measure immediately post-lysis • Fresh Cord Blood: Keep prepared samples on wet ice and measure within 1 hour of lysing • Thawed Frozen Cord Blood: Measure immediately post-lysis | • Fresh Leukapheresis Products: Acquire within 1 hour of lysing • Mobilized Peripheral Blood: Acquire within 1 hour of lysing • Fresh Cord Blood: Acquire within 1 hour of lysing • Thawed Cord Blood: Acquire immediately post-lysis |
| Interfering conditions | No interference was observed in samples with up to 7.5mg/mL of albumin. No interference was observed in samples with up to 10mg/dL of bilirubin. No interference was observed in samples with up to 550μg/mL of cyclophosphamide. No interference was observed in samples with up to 50mg/dL of hemoglobin. No interference was observed in samples with up to 0.25μg/mL of doxorubicin. No interference was observed in samples with up to 60ng/mL of G-CSF. No interference was observed in samples with up to 250mg/dL of intralipid. No interference was observed in samples with up to 20μg/mL of paclitaxel. No interference was observed in samples with up to 1% of gamma globulin. | No interference was observed in samples with up to 60g/dL of albumin. No interference was observed in samples with up to 40mg/dL of bilirubin. No interference was observed in samples with up to 55mg/dL of cyclophosphamide. No interference was observed in samples with up to 1g/dL of hemoglobin. No interference was observed in samples with up to 1.932μg/mL of doxorubicin. No interference was observed in samples with up to 60ng/mL of G-CSF. No interference was observed in samples with up to 1,140mg/dL of intralipid. No interference was observed in samples with up to 10.8μg/mL of paclitaxel. |
5
{6}
Table 3 Instruments
| | ADAMII | BD FACSLyric™ |
| --- | --- | --- |
| Light Source | Green LED (525nm) Blue LED (488nm) | Blue laser: 488 nm, 20 mw Red laser: 640 nm, 40 mw Violet laser: 405 nm, 40 mw |
| Fluorescence filters | FITC Ex 466/40, Em 525/50 PE Ex 510/42, Em 572/28 PerCP Ex 525/50, Em 650LP | FITC 527/32 PE 586/42 PerCP 700/54 PE-Cy7 783/56 APC 660/10 APC-Cy7 783/56 |
| Method of Detection | High-sensitivity monochrome CCD | Photodiode and PMT |
| Objective Lens | 10X objective and 0.7X tube lens | Flow cell lens: 1.2 NA |
| Stage | Automated X-Y-Z stage | Manual, Automated universal loader (optional) |
| Exported formats: Image | JPEG | N/A |
| Electronic input | 12V DC, 5.0A | 2 A, 200 W |
| Operating Power | 100 – 240 VAC, 1.5 A 50/60 Hz | 100 – 240 V ± 10% VAC 50 – 60 ± 10% Hz |
| Operating environment | 5 – 40°C, 20 -95% | 15 - 30°C, 15 – 85% |
| Weight | 19.3 kg | ~ 70 kg |
| Dimensions | 30W x 42D x 37H cm | 107W x 58D x 58H cm |
6
{7}
## G. Summary of Performance Data:
### 1. Summary of Method Comparison/Accuracy
| ADAMII-CD34 Kit compared to BD Stem Cell Enumeration Kit Pooled Data | | | |
| --- | --- | --- | --- |
| Parameters | N | Relative Difference to Predicate (%) | |
| | | Mean Relative Bias | 95% CI |
| viable CD34+ cell counts [cells/μL] | 1089 | 1.14% | 0.44% to 1.84% |
| ratio of viable CD34 out of viable CD45 [%] | 1089 | 1.37% | 0.57% to 2.16% |
| viable CD45+ cell count [1,000 cells/μL] | 1089 | -0.17% | -0.70% to 0.37% |
| ADAMII-CD34 Kit Regression Analysis | | | | |
| --- | --- | --- | --- | --- |
| Parameter | N | R^{2} | Slope (95% CI) | Intercept (95% CI) |
| **Pooled data** | | | | |
| CD34 (cells/μL) | 1089 | 0.99 | 1.00 (1.00-1.01) | 0.05 (-0.2-0.36) |
| %CD34 in CD45 | 1089 | 0.98 | 1.02 (1.01-1.03) | 0.00 (0.00-0.00) |
| CD45 (1000 cells/μL) | 1089 | 0.99 | 0.99 (0.98-0.99) | 0.18 (0.10-0.28) |
| **Thawed HPC-A** | | | | |
| CD34 (cells/μL) | 176 | 0.98 | 1.00 (0.98-1.01) | 3.88 (-0.75-9.15) |
| % CD34 in CD45 | 176 | 0.95 | 1.03 (0.99-1.07) | 0.01 (-0.01-0.03) |
| CD45 (1000 cells/ μL) | 176 | 0.98 | 0.94 (0.91-0.98) | 1.16 (-0.03-2.50) |
### 2. Summary of Precision
| | Clinical samples, CV [%] | | | | |
| --- | --- | --- | --- | --- | --- |
| | | | Site1 | Site2 | Site3 |
| Thawed HPC-A | Viable CD34 counts | Low | 9.65 | 8.76 | 10.22 |
| | | Mid1 | 10.46 | 7.61 | 9.65 |
| | | Mid2 | 7.25 | 5.26 | 9.31 |
| | | High | 7.15 | 5.46 | 9.11 |
| | Viable CD34 % | Low | 9.66 | 6.45 | 7.75 |
| | | Mid1 | 8.50 | 9.34 | 8.06 |
| | | Mid2 | 8.40 | 7.44 | 6.78 |
| | | High | 6.60 | 6.04 | 6.82 |
7
{8}
# 3. Summary of Linearity
• We have established following linearities.
○ CD34
▪ MPB: 1 ~ 100 cells/μL
▪ HPC-A: 1 ~ 1,000 cells/μL
▪ Cord blood: 1 ~ 120 cells/μL
# 4. Summary of Sample Stability
For thawed HPC-A samples, the age of stored sample and the age of stained sample are immediately after thawing and lysing, respectively.
# 5. Summary of Analytical Sensitivity
For thawed HPC-A samples, a LoB of 1.38 cells/μL, LoD of 8.15 cells/μL, and LoQ of 40.06 CD34 cells/μL for viable CD34 counts were established.
# H. Proposed Labeling
The labeling complies with 21 CFR §809.10. Symbols used in labeling comply with ISO 15223- 1:2016 Medical Devices – Symbols to be used with medical device labels, labelling and information to be supplied – Part 1: General requirements.
# I. Compliance with standards and guidelines
Testing complied with the following guidelines except where modifications were required by FDA:
• EN 61010-1:2010 (ed 3) Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 1: General Requirements
• EN 61010-2-081:2001 (ed 1) +A1:2003 Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-081: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes
• EN 61010-2-101:2002 (ed 1) Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment
• EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use. EMC requirements. General requirements
• EN 61326-2-6:2013 Electrical equipment for measurement, control and laboratory use – EMC requirements – Part 2-6: Particular requirements – In vitro diagnostic (IVD) medical equipment
• CLSI EP05-A3, Evaluation of Precision Performance of Quantitative Measurement Methods
• CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures; A Statistical Approach; Approved Guideline
• CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline
• CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guidelines
8
{9}
- CLSI EP09-A3, Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline (*withdrawn but current at time of testing*)
# **J. Conclusion**
The submitted information in this premarket notification is complete and supports a substantial equivalence determination when compared to the predicate device.
9
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.