Peripheral blood processing device for wound management
BK241079 · Innoveren Scientific, Inc. Cincinnati, OH · PMQ · Mar 28, 2025 · Hematology
Device Facts
Record ID
BK241079
Device Name
Peripheral blood processing device for wound management
Applicant
Innoveren Scientific, Inc. Cincinnati, OH
Product Code
PMQ · Hematology
Decision Date
Mar 28, 2025
Decision
SESE
Regulation
21 CFR 864.9245
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SkinDisc™ Wound System is designed to be used at point-of-care for the safe and rapid preparation of platelet rich plasma (PRP) and platelet poor plasma (PPP), from a small sample of a patient's own peripheral blood. Under the supervision of a healthcare professional, the SkinDisc™ is topically applied for the management of exuding cutaneous wounds, such as leg ulcers, pressure ulcers and diabetic ulcers, and mechanically or surgically debrided wounds.
Device Story
Point-of-care system for autologous platelet-rich plasma (PRP) and platelet-poor plasma (PPP) preparation; utilizes patient's peripheral blood; blood processed via density-based separation in table-top swinging bucket centrifuge; kit includes Plasmax Plasma Concentrator, GPS III Platelet Concentrate Separation Kit, ACD-A, and accessories; clinician combines prepared PRP/PPP with exogenous Calcium Chloride Thrombin (not supplied) to create wound dressing; applied topically by healthcare professional to exuding cutaneous wounds; facilitates wound management; benefits include accelerated healing and closure of chronic wounds.
Clinical Evidence
Single-center, open-label, randomized controlled trial (n=19) comparing SkinDisc to standard of care (SOC) for diabetic foot ulcers (DFU). 10 subjects in SkinDisc group, 9 in SOC. Results at 12 weeks: 100% DFU size reduction in SkinDisc group vs 49.2% in SOC. Average closure time 8.5 weeks for SkinDisc; only 1 patient achieved closure in SOC group by 12 weeks. No treatment-related adverse events reported.
Technological Characteristics
System components include medical grade polymer tubes, Plasmax Plasma Concentrator, and GPS III Platelet Concentrate Separation Kit. Operates via density-based separation using a swinging bucket centrifuge (3200 rpm for 15 mins). Requires external Calcium Chloride Thrombin. Biocompatibility per ISO-10993. Single-use, sterile kit assembly.
Indications for Use
Indicated for management of exuding cutaneous wounds, including leg, pressure, and diabetic ulcers, and mechanically or surgically debrided wounds in patients requiring autologous PRP/PPP topical application.
Regulatory Classification
Identification
An automated blood cell separator is a device that uses a centrifugal or filtration separation principle to automatically withdraw whole blood from a donor, separate the whole blood into blood components, collect one or more of the blood components, and return to the donor the remainder of the whole blood and blood components. The automated blood cell separator device is intended for routine collection of blood and blood components for transfusion or further manufacturing use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is a guidance for industry and FDA staff entitled “Class II Special Controls Guidance Document: Automated Blood Cell Separator Device Operating by Centrifugal or Filtration Separation Principle.”
Plasmax Plasma Concentrator with GPS® II Mini Platelet Concentrate Separation Kit With ACD-A (BK070003)
Plasma Concentrator GPS® III Mini Platelet Concentrate Separation Kit With ACD-A (BK050016)
GPS® III Platelet Concentrate Separation Kit With ACD-A (BK070026)
Submission Summary (Full Text)
{1}
# **510(k) SUMMARY**
**For**
**SkinDisc™ Lite - Wound System and the SkinDisc™ - Wound System**
**Submitter Information:**
Innoveren Scientific, Inc.
151 W 4th St Ste 223
Cincinnati, OH 45202
Contact Person & Official Correspondent: Tanya Rhodes
Telephone Number: (727) 458-4291
Date Prepared: 2.12.2025
**Device Information:**
| **FDA Product Code:** | **PMQ** |
| --- | --- |
| FDA Regulation Number: | 21 CFR 864.9245 |
| FDA Classification Name: | Automated Blood Cell Separator |
| Common Name | Peripheral Blood Processing Device for Wound Management |
| Classification Panel: | Hematology |
| FDA Classification: | Class II |
| Device Name: | SkinDisc Lite; SkinDisc |
| **Predicate Device Name** | **510(k)** | **Company Name** |
| --- | --- | --- |
| REGENKIT-Wound Gel-2 | BK210661 | RegenLab SA |
| **Reference Device Name** | | |
| Plasmax Plasma Concentrator with GPS® II Mini Platelet Concentrate Separation Kit With ACD-A | BK070003 (Special) | Biomet Biologics, Inc. |
| Plasma Concentrator GPS® III Mini Platelet Concentrate Separation Kit With ACD-A, GPS® III Platelet Concentrate Separation Kit With ACD-A, GPS® III Platelet Concentrate | BK050016 BK070026 (Special) | Biomet Biologics, Inc. |
{2}
# **Description of Device:**
SkinDisc™ and SkinDisc™ Lite– Wound Systems are single use medical devices used to rapidly prepare a platelet rich plasma (PRP) and platelet poor plasma (PPP) from a small volume of the patient's own peripheral blood that is drawn at the time of treatment. The device kit consists of a disposable separator and accessories to be used with a table-top, swinging bucket centrifuge.
The PRP and PPP are obtained by density using the Zimmer-Biomet centrifuge, according to centrifuge operating instructions. Under the supervision of a healthcare professional, PRP and PPP are prepared and combined with commercially available Calcium Chloride Thrombin (not supplied in the kit) to produce (SkinDisc™/SkinDisc™Lite) which are suitable for the management of exuding cutaneous wounds, such as leg ulcers, pressure ulcers, and diabetic ulcers, and mechanically or surgically debrided wounds.
The SkinDisc™ Wound System contains a Plasmax Plus Plasma Concentrator (60ml), a GPS® III Platelet Concentrate Separation Kit, ACD-A and blood draw components. 60ml of blood produces approximately ~26 ml of SkinDisc for use in the treatment of wounds.
The SkinDisc™ Lite – Wound System contains a Plasmax Plasma Concentrator (30ml), a GPS® III Mini Platelet Concentrate Separation Kit, ACD-A and blood draw components. 30ml of blood produces approximately ~13 ml of SkinDisc for use in the treatment of wounds.
# **Intended Use/Indications For Use:**
The SkinDisc™ Wound System is designed to be used at point-of-care for the safe and rapid preparation of platelet rich plasma (PRP) and platelet poor plasma (PPP), from a small sample of a patient's own peripheral blood. Under the supervision of a healthcare professional, the SkinDisc™ is topically applied for the management of exuding cutaneous wounds, such as leg ulcers, pressure ulcers and diabetic ulcers, and mechanically or surgically debrided wounds.
# **Comparison of Technological Characteristics:**
SkinDisc™ Wound Systems and the predicate device are both designed for the preparation of PRP by simple centrifugation of a small sample of the patient's own blood. The subject and predicate devices are substantially equivalent based on the following similar technological elements:
- All are used at the patients point of care to separate and prepare autologous blood, blood fractions in a single procedure to be used clinically
- All separate blood/blood components by centrifugation/density.
- Similar blood volumes are used/Similar blood component gel volumes produced.
- All are based on phlebotomy with the provided blood collection set.
- Liquid PRP collection)
- PRP gel that is used as a wound management dressing
- Same indications for use statements.
{3}
A comparison of the subject and predicate device is provided in the following table.
| Technological Characteristic | Subject Device: SkinDisc System | Predicate Device: RegenKit WG-2 |
| --- | --- | --- |
| **Regulatory Information** | | |
| **510(k) Number** | BK241079 (new) | BK210661 |
| **Device Name** | SkinDisc Lite; SkinDisc | RegenKit – Wound Gel - 2 |
| **Manufacturer** | Innoveren Therapeutics | Regen Lab SA |
| **Regulation Number** | 21 CFR 864.9245 | 21 CFR 864.9245 |
| **Device Class** | Class II | Class II |
| **Product Code** | PMQ | PMQ |
| **Classification Panel** | Hematology | Hematology |
| **Intended Use** | | |
| **Indications For Use** | The SkinDisc™ Wound System is designed to be used at point-of-care for the safe and rapid preparation of platelet rich plasma (PRP) and platelet poor plasma (PPP), from a small sample of a patient's own peripheral blood. Under the supervision of a healthcare professional, the SkinDisc™ is topically applied for the management of exuding cutaneous wounds, such as leg ulcers, pressure ulcers and diabetic ulcers, and mechanically or surgically debrided wounds. | RegenKit®-Wound Gel-2 is designed to be used at point-of-care for the safe and rapid preparation of platelet-rich plasma (PRP) gel (RegenWound Gel, RWG) from a small sample of a patient's own peripheral blood. Under the supervision of a healthcare professional, the Regen Wound Gel is topically applied for the management of exuding cutaneous wounds, such as leg ulcers, pressure ulcers and diabetic ulcers, and mechanically or surgically debrided wounds. |
| **Prescription Status** | Rx | Rx |
| **Primary Technological Characteristics** | | |
| **Kit Components** | **Supplied: SkinDisc System:** GPS III Separator Plasma Concentrator 60ml syringe 10ml syringe | **Supplied: RegenKit-Wound Gel-2** 2 x Regen™ BCT tubes 1 x Regen™ ATS tube |
{4}
| | 18Ga needle set Adhesive tape Alcohol pads **Required commercially available components, not supplied in the kits:** Small sterile container, with a flat bottom, and a minimal capacity of 15 mL (0.5 oz) | **Required commercially available components, not supplied in the kits:** Small sterile container, with a flat bottom, and a minimal capacity of 15 mL (0.5 oz). Tube holder Accessories for phlebotomy and wound dressing including, but not limited to, tourniquet, sterile alcohol prep pads, sterile adhesive 1 x Safety–Lok™ Butterfly needle 1 x Collection holder (blue top) 1 x Transfer device (pink top) 2 x 18G 1 ½ Red transfer needles 2 x 5 ml Luer-Lok™ syringes. |
| --- | --- | --- |
| **Reagents** | **Not supplied:** USP Calcium Chloride Thrombin 10% | **Not supplied:** USP Calcium Gluconate 10% or USP Calcium Chloride 10% |
| **Blood Collection** | The GPS III Platelet Separator - 1x medical grade polymer tube The Plasma Concentrator – 1x medical grade polymer tube (with porous polyacrylamide desalting beads). | Regen BCT – 2 x class 1 cerium III borosilicate glass tubes Regen ATS tube (with coagulant). 1x class 1 cerium III borosilicate glass tube |
| **Anticoagulant** | 30ml bottle ACDA | 4% sodium citrate solution |
| **Centrifugation** | Zimmer Biomet Centrifuge 15 mins x 3200 rpm for PRP/PPP 2 minutes x 2000 rpm for PPPc | Suitable clinical centrifuge with swinging buckets 5 minutes x 1(500 x g) for PRP (BCT tubes) Up to 2 x 5 minutes for PPP / PPPc (ATS tubes) |
| **Kit Sterilization** | Kit assembly of sterile components | Sterilized by Gamma irradiation |
| **Shelf Life** | 24 months | 24 months |
{5}
Innoveren has demonstrated that the difference in technological characteristics of the subject device and the predicate device does not raise new questions of safety or effectiveness.
## Summary of Non-Clinical Testing and Risk Analysis:
### Biocompatibility Testing:
Various studies were conducted in accordance with the FDA guidance “Use of International Standard ISO-10993, ‘Biological Evaluation of Medical Devices Part 1: Evaluation and Testing”. The battery of testing included the following tests:
- Cytotoxicity
- Sensitization
- Irritation
- Acute Systemic toxicity
- Hemocompatibility
- Rabbit Pyrogen study
The components of SkinDisc™ Systems are considered blood contacting (either circulating blood or indirect blood path) for a duration of less than 24 hours.
### Performance Testing:
Non-clinical bench tests to evaluate performance of the SkinDisc system was conducted, including, cell recovery efficiency (i.e., red blood cell count, white blood cell count, platelet count), moisture content, clot visual examination and clot handling characteristics. In all instances, the SkinDisc Systems worked as intended and the performances observed were as expected and considered equivalent to the performance of predicate device.
### Clinical Testing:
A summary from a single center, open label, randomized controlled clinical trial was performed and provided in this submission to evaluate the safety and efficacy of the autologous platelet gel as compared to standard of care treatment of diabetic foot ulcers (DFU). The trial enrolled subjects with type 1 and type 2 diabetes mellitus with one or more DFUs with a wound surface area up to 25cm²
19 subjects were treated (10 in SDL treatment Group and 9 in SOC control group).
The results of the study showed that at week 12 the DFU % reduction in ulcer size was 100% in the SkinDisc treatment group compared to 49.2% in the SOC control group. The average DFU closure time was 8.5 weeks in the SkinDisc treatment group with only 1 patient achieving closure by 12 weeks in the SOC control group. The SkinDisc device was well tolerated, with no treatment-related adverse events.
{6}
### Conclusion Regarding Substantial Equivalence:
The SkinDisc™ Systems have the same intended use, incorporate the same fundamental technology, and have similar indications for use as the predicate devices, use similar material and energy sources, and have a similar design as the predicate devices.
Based on this, Innoveren has demonstrated that the SkinDisc System is substantially equivalent to the predicate device and is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.