← Product Code [MMH](/productcode/MMH) · BK241069

# Blood Bank Information Management System (BBIMS) v3.01 (BK241069)

_Carter BloodCare Bedford, TX · MMH · Jul 9, 2024 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/BK241069

## Device Facts

- **Applicant:** Carter BloodCare Bedford, TX
- **Product Code:** [MMH](/productcode/MMH.md)
- **Decision Date:** Jul 9, 2024
- **Decision:** SESE
- **Regulation:** 21 CFR 864.9165
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** Software as a Medical Device

## Indications for Use

Blood Bank Information Management System (BBIMS) v3.01 is a Blood Establishment Computer Software that is intended for use by trained healthcare professionals for the following blood manufacturing activities: - Provides a donor-administered or staff entered health history questionnaire - Transfers donor responses collected from the progressive web application to the system's network - Provides a staff entered physical assessment - Provides eligibility information regarding the suitability of a donor making a donation - Defers donors when they are prohibited from making a donation - Transfers donor information to Vista® Information System for donation optimization and receives procedure data information from Vista (when using certain automated blood collection systems) - Provides for maintenance and auditing of donor, donation and unit information - Provides for generation of donor notification letters - Provides for defining component preparation, modification and maintenance information - Provides for defining pooled products (e.g., platelets, cryoprecipitate) into single components - Transfers test orders and receives test results for all units processed, either through manual entry and/or donor sample testing software systems interfaces - Provides for labeling components based on Codabar and ISBT standards using HemaTrax® Unity Print Server to print 2x2, 4x2 and 4x4 labels - Provides for storing patient information during the manual crossmatch/consultation process - Provides for ordering products and storing information on inventories, orders, shipments, returns and imports

## Device Story

BBIMS v3.01 is a blood establishment computer software system used by trained healthcare professionals in blood bank settings. It manages the end-to-end blood donation lifecycle: donor health history questionnaires (via web app), physical assessments, eligibility screening, and deferral tracking. The system integrates with Vista Information Systems for donation optimization and interfaces with laboratory systems for test order/result management. It supports component preparation, pooling, and labeling (Codabar/ISBT standards) via HemaTrax Unity Print Server. The system maintains audit trails for donor/unit records and manages inventory, shipping, and patient crossmatch data. By centralizing donor and unit information, it ensures regulatory compliance, accurate labeling, and efficient blood product management, ultimately supporting safe blood supply operations.

## Clinical Evidence

Bench testing only. Verification and validation testing performed at unit, integration, and functional/system levels. Testing included program packages and risk control measures using established software development procedures to ensure safety-critical items and implementation hazards were addressed.

## Technological Characteristics

Blood establishment computer software; supports Codabar and ISBT labeling standards; integrates with HemaTrax Unity Print Server; interfaces with Vista Information Systems and external laboratory testing systems; utilizes progressive web application for donor data collection; network-based architecture.

## Regulatory Identification

Blood establishment computer software (BECS) is a device used in the manufacture of blood and blood components to assist in the prevention of disease in humans by identifying ineligible donors, by preventing the release of unsuitable blood and blood components for transfusion or for further manufacturing into products for human treatment or diagnosis, by performing compatibility testing between donor and recipient, or by performing positive identification of patients and blood components at the point of transfusion to prevent transfusion reactions. This generic type of device may include a BECS accessory, a device intended for use with BECS to augment the performance of the BECS or to expand or modify its indications for use.

## Special Controls

*Classification.* Class II (special controls). The special controls for these devices are:(1) Software performance and functional requirements including detailed design specifications (
*e.g.,* algorithms or control characteristics, alarms, device limitations, and safety requirements).(2) Verification and validation testing and hazard analysis must be performed.
(3) Labeling must include:
(i) Software limitations;
(ii) Unresolved anomalies, annotated with an explanation of the impact on safety or effectiveness;
(iii) Revision history; and
(iv) Hardware and peripheral specifications.
(4) Traceability matrix must be performed.
(5) Performance testing to ensure the safety and effectiveness of the system must be performed, including when adding new functional requirements (
*e.g.,* electrical safety, electromagnetic compatibility, or wireless coexistence).

## Predicate Devices

- Blood Bank Information Management System (BBIMS) v2.06.00 ([BK180246](/device/BK180246.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Carter BloodCare

## 510(k) SUMMARY (BK241069)

**Date Summary was Prepared:** 06-28-2024

### Contact Details:

Applicant Name: Carter BloodCare
Applicant Address: 2205 Highway 121 South, Bedford, TX 76021 United States
Applicant Contact: Shankar Goudar, Chief Information Officer
Applicant Contact Telephone: 817-412-5344
Applicant Contact Email: sgoudar@carterbloodcare.org

### Device Name:

Trade Name: Blood Bank Information Management System (BBIMS) v3.01
Classification Name: Blood Establishment Computer Software and Accessories
Device Class: Class II
Product Code: MMH
Panel: Hematology
Regulation Number: 21 CFR § 864.9165

### Legally Marketed Predicate Device:

Predicate #: BK180246
Predicate Trade Name: Blood Bank Information Management System (BBIMS) v2.06.00
Product Code: MMH

### Device Description Summary:

BBIMS v3.0,1 with its QuickScreen component, is a full-service Blood Bank Information Management System that includes the following features:

- Collection Services
- Donor Relations
- Donation/Unit Entry & Maintenance
- Component Preparation & Maintenance
- Test Results Processing
- Component Labeling
- Reference Laboratory Services
- Product Orders, Shipping, & Inventory
- General Utilities
- Record Auditing Information
- System Administration

### Intended Use/indications for Use:

Blood Bank Information Management System (BBIMS) v3.01 is a Blood Establishment Computer Software that is intended for use by trained healthcare professionals for the following blood manufacturing activities:

- Provides a donor-administered or staff entered health history questionnaire
- Transfers donor responses collected from the progressive web application to the system's network
- Provides a staff entered physical assessment
- Provides eligibility information regarding the suitability of a donor making a donation
- Defers donors when they are prohibited from making a donation
- Transfers donor information to Vista® Information System for donation optimization and receives procedure data information from Vista (when using certain automated blood collection systems)

CarterBloodCare.org | 800-366-2834 | 2205 Highway 121, Bedford, TX 76021

Page 1 of 2

{1}

Carter BloodCare

- Provides for maintenance and auditing of donor, donation and unit information
- Provides for generation of donor notification letters
- Provides for defining component preparation, modification and maintenance information
- Provides for defining pooled products (e.g., platelets, cryoprecipitate) into single components
- Transfers test orders and receives test results for all units processed, either through manual entry and/or donor sample testing software systems interfaces
- Provides for labeling components based on Codabar and ISBT standards using HemaTrax® Unity Print Server to print 2x2, 4x2 and 4x4 labels
- Provides for storing patient information during the manual crossmatch/consultation process
- Provides for ordering products and storing information on inventories, orders, shipments, returns and imports

Indications for Use Comparison:

BBIMS v3.01 and the predicate (BK180246) are Blood Establishment Computer Software intended for use by trained blood bank staff with a self-administered questionnaire intended to be used by donors. BBIMS v3.01 has the same indications of use with the exception that interfaces with LifeTrak will no longer be supported. LifeTrak is a third-party system and does not contribute to the BBIMS design, functions, or operations.

Technological Comparison:

BBIMS v3.01 has the same technological characteristics in regard to design, function, and application as the corresponding parts of the referenced predicated device. Testing for BBIMS v3.01 verified the system is safe and effective and performs as well or better than the predicate device when utilized within its intended use.

Non-Clinical and/or Clinical Test Summary & Conclusions:

Verification and validation testing was performed at the unit, integration, and functional/system levels including verification testing of program packages and risk control measures. Test case matrices were developed using established procedures for software development.

Carter BloodCare ensures that all safety critical items have been thoroughly tested and can demonstrate that all methods of control for intended use and general implementation hazards have been tested.

CarterBloodCare.org | 800-366-2834 | 2205 Highway 121, Bedford, TX 76021
Page 2 of 2

---

**Source:** [https://fda.innolitics.com/device/BK241069](https://fda.innolitics.com/device/BK241069)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
