AQUIOS CL Flow Cytometer: The AQUIOS CL Flow Cytometer is intended for use with in vitro diagnostic flow cytometric applications using up to four fluorescent detection channels using a blue (488 nm) laser, two light scatter detection channels, and electronic volume (EV). AQUIOS Flow Cytometry Software may be run on an independent computer workstation for offline analysis of results generated by the AQUIOS CL Flow Cytometer with its monoclonal antibody reagents. The offline analysis must be performed in accordance with the product labeling. AQUIOS STEM Kit: The AQUIOS STEM Kit is an in vitro diagnostic medical device intended to be used by laboratory professionals for the simultaneous identification and enumeration of percent (%) CD34+ cell population, CD34+ cells/ μL, and CD45+ cells/ μL in fresh peripheral or mobilized peripheral whole blood, fresh or thawed bone marrow, fresh or thawed apheresis products, and fresh or thawed cord blood on the automated AQUIOS CL Flow Cytometry System. AQUIOS STEM CD34 Control Cells: AQUIOS STEM CD34 Control Cells Level 1 and Level 2 are in vitro diagnostic medical devices intended to be used by laboratory professionals as process controls for the verification of the parameters CD34 and CD45 as part of the AQUIOS STEM System on the automated AQUIOS CL Flow Cytometer System. AQUIOS STEM CD34 Control Cells are liquid preparations of stabilized human leukocytes (lymphocytes, monocytes, and granulocytes) and erythrocytes that have lysing, light scatter, antigen expression, and antibody staining properties representative of those found in human whole blood specimens.
Device Story
AQUIOS STEM System performs automated flow cytometric identification and enumeration of hematopoietic stem/progenitor cells (HPC). Input: fresh/thawed whole blood, bone marrow, apheresis, or cord blood samples. Process: automated on-board sample preparation (staining with CD45-FITC/CD34-PE, 7-AAD viability dye, lysing); flow cytometric analysis using 488nm blue laser, light scatter, and electronic volume. Software: AQUIOS STEM Software automatically gates populations per ISHAGE guidelines; calculates viable CD34+/CD45+ counts using STEM-Count Fluorospheres. Output: percent CD34+, absolute CD34+ cells/μL, absolute CD45+ cells/μL. Used in clinical laboratories by professionals. Results assist in clinical decision-making regarding stem cell transplantation/therapy. Benefits: automated, standardized workflow reduces manual error, improves reproducibility, and ensures consistent cell enumeration.
Clinical Evidence
Bench testing only. Studies included: linearity (fluorescence and assay), carryover, detection capability (LoB, LLoD, LLoQ), specimen age stability, anticoagulant equivalency, and precision (repeatability/reproducibility). Method comparison against predicate stemCXP system demonstrated acceptable substantial equivalence for viable CD45+ cells/μL, viable CD34+ cells/μL, and viable percent (%) CD34+. Reference intervals were established per CLSI EP28-A3c.
Indicated for laboratory professionals to identify and enumerate CD34+ and CD45+ cell populations in fresh or thawed peripheral blood, bone marrow, apheresis products, and cord blood using the AQUIOS CL Flow Cytometry System.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
{1}
# **510(k) Summary for
AQUIOS STEM SYSTEM with AQUIOS CL Flow Cytometer**
# **510(k) Owner / Submitter Information**
Name: Beckman Coulter Inc.
Address: 11800 SW 147$^{th}$ Ave., Miami, FL 33196
Phone #: (b) (6)
Contact Person: Patricia Zenelaj
Email Address: (b) (6)@beckman.com
Date Submitted: June 1$^{st}$, 2023
# **Proposed Device**
# **Information**
Trade Name: **AQUIOS CL Flow Cytometer**
Common Name: AQUIOS CL
Classification Name: Automated differential cell counter (21 CFR 864.5220)
Classification: Class II
Product Code: OYE
Panel: Hematology
Trade Name: **AQUIOS STEM Software**
Common Name: AQUIOS STEM Software
Classification Name: Automated differential cell counter (21 CFR 864.5220)
Classification: Class II
Product Code: OYE
Panel: Hematology
Trade Name: **AQUIOS STEM Kit**
Common Name: AQUIOS STEM Kit
Classification Name: Automated differential cell counter (21 CFR 864.5220)
Classification: Class II
Product Code: OYE
Panel: Hematology
Trade Name: **AQUIOS STEM CD34 Control Cells**
Common Name: AQUIOS STEM Control Cells Level 1 and Level 2
Classification Name: Hematology quality control mixture (21 CFR 864.8625)
Classification: Class II Exempt, based on regulation 21 CFR 864.8625 which states that Quality Control materials are now exempt from premarket notification subject to the limitations under 21 CFR 862.9
Product Code: JPK
Panel: Hematology
---510(k) Submission for AQUIOS STEM System
510(k) Summary
Page 1 of 18
Confidential - Company Proprietary
{2}
## **Predicate Device Information**
| Manufacturer | Predicate | 510(k) Number | Product Code | Classification |
| --- | --- | --- | --- | --- |
| Beckman Coulter Inc | Stem-Kit Reagent | BK210639 BK040032 | GKZ** | Class II (864.5220) Automated differential cell counter |
| | FC500 Flow cytometer | K182886 | GKZ** | |
| | AQUIOS CL Flow Cytometer | K141932 | OYE | |
| | stemCXP software | BK210639 BK040055 | GKZ** | |
| | Stem-Trol control cells | BK040032 | JPK* | Class II Exempt (864.8625) Hematology Quality Control Mixture |
* Based on regulation 21 CFR 864.8625 which states that Quality Control materials are exempt from premarket notification subject to the limitations under 21 CFR 862.9.
** GKZ was the product code for flow cytometer, reagents, controls and software until new product code OYE which is specific to Flow Cytometric products was established.
## **Device Description**
The AQUIOS CL Flow Cytometry System is composed of the following components:
- AQUIOS CL Flow Cytometer
- AQUIOS STEM Software
- AQUIOS STEM Kit Reagent Kit
- AQUIOS STEM CD34 Control Cells
The AQUIOS CL Flow Cytometer is intended for use with in vitro diagnostic flow cytometric applications using up to four fluorescent detection channels using a blue (488 nm) laser, two light scatter detection channels, and electronic volume (EV).
AQUIOS STEM Kit contains the following reagents:
- AQUIOS STEM Kit CD45-FITC / CD34-PE – 2 vials
- AQUIOS STEM Kit CD45-FITC / IsoClonic Control-PE – 1 vial
- AQUIOS STEM Kit 7-AAD Viability Dye – 3 vials
- AQUIOS STEM Kit Lysing Solution – 3 vials
- AQUIOS STEM Kit STEM-Count Fluorospheres – 2 vials
The AQUIOS STEM Kit is designed to identify human hematopoietic stem and progenitor cells (HPC) using the following criteria: “True CD34+ cells (a) express CD34 antigen, (b) express CD45 antigen with staining intensity characteristic of blast cells (i.e., readily detectable but at lower levels than lymphocytes and
510(k) Submission for AQUIOS STEM System
510(k) Summary
Page 2 of 18
Confidential - Company Proprietary
{3}
monocytes), and (c) exhibit low side-angle and low to intermediate forward angle light scatter characteristics of blast cells". Exclusion of dead CD34+ HPC from viable CD34+ HPC enumeration is achieved using the 7-AAD Viability Dye.
The AQUIOS STEM Software is designed for the AQUIOS CL flow cytometer. It includes the algorithms and test definitions that provide automated analysis and results using AQUIOS STEM reagents. AQUIOS STEM Software automatically generates gates and regions based on ISHAGE Guidelines to identify populations of interest. The software enables enumeration of Viable CD34+ and CD45+ cell count based on the identified population of interest and AQUIOS STEM-Count Fluorophore counts.
The AQUIOS Flow Cytometry System also offers an optional standalone offline workstation. This workstation is identical to the workstation that is physically connected to the instrument and can be used for off-line analysis of results generated by the AQUIOS CL Flow Cytometer with AQUIOS STEM Kit reagents and AQUIOS STEM software according to the product labeling.
# **Intended Use:**
# AQUIOS CL Flow Cytometer
The AQUIOS CL Flow Cytometer is intended for use with in vitro diagnostic flow cytometric applications using up to four fluorescent detection channels using a blue (488 nm) laser, two light scatter detection channels, and electronic volume (EV).
AQUIOS Flow Cytometry Software may be run on an independent computer workstation for offline analysis of results generated by the AQUIOS CL Flow Cytometer with its monoclonal antibody reagents. The offline analysis must be performed in accordance with the product labeling.
# AQUIOS STEM Kit
The AQUIOS STEM Kit is an in vitro diagnostic medical device intended to be used by laboratory professionals for the simultaneous identification and enumeration of percent (%) CD34+ cell population, CD34+ cells/ μL, and CD45+ cells/ μL in fresh peripheral or mobilized peripheral whole blood, fresh or thawed bone marrow, fresh or thawed apheresis products, and fresh or thawed cord blood on the automated AQUIOS CL Flow Cytometry System.
# AQUIOS STEM CD34 Control Cells
AQUIOS STEM CD34 Control Cells Level 1 and Level 2 are in vitro diagnostic medical devices intended to be used by laboratory professionals as process controls for the verification of the parameters CD34 and CD45 as part of the AQUIOS STEM System on the automated AQUIOS CL Flow Cytometry System.
AQUIOS STEM CD34 Control Cells are liquid preparations of stabilized human leukocytes (lymphocytes, monocytes, and granulocytes) and erythrocytes that have lysing, light scatter, antigen expression, and antibody staining properties representative of those found in human whole blood specimens.
510(k) Submission for AQUIOS STEM System
510(k) Summary
Page 3 of 18
Confidential - Company Proprietary
{4}
The kit contains 2 levels of CD34 positive cell controls with approximately 10 CD34+ cells/μL (level 1) and approximately 30 CD34+ cells/μL (level 2) that are processed in the same manner as whole blood. This allows verification of application performance. AQUIOS STEM CD34 Control Cells Level 1 and Level 2 are quantitative controls used in combination with the AQUIOS STEM Kit as part of the AQUIOS STEM System.
---510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 4 of 18
{5}
# **Substantial Equivalence Comparison between Subject and Predicate Devices**
In this submission, AQUIOS STEM Kit System with AQUIOS CL Flow Cytometer (subject device) is compared with Stem Kit Reagents and stemCXP system with the FC500 Flow Cytometer (predicate device). A comparison of the similarities and differences between the subject device and the predicate device is presented in Table 1.
Table 1 Comparison Between Subject and Predicate Device
| | Predicate | Subject Device | Comments |
| --- | --- | --- | --- |
| Characteristic | Stem Kit Reagents and stemCXP system with the FC500 Flow Cytometer | AQUIOS STEM Kit System with AQUIOS CL Flow Cytometer | |
| Intended Use | Stem-Kit Reagents consist of a two-color fluorescent (FITC, PE) murine monoclonal antibody reagent, a two-color murine fluorescent (FITC, PE) isoclonic control, an absolute count reagent, a cell viability reagent, and a lysing reagent. It is intended “For In Vitro Diagnostic Use” for the simultaneous identification and enumeration of CD45+ and dual-positive CD45+ CD34+ cell population percentages and absolute counts in biological specimens by flow cytometry. Biological specimens include fresh normal or mobilized peripheral whole blood, and fresh or thawed apheresis products, cord blood and bone marrow. Cell population measurements may be obtained using either an automated method or a manual method for gating and analysis on the FC500. Refer to this Stem-Kit Reagents package insert for complete instructions if using the manual method. Refer to the stemCXP System Guide provided with the stemCXP 2.0 software kit and stemCXP version 2.0, WIN 7 for complete instructions if using the automated method. | The AQUIOS STEM Kit is an in vitro diagnostic medical device intended to be used by laboratory professionals for the simultaneous identification and enumeration of percent (%) CD34+ cell population, CD34+ cells/ μL, and CD45+ cells/ μL in fresh peripheral or mobilized peripheral whole blood, fresh or thawed bone marrow, fresh or thawed apheresis products, and fresh or thawed cord blood on the automated AQUIOS CL Flow Cytometry System. | Subject Device intended use reflects the use of AQUIOS STEM system on AQUIOS CL flow cytometer. Rest of the intended use is updated for clarity. Context of use remains the same. |
| Device Classification | Stem CXP System on FC500 Flow Cytometer: Class II, 864.5220 | AQUIOS CL Flow Cytometer System: Class II, 864.5220 | Same |
510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 5 of 18
{6}
| | Predicate | Subject Device | Comments |
| --- | --- | --- | --- |
| Characteristic | Stem Kit Reagents and stemCXP system with the FC500 Flow Cytometer | AQUIOS STEM Kit System with AQUIOS CL Flow Cytometer | |
| | Stem-Kit Reagent: Class II, 864.5220 Stem-Trol Cells: 21 CFR 864.8625 | AQUIOS STEM Kit: Class II, 864.5220 AQUIOS STEM Control Cells: 21 CFR 864.8625 | |
| Product Code | Stem CXP System on FC500 Flow Cytometer: GKZ Stem-Kit Reagent: GKZ Stem-Trol Cells: JPK | AQUIOS CL Flow Cytometer System: OYE AQUIOS STEM Kit: OYE AQUIOS STEM Control Cells: JPK | GKZ was the product code for flow cytometer, reagents, controls and software until new product code OYE which is specific to Flow Cytometric products was established. |
| **Instrument Similarities and Differences** | | | |
| Sample Analysis | - Principle of analysis – Flow cytometry - Detection hardware – Lasers, fluidics, optics, electronics - Sample analysis pathway | Same | Same |
| Optics | Laser light delivered by mirrors, prisms, and lenses | Same | Same |
| Workstation | Software functionality to allow – - Patient data management – storage, review, reporting to LIS - Control data management – storage, review, reporting - System configuration management - System service test and adjustment procedures | Same | Same |
| System Configuration | FC500 System with PC-based workstation running stemCXP application-specific software | AQUIOS System with Reagent Cart and All-in-One Touch Screen Computer loaded with AQUIOS System Software, Database, and AQUIOS STEM Software | Substantially Equivalent |
| Operating System | Microsoft Windows 7 | Microsoft Windows 10 | Substantially Equivalent |
| Controlling software | CXP System Software | AQUIOS System Software | Substantially Equivalent |
510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 6 of 18
{7}
| | Predicate | Subject Device | Comments |
| --- | --- | --- | --- |
| Characteristic | Stem Kit Reagents and stemCXP system with the FC500 Flow Cytometer | AQUIOS STEM Kit System with AQUIOS CL Flow Cytometer | |
| Lasers | Blue – 488 nm argon ion Laser | Blue – 488 nm Blue Laser | Substantially Equivalent |
| Optics | FC500 flow cell Laser light delivered by mirrors, prisms, and lenses. Emitted light delivered by mirrors | AQUIOS flow cell Laser light delivered by mirrors, prisms, and lenses. Emitted light delivered by mirrors | Substantially Equivalent |
| Sample Preparation | Manual | Automated On-board | Substantially Equivalent |
| Specimen Introduction | Manual | Specimens are loaded in the Autoloader using cassettes / or the single loader as single specimen tubes | Substantially Equivalent |
| Prepared Sample Introduction | Single 12X75 mm test tubes | 96 well plate | Substantially Equivalent |
| Specimen Identification | Manual entry into Worklist | Barcode – positive sample identification or manual entry | Substantially Equivalent |
| Prepared Sample Identification | Worklist and Carousel position | Software tracked by well position. | Substantially Equivalent |
| Standardization | • Automatic voltage/gain adjustments • Compensation matrix calculation | A verification check of standardization is performed with Flow Check and controls | Substantially Equivalent |
| Sample Analysis | • Principle of analysis – Flow cytometric Detection hardware: Lasers, fluidics, optics, electronics. Sample analysis pathway: Manual gating of cellular populations per instrument IFU or stemCXP automated gating of cellular populations based on the ISHAGE gating strategy | Same • Principle of analysis - Flow cytometric Detection hardware: Lasers, fluidics, optics, electronics. Sample analysis pathway: Automated gating of cellular population by AQUIOS STEM System Software. The software only allows gates/region readjustments by the user. Based on the ISHAGE gating strategy | Same |
| Count Method | Cellular events are compared to Stem Count Fluorospheres events. Automated method: Absolute counts are determined by FC500 stemCXP | Same Absolute counts are determined by AQUIOS STEM Software | Same |
510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 7 of 18
{8}
| | Predicate | Subject Device | Comments |
| --- | --- | --- | --- |
| Characteristic | Stem Kit Reagents and stemCXP system with the FC500 Flow Cytometer | AQUIOS STEM Kit System with AQUIOS CL Flow Cytometer | |
| Photomultiplier Tubes (PMTs) / Colors | - Forward Scatter – Solid state detector - Side Scatter Detector – photodiode - 4 Fluorescent Detectors –Photomultipliers - 525nm BP Filter - 575nm BP Filter - 620nm BP Filter - 675nm BP Filter | - FS Solid State Detector - SS Solid State Detector – (488LP Dichroic) - 4 Fluorescent Detectors - 525nm BP Filter - 575nm BP Filter - 620nm BP Filter - 675nm BP Filter | Substantially Equivalent |
| Detectors/Parameters | Six (FS, SS, FL1-FL4) | Seven (EV, FS, SS, FL1 – FL4) | Substantially Equivalent |
| Off-line analysis software | Analysis of Listmode files with stemCXP | Stand-alone workstations with AQUIOS STEM Software to review data offline | Substantially Equivalent |
| **Reagent and Controls Similarities and differences** | | | |
| Reagent Form | Liquid form | Same Liquid form | Same |
| Storage Conditions | 2 – 8 °C | Same except for lysing solution storage 2 – 8 °C Lysing solution to be stored at 18-25°C upon receipt | Substantially Equivalent |
| Set-up Reagents Standardization | Flow Set | Set at installation by Service. Requires only standardization verification via the AQUIOS STEM Controls. | Substantially Equivalent |
| Verification of Optical alignment and fluidics | Flow Check Fluorospheres | Same Flow Check Fluorospheres | Same |
| Control cell assay parameter | One Control Level: Positive | Two Control Levels | Substantially Equivalent |
| Reagent Assay | Stem Reagent Kit: - CD45-FITC / CD34-PE - CD45-FITC / IsoClonic Control-PE - Stem-Count Fluorospheres. - 7-AAD Viability Dye - 10X NH4Cl Lysing Solution | Same except for lysing solution AQUIOS STEM Reagent Kit: - AQUIOS STEM Kit CD45-FITC / CD34-PE - AQUIOS STEM Kit CD45-FITC / IsoClonic Control-PE | Substantially Equivalent |
510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 8 of 18
{9}
| | Predicate | Subject Device | Comments |
| --- | --- | --- | --- |
| Characteristic | Stem Kit Reagents and stemCXP system with the FC500 Flow Cytometer | AQUIOS STEM Kit System with AQUIOS CL Flow Cytometer | |
| | | • AQUIOS STEM Kit STEM-Count Fluorospheres • AQUIOS STEM Kit 7-AAD Viability Dye • AQUIOS STEM Kit Lysing Solution | |
| Clones | Specificity CD45 CD34 Clone J33 581 | Same Specificity CD45 CD34 Clone J33 581 | Same |
| Sample Preparation with Monoclonal Antibodies | Manually prepared by pipetting and adding specimen to reagents, incubating, lysing, adding Fluorospheres | On-Board Automated Preparation of the same steps as the predicate. | Substantially Equivalent |
| Lyse Reagents | Ammonium Chloride NH4CL | AQUIOS STEM Kit Lysing Solution | |
| Reagents for Color Compensation | Quick Comp 2 | Set at installation by Service and verified with the AQUIOS STEM control cells | Substantially Equivalent |
| Closed Vial Reagent Stability | 365 Days | 325 Days | Substantially Equivalent |
| Open Vial Reagent Stability Claim | 30 Days | 75 Days | Substantially Equivalent |
| On Board Stability | N/A | 70 Hours | N/A |
| Specimen Stability | • Whole Blood, and Mobilized Whole Blood - 20 hours at room temperature (18 - 25°C) • Cord Blood - 24 hours at room temperature (18 - 25°C) • Apheresis products and Bone Marrow - 24 hours at 2- 8°C | • Whole Blood, Mobilized Whole Blood, and Cord Blood - 24 hours at room temperature (18 - 25°C) • Bone Marrow and Apheresis 24 hours at 2- 8°C | Substantially Equivalent |
| Prepared Specimen Stability | • 60 minutes on ice after the addition of Stem Count Fluorospheres for Apheresis, Cord Blood, Bone Marrow Samples • 45 minutes on ice after the addition of Stem Count Fluorospheres for Whole Blood and Mobilized Whole Blood | Prepared Specimen Stability not applicable. After Lyse incubation of 15 - 20 minutes, the AQUIOS STEM Fluorospheres are added, and the sample is immediately analyzed | N/A |
| **Software Similarities and differences** | | | |
| System Configuration | FC500 System with PC-based workstation running stemCXP application-specific software | AQUIOS System with Reagent Cart and All-in-One Touch Screen Computer workstation running AQUIOS STEM application-specific software | Substantially Equivalent |
510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 9 of 18
{10}
| | Predicate | Subject Device | Comments |
| --- | --- | --- | --- |
| Characteristic | Stem Kit Reagents and stemCXP system with the FC500 Flow Cytometer | AQUIOS STEM Kit System with AQUIOS CL Flow Cytometer | |
| Operating System | Microsoft Windows WIN7 | Microsoft Windows WIN10 | Substantially Equivalent |
| Programming Language | C++ | C# VB.Net | Substantially Equivalent |
| Workstation Interface | USB to fiber optic | USB to Serial | Substantially Equivalent |
| Data Acquisition | Digital Pulse Processing | SAME | Same |
| General | Software functionality to allow: • Patient data management – storage, review, reporting • Control data management – storage, review, reporting • System configuration management System service test and adjustment procedures. | SAME | Same |
| Data Processing | • Region/gates evaluation • Statistics generation Export of results to MS Excel/PDFfile | SAME | Same |
| Standardization | • Automatic voltage/gain adjustments | Automatic gain calibration using target value | Substantially Equivalent |
| User interface | • Plots and reports | SAME | Same |
| Control Software | • The FC500 acquisition software enables the user to acquire data from the cytometer and to analyze, display, print and export acquired listmode data. The embedded software resides in the FC500 instrument. It controls the instrument functionality including the multi-carousel loader (MCL) for sample introduction. • The embedded software controls the instruments' lasers, acquisition system and fluidics. The instrument fluidics aspirates the sample and performs instrument maintenance functions such as startup/shutdown. The embedded software also captures and provides the data to the workstation for processing. | • The Aquios software enables the user to acquire data from the Aquios flow cytometer instrument and to analyze, display, print and export acquired fcs file. • The Aquios software controls the instrument functionality including the cassettes multi-loader for sample introduction. • The Aquios software controls the instruments' lasers, acquisition system and fluidics. The instrument fluidics aspirates the sample and performs instrument maintenance functions such as startup/shutdown. | Substantially Equivalent |
510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 10 of 18
{11}
| Characteristic | Predicate | Subject Device | Comments |
| --- | --- | --- | --- |
| | Stem Kit Reagents and stemCXP system with the FC500 Flow Cytometer | AQUIOS STEM Kit System with AQUIOS CL Flow Cytometer | |
| | | The Aquios Software also captures and provides the data to the workstation for processing. | |
| Data Processing | - Region and Gate processing - Statistics Generation - Export to Excel - Incoming data stream formatted into per-parameter vectors for improved performance of lin/log data transformations and compensation calculations | SAME | Same |
| User Account Control | - Multiple role assignments - User account information stored in password protected database. | SAME | Same |
| Report generator | - Panel reports & QC | SAME | Same |
510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 11 of 18
{12}
# Summary of AQUIOS CL Flow Cytometer Performance Testing
| Study | Testing Approach | FDA Guidance Documents | Standards/ References | Testing Results |
| --- | --- | --- | --- | --- |
| Instrument Linearity | Verify fluorescence detection is linear using standard AQUIOS CL settings. | None | None | Linearity of fluorescence measurements was demonstrated. |
| Laser Performance Characteristics | Verify stability of the laser performance of the AQUIOS CL flow cytometer over time. | None | None | Analysis of the data collected demonstrates that the AQUIOS CL laser performance is stable over time. |
| Instrument Carryover | To verify instrument carryover using SPHERO AccuCount Ultra-Rainbow Fluorescent Particles meets performance specifications. | Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells - Carryover (Section 12) | CLSI H26-A2; FDA Standards Recognition #7-210 | Analysis of the data collected demonstrates that the AQUIOS CL meets the whole blood carryover performance requirements. |
# Summary of AQUIOS STEM Kit Reagents Characterization and Analytical Testing:
| Study | Objective | FDA Guidance Documents | Standards/ References | Testing Results |
| --- | --- | --- | --- | --- |
| Antibody Specificity | Identify specific antibody clones used for the manufacturing of the AQUIOS STEM Kit Reagents. | None | None | Specific antibody clones that are used in the AQUIOS STEM Kit Reagents are identified. |
| Dose Titration | Determine the optimal dose (in μg/test) of each antibody conjugates and nucleic acid DNA dye 7AAD used in AQUIOS STEM Reagent Kit | None | None | Optimal dosage for antibody conjugates and 7-AAD determined. |
510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 12 of 18
{13}
| Study | Objective | FDA Guidance Documents | Standards/ References | Testing Results |
| --- | --- | --- | --- | --- |
| Antibody Staining and Lyse incubation Range | Verify the acceptable performance of varying antibody reagent and lyse incubation times on the AQUIOS CL Flow Cytometry System when processing multiple samples in parallel. | None | None | Acceptable staining and lyse incubation times verified. |
| Frozen specimen handling | Define the maximum time between the thawing of a frozen sample and the sample processing with the AQUIOS STEM SYSTEM on the AQUIOS CL Flow Cytometer | None | None | The study established the maximum time between the thawing of a frozen sample and the sample processing with the AQUIOS STEM system. |
| Lot to Lot Reproducibility - Reagent | Verify acceptable variability of multiple lots of AQUIOS STEM Kit | None | CLSI EP05-A3, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Third Edition. FDA Standards Recognition #7-251 | Analysis of the data collected demonstrates that the AQUIOS STEM Kit reagents have acceptable lot variability performance. |
| Lot to Lot Reproducibility – Control Cells | Verify acceptable variability of multiple lots of AQUIOS STEM CD34 Control Cells | None | CLSI EP05-A3, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Third Edition. FDA Standards Recognition #7-251 | Analysis of the data collected demonstrates that the AQUIOS STEM CD34 Control Cells have acceptable lot variability performance. |
| Reagent Stability | Demonstrate the shelf life of the AQUIOS STEM Kit reagents. | None | CLSI EP25-A – Evaluation of Stability of In Vitro Diagnostic Reagents. Approved Guideline. FDA Standards Recognition #7-235 | Analysis of the data collected demonstrates that the AQUIOS STEM Kit reagents meet performance requirements in support of the product’s stability claims. |
510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 13 of 18
{14}
| Study | Objective | FDA Guidance Documents | Standards/ References | Testing Results |
| --- | --- | --- | --- | --- |
| Interference | Determine the impact of interfering substances on the results from AQUIOS STEM SYSTEM. | None | CLSI EP07 Edition A3: Interference testing in clinical chemistry. Third edition. April 2018. FDA Recognition Number 7-275 CLSI EP37: Supplemental Tables for interference testing in clinical chemistry, 1st edition-April 2018. FDA Recognition Number 7-284 | Impact of interfering substances on AQUIOS STEM SYSTEM results determined. |
| Analytical Measuring Interval | Establish the Analytical Measuring Interval for the AQUIOS STEM Kit reagents when analyzed on the AQUIOS CL Flow Cytometry system. | None | • CLSI EP17-A2. Evaluation of the detection capability for clinical laboratory measurement procedures. Approved guideline. FDA Recognition Number 7-233 • CLSI EP-06 Ed2 Nov 2020 Evaluation of Linearity of Quantitative Procedures: Approved Guideline. FDA Recognition Number 7-306 | Analytical Measuring Intervals for each of the markers in AQUIOS STEM SYSTEM have been established for use with the AQUIOS CL flow cytometer. |
# Summary of AQUIOS STEM Kit System Performance Testing:
| Study | Testing Approach | FDA Guidance Documents | Standards/ References | Testing Results |
| --- | --- | --- | --- | --- |
| Assay Linearity | Verify the linear range of absolute count of CD34+ cells/μL and the absolute count of CD45+/μL using the AQUIOS STEM Kit reagents on | Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for | CLSI EP-06 Ed2 Nov 2020 Evaluation of Linearity of Quantitative Procedures: Approved Guideline. FDA Recognition Number 7-306 | Analysis of the data collected demonstrates that the AQUIOS STEM SYSTEM on AQUIOS CL Flow Cytometer meets the linearity performance requirements. |
510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 14 of 18
{15}
| Study | Testing Approach | FDA Guidance Documents | Standards/ References | Testing Results |
| --- | --- | --- | --- | --- |
| | the AQUIOS CL flow cytometer. | Immature or Abnormal Blood Cells - Linearity (Section 11) | | |
| Assay Carryover – Specimen and Reagent | Verify carryover of specimens and reagents on the AQUIOS CL meets performance specifications. | Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells - Carryover (Section 12) | CLSI H26-A2 Validation, Verification, and Quality Assurance of Automated Hematology Analyzers, Approved Standard - 2nd Edition; June 2010; FDA Standards Recognition # 7-210 ICSH Guideline Clin Lab Haematol, 16(2):157-174, 1994 | Analysis of the data collected demonstrates that the AQUIOS STEM SYSTEM meets carryover performance requirements. |
| Detection Capability | To establish and verify Limit of Blank (LoB), Lower Limit of Detection (LLoD), Lower Limit of Quantitation (LLoQ) for the AQUIOS STEM Kit reagents when analyzed on the AQUIOS CL instrument. | None | CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition; FDA Standards Recognition #7-233 | Analysis of the data collected established the Limit of Blank, Limit of Detection, and Limit of Quantitation values for each AQUIOS STEM Kit reagent when evaluated on the AQUIOS CL Flow Cytometry System for Detection Capability. |
| Specimen Age Stability and Acceptable Lyse Incubation Duration | Stability of various specimens was evaluated to verify specimen and acceptable lyse incubation duration. | Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells - Specimens (Section 13) | ICSH Guideline: Am J Clin Path, 100(4)371-372, 1993. | Analysis of the data collected demonstrates that the AQUIOS STEM SYSTEM when analyzed on the AQUIOS CL meet the requirements for specimen and lyse incubation duration. |
| Anticoagulant Equivalency | Demonstrate the equivalent performance of whole blood, mobilized peripheral whole blood, bone marrow, apheresis product, and cord blood specimens when collected in different anticoagulants K2EDTA, K3EDTA, Heparin, | Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells - Accuracy (Section 8) | CLSI EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples - Third Edition. FDA Standards Recognition 7-296 | Analysis of the data collected demonstrated equivalence of performance of various clinical specimens across anti coagulants when processed with AQUIOS STEM SYSTEM on AQUIOS CL Flow Cytometer. |
510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 15 of 18
{16}
| Study | Testing Approach | FDA Guidance Documents | Standards/ References | Testing Results |
| --- | --- | --- | --- | --- |
| | Heparin/ACD-A mix, ACD-A, and CPD when processed with the AQUIOS STEM Kit on the AQUIOS CL instrument. | | | |
| AQUIOS STEM Test Option equivalency study | Demonstrate the equivalent performance of AQUIOS STEM system in enumerating viable CD34+ cells/μL, percent (%) CD34+ and the count of viable CD45+ cells/μL when the samples are processed across these AQUIOS STEM system analysis options. | None | CLSI EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples - Third Edition | Equivalent performance of the three AQUIOS STEM system analysis options demonstrated. |
| Precision Study using Control Material | Demonstrate system precision for total CD45+ cells/μL), total CD34+ cells/μL and total percent (%) CD34+ using control material as a surrogate for a stabilized sample | Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells - Precision (Section 9) | CLSI EP5-A3, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Third Edition. FDA Standard Recognition # 7-251 | Analysis of the data collected demonstrates that the AQUIOS STEM SYSTEM on AQUIOS CL meets performance requirements for repeatability and reproducibility using control material. |
| Precision Study using Clinical Specimens (Specimen Repeatability) | Demonstrate assay precision for viable CD45+ cells/μL, viable CD34+ cells/μL and viable percent (%) CD34+ using clinical specimens | Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells - Precision (Section 9) | CLSI H26-A2 Validation, Verification, and Quality Assurance of Automated Hematology Analyzers, Approved Standard – 2nd Edition; June 2010; FDA Standards Recognition # 7-210 | Analysis of the data collected demonstrates that the AQUIOS STEM SYSTEM on AQUIOS CL meets repeatability performance requirements for clinical specimens. |
510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 16 of 18
{17}
| Study | Testing Approach | FDA Guidance Documents | Standards/ References | Testing Results |
| --- | --- | --- | --- | --- |
| Precision Study using Clinical Specimens (frozen apheresis) | Demonstrate site-to-site system precision for viable CD34+ cells/μL using frozen apheresis specimens | Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells - Precision (Section 9) | CLSI EP5-A3, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Third Edition. FDA Standard Recognition # 7-251 | Analysis of the data collected demonstrates that the AQUIOS STEM Kit on AQUIOS CL meets performance requirements for repeatability and reproducibility for frozen apheresis samples. |
| Measurement Procedure Comparison (Method comparison to predicate) | Demonstrate the substantial equivalency of the AQUIOS STEM System for the simultaneous identification and enumeration of viable CD45+ cells/μL, viable CD34+ cells/μL and viable percent (%) CD34+ | Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells - Accuracy (Section 8) | CLSI EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guidelines – Third Edition #7-296 | Analysis of the data collected demonstrates that the AQUIOS STEM Kit reagent shows acceptable substantial equivalence as compared to the predicate stemCXP SYSTEM. |
| Reference Interval | Establish adult reference interval for viable CD34+ cells/μL, viable percent (%) CD34+, and viable CD45+ cells/μL on the AQUIOS STEM Kit system. | Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells – Reference Values (Section 14) | CLSI EP28-A3c, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory. Third Edition FDA Standard Recognition # 7-224 | Adult reference intervals were established. |
510(k) Submission for AQUIOS STEM System
510(k) Summary
Confidential - Company Proprietary
Page 17 of 18
{18}
# **Substantial Equivalence Conclusion to Demonstrate Safety, Effectiveness & Equivalent Performance to Predicate:**
In conclusion, the AQUIOS STEM System as described in this submission is substantially equivalent in terms of safety and effectiveness to its predicate devices.
This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.
---510(k) Submission for AQUIOS STEM System
510(k) Summary
Page 18 of 18
Confidential - Company Proprietary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.