AMICUS Separator System

BK230824 · Fresenius Kabi AG Lake Zurich, IL · GKT · Apr 21, 2023 · Hematology

Device Facts

Record IDBK230824
Device NameAMICUS Separator System
ApplicantFresenius Kabi AG Lake Zurich, IL
Product CodeGKT · Hematology
Decision DateApr 21, 2023
DecisionSESE
Regulation21 CFR 864.9245
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The AMICUS Separator System is an automated blood cell separator intended for use in the collection of blood components and mononuclear cells. The AMICUS Separator System is an automated blood cell separator intended for use in therapeutic apheresis applications and may be used to perform therapeutic plasma exchange (TPE). The AMICUS Separator System is an automated blood component separator intended for use in therapeutic apheresis applications and may be used to perform red blood cell exchange, depletion, and depletion/exchange (RBCX) procedures.

Device Story

Continuous-flow centrifugal separator; draws whole blood from donor/patient; separates components by density; collects target components; returns remainder to donor/patient. Operates via pumps, clamps, and valves; controlled by operator via touch screen. Used in clinical settings for therapeutic apheresis or blood component collection. Software v6.1 and SPIKESMART hardware provide risk controls to prevent misconnection of solution containers during functionally closed kit procedures. Output includes collected blood components; system provides audible/visual alerts for monitoring. Benefits include automated, efficient component separation and therapeutic exchange capabilities.

Clinical Evidence

Bench testing only. Verification activities included functional and performance testing of new functionally closed kits, AMICUS Software v6.1, and SPIKESMART. Systems and software verification demonstrated functional equivalence to the predicate system.

Technological Characteristics

Continuous-flow centrifugal separator; uses pumps, clamps, and valves for fluid management. Disposable sterile fluid path kits. Touch screen interface. Software v6.1. SPIKESMART saline hook component. Connectivity: Standalone instrument. Sterilization: Not specified (disposables are sterile).

Indications for Use

Indicated for therapeutic plasma exchange (TPE); red blood cell exchange (RBCX) and depletion/exchange for sickle cell disease in adults and children; collection of platelets (leukocytes reduced), mononuclear cells, plasma (fresh frozen, PF24, PF24RT24, source), and red blood cells (leukocytes reduced).

Regulatory Classification

Identification

An automated blood cell separator is a device that uses a centrifugal or filtration separation principle to automatically withdraw whole blood from a donor, separate the whole blood into blood components, collect one or more of the blood components, and return to the donor the remainder of the whole blood and blood components. The automated blood cell separator device is intended for routine collection of blood and blood components for transfusion or further manufacturing use.

Special Controls

*Classification.* Class II (special controls). The special control for this device is a guidance for industry and FDA staff entitled “Class II Special Controls Guidance Document: Automated Blood Cell Separator Device Operating by Centrifugal or Filtration Separation Principle.”

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1} FRESENIUS KABI Fresenius Kabi Three Corporate Drive Lake Zurich, Illinois 60047 T 847-550-2300 T 888-391-6300 www.fresenius-kabi.us ## 510(k) SUMMARY ### Date Prepared 21 March 2023 ### Owner/Operator Owner/ Operator# 9027285 Fresenius Kabi AG Else-Kröner-Str. 1 61352 Bad Homburg Germany ### Contact Person Molly Mevis Specialist II, Regulatory Affairs Transfusion Medicine and Cell Therapies Fresenius Kabi MedTech Three Corporate Drive Lake Zurich, IL 60047 847-320-1584 (b) (6) @fresenius-kabi.com ### Device Trade Name AMICUS Separator System ### Common Name/Usual Name: Automated Separator, Blood Cell and Plasma, Therapeutic Automated Blood Cell Separator (Centrifugal Separation Principle) ### Classification Name 21 CFR 864.9245 Automated Blood Cell Separators Automated blood cell separators which are based on centrifugation type technology have been classified by the Center for Biologics Evaluation and Research as Class II devices with Special Controls (Docket 2005N-0017, Final Rule, 30-Nov-07; updated March 2011 OMB Control No. 0910-0594). Page 1 of 6 {2} FRESENIUS KABI Fresenius Kabi Three Corporate Drive Lake Zurich, Illinois 60047 T 847-550-2300 T 888-391-6300 www.fresenius-kabi.us Automated separators, used for separation of blood cells and plasma for therapeutic purposes, have not been classified under a regulation by the Center for Devices and Radiological Health due to pre-amendment status. ### **Product Code and Classification Panel** LKN (Gastroenterology/Urology panel) - Unclassified (due to pre-amendment status) GKT (Hematology panel) - Separator, Automated, Apheresis ### **Legally Marketed Device under which Substantial Equivalence is Being Claimed** Fresenius Kabi is claiming substantial equivalence with the currently cleared AMICUS Separator System. The AMICUS Separator System is comprised of the AMICUS Separator instrument and a disposable AMICUS apheresis kit specific to the procedure being performed. The AMICUS Separator System was most recently cleared for market by CDRH under K200530 on September 11, 2020, and by CBER under BK170099 on October 10, 2017. The use of AMICUS Kits with a PAS connector was previously cleared under BK090065 (March 4, 2010) and the use of functionally closed AMICUS Platelet Kits was previously cleared under BK150288 (August 28, 2015). ### **Device Description** The AMICUS Separator System is comprised of the AMICUS Separator instrument and a disposable apheresis kit specific to the procedure being performed. The AMICUS Separator is a continuous-flow, centrifugal device that operates using pumps, clamps, and valves to move donor/patient blood through a single-use, sterile fluid path disposable kit. The Separator draws whole blood from a donor/patient, separates the blood into its components, collects one or more of the blood components, and returns the remainder of the blood components to the donor/patient. The cells are centrifugally separated within the kit by density differences. The operator is responsible for connecting and monitoring the donor/patient and operating and monitoring the AMICUS Separator during the procedure. The operator controls the Separator through a touch screen. When necessary, the operator is alerted of problems with information on the screen and corresponding audible alerts. Once the cell separation and collection is complete, the operator disconnects the donor/patient, removes the kit from the top panel of the Separator, and disposes of the kit per institutional SOPs. The kit is packaged in a recyclable plastic tray. ### **Statement of Intended Use** The AMICUS Separator System is an automated blood cell separator intended for use in the collection of blood components and mononuclear cells. Page 2 of 6 {3} **FRESENIUS KABI** **Fresenius Kabi** Three Corporate Drive Lake Zurich, Illinois 60047 T 847-550-2300 T 888-391-6300 www.fresenius-kabi.us The AMICUS Separator System is an automated blood cell separator intended for use in therapeutic apheresis applications and may be used to perform therapeutic plasma exchange (TPE). The AMICUS Separator System is an automated blood component separator intended for use in therapeutic apheresis applications and may be used to perform red blood cell exchange, depletion, and depletion/exchange (RBCX) procedures. ### **Indications for Use** Depending on the AMICUS Separator System disposable used in the therapeutic apheresis procedure, the AMICUS Separator System has been cleared for the following: The AMICUS Separator System is an automated blood cell separator indicated to perform therapeutic plasma exchange (TPE). (K111702) The AMICUS Exchange Kit is indicated for use in therapeutic plasma exchange (TPE). The kit is for use with the AMICUS separator. (K111702) The AMICUS Separator System is an automated blood component separator indicated to perform red blood cell exchange (RBCX), including exchange and depletion/exchange procedures, for the transfusion management of sickle cell disease in adults and children. (K180615) The AMICUS Exchange Kit – Therapeutics is indicated for use in therapeutic plasma exchange (TPE) and red blood cell exchange (RBCX). The kit is for use with the AMICUS separator. (K111702, K180615) The waste transfer set is indicated for use in red blood cell exchange (RBCX). The set is for use with the AMICUS separator. (K180615) The Blood Component Filter Set with Vented Spike and Luer Adapter is indicated for the administration of blood and blood components during a therapeutic plasma exchange (TPE) or red blood cell exchange (RBCX) therapeutic apheresis procedure. The set is for use with the AMICUS separator. (K111702, K180615) The AMICUS Separator System is an automated blood cell separator indicated for the collection of blood components and mononuclear cells. Depending on the AMICUS Separator System apheresis kit used in the collection of products, the AMICUS Separator System has been cleared to collect: Platelets Pheresis, Leukocytes Reduced (single, double, or triple units) (BK960005, BK990009) Page 3 of 6 {4} FRESENIUS KABI Fresenius Kabi Three Corporate Drive Lake Zurich, Illinois 60047 T 847-550-2300 T 888-391-6300 www.fresenius-kabi.us Platelets Pheresis, Leukocytes Reduced, Platelet Additive Solution (InterSol) (single, double, or triple units) (BK090065) Red Blood Cells, Leukocytes Reduced (by apheresis) (BK000039) Mononuclear Cells (BK000047) Plasma (BK960005, BK120041) - Fresh Frozen Plasma - Must be prepared and placed in a freezer at -18° C or colder within eight hours after phlebotomy. - Plasma Frozen Within 24 Hours After Phlebotomy (PF24) - Must be stored at 1-6° C within eight hours after phlebotomy and placed in a freezer at -18° C or colder within 24 hours after phlebotomy. - Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma. - Plasma Frozen Within 24 Hours After Phlebotomy (PF24) Held at Room Temperature Up to 24 Hours After Phlebotomy (PF24RT24) - Can be stored at room temperature for up to 24 hours after phlebotomy. Product must be placed in a freezer at -18° C or colder within 24 hours after phlebotomy. - Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma. - Source Plasma The device is designed to collect products while maintaining an extracorporeal volume at or below 10.5 ml/kg and a donor post-platelet count greater than or equal to 100,000 platelets/microliter. (BK960005 and BK990009) Platelets Pheresis (single, double, or triple units) may be manufactured from products that do not meet leukocyte reduction product standards. This does not apply to Platelet Pheresis, Platelet Additive Solution (InterSol) (single, double, or triple units). (BK990009 and BK090065) Page 4 of 6 {5} **FRESENIUS KABI** **Fresenius Kabi**^{}[] Three Corporate Drive^{}[] ^{}[] Lake Zurich, Illinois 60047^{}[] ^{}[] T 847-550-2300^{}[] ^{}[] T 888-391-6300^{}[] ^{}[] www.fresenius-kabi.us The AMICUS platelet storage container is cleared to store Platelets Pheresis, Leukocytes Reduced in 100% plasma for up to seven days. Additionally, for platelet units stored past five days and through seven days, every product must be tested with a bacterial device cleared by the FDA and labeled as a “safety measure.” (BK050038 and BK150242) ### **Technological Comparison as Compared to the Predicate Device** The technological characteristics of the AMICUS Separator are standardized across procedures and remain the same as the predicate AMICUS Separator System. This includes the centrifuge system, separation technology, fluid control system, and anticoagulant management system which monitors and controls the delivery of anticoagulant solution based on operator parameters. The physical design of the AMICUS instrument includes three main areas of operation (touch screen, top panel, and centrifuge) that have the same technological characteristics as the predicate AMICUS device. There is also no change to the scientific technology or principles of operation of the AMICUS Separator System due to the introduction of the AMICUS SPIKESMART system for use with the new functionally closed kits. The data management capabilities of the AMICUS Separator are unchanged, and performance specifications are not impacted by AMICUS Software version 6.1. The two new functionally closed disposable kits (4R2358 and 4R2359) do not introduce any new components or materials to the currently cleared AMICUS apheresis kit family. These kits are being created to offer a new kit configuration to customers and have no impact on the AMICUS Separator System fundamental technology and principles of operation. ### **Modification to the Existing Device** The primary change to the currently cleared AMICUS apheresis kit family is the introduction of two new functionally closed apheresis kits with PAS spikes. Product labeling has been created for the new disposable kit product codes based on the currently cleared AMICUS apheresis kit labeling. The primary change to the AMICUS Separator covered under this submission is the introduction of AMICUS Software version 6.1 and AMICUS SPIKESMART, intended to support the use of the new functionally closed kits. AMICUS Software version 6.1 includes a new “Functionally Closed Kit” feature with additional risk control measures for platelet collection procedures, and a new optional transfer state for the MNC collection procedure. AMICUS SPIKESMART includes a replacement saline hook component designed to be specifically compatible with certain saline containers. The purpose of the added software risk controls and AMICUS SPIKESMART is to further reduce the risk of the operator misconnecting solution containers to the incorrect lines during platelet collection procedures with a functionally closed kit. Additionally, the Page 5 of 6 {6} **FRESENIUS KABI** **Fresenius Kabi** Three Corporate Drive Lake Zurich, Illinois 60047 T 847-550-2300 T 888-391-6300 www.fresenius-kabi.us software has been updated to allow for an alternate MNC transfer state option. The purpose of this feature is to provide the customer with more flexibility and procedural control. A new version of the currently cleared AMICUS Software version 6.0 Operator’s Manual has also been created to include instructions for operation with the new Functionally Closed Kit feature. This version of the Operator’s Manual is based on version 6.0 and is titled 6.1. ### **Performance Data** Performance testing and data submitted under previously cleared AMICUS Apheresis System filings remains valid for demonstrating performance of the AMICUS Separator System. Additional functional and performance testing was conducted to verify the use of the new functionally closed AMICUS apheresis kits with the AMICUS Software version 6.1 and SPIKESMART. Systems and software verification testing was also performed to demonstrate the functional equivalence of the proposed AMICUS Separator System device. The results of all verification testing demonstrate that the proposed AMICUS Separator System performs as intended in a safe and effective manner. ### **Conclusion** Based on the verification activities performed, the AMICUS Separator System modified with the new kit configurations, software version 6.1, and SPIKESMART, is substantially equivalent to the currently marketed AMICUS Separator System. Page 6 of 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...