Erytra Eflexis is a fully-automated analyzer designed to automate in vitro immunohematological testing of human blood utilizing DG Gel 8 cards technology, including Blood Grouping, Antigen Typing, Antibody Screening, Antibody Identification, Antibody Titration, Compatibility Tests, and Direct Antiglobulin Tests. As a standalone analyzer or interfaced to the customer's Laboratory Information System (LIS), Erytra Eflexis automates test processing functions and data management requirements using DG Gel 8 cards and digital image processing.
Device Story
Erytra Eflexis is a fully-automated analyzer for gel immunohematology testing; processes human blood samples using DG Gel 8 cards. System automates sample/reagent loading, incubation, centrifugation, and digital image analysis of microtubes. Operated by laboratory personnel in clinical, transfusion, or donation centers; functions as standalone or LIS-interfaced unit. Software interprets results based on predefined rules; provides objective, standardized analysis to eliminate manual transcription errors and reduce operator contamination risk. Output includes automated test results for blood grouping and antibody identification, facilitating clinical decision-making for transfusion compatibility. Version 3.0.0 introduces architectural updates, including Adobe AIR integration, unified Manager/Executor applications, and improved LIS communication.
Clinical Evidence
No clinical data provided. Verification and validation were conducted per internal life cycle and software validation protocols. Regression analysis and bench testing confirmed that software version 3.0.0 changes do not impact instrument performance characteristics compared to the predicate.
Technological Characteristics
Automated analyzer; uses DG Gel 8 cards; digital image processing for microtube analysis. Hardware includes 3 incubators, 2 centrifuges, and 1 pipetting unit. Connectivity via LIS (unidirectional/bidirectional). Software runs on Microsoft Windows 10 IoT Enterprise. Access control complies with 21 CFR Part 11 with encrypted passwords. Reagent red cell suspension maintained by rotation.
Indications for Use
Indicated for in vitro immunohematological testing of human blood, including blood grouping, antigen typing, antibody screening/identification/titration, compatibility testing, and direct antiglobulin tests in clinical, transfusion, and donation laboratory settings.
Regulatory Classification
Identification
An automated blood grouping and antibody test system is a device used to group erythrocytes (red blood cells) and to detect antibodies to blood group antigens.
{1}
Diagnostic Grifols, S.A.
510(k) Premarket Notification
Erytra Eflexis
5 - 1
## Section 5 – 510(k) Summary
### Submitter’s Details:
Name Diagnostic Grifols S.A.
Address Passeig Fluvial, 24, Parets del Valles
Barcelona, 08150, Spain
Establishment Registration 3002772505
Number:
Contact Person: Elvira Estapé Egea
Regulatory Affairs Manager
elvira.estape@grifols.com
Phone: (34) 670-922-237
Date of Summary: March 29th, 2023
### Name of Device:
Trade Name: Erytra Eflexis
Classification Name: Automated Blood Grouping and Antibody Test System
Device Class: II
Product Code: KSZ
Regulation Number: 21 CFR 864.9175
### Identification of the Legally Marketed Device (Predicated Device):
Trade Name: Erytra Eflexis
Classification Name: Automated Blood Grouping and Antibody Test System
510(k) Number: BK220752
Device Class: II
Product Code: KSZ
Regulation Number: 21 CFR 864.9175
Clearance Letter November 1, 2022
### Device Description:
Erytra Eflexis is designed to automate all necessary operations and procedures to process gel immunohematology tests, allowing laboratories to:
1. Create test profiles and optimize profile implementation in the shortest time and with the most accurate results.
2. Increase process safety and traceability by eliminating possible identification and transcription errors.
3. Increase analytical reliability by standardizing all steps, eliminating possible handling and processing errors, and interpreting the results with objective criteria.
4. Reduce the danger of contamination for operators by reducing operator interaction with
{2}
Diagnostic Grifols, S.A.
510(k) Premarket Notification
Erytra Eflexis
5 - 2
the samples and reagents during the analytical process. Operator interaction is limited to the loading and unloading of the analyzer.
In addition, Erytra Eflexis adapts to the needs and differing operational workflows in immunohematology laboratories, donation centers, transfusion centers, and clinical testing laboratories, as well as different work rhythms (routine, emergency) and the flow of samples processed over different shifts.
Erytra Eflexis automates the following gel immunohematology tests:
- ABO Red Cell and Serum Grouping.
Rh(D) Typing.
Antigen Typing.
Antibody detection.
Antibody identification.
Antibody titration.
- Direct Antiglobulin test.
- Compatibility Tests (Crossmatching).
# **Indications for Use:**
Erytra Eflexis is a fully-automated analyzer designed to automate in vitro immunohematological testing of human blood utilizing DG Gel 8 cards technology, including Blood Grouping, Antigen Typing, Antibody Screening, Antibody Identification, Antibody Titration, Compatibility Tests, and Direct Antiglobulin Tests.
As a standalone analyzer or interfaced to the customer's Laboratory Information System (LIS), Erytra Eflexis automates test processing functions and data management requirements using DG Gel 8 cards and digital image processing.
{3}
Diagnostic Grifols, S.A.
510(k) Premarket Notification
Erytra Eflexis
5-3
Comparison to Predicate Device:
| Parameter | Predicate Device Diagnostic Grifols S.A. Erytra Eflexis with software version 1.2.5 (BK220752) | Subject Device Diagnostic Grifols S.A. Erytra Eflexis with new software version 3.0.0 |
| --- | --- | --- |
| General | | |
| Indications for Use Statement | Erytra Eflexis is a fully-automated analyzer designed to automate in vitro immunohematological testing of human blood utilizing DG Gel 8 cards technology, including Blood Grouping, Antigen Typing, Antibody Screening, Antibody Identification, Antibody Titration, Compatibility Tests, and Direct Antiglobulin Tests. As a standalone analyzer or interfaced to the customer’s Laboratory Information System (LIS), Erytra Eflexis automates test processing functions and data management requirements using DG Gel 8 cards and digital image processing. | Erytra Eflexis is a fully-automated analyzer designed to automate in vitro immunohematological testing of human blood utilizing DG Gel 8 cards technology, including Blood Grouping, Antigen Typing, Antibody Screening, Antibody Identification, Antibody Titration, Compatibility Tests, and Direct Antiglobulin Tests. As a standalone analyzer or interfaced to the customer’s Laboratory Information System (LIS), Erytra Eflexis automates test processing functions and data management requirements using DG Gel 8 cards and digital image processing. |
| Classification | II | II |
| Product Code | KSZ | KSZ |
| Regulation number | 21 CFR 864.9175 | 21 CFR 864.9175 |
| Common name | Automated Blood Grouping and Antibody Test System | Automated Blood Grouping and Antibody Test System |
| Test performed | - ABO Red Cell and Serum Grouping - Rh(D) Typing - Antigen Typing - Antibody detection - Antibody identification - Antibody titration - Direct Antiglobulin test - Compatibility Tests (Crossmatching) | - ABO Red Cell and Serum Grouping - Rh(D) Typing - Antigen Typing - Antibody detection - Antibody identification - Antibody titration - Direct Antiglobulin test - Compatibility Tests (Crossmatching) |
| Primary components | Analyzer Computer Software Optional hand-held bar code scanner Optional printer | Analyzer Computer Software Optional hand-held bar code scanner Optional printer |
| Specimen Types | Plasma, Serum and Red Blood Cells. | Plasma, Serum and Red Blood Cells. |
{4}
Diagnostic Grifols, S.A.
510(k) Premarket Notification
Erytra Eflexis
5 - 4
| Parameter | Predicate Device Diagnostic Grifols S.A. Erytra Eflexis with software version 1.2.5 (BK220752) | Subject Device Diagnostic Grifols S.A. Erytra Eflexis with new software version 3.0.0 |
| --- | --- | --- |
| **Reagents** | Erytra Eflexis is used with Diagnostic Grifols DG Gel 8 cards, Medion Grifols Diagnostics Reagent Red Blood Cells and Validated Antisera Reagents. | Erytra Eflexis is used with Diagnostic Grifols DG Gel 8 cards, Medion Grifols Diagnostics Reagent Red Blood Cells and Validated Antisera Reagents. |
| **Positive identification of samples and reagents** | Yes | Yes |
| **Throughput** | 36 samples (ABO/Rh cards) per hour, including forward & reverse group | 36 samples (ABO/Rh cards) per hour, including forward & reverse group |
| **Useful life** | 7 years, considering 8 hours a day of continuous operation at maximum throughput (using ABO/Rh DG Gel 8 cards), 250 days a year. | 7 years, considering 8 hours a day of continuous operation at maximum throughput (using ABO/Rh DG Gel 8 cards), 250 days a year. |
| **Hardware** | | |
| **Reagent Red Cell suspension** | Maintained by Rotation Movement | Maintained by Rotation Movement |
| **Barcode Symbologies** | - Interleaved 2 of 5. - Code 3 of 9. - Codabar. - EAN 8 / EAN 13 / JAN 8 / JAN 13. - Codes 128 A, B & C (ISBT 128). - Others under configuration. | - Interleaved 2 of 5. - Code 3 of 9. - Codabar. - EAN 8 / EAN 13 / JAN 8 / JAN 13. - Codes 128 A, B & C (ISBT 128). - Others under configuration. |
| **Sample loading capacity** | 72 tubes simultaneously | 72 tubes simultaneously |
| **Reagent loading capacity** | 2 removable, independent and random-access racks. | 2 removable, independent and random-access racks. |
| **Sample/Reagent Dispensing (pipetting) Unit** | 1 | 1 |
| **Card loading capacity** | 200 cards | 200 cards |
| **Incubator** | 3 independent incubators | 3 independent incubators |
| **Centrifuge** | 2 independent centrifuges | 2 independent centrifuges |
| **System solution and waste containers** | 1 container for Grifols Wash Solution A 1 container for Grifols Wash Solution B 2 containers for waste solutions that can be configured to become Wash Solution containers if external drain is used. 1 disposable container for processed DG Gel 8 cards | 1 container for Grifols Wash Solution A 1 container for Grifols Wash Solution B 2 containers for waste solutions that can be configured to become Wash Solution containers if external drain is used. 1 disposable container for processed DG Gel 8 cards |
{5}
Diagnostic Grifols, S.A.
510(k) Premarket Notification
Erytra Eflexis
5-5
| Parameter | Predicate Device Diagnostic Grifols S.A. Erytra Eflexis with software version 1.2.5 (BK220752) | Subject Device Diagnostic Grifols S.A. Erytra Eflexis with new software version 3.0.0 |
| --- | --- | --- |
| Firmware/Software | | |
| Operating System | Microsoft Windows 10 IoT Enterprise | Microsoft Windows 10 IoT Enterprise |
| Software version | 1.2.5 | 3.0.0 |
| Software applications | Executor, Manager, Importer, Technical Service, Diagnostic, System Tools and Diana Server. | Eflexis, Importer, Technical Service, Diagnostic, System Tools and Diana Server. |
| Interfaced with Laboratory Information System (LIS) | Yes - Unidirectional. - Bidirectional. | Yes - Unidirectional. - Bidirectional. |
| Capability to process STAT samples | Yes | Yes |
| QC procedures implemented | Yes | Yes |
| Manual Entry of Sample IDs | Requires Double Blind Entry | Requires Double Blind Entry |
| Incubator temperature | Incubators can be set up independently at 37°C or 24°C. Temperature accuracy at 37°C: ±1°C Temperature accuracy at 24°C: ±3°C | Incubators can be set up independently at 37°C or 24°C. Temperature accuracy at 37°C: ±1°C Temperature accuracy at 24°C: ±3°C |
| Relative centrifugal force | 127g +/- 12g at 4.6mm from the bottom of the microtube. | 127g +/- 12g at 4.6mm from the bottom of the microtube. |
| Centrifugation time | 9.0 minutes +/- 0.5min (measured from the starting to accelerate moment to the starting to brake moment). | 9.0 minutes +/- 0.5min (measured from the starting to accelerate moment to the starting to brake moment). |
| Automated results reading | Yes. Microtube Digital Image Analysis. | Yes. Microtube Digital Image Analysis. |
| Automated test interpretation | Yes. According to predefined rules stated in a definition file. | Yes. According to predefined rules stated in a definition file. |
| Access control | User access limited to authorized individuals with specific privileges. Every user is unique and has a user identifier and specific password. User's passwords are encrypted. Complies with 21CFR11. | User access limited to authorized individuals with specific privileges. Every user is unique and has a user identifier and specific password. User's passwords are encrypted. Complies with 21CFR11. |
| Barcode Reading | Sample Identification Reagent Lot N°. and Expiration Date | Sample Identification Reagent Lot N°. and Expiration Date |
{6}
Diagnostic Grifols, S.A.
510(k) Premarket Notification
Erytra Eflexis
5 - 6
## Main functionalities introduced with software version 3.0.0
The new software version 3.0.0 introduces both new functionalities, major architecture changes and resolution of defects and complaints. The main functionalities are described in Section 16 - Software. A detailed Revision History table including all changes from version 1.2.5 to 3.0.0 is included in Attachment 1.
The main functionalities introduced in software version 3.0.0 are:
- Implement use of Adobe AIR technology
- To create one single application for Manager and Executor applications
- Implement reagents stability on board functionality
- Implement the new planification of techniques in incubators and centrifuges
- Implement the management of reusable cards area
- Modify the decontamination process of the instrument
- Modify Eflexis Diagnostic application
- Improve the communication with LIS
- Implement new Backup and Restore applications
## Performance:
Verification and Validation activities were conducted following the instruction DI.INS.301_ING “Life and Cycle and Software Validation”. Regression analysis and Verification testing data showed no impact on the instrument’s performance characteristics by the changes described in this submission.
## Conclusions:
Diagnostic Grifols S.A. concludes, based on all the information submitted and discussed in this submission and in this summary that Erytra Eflexis with the new software version 3.0.0, when used for the defined indications for use, performs as well as or better than the legally marketed predicate device Erytra Eflexis (BK220752) running under software 1.2.5.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.