BK230807 · Genesis Biologics, Inc. · ORG · May 10, 2023 · Hematology
Device Facts
Record ID
BK230807
Device Name
Adaptive Autologous Processing System
Applicant
Genesis Biologics, Inc.
Product Code
ORG · Hematology
Decision Date
May 10, 2023
Decision
SESE
Regulation
21 CFR 864.9245
Device Class
Class 2
Indications for Use
The Adaptive Autologous Processing System™ (AAPS) is indicated to be used for the safe and rapid preparation of autologous platelet-rich plasma (PRP) from a small sample of peripheral blood at the patient point-of-care. The PRP is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics.
Device Story
Single-use, sterile syringe assembly for autologous blood component separation; utilizes desktop centrifuge for density-based separation. Input: whole blood sample. Operation: blood drawn into main chamber; centrifugation separates plasma, buffy coat, and red blood cells; hollow threaded plunger with control valve allows aspiration of plasma/buffy coat into secondary chamber; red blood cells dispelled. Output: concentrated PRP for mixing with bone graft. Used at point-of-care by clinicians. Benefits: improved handling characteristics of bone graft materials for surgical applications.
Clinical Evidence
Bench testing only. Comparative study evaluated platelet concentrates from subject vs. predicate device. Metrics included platelet yields, recoveries, average concentration, pH, p-selectin expression, ADP-stimulation response, platelet function, aggregation, hypotonic stress response, and bone graft retention. All validations and qualifications met predetermined acceptance criteria.
Technological Characteristics
Syringe assembly with hollow threaded plunger and locking collar. Materials: Polycarbonate, Thermoplastic elastomer, silicone, Polyoxymethylene Copolymer. Principle: Density-based centrifugal separation. Energy: Manual. Connectivity: None. Sterilization: Ethylene oxide (ISO 11135, SAL 10⁻⁶). Form factor: 25mL syringe.
Indications for Use
Indicated for patients requiring autologous platelet-rich plasma (PRP) preparation from peripheral blood at point-of-care for mixing with bone graft materials to improve handling during application to bony defects.
Regulatory Classification
Identification
An automated blood cell separator is a device that uses a centrifugal or filtration separation principle to automatically withdraw whole blood from a donor, separate the whole blood into blood components, collect one or more of the blood components, and return to the donor the remainder of the whole blood and blood components. The automated blood cell separator device is intended for routine collection of blood and blood components for transfusion or further manufacturing use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is a guidance for industry and FDA staff entitled “Class II Special Controls Guidance Document: Automated Blood Cell Separator Device Operating by Centrifugal or Filtration Separation Principle.”
{1}
# 510(k) Summary
Prepared: May 9, 2023
| I. | **Company** | Genesis Biologics, Inc. 16782 Von Karman Ave., Suite 17 Irvine, CA 92606 United States |
| --- | --- | --- |
| II. | **Contact** | Trevor J. Denbo Submissions Consultant +1.949.424.7607 |
| III. | **Product Trade Name** | Adaptive Autologous Processing System |
| IV. | **Common Name** | Platelet and Plasma Separator for Bone Graft Handling |
| V. | **Classification Name** | Automated blood cell separator |
| VI. | **Regulation Number** | 21 CFR 864.9245 |
| VII. | **Device Class** | Class II |
| VIII. | **Classification Product Code** | ORG |
| IX. | **Predicate Device** | BK190406 Arthrex Double Syringe (ACP)® System |
# X. **Description**
The *Adaptive Autologous Processing System* is a sterile, single use, syringe assembly that is intended for separating and concentrating blood components along with the use of a desktop centrifuge. It enables blood to be separated and aspirated in a single device after centrifugation.
The *Adaptive Autologous Processing System* consists of a syringe with a hollow threaded plunger that provides a secondary chamber and locking collar that engages the thread plunger for the controlled collection of whole blood and the separation of platelet rich plasma. A sample is introduced to the main chamber of the syringe, the locking collar is engaged, and centrifugation is performed. The whole blood is separated into discrete gradient layers based upon density with the plasma located on top and the buffy coat suspended directly below followed by the red blood cells at the distal end of the syringe. The secondary chamber located inside the plunger is opened through a control valve and with a twisting motion downward the upper plasma layer securely enters the secondary chamber until the buffy coat is captured. Once the plasma and buffy coat are secured the secondary chamber is closed and the red blood cells and residual hematocrit are dispelled out of the syringe.
The threaded plunger with its secondary hollow chamber enables controlled capture of the plasma for forms a stable barrier between the plasma layers and the hematocrit components.
The centrifuge and blood draw accessories are not included with the system and is not a part of the present submission.
{2}
# XI. Indications for Use
The Adaptive Autologous Processing System™ (AAPS) is indicated to be used for the safe and rapid preparation of autologous platelet-rich plasma (PRP) from a small sample of peripheral blood at the patient point-of-care. The PRP is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics.
# XII. Summary of the Technological Characteristics
The Adaptive Autologous Processing System is an improved syringe assembly, intended for the collection of whole blood and the separation of platelet rich plasma in a single device after they have been centrifuged. The Adaptive Autologous Processing System consists of a syringe with a hollow threaded plunger that provides a secondary chamber and locking collar that engages the thread plunger for the controlled collection of whole blood and the separation of platelet rich plasma
The fundamental scientific technology, design, materials of construction, processing methods and mechanism of operation of the subject device are identical to the predicate device. All devices are provided as sterile concentrating systems, designed to concentrate and aid in separation of a starting source material (blood) by density using a centrifuge. All devices include a single-use, disposable receptacle (e.g. concentrating device, separator, centrifuge tube assembly, etc.) that is designed to accept a volume of blood, and then undergo centrifugal processing to obtain platelet rich plasma (PRP). The table below summarizes the comparison of characteristics between the subject and predicate device.
| | Arthrex Double Syringe (ACP) (BK190406) | Adaptive Autologous Processing System (Subject Device) |
| --- | --- | --- |
| Indications for Use | The Arthrex Double Syringe (ACP) Kit is indicated for the safe and rapid preparation of autologous platelet rich plasma (PRP) from a small sample of peripheral blood at the patient's point of care. The PRP is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics. | The Adaptive Autologous Processing System™ (AAPS) is indicated to be used for the safe and rapid preparation of autologous platelet-rich plasma (PRP) from a small sample of peripheral blood at the patient point-of-care. The PRP is mixed with autograft and/or allograft bone prior to application to a bony defect for improving handling characteristics. |
| Regulation | 21 CFR 864.9245 | Identical |
| Device Class | II | Identical |
| Device Code | ORG | Identical |
| System Components | • Upper syringe • Lower syringe • Threaded cap | • Syringe assembly • Threaded cap |
| Materials | Polypropylene, Thermoplastic elastomer, Silicone, ACD-A | Polycarbonate, Thermoplastic elastomer, silicone, Polyoxymethylene Copolymer |
| Principle of Operation | Separation of blood based on density | Identical |
| Energy source | Manual | Identical |
| Method of Processing | Centrifugation | Identical |
| Centrifuge Device | General purpose centrifuge | Identical |
| Sterile | Yes | Yes |
| Sizes | 15 mL | 25mL |
{3}
The Reference Device used for this submission, BK200449 Healeon Duet, has the identical design in every way. Only differences are labeling, by necessity to account for the difference in legal manufacturer and some subtle differences in the instructions for use:
- include a second spin option,
- change spin from 10 minutes @ 400RCF to 2 minutes @ 1500 RCF, and
- add a dispelling of hematocrit step.
# XIII. Discussion of the Non-Clinical Testing
# Biocompatibility Testing
Biocompatibility testing on the patient contacting materials of the device was conducted in accordance with ISO 10993-1 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process. Per ISO 10993-1, the Adaptive Autologous Processing System is categorized as an externally communicating device, with limited exposure (contact < 24 hours) with indirect blood contact. Testing included cytotoxicity (per ISO 10993-5), sensitization & intracutaneous reactivity (per ISO 10993-10), acute systemic toxicity (per ISO 10993-11), pyrogenicity per USP: 151 and USP 41-NF 36.
# Sterilization Validation
A sterilization dose qualification using the half-cycle method in compliance with ISO 11135:2014 was performed to validate an ethylene oxide sterilization process for Adaptive Autologous Processing System. Results demonstrated that the product is reliably sterilized to a 10⁻⁶ sterility assurance level (SAL) using these predetermined parameters. Genesis Biologics, Inc. intends to use this same procedure to increase sterilization efficiency, such as load capacity, configuration, or location post clearance.
# Transport and Shelf-life
A transportation validation per ASTM D4169 and a packaging shelf life validation per ISO 11607 was conducted using accelerated aging to demonstrate that the package is designed, manufactured, and packed in such a way that the characteristics and performances of the packaging during the intended use will not adversely be affected during the full life cycle of the device; that there is assurance that the device is sterile when placed in the market and will remain sterile, under the storage and transport conditions laid down, until the protective packaging is opened; and that the product will be kept without deterioration at the high level of cleanliness so as to minimize the risk of microbial contamination.
# Usability / Human Factors
A validation was conducted to demonstrate that for Adaptive Autologous Processing System user interface was designed and engineered to maximize the likelihood that this product will be safe and effective for the intended users, uses, and use environments.
# Predicate Equivalency Evaluation
A study was conducted to compare platelet concentrates produced by the for Adaptive Autologous Processing System and those produced by the predicate. Parameters included platelet yields, platelet recoveries, average PRP platelet concentration, pH, p-selectin expression on resting platelets, platelet response to ADP-stimulation, platelet function, platelet aggregation, hypotonic stress response, and Bone graft material retention. This evaluation demonstrated that the platelet concentrates obtained by the Adaptive Autologous Processing System is substantially equivalent to those of the predicate.
{4}
Additionally, the *Adaptive Autologous Processing System* and the predicate possess the same indications, device class, device code, principles of operation, methods of processing, sterility assurance level, and biocompatibility. The subject device uses components readily available to the user, and the size, where the larger size of the subject device provides increased usability. None of the differences negatively impact the device’s substantial equivalence when compared to the predicate. All validations, verifications, and qualifications passed the predetermined acceptance criteria. The *Adaptive Autologous Processing System* therefore has been shown to be substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.