Procedure Tray - Lipoaspirate fat harvest and injection kit
BK220775 · Hydra Srl Italy · QUB · Jan 4, 2023 · General, Plastic Surgery
Device Facts
Record ID
BK220775
Device Name
Procedure Tray - Lipoaspirate fat harvest and injection kit
Applicant
Hydra Srl Italy
Product Code
QUB · General, Plastic Surgery
Decision Date
Jan 4, 2023
Decision
SESE
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
LG PT 15, LG PT 14, LG PT 12, LG PT 04, LG PT 03, LG PT 02 are intended for removal and transfer/reimplantation of autologous adipose tissue without further manipulation for aesthetic body contouring, within the same procedure. In combination with the Lipogems System these devices are intended for use in the following surgical specialties: orthopedic surgery, arthroscopic surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and plastic and reconstructive surgery.
Device Story
Procedure Tray (LG PT series) is a sterile, single-use kit containing stainless steel cannulas (infiltration, aspiration, injection), plastic syringes, and Luer lock connectors; used for manual harvesting, transfer, and re-implantation of autologous adipose tissue. Operated by surgeons in clinical settings; components housed in thermoformed tray to facilitate ergonomic workflow and decanting. Device used in conjunction with Lipogems process device for mechanical micro-fragmentation. No active components (e.g., pumps) permitted. Output is processed adipose tissue for re-implantation; benefits include standardized, easy-to-locate components for fat grafting procedures.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing and biocompatibility evaluation (cytotoxicity, intracutaneous reactivity, acute systemic toxicity, LAL test, pyrogen test, and delayed hypersensitivity test). All biocompatibility results were compliant.
Technological Characteristics
Components: stainless steel cannulas (17G, 13G, 20G) and medical-grade plastic syringes/connectors. Sterilization: Ethylene Oxide (EtO). Form factor: thermoformed tray with defined slots for syringes. Connectivity: None (manual, standalone). Shelf-life: 5 years. Biocompatibility: ISO 10993-1 compliant.
Indications for Use
Indicated for patients requiring removal and transfer/reimplantation of autologous adipose tissue for aesthetic body contouring or surgical procedures in orthopedic, arthroscopic, neurosurgical, gastrointestinal, urological, general, gynecological, thoracic, laparoscopic, and plastic/reconstructive surgery.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{0}
# Hydra
510(k) Number: BK220775 December, 2 2022
# 510(k) Summary
## I. SUBMITTER
Company name: Hydra Srl
Company address: Via Savonarola n° 4
41037 Mirandola (MO)
Italy
Email: (b) (6) @hydra-srl.com - (b) (6) @hydra-srl.com
Contact person: Serena Coronati – (b) (6) @coronaticonsulting.it
Date of submission: September 27, 2022
## II. DEVICE
Name of Device: PROCEDURE TRAY
Common or Usual Name: PROCEDURE TRAY
Proprietary Name /Device trade name: LG PT 15, LG PT 14, LG PT 12, LG PT 04, LG PT 03, LG PT 02
Classification Name: 21 CFR 878.5040 Suction lipoplasty system
Regulatory Class: II
Product Code: QUB
## III. PREDICATE DEVICE
Name of predicate device: KTA Adipose Treatment Kit
Manufacturer: Stemcis sas / Jointechlabs, Inc.
510(k) number: K190386 / BK220718-0
This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
## IV. DEVICE DESCRIPTION
The medical device 'Procedure Tray' (trade name LG PT XX) is a sterile, single-use kit that contain a selection of cannulas (for infiltration, fat harvesting and fat re-injection), syringes and transfer luer lock adaptors in one easy to use and locate kit.
The medical device is used in procedure for subcutaneous fat harvesting and re-implant in combination with 'Lipogems process device' manufactured by Lipogems International Spa through mechanical micro-fragmentation.
Section 05 – Page 1
{1}
# Hydra
510(k) Number: BK220775 December, 2 2022
The medical procedure includes the following steps:
- Harvesting of adipose tissue.
- Processing of the harvested tissue (involves only Lipogems process device)
- Injection of the tissue to the treatment site.
Six different configurations are available in order to meet patients' needs:
| REF / Code | Description | Kit composition | Quantity |
| --- | --- | --- | --- |
| LG PT 12 | Lipoaspirate fat harvest and injection kit - 12 pieces (without cannulas) | - 20 ml aspiration syringe - 1 ml injection syringe - 10 ml syringe - 50 ml syringe - Luer Lock female to female connector - Tray L | 1 3 4 2 2 1 |
| LG PT 14 | Lipoaspirate fat harvest and injection kit - 14 pieces (2 cannulas) | - 20 ml aspiration syringe - 1 ml injection syringe - 10 ml syringe - 50 ml syringe - Luer Lock female to female connector - 17G infiltration cannula - 13G aspiration cannula - Tray L | 1 3 4 2 2 1 1 1 |
| LG PT 15 | Lipoaspirate fat harvest and injection kit - 15 pieces (3 cannulas) | - 20 ml aspiration syringe - 1 ml injection syringe - 10 ml syringe - 50 ml syringe - Luer Lock female to female connector - 17G infiltration cannula - 13G aspiration cannula - 20G injection cannula - Tray L | 1 3 4 2 2 1 1 1 1 |
| LG PT 04 | Lipoaspirate fat harvest and injection kit - 4 pieces (3 cannulas) | - 20 ml aspiration syringe - 17G infiltration cannula - 13G aspiration cannula - 20G injection cannula - Tray S | 1 1 1 1 1 |
Section 05 – Page 2
{2}
Hydra
510(k) Number: BK220775 December, 2 2022
| REF / Code | Description | Kit composition | Quantity |
| --- | --- | --- | --- |
| LG PT 03 | Lipoaspirate fat harvest and injection kit - 3 pieces (3 cannulas) | - 17G infiltration cannula - 13G aspiration cannula - 20G injection cannula - Tray S | 1 1 1 1 |
| LG PT 02 | Lipoaspirate fat harvest and injection kit - 2 pieces (2 cannulas) | - 17G infiltration cannula - 13G aspiration cannula - Tray S | 1 1 1 |
**Table 1** – Brief description of the available configurations of "Procedure Tray"
Components brief description:
- 20 ml Suction Syringe with Male Luer Lock Connector and Four Position Plunger Lockout Device.
- 50 ml syringes with male Luer Lock connector.
- Female-female Luer Lock connectors, preassembled on the 50 ml syringes, designed to allow the transfer of biological material from the collection syringe to those necessary to use the processing device and from the latter to those necessary for the execution of the plant.
- 10 ml syringes with male Luer Lock connector, intended for the infiltration of the anaesthetic solution, for the loading of the processing device and for the extraction of the material processed by it.
- Syringes of 1 ml intended for replanting the microfragmented tissue.
- 17Gx185 mm cannula for infiltration with Luer-Lock connector, equipped with rounded tip and 6 orifices on its lateral surface, equipped with protective cap.
- 13Gx185 mm suction cannula with Luer-Lock connector, equipped with rounded tip and 5 suction orifices on its lateral surface, equipped with protective cap.
- 20Gx70 mm injection cannula, equipped with a rounded tip and an elongated lumen on the lateral surface, equipped with a protective cap.
Section 05 – Page 3
{3}
Hydra
510(k) Number: BK220775 December, 2 2022
All cannulas contained in LG PT 14, LG PT 15, LG PT 02, LG PT 03 and LG PT 04 “Procedure Tray” are cannulas with a rounded tip for manual use and are disposable, intended for use in general surgical procedures and particularly suitable for the collection and transfer of autologous adipose tissue. They should be used exclusively with the manual syringes contained in the same procedural tray. Each cannula is provided with a protector sheath and identified by a colour code:
Infiltration cannula 17Gx185 mm: Red
Aspiration cannula 13Gx185 mm: Blue
Injection cannula 20Gx70 mm: Yellow
The cannulas must not be used in combination with active devices of any type such as, for example, motorized suction or injection pumps.
The tray is produced by a mould of thermoformable material. Procedure tray allows each component to be housed in a defined position improving usability and ergonomic requirements during surgical procedure. Procedure tray is available in two different dimensions: Tray L used for codes LG PT 15, LG PT 14 and LG PT 12 and Tray S used for codes LG PT 02 LG PT 03 and LG PT 04.
The procedure tray provides slots for the decanting phase (i.e., to separate the aqueous phase, or emulsified oils, from the lipid phase of the content); in particular four slots for the vertical positioning of the 10 ml syringes and an inclined position for the 50 ml syringe.
“Procedure Tray” is sterile (sterilized with ethylene oxide gas) and it is intended for single patient use only.
# V. INDICATIONS FOR USE
LG PT 15, LG PT 14, LG PT 12, LG PT 04, LG PT 03, LG PT 02 are intended for removal and transfer/reimplantation of autologous adipose tissue without further manipulation for aesthetic body contouring, within the same procedure. In combination with the Lipogems System these devices are intended for use in the following surgical specialties: orthopedic surgery, arthroscopic surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and plastic and reconstructive surgery.
The indications for use are the same of the predicate device.
Section 05 – Page 4
{4}
# Hydra
510(k) Number: BK220775 December, 2 2022
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| | Procedure Tray | Predicate Device BK220718-0/ K190386 |
| --- | --- | --- |
| **General description of device** | The medical device 'Procedure Tray' (trade name LG PT xx) contains a selection of cannulas (for infiltration, fat harvesting and fat reinjection), syringes and transfer luer lock adaptors in one easy to use and locate kit that is available in 6 different versions. | The KTA Adipose Treatment Kit range consists in a family of products which are designed to harvest, process with minimal manipulation and reinject small – KTAMICRO, KTAEASY, KTAMY – to large – KTAMACRO, KTAMACROMED, KTASPIN – volumes of autologous fat. Components: syringes with removable plunger, centrifuge tubes with filter, aspiration & fat injection cannula, fat injection syringes, skin puncture needles, and oil extraction syringe & needle. |
| **Device components** | Components of the most complete version of 'Procedure Tray' LG PT 15: 20 ml aspiration syringe, 1 ml injection syringe, 10 ml syringe, 50 ml syringe, Luer Lock female to female connector, 17G infiltration cannula, 13G aspiration cannula, 20G injection cannula, Tray. | Components of KTAMACRO: 1 infiltration cannula (Ø 2.1 mm x 250 mm), 2 harvesting cannula 20 holes (Ø 3.0 x 3.5 mm x 250 mm) and (Ø 3.0 x 3.5 mm x 200 mm), 6 incremental syringes – 60 ml, 2 injection cannulas – 14G (Ø 1.7 – 2.1 mm x 200 mm) and (Ø 1.7 – 2.1 mm x 130 mm), 10 injection syringes – 10 ml, 2 needles – 18G / 18G NOKOR, 8 Luer-Lock caps, 1 nut to secure the rotor, 2 Luer-Lock connectors. |
Section 05 – Page 5
{5}
Hydra
510(k) Number: BK220775 December, 2 2022
| | Procedure Tray | Predicate Device BK220718-0/K190386 |
| --- | --- | --- |
| **Representation of device** | | |
| **Materials of components** | Syringes and connectors are made of plastic materials, needles and cannula are made of stainless steel. | Syringes composed of medical grade plastics, needles and cannula composed of steel. |
| **Sterilization** | Ethylene Oxide Gas (EtO) | Ethylene Oxide Gas (EtO) |
| **Disposable/Reusable** | Disposable, single use only | Disposable, single use only |
| **Shelf-life** | 5 years | 3 years |
| **Packaging** | The tray is packaged in a pouch made of LDPE transparent with insert in Tyvek 1073B | Tyvek sealing |
**Table 2** – Comparison of technological characteristics between “Procedure Tray” and predicate device
“Procedure Tray” manufactured by Hydra Srl and the predicate device have the same intended use, general configuration, sterilization method, components materials and are both single use devices.
Technical differences are the following:
- “Procedure Tray” contains 3 cannulas: 17G, 13G and 20G while predicate device contains infiltration cannula (Ø 2.1 mm x 250 mm), harvesting cannula 20 holes (Ø 3.0 x 3.5 mm x 250 mm) and (Ø 3.0 x 3.5 mm x 200 mm), injection cannulas – 14G (Ø 1.7 – 2.1 mm x 200 mm) and (Ø 1.7 – 2.1 mm x 130 mm), needles – 18G / 18G NOKOR.
- “Procedure Tray” contains 20 ml aspiration syringe, 1 ml injection syringe, 10 ml syringe, 50 ml syringe, Luer Lock female to female connector while predicate device contains incremental syringes – 60 ml, injection syringes – 10 ml, Luer-Lock caps, nut to secure the rotor, Luer-Lock connectors.
- Shelf-life of “Procedure Tray” is 5 years while shelf-life of predicate device is 3 years.
Section 05 – Page 6
{6}
Hydra
510(k) Number: BK220775 December, 2 2022
# VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# **Biocompatibility testing**
The biocompatibility evaluation for “Procedure Tray” was conducted in accordance with the guideline “Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process”.
The device was classified as an external communicating device in contact with tissue for a limited contact duration (≤24h).
The following tests were performed:
- Cytotoxicity
- Intracutaneous reactivity
- Acute systemic toxicity
- Lal test
- Pyrogen test
- Delayed hypersensitivity test (GPMT)
All results were compliant.
# VIII. CONCLUSIONS
“Procedure Tray” is substantially equivalent to the predicate devices in Indications for Use, composition and design. It is as safe and as effective as the predicate. Hydra Srl considers “Procedure Tray” to be substantially equivalent to the predicate device.
Section 05 – Page 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.