Galileo Echo

BK210608 · Immucor, Inc. Norcross, GA · KSZ · Jan 8, 2025 · Hematology

Device Facts

Record IDBK210608
Device NameGalileo Echo
ApplicantImmucor, Inc. Norcross, GA
Product CodeKSZ · Hematology
Decision DateJan 8, 2025
DecisionSESE
Regulation21 CFR 864.9175
Device ClassClass 2

Indications for Use

Echo Lumena: The Echo Lumena is a microprocessor-controlled instrument designed to fully automate immunohematology in vitro diagnostic testing of human blood. The Echo Lumena automates test processing, result interpretation, and data management functions. The Echo Lumena is designed to automate standard immunohematology assays using a micro-well strip-based platform. Assays include ABO and Rh (D) typing, detection/identification of IgG red blood cell antibodies, compatibility testing, and red blood cell phenotyping. The Echo Lumena is a closed system intended for use only with the reagents described in the Echo Lumena Operator Manual, Regional Attachment – North America- Echo Lumena Assay Reagents and Cutoffs. Galileo Echo: The Galileo Echo is a microprocessor-controlled instrument designed to fully automate immunohematology in vitro diagnostic testing of human blood. The Galileo Echo automates test processing, result interpretation and data management functions. The Galileo Echo is designed to automate standard immunohematology assays using a micro-well strip-based platform. Assays include ABO and Rh (D) typing, detection/identification of IgG red blood cell antibodies, compatibility testing and red blood cell phenotyping. The Galileo Echo is a closed system intended for use only with the reagents described in the Galileo Echo Operator Manual, Regional Attachment – North America- Galileo Echo Assay Reagents and Cutoffs.

Device Story

Automated blood bank analyzer; performs immunohematology assays via hemagglutination and solid phase red cell adherence (SPRCA). System inputs: blood samples, liquid reagents, micro-well strips. Robotic arm moves samples/reagents through incubation, washing, centrifugation, and reading stations. CCD camera captures micro-well images; software performs multi-feature image analysis to calculate reaction values; assigns results based on predefined criteria. Used in clinical laboratories; operated by trained personnel. Provides automated result interpretation and data management; integrates with laboratory information systems. Benefits: walk-away automation, standardized testing, reduced manual error.

Clinical Evidence

Method comparison study conducted at two external sites and one internal site. Compared Automated C3d DAT assay on Galileo Echo/Echo Lumena against a comparator method. Total sample size: 1,143 (initial) and 1,141 (resolved). Initial PPA: 95.17% (90.16% CI); NPA: 99.32% (98.73% CI). Resolved PPA: 100% (97.70% CI); NPA: 99.33% (98.74% CI).

Technological Characteristics

Microprocessor-controlled robotic instrument; micro-well strip-based platform. Sensing: CCD camera for digital image analysis. Connectivity: LIS interface. Software: Microsoft Windows 10. Closed system; requires specific Immucor reagents. Automated incubation, washing, and centrifugation.

Indications for Use

Indicated for automated in vitro diagnostic immunohematology testing of human blood, including ABO/Rh typing, IgG antibody detection/identification, compatibility testing, red blood cell phenotyping, and C3d DAT assays.

Regulatory Classification

Identification

An automated blood grouping and antibody test system is a device used to group erythrocytes (red blood cells) and to detect antibodies to blood group antigens.

Predicate Devices

Submission Summary (Full Text)

{1} Traditional 510(k) Premarket Notification IMMUCOR ## 7.0 510(k) Summary ### Date Prepared June 4, 2021 ### 510(k) Owner Immucor, Inc. 3130 Gateway Drive Norcross, Georgia 30071 Establishment Registration Number: 1034569 ### Contact Information Name of Contact: Steven Appel Phone Number: 770-440-2051 Fax Number: 770-242-8930 ### Device Name | Trade/Device Name: | Echo Lumena® | Galileo Echo® | | --- | --- | --- | | Common Name: | Automated Blood Bank Analyzer | Automated Blood Bank Analyzer | | Classification Name: | Automated blood grouping and antibody test system | Automated blood grouping and antibody test system | | Unique Device Identifier (UDI): | 10888234002338 | 10888234001584 | ### Device Class | Trade/Device Name: | Echo Lumena® | Galileo Echo® | | --- | --- | --- | | Regulatory Class: | II | II | | Product Code: | KSZ | KSZ | | Regulation Number: | 21CFR§864.9175 | 21CFR§864.9175 | | Classification Advisory Committee: | Hematology | Hematology | | Review Advisory Committee: | Hematology | Hematology | ### Predicate Device Information | Trade/Device Name: | Echo Lumena® | Galileo Echo® | | --- | --- | --- | | Clearance: | BK200483 | BK200483 | | Date Cleared: | June 22, 2020 | June 22, 2020 | ### Device Description Galileo Echo (v.2.0) is an upgrade of the Galileo Echo to bring that instrument up to the design and performance specifications of the Echo Lumena by replacement of camera and software components; both instruments are technologically and functionally identical. The Galileo Echo (v.2.0) and Echo Lumena (hereafter "Echo" will refer to either instrument) are designed to automate standard immunohematology assays and to operate as walk-away systems, meaning you can leave the Echo to operate independently for periods of time. Several unified principles have been integrated into the Echo system to support and simplify the system's overall operation. The Echo uses two (2) basic methods of blood testing, namely hemagglutination and solid phase red cell adherence (SPRCA). Hemagglutination is used for red blood cell typing assays which includes ABO and Rh(D) typing and red blood cell phenotyping. Solid phase is used for antibody detection and identification assays, compatibility testing and weak D. The Echo is a closed system and can only be used with specified Immucor products. The Echo is a robotic instrument programmed to move micro-well strips, liquid reagent fluids, and blood Traditional 510(k) Premarket Notification Modification to Echo Lumena and Galileo Echo by Addition of Automated C3d DAT Assay Page 61 of 128 {2} Traditional 510(k) Premarket Notification IMMUCOR sample fluids to different processing areas for a given assay in the correct sequence, such as the incubator, the micro-well washing station, the centrifuge, and the reader. The Echo micro-well reader uses a CCD camera to capture an image of the micro-well. The Echo software calculates a reaction value for each well based on a multi-feature image analysis. The Echo then assigns a result and interpretation to the well based on predefined criteria associated with the calculated reaction value. Some assay protocols require multiple test wells for a given blood sample interpretation, such as ABO and Rh (D) typing. The Echo uses software to drive its mechanics and data processing. The operator uses hardware in combination with the software to operate and maintain the Echo. ## Intended Use The intended use of the modified devices, as described in the labeling, has not changed as a result of the modifications. ### Echo Lumena: The Echo Lumena is a microprocessor-controlled instrument designed to fully automate immunohematology in vitro diagnostic testing of human blood. The Echo Lumena automates test processing, result interpretation, and data management functions. The Echo Lumena is designed to automate standard immunohematology assays using a micro-well strip-based platform. Assays include ABO and Rh (D) typing, detection/identification of IgG red blood cell antibodies, compatibility testing, and red blood cell phenotyping. The Echo Lumena is a closed system intended for use only with the reagents described in the Echo Lumena Operator Manual, Regional Attachment – North America- Echo Lumena Assay Reagents and Cutoffs. ### Galileo Echo: The Galileo Echo is a microprocessor-controlled instrument designed to fully automate immunohematology in vitro diagnostic testing of human blood. The Galileo Echo automates test processing, result interpretation and data management functions. The Galileo Echo is designed to automate standard immunohematology assays using a micro-well strip-based platform. Assays include ABO and Rh (D) typing, detection/identification of IgG red blood cell antibodies, compatibility testing and red blood cell phenotyping. The Galileo Echo is a closed system intended for use only with the reagents described in the Galileo Echo Operator Manual, Regional Attachment – North America- Galileo Echo Assay Reagents and Cutoffs. ## Technological Comparison to Predicate Device Below is a summary of the technological characteristics of modified Echo Lumena and Galileo Echo (proposed devices) compared to the predicate devices (BK200483). | Properties/Features/ Characteristics | Predicates (BK200483) | Modified Echo Lumena and Galileo Echo | Comparison | | --- | --- | --- | --- | | Common Name | Automated Blood Bank Analyzer | Automated Blood Bank Analyzer | Identical | | Classification | Class II | Class II | Identical | | Regulation | 21 CFR 864.9175 | 21 CFR 864.9175 | Identical | | Intended Use | Automated immunohematology analyzer for in vitro diagnostic use | Automated immunohematology analyzer for in vitro diagnostic use | Identical | Traditional 510(k) Premarket Notification Modification to Echo Lumena and Galileo Echo by Addition of Automated C3d DAT Assay Page 62 of 128 {3} Traditional 510(k) Premarket Notification IMMUCOR | Properties/Features/Characteristics | Predicates (BK200483) | Modified Echo Lumena and Galileo Echo | Comparison | | --- | --- | --- | --- | | Indications for Use | Microprocessor controlled instrument designed to fully automate immunohematology in vitro diagnostic testing of human blood. Automates test processing, result interpretation, and data management functions. Designed to automate standard immunohematology assays using a micro-well strip-based platform. Assays include ABO and Rh (D) typing, detection/identification of IgG red blood cell antibodies, compatibility testing, and red blood cell phenotyping. | Microprocessor controlled instrument designed to fully automate immunohematology in vitro diagnostic testing of human blood. Automates test processing, result interpretation, and data management functions. Designed to automate standard immunohematology assays using a micro-well strip-based platform. Assays include ABO and Rh (D) typing, detection/identification of IgG red blood cell antibodies, compatibility testing, and red blood cell phenotyping. | Identical | | Tests Performed | ABO & RH Typing Antibody Screen Antibody Identification IgG Crossmatch Direct Antiglobulin Antigen Typing | ABO & RH Typing Antibody Screen Antibody Identification IgG Crossmatch Direct Antiglobulin Antigen Typing | Identical | | Test Reactions Reading | Digital Image Analysis | Digital Image Analysis | Identical | | User Interface | By Computer Workstation | By Computer Workstation | Identical | | System Security requires user passwords for access | Yes | Yes | Identical | | Testing performed on Plasma | Yes | Yes | Identical | | Testing performed on serum | Yes | Yes | Identical | | Barcode read on reagent and samples to confirm presence and location on the instrument | Yes | Yes | Identical | | Barcode read of reagent lot number and expiration date | Yes | Yes | Identical | | Manual entry of sample or reagent barcode requiring double blind entry | Yes | Yes | Identical | | Acceptable reagent vial size | 10 mL | 10 mL | Identical | | Sample and reagent volume verification at aspiration | Yes | Yes | Identical | | Programmed to track volume or usage of each reagent vial or plate | Yes | Yes | Identical | | Prepares sample red cell suspension | Yes | Yes | Identical | | Multiple vials of same reagent can be loaded on instrument. When empty instrument switches to second vial | Yes | Yes | Identical | | Maintains Reagent Red Cell suspensions by agitation | Yes | Yes | Identical | | Walk away testing capability | Yes | Yes | Identical | | Instrument will discontinue operation if liquid waste is full | Yes | Yes | Identical | | Incubation duration and temperature are monitored | Yes | Yes | Identical | Traditional 510(k) Premarket Notification Modification to Echo Lumena and Galileo Echo by Addition of Automated C3d DAT Assay Page 63 of 128 {4} Traditional 510(k) Premarket Notification IMMUCOR | Properties/Features/ Characteristics | Predicates (BK200483) | Modified Echo Lumena and Galileo Echo | Comparison | | --- | --- | --- | --- | | Centrifuge performs at a consistent rpm range and duration | Yes | Yes | Identical | | Can be interfaced to laboratory information systems | Yes | Yes | Identical | | Camera Model/ module | Lumenera Camera | Lumenera Camera | Identical | | Software Operating System | Microsoft Windows 10 | Microsoft Windows 10 | Identical | | Algorithm interpretation and thresholds | Interpretation by camera imaging | Interpretation by camera imaging | Identical | | Automated C3d DAT Assay | No | Yes | New | ### Clinical Performance Method comparison studies were performed externally at two sites and internally at Immucor, Inc. Specimens were tested with the Automated C3d DAT assay on Galileo Echo/Echo Lumena and the comparator method. Samples with initial equivocal results were retest and the results used for calculations. Test results were evaluated for agreement with the comparator method. Results are summarized in the following tables: Note: Agreement between methods does not indicate which method is correct. | Initial C3d DAT | Comparator | | Agreement | | | | --- | --- | --- | --- | --- | --- | | | | Positive | | | Negative | | Galileo Echo^{#} & Echo Lumena | Positive | 99* | 7 | Positive % Agreement | 95.17% | | | | | | PPA (95% Lower 1-sided CI) | 90.16% | | | Negative | 5 | 1,032 | Negative % Agreement | 99.32% | | | | | | NPA (95% Lower 1-sided CI) | 98.73% | $^{#}$ Galileo Echo software version 2.1. Positive samples include 86 C3d-positive contrived samples. * One contrived sample tested repeat equivocal and was removed from PPA calculation. | Resolved C3d DAT | Comparator | | Agreement | | | | --- | --- | --- | --- | --- | --- | | | | Positive | | | Negative | | Galileo Echo^{#} & Echo Lumena | Positive | 99 | 7 | Positive % Agreement | 100% | | | | | | PPA (95% Lower 1-sided CI) | 97.70% | | | Negative | 0* | 1,035 | Negative % Agreement | 99.33% | | | | | | NPA (95% Lower 1-sided CI) | 98.74% | $^{#}$ Galileo Echo software version 2.1. * Excludes 3 samples reacting 1+ or weaker with the comparator reagent per limitation in Operator Manuals. ### Basis for Claim of Substantial Equivalence This Traditional 510(k) is submitted to modify legally marketed devices (predicates). Immucor, Inc. is the holder of the 510(k)s for the predicate devices. The Indications for Use of the proposed devices are unchanged from the legally marketed devices (predicates). The intended use of the modified devices, as described in the labeling, has not changed as a result of the modifications. Finally, fundamental scientific technology of the proposed devices is unchanged from the legally marketed devices (predicates). There are no significant differences between the modified Traditional 510(k) Premarket Notification Modification to Echo Lumena and Galileo Echo by Addition of Automated C3d DAT Assay Page 64 of 128 {5} Traditional 510(k) Premarket Notification IMMUCOR instruments and the predicates as relates to Intended Use or Principle of Operation. With the exception of the addition of the automated C3d DAT Assay, the predicate and modified devices are identical. The modified Echo Lumena and Galileo Echo instruments are as safe and effective as the currently marketed predicates under BK200483. Traditional 510(k) Premarket Notification Modification to Echo Lumena and Galileo Echo by Addition of Automated C3d DAT Assay Page 65 of 128
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...